The Paltop Conical Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. Narrow diameter implants are intended for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited by the adjacent teeth and roots. The Paltop Conical Implant System is also indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
Paltop Conical Implant System consists of endosseous root-form dental implants, abutments, and retaining screws; used for single-unit, multi-unit, and overdenture restorations. Implants are surgically placed into jaw bone by dental professionals; abutments are secured to implants to support prosthetic teeth. System includes various configurations: straight, angulated, multi-unit, and temporary abutments. Implants feature SLA (sand-blasted, large grit, acid-etched) surface treatment. Components are fabricated from Grade 23 titanium alloy (ASTM F136) or PEEK. Device provides mechanical support for prosthetic restorations, restoring patient chewing function. Clinical benefit is the replacement of missing teeth via stable osseointegrated support.
Clinical Evidence
No clinical data were included in this submission. Bench testing included fatigue testing per ISO 14801:2016, biocompatibility evaluation per ISO 10993-1, and endotoxin testing per USP<85>/<161>. MRI safety was assessed via scientific rationale and literature.
Technological Characteristics
Endosseous screw-type implants with internal hex connection and coronal conical taper. Materials: Grade 23 Titanium Alloy (ASTM F136), PEEK. Surface: SLA (Sand-blasted, Large grit, Acid-Etched). Sterilization: Gamma irradiation (sterile components) or steam (non-sterile components). Dimensions: Diameters 3.25-5.0 mm; lengths 8-16 mm.
Indications for Use
Indicated for surgical and restorative placement in upper or lower jaw bone to support prosthetic devices (artificial teeth) for chewing function restoration. Narrow diameter implants indicated for mandibular central/lateral incisors and maxillary lateral incisors in partially edentulous jaws with limited horizontal space. Indicated for immediate loading given good primary stability and appropriate occlusal loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Paltop Advanced Dental Solutions, Ltd % Chris Brown Manager Aclivi, LLC 3250 Brackley Drive Ann Arbor, Michigan 48105
Re: K220200
Trade/Device Name: Paltop Conical Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: April 28, 2022 Received: April 29, 2022
Dear Chris Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K220200
Device Name Paltop Conical Implant System
#### Indications for Use (Describe)
The Paltop Conical Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. Narrow diameter implants are intended for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited by the adjacent teeth and roots. The Paltop Conical Implant System is also indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K220200 Paltop Advanced Dental Solutions, Ltd Paltop Conical Implant System
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Paltop Advanced Dental Solutions, Ltd<br>Hashita 5, Industrial Park<br>Caesarea 3088900<br>Israel | |
|-------------------|---------------------------------------------------------------------------------------------------|-------------------|
| | Telephone: | +(972) 4-627 1711 |
| | Fax: | +(972) 4-627 5363 |
| Official Contact | Zina Gurgov, Director of QA/RA<br>Email: zgurgov@keystonedental.com | |
| Date submitted: | 5/25/2022 | |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | Paltop Conical Implant System | |
|----------------------------|----------------------------------------------------------------------------------------------|--|
| Common Name: | Implant, Dental, Endosseous, Root-Form | |
| Classification Name: | Endosseous dental implant | |
| Classification Regulation: | 21 CFR 872.3640 | |
| Device Class: | Class II | |
| Product Code: | DZE, NHA | |
| Review Panel: | Dental | |
| Reviewing Branch: | Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1)<br>Dental Devices (DHT1B) | |
# PREDICATE DEVICE INFORMATION
The devices within this submission are substantially equivalent in indications, intended use and design principles to the following primary predicate and reference devices:
| 510(k) | Primary Predicate Device Name | Company Name |
|---------|----------------------------------------|----------------------------------|
| K210117 | Paltop Narrow Implant | Paltop Advanced Dental Solutions |
| 510(k) | Reference Device Name | Company Name |
| K112795 | Paltop Advanced Dental Solution System | Paltop Advanced Dental Solutions |
{5}------------------------------------------------
### DEVICE DESCRIPTION
The purpose of this submission the marketing clearance for the Paltop Conical Implant System which comprises endosseous root-form dental implants, mating abutment screws, and other associated components for single-unit, multi-unit, and overdenture restorations.
Endosseous dental implants are surgically implanted into a patient's mouth to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. Endosseous dental implant abutments are secured to dental implants with a retaining screw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function.
The Paltop Conical Implant System implants are one- and two-stage endosseous screw type dental implants with associated compatible abutments, screws, and other associated accessory components. The Paltop Conical Implant System prosthetic components include healing caps, and multi-unit abutments which are available in straight or angulated configurations. Prosthetic devices used with the dental implant abutments in this submission may be screw-retained or cemented.
The implants, titanium abutment screws are fabricated from a Titanium-6 Aluminum 4 Vanadium ELI titanium alloy (Grade 23) which conforms to ASTM F136, Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). The Subject device implants are surface treated with SLA (Sand-blasted, Large Grit, Acid Etched). Immediate temporary abutments include a PEEK (Polyether ether Ketone) material sleeve, and the SAS abutment includes a PEEK healing cap/coping.
All implants and prosthetic components are one-time use device components are provided sterile and sterilized by gamma irradiation except for Single-Unit and Multi-Unit copings, SAS Abutment components and all replacement screws which are provided non-sterile. Devices provided as non-sterile are sterilized by steam.
| Implant<br>Type | Implant Body<br>Diameter (mm) | Implant Platform<br>Diameter (mm) | Lengths (mm) |
|--------------------|-------------------------------|-----------------------------------|---------------------|
| Dynamic<br>Conical | Ø 3.25 | Ø 3.25 | 10, 11.5, 13, 16 |
| | Ø 3.75 | Ø 3.75 | 8, 10, 11.5, 13, 16 |
| | Ø 4.2 | Ø 4.2 | 8, 10, 11.5, 13, 16 |
| | Ø 5.0 | Ø 5.0 | 8, 10, 11.5, 13, 16 |
# Paltop Conical Implant System - Implant Sizes
#### Paltop Conical Implant System - Abutment Types
| | Subject Device Abutment Designs | | | | | | | | | |
|-----------------------------|---------------------------------|----------|-------------------------|------------------------|----------------------|---------|-------------------------|------------------------|---------|---------------|
| Implant<br>Diameter<br>(mm) | Healing<br>Caps | Straight | Straight<br>Single-Unit | Straight<br>Multi-Unit | Angled<br>Multi-Unit | Angled | Temporary | Temporary<br>Immediate | SAS | Straight Ball |
| 3.25 | X | X | X | X | X | X | X | X | X | X |
| 3.75 | X | X | X | X | X | X | X | X | X | X |
| 4.2 | X | X | X | X | X | X | X | X | X | X |
| 5.0 | X | X | X | X | X | X | X | X | X | X |
| | Non-<br>indexed | Indexed | Non-<br>indexed | Non-<br>Indexed | Indexed | Indexed | Indexed,<br>Non-Indexed | Non-<br>Indexed | Indexed | Non-Indexed |
| Material | Grade 23 Titanium Alloy | | | | | | | | | |
| Finish | Gold<br>Anodize | | | | | | None | | | |
{6}------------------------------------------------
### INDICATIONS FOR USE
The Paltop Conical Implant System is in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. Narrow diameter implants are intended for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited by the adjacent teeth and roots. The Paltop Conical Implant System is also indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
### EQUIVALENCE TO MARKETED DEVICE
The Subject device is highly similar to the Primary Predicate device with respect to Indications for Use and technological principles. The Comparison tables below compare the Indications for Use and Technological Characteristics of the Subject and Primary Predicate/Reference devices.
| Device | Indications for Use Statement |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device<br>Paltop Conical Implant<br>System | The Paltop Conical Implant System is indicated for use in surgical and restorative applications for placement in the bone<br>of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the<br>patient's chewing function. Narrow diameter implants are intended for placement in the mandibular central, lateral<br>incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited by the<br>adjacent teeth and roots. The Paltop Conical Implant System is also indicated for immediate loading when good primary<br>stability is achieved and with appropriate occlusal loading. |
| Primary Predicate<br>Device<br>Paltop Narrow Implant<br>(K210117) | The Paltop Narrow Implant is indicated for use in surgical and restorative applications for placement in the mandibular<br>central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is<br>limited by the adjacent teeth and roots, to provide support for prosthetic devices, such as artificial teeth, in order to<br>restore the patient's chewing function. The Paltop Narrow Implant is indicated also for immediate loading when good<br>primary stability is achieved and with appropriate occlusal loading. |
| Reference Device<br>Paltop Advanced Dental<br>Solutions System<br>(K112795) | The Paltop Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of<br>the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's<br>chewing function. The Paltop Dental Implant System is indicated also for immediate loading when good primary stability<br>is achieved and with appropriate occlusal loading. |
# Comparison of Indications for Use Statements
The Subject, Primary Predicate and Reference devices have highly similar Indications for Use, differing primarily in device name. The Primary Predicate device was a narrow diameter implant submission and the Reference device included larger diameter implants. The Subject device Indications for Use statement combines the relevant features of both diameter configuration implants into a single statement as the Subject device encompasses both Narrow and standard diameter implants and includes the trade name of the Subject device. These minor differences do not raise new questions of safety or effectiveness as all the Indications for Use statements express equivalent intended use.
| Design Parameter | Subject Device | Primary Predicate Device | Reference Device |
|-----------------------------------|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| | Paltop Conical Implant System<br>Paltop Advanced Dental Solutions | Paltop Narrow Implant (K210117)<br>Paltop Advanced Dental Solutions | Paltop Advanced Dental Solutions System (K112795)<br>Paltop Advanced Dental Solutions |
| Regulation # | 21 CFR 872.3640 | 21 CFR 872.3640 | 21 CFR 872.3640 |
| Product Code | DZE, NHA | DZE, NHA | DZE, NHA |
| Classification | Class II | Class II | Class II |
| Materials | Implants/Screws - Titanium Ti-6Al-4V ELI<br>Abutments- Titanium Ti-6Al-4V ELI, PEEK | Implants/Screws - Titanium Ti-6Al-4V ELI<br>Abutments- Titanium Ti-6Al-4V ELI, PEEK | Implants/Screws - Titanium Ti-6Al-4V ELI<br>Abutments- Titanium Ti-6Al-4V ELI or PEEK |
| Reason for<br>Predicate/Reference | Not Applicable | Narrow diameter implant, Implant length,<br>implant design, implant modified surface,<br>sterilization, biocompatibility, how provided | Standard diameter implants, implant lengths |
#### Comparison of Technological Characteristics
{7}------------------------------------------------
### Implant Design
| | Subject Device | Primary Predicate Device | Reference Device | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|-----------------------------------------|--|--|--|--|--|--|----------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|
| Design Parameter | Paltop Conical Implant System<br>Paltop Advanced Dental Solutions | Paltop Narrow Implant (K210117)<br>Paltop Advanced Dental Solutions | Paltop Advanced Dental Solutions System (K112795)<br>Paltop Advanced Dental Solutions | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | Endosseous screw-type implant with<br>internal connection. Parallel coronal and<br>midsection, micro threads on neck,<br>reverse buttress thread in mid-section<br>tapering to an active/cutting apex.<br>Platform switching taper on implant top-<br>level. | Dynamic -<br>Endosseous screw-type implant with<br>internal connection. Parallel coronal and<br>midsection, micro threads on neck,<br>reverse buttress thread in mid-section<br>tapering to an active/cutting apex. | Dynamic -<br>Endosseous screw-type implant with internal<br>connection. Parallel coronal and midsection, micro<br>threads on neck, reverse buttress thread in mid-<br>section tapering to an active/cutting apex. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Dynamic Conical<br>D = Implant Body Diameter<br>IP = Implant Platform Diameter | D IP Lengths 3.25 3.25 10, 11.5, 13, 16 3.75 3.75 8, 10, 11.5, 13, 16 4.2 4.2 8, 10, 11.5, 13, 16 5.0 5.0 8, 10, 11.5, 13, 16 | | | | | | | | | | | | | | | | D IP Lengths 3.0 3.0 10, 11.5, 13, 16 | | | | | | | D IP Lengths 3.75 3.75 8, 10, 11.5, 13, 16 4.2 4.2 8, 10, 11.5, 13, 16 5.0 5.0 8, 10, 11.5, 13, 16 | | | | | | | | | | | | |
| Mode of Operation | Provide support for prosthetic devices,<br>such as artificial teeth, to restore the<br>patient's chewing function. | Provide support for prosthetic devices,<br>such as artificial teeth, to restore the<br>patient's chewing function. | Provide support for single and multi-unit<br>prostheses to restore the patient's chewing<br>function. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Implant Material | Grade 23 Titanium Alloy | Grade 23 Titanium Alloy | Grade 23 Titanium Alloy | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Implant Surface<br>Treatment | Sand-blasted, Large grit, Acid-Etched<br>(SLA) | Sand-blasted, Large grit, Acid-Etched<br>(SLA) | Sand-blasted, Large grit, Acid-Etched (SLA) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Sterilization Method | Gamma Sterilization | Gamma Sterilization | Gamma Sterilization | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Implant/<br>Abutment Interface | Hex Internal interface with coronal<br>conical taper | Hex Internal interface | Hex Internal interface | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
The Subject device implant design material and external threads, and implant lengths are the Primary Predicate and Reference devices. The range of Subject device implant diameters are supported by the Primary Predicate and Reference devices. Implant surface treatment materials and process is the same for the Subject and Primary Predicate devices. The Subject device implants are sterilized by gamma radiation and packaged the same as the Primary Predicate device K210117. The Subject device implants are single use, single user, the same as the Primary Predicate device. The introduction of the Subject device platform switching, and the coronal taper of the implant/abutment interface connection does not affect substantial equivalence nor change the intended use of the devices. Differences in the implant/abutment connections between the Subject and Primary Predicate and Reference devices do not change the intended use and have been mitigated through non-clinical bench performance testing.
{8}------------------------------------------------
# Technological Characteristics Comparison Table - Abutments
| Comparison | Subject Device<br>Paltop Conical Implant System<br>Paltop Advanced Dental Solutions | Primary Predicate Device<br>Paltop Narrow Implant (K210117)<br>Paltop Advanced Dental Solutions | Reference Device<br>Paltop Advanced Dental Solutions System (K112795)<br>Paltop Advanced Dental Solutions | |
|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--|
| Product Code | DZE, NHA | DZE, NHA | DZE, NHA | |
| Regulation | 872.3640, 872.3630 | 872.3640, 872.3630 | 872.3640, 872.3630 | |
| Intended Use | Functional and esthetic rehabilitation of the | Functional and esthetic rehabilitation of the | Functional and esthetic rehabilitation of the edentulous | |
| | edentulous maxilla and mandible | edentulous maxilla and mandible | maxilla and mandible | |
| Reason for Predicate/Reference<br>Sterilization method - Sterile components | Not Applicable<br>Gamma Sterilization | Use with small diameter implant, abutment designs<br>Gamma Sterilization | Use with standard diameter implants, abutment designs<br>Gamma Sterilization | |
| Sterilization method - non-Sterile | | | Gamma Sterilization | |
| components | Steam sterilization | Steam sterilization | Steam sterilization | |
| Abutment/Screw Material | Grade 23 Titanium Alloy<br>no finish unless otherwise specified | Grade 23 Titanium Alloy<br>no finish unless otherwise specified | Grade 23 Titanium Alloy<br>no finish unless otherwise specified | |
| Use with Implant diameters | 3.25 mm, 3.75 mm, 4.2 mm, 5.0 mm | 3.0 mm, 3.25 mm | 3.75 mm, 4.2 mm, 5.0 mm | |
| Cover Screws (supplied | | | | |
| w/implants) | PD - 3.25 mm, 3.75 mm, 4.2 mm, 5.0 mm | PD - 3.0 mm, 3.25 mm | PD - 3.75 mm, 4.2 mm, 5.0 mm | |
| Healing Caps | Design<br>GH<br>PD<br>Max<br>CA<br>PH | Design<br>GH<br>PD<br>Max<br>CA<br>PH | Design<br>GH<br>PD<br>Max<br>CA<br>PH | |
| | Straight<br>2<br>3<br>0<br>1 | Straight<br>2,3,5<br>4<br>0<br>0.25 | Straight<br>1,2,3<br>4.5<br>0<br>1 | |
| | Straight<br>2, 3, 5<br>4.5<br>0<br>1 | Concave<br>4,5,6,7<br>4.4<br>0<br>1 | Concave<br>1,2,3<br>5.5<br>0<br>1 | |
| | Concave<br>2,3,4,5,7<br>4.5<br>0<br>1 | Blue anodized finish | No finish | |
| | Concave<br>2,3,5,7<br>6.0<br>0<br>1 | | | |
| | Gold anodized finish | | | |
| Straight Abutments | Design<br>GH<br>PD<br>Max<br>CA<br>PH | n/a | Design<br>GH<br>PD<br>Max<br>CA<br>PH | |
| | Knife-edge<br>0.8<br>3.8<br>0<br>8 | | Narrow<br>0<br>3.9<br>0<br>8 | |
| | Concave<br>1,2,3,4,5<br>4.5<br>0<br>7.5 | | Straight<br>1,2,3<br>4.5, 5.5<br>0<br>8 | |
| | Concave<br>1,2,3,4,5<br>6.0<br>0<br>7.5 | | Anatomic<br>1,2,3<br>4.7<br>0<br>8 | |
| Straight Single-Unit Abutment | Design<br>GH<br>PD<br>Max<br>CA<br>PH | Design<br>GH<br>PD<br>Max<br>CA<br>PH | | |
| | w/Temp<br>Cylinder<br>1,2 3,4,5<br>4.5<br>0<br>5<br>(min) | w/Temp<br>Cylinder<br>1,2 3,4<br>4.5<br>0<br>5<br>(min) | | |
| | w/Interface<br>Coping<br>1,2 3,4,5<br>4.5<br>0<br>5.5 | w/Interface<br>Coping<br>1,2 3,4<br>4.5<br>0<br>5.5 | n/a | |
| | w/Healing<br>Cap<br>1,2 3,4,5<br>4.5<br>0<br>3.5 | w/Healing<br>Cap<br>1,2 3,4<br>4.5<br>0<br>3.5 | | |
| Straight Multi-Unit Abutment | Design<br>GH<br>PD<br>Max<br>CA<br>PH | Design<br>GH<br>PD<br>Max<br>CA<br>PH | Design<br>GH<br>PD<br>Max<br>CA<br>PH | |
| | w/Temp<br>Cylinder<br>1,2 3,4,5<br>5<br>0<br>5<br>(min) | w/Temp<br>Cylinder<br>1,2 3,4,5<br>5<br>0<br>5<br>(min) | w/Ti-<br>Abutment<br>(Cylinder)<br>1,2 3<br>4.5<br>0<br>n/s | |
| | w/Interface<br>Coping<br>1,2 3,4,5<br>5<br>0<br>4.5<br>(min) | w/Interface<br>Coping<br>1,2 3,4,5<br>5<br>0<br>4.5<br>(min) | w/Healing<br>Cap<br>1,2 3<br>4.5<br>0<br>3.3 | |
| | w/Healing<br>Cap<br>1,2 3,4,5<br>5<br>0<br>4.5 | w/Healing<br>Cap<br>1,2 3,4,5<br>5<br>0<br>4.5 | | |
| Angulated Multi-Unit Abutment | Design<br>GH<br>PD<br>CA<br>PH | Design<br>GH<br>PD<br>CA<br>PH | Design<br>GH<br>PD<br>CA<br>PH | |
| | 17°w/<br>Temp Cylinder<br>3,4.5<br>5<br>17<br>5<br>(min) | 17°w/<br>Temp Cylinder<br>3<br>5<br>17<br>5<br>(min) | 17°w/<br>Ti-Abutment<br>(Cylinder)<br>2,3<br>4.5<br>17<br>n/s | |
| | 17°w/<br>Interface<br>Coping<br>3,4.5<br>5<br>17<br>4.5<br>(min) | 17°w/<br>Interface<br>Coping<br>3<br>5<br>17<br>4.5<br>(min) | 17°w/ Healing<br>Cap<br>2,3<br>4.5<br>17<br>3.3 | |
| | 17°w/ Healing<br>Cap<br>3,4.5<br>5<br>17<br>4.5 | 17°w/ Healing<br>Cap<br>3<br>5<br>17<br>4.5 | | |
| | Design<br>GH<br>PD<br>CA<br>PH | | | |
| | 30°w/<br>Temp Cylinder<br>4<br>5<br>30<br>5<br>(min) | | | |
| | 30°w/<br>Interface<br>Coping<br>4<br>5<br>30<br>4.5<br>(min) | | | |
| | 30°w/ Healing<br>Cap<br>4<br>5<br>30<br>4.5 | | | |
| Angulated Abutment | Design<br>GH<br>PD<br>CA<br>PH | n/a | Design<br>GH<br>PD<br>CA<br>PH | |
| | Knife-edge<br>0.8<br>4.5<br>20<br>8 | | Angulated<br>1<br>4.75<br>15, 25<br>8 | |
| | Concave<br>1,2,3,4,5<br>4.5<br>20<br>7.5 | | Anatomic<br>1,2,3<br>4.75,5<br>15, 25<br>8 | |
| | | | | |
{9}------------------------------------------------
Image /page/8/Figure/0 description: This image is a comparison chart of dental implants and abutments from Paltop Advanced Dental Solutions. The chart compares the Paltop Conical Implant System (Subject Device) to the Paltop Narrow Implant (Primary Predicate Device) and the Paltop Advanced Dental Solutions System (Reference Device). The chart provides information on the design, gingival height (GH), prosthetic diameter (PD), correction angle (CA), and platform height (PH) for various abutments, including temporary abutments, SAS abutments, ball abutments, and straight abutments. The chart also includes information on the temporary use of the abutments, such as up to 90 days or up to 30 days.
PD - Prosthetic Diameter PH – Post Height
Overall, the Subject device abutments are highly similar to the Primary Predicate and Reference devices. Abutment designs are the same in principle, and the introduction of the SAS Abutment is considered a minor variation and combination of the Temporary, Immediate abutment, healing cap and straight abutment concepts. Critical abutment dimensions, such as the Gingival Height, Prosthetic Diameter, Post Correction Angle, and Post Heights are highly similar between the Subject, Primary Predicate and Reference device abutment designs. Subject device Abutments and screws are fabricated from the same materials as the Primary Predicate device. Sterilization and packaging of the sterile Subject device abutments and screws are the Primary Predicate device. Cleaning and sterilization of nonsterile Subject device abutments and screws are the Primary Predicate device.
# Cover Screws
The Subject device Cover Screws are highly similar, combining the platform diameters of the Primary Predicate and Reference device Cover Screws.
# Healing Caps
The Subject device Healing Caps are highly similar to the Primary Predicate and Reference device Healing Caps, with only slight differences in gingival height, anodization color and prosthetic diameter dimensions, and the implant connection.
# Straight Abutments
The Subject device Straight Abutments are highly similar to the Reference device Straight Abutments, with only slight differences in gingival height, prosthetic diameter and post height dimensions, and implant connection.
{10}------------------------------------------------
# Straight Single-Unit Abutments
The Subject device Straight Single-Unit Abutments are highly similar to the Primary Predicate device Straight Single-Unit Abutments, adding one additional gingival height option and the new implant connection. Interface copings, temporary cylinders, and healing cap accessories as part of two-part abutments are supported by the sponsor's Primary Predicate device. The use of Single-Unit Abutment accessories including limitations for hand modification of cylinders is the same as the Primary Predicate device.
# Multi-Unit Abutments
The Subject device Straight Multi-Unit Abutments are the Primary Predicate device Straight Multi-Unit Abutments, differing only in the implant connection.
The Subject device 17° Angulated Multi-Unit Abutments are the Primary Predicate device 17° Angulated Multi-Unit Abutments, including one additional gingival height option and differing in the implant connection. The Subject device 30° Angulated Multi-Unit Abutment is similar to the Primary Predicate device 17° Angulated Multi-Unit Abutment, with a larger correction angle and higher gingival height. These specific changes are supported through non-clinical performance testing of the Subject device.
Interface copings, temporary cylinders, and healing cap accessories as part of two-part abutments are supported by the sponsor's Primary Predicate and Reference devices. The use of Multi-Unit Abutment accessories including limitations for hand modification of cylinders is the same as the Primary Predicate device.
### Angulated Abutments
The Subject device Angulated Abutments are highly similar to the Reference device Angulated Abutments, with only slight differences in gingival height and maximum correction and differing in implant connection.
# Temporary Abutments
The Subject device Titanium Temporary Abutments (engaging) are highly similar to the Primary Predicate device Titanium Temporary Abutments (engaging) differing only in implant connection.
The Subject device Temporary, Immediate Abutments are highly similar to the Primary Predicate device Temporary Immediate, Abutments, with only slight differences in available prosthetic diameters and differing in implant connection.
# SAS Abutments
The Subject device SAS Abutments are similar to the Primary Predicate device Temporary Immediate Abutment and healing caps, and the Reference device Straight abutments, with several slight differences. The Subject device SAS abutment combines the Primary Predicate device temporary immediate abutment technology, where a PEEK sleeve can be used as part of chairside fabricated temporary prosthesis. Otherwise, the PEEK cap is cemented to the SAS abutment and used as a healing abutment. The Subject device PEEK material is the same as the Primary Predicate device PEEK material and is subject to the same time of use limitations. It uses the Reference device Straight abutment technology to use the abutment portion of the SAS abutment as a permanent straight abutment to support a cementretained restoration. Slight differences in Gingival Height, Post Height, and implant connection are present, but are supported and mitigated through non-clinical performance testing of the Subject device.
#### Ball Abutments
The Subject device Ball Abutments are highly similar to the Primary Predicate and Reference device Ball Abutments, with only slight differences due to two additional gingival height dimensions offered for the implant connection.
{11}------------------------------------------------
Overall, minor differences in the Subject device implant external thread designs, dimensions, correction angles and implant connection do not affect substantial equival Height dimensions of the Subject device abutment components (0.5-7 mm) are highly similar to and encompassed by the Primary Predicate and Reference devices (0.5-7 mm). The Prosthetic diameters of the Subject device abutment components (3-6 mm) are highly similar to the Primary Predicate and Reference devices (3.9-5.5 mm).
Any differences in implant abutment designs or dimensions have been mitigated and demonstrated to be suitable for intended use through non-clinical bench performance testing.
# NON-CLINICAL PERFORMANCE TEST DATA
Fatigue testing was performed according to the requirements of ISO 14801:2016, Dentistry – Implants – Dynamic loading test for Endosseous Dental Implants. The worst-case scenario was chosen based on the FDA Guidance, Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments.
The Subject devices have the same nature of body contact duration, material formulation and sterilization methods compared to the sponsor's Primary Predicate and Reference devices.
Biological Evaluation of the Subject device was performed according to ISO 10993-1.
Test results from ISO 10993-5 Cytotoxicity testing is leveraged from the sponsor's Primary Predicate and Reference devices to support suitable biocompatibility of the Subject device.
Endotoxin testing on the Subject device or suitable test specimens was performed following USP<85> and USP<161> according to the sponsor's endotoxin sampling plan.
Test results and Sterilizations performed for the sponsor's Primary Predicate devices is leveraged demonstrate suitable sterilization of the Subject device sterile components.
Sterilization validations for non-sterile components which may be modified and require subsequent sterilization and is leveraged from the Primary Predicate device.
The results of the non-clinical testing demonstrate conformance with testing requirements and support a finding of substantial equivalence with respect to the Subject and Primary Predicate device.
Non-clinical worst-case MRI review was performed to evaluate the Subject device components in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
No clinical data were included in this submission.
{12}------------------------------------------------
### CONCLUSION
Overall, the Indications for Use statements for the Subject and Primary Predicate devices are highly similar.
Overall, the Technological Characteristics, mode of operation and materials of the Subject device are the same or highly similar to that of the Primary Predicate device. Slight differences in design dimensions do not affect the intended use of the device and are mitigated or supported through non-clinical performance testing results. ISO 14801 mechanical performance testing performed on worst-case constructs of the Subject device to demonstrate suitability for intended use of the Subject device implant platform, gingival height, and post correction angles combinations.
Overall, the Subject device and Primary Predicate devices have been demonstrated to be Substantially Equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.