Retrospective clinical CT image datasets from multiple U.S. and international institutions
The retrospective clinical dataset was used to validate the accuracy and reliability of the AI/ML-based TAVR Planning Assist feature by comparing automated measurements against expert-generated ground truth.
Retrospective validation of aortic root measurements; Retrospective performance validation study
Patients undergoing TAVR planning; includes diverse demographics, scanner types, and calcification severities.; Sample Size: 89; Number of Sites: Multiple U.S. and international institutions
Expert manual measurements (ground truth) generated using the FDA-cleared predicate device
Intraclass correlation coefficients (ICC) and Bland-Altman bias for aortic root and LVOT measurements
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Aortic annulus measurements
Deep learning algorithms for segmentation and landmark detection
ICC > 0.90
ICC 0.91–0.95
218 cases
—
89 independent TAVR CT planning cases
2 (board-certified cardiac CT readers)
Coronary artery height measurements
Deep learning algorithms for segmentation and landmark detection
ICC > 0.80
ICC 0.91–0.94
218 cases
—
89 independent TAVR CT planning cases
2 (board-certified cardiac CT readers)
Aortic root angle
Deep learning algorithms for segmentation and landmark detection
ICC > 0.90
ICC 0.97
218 cases
—
89 independent TAVR CT planning cases
2 (board-certified cardiac CT readers)
Left ventricular outflow tract (LVOT) measurements
Deep learning algorithms for segmentation and landmark detection
ICC > 0.90
ICC 0.86–0.93
218 cases
—
89 independent TAVR CT planning cases
2 (board-certified cardiac CT readers)
Sinus of Valsalva measurements
Deep learning algorithms for segmentation and landmark detection
ICC > 0.90
ICC 0.96–0.97
218 cases
—
89 independent TAVR CT planning cases
2 (board-certified cardiac CT readers)
Indications for Use
Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning is a non-invasive post-processing application designed to assist medical professionals with the assessment of the aortic valve and in pre-operational planning and post-operative evaluation of transcatheter aortic valve replacement procedures. Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning includes general functionality such as: • The software processes CT (computed tomography) image data to provide 3D segmentation of heart structures and vessels relevant to approach planning. • The user can review the 2D and 3D images to select and plan the delivery path. • The user can determine C-arm angles for use during the procedure. • The user can verify and adjust the results of segmentation and cross-section measurements. • The software provides visualization techniques such as volume rendering, MIP, MPR and curved MPR. • The user can identify and edit contours and the centerline automatically or manually. • The user can generate a report with relevant approach planning data and measurements for device sizing. • The software can provide visualization of calcium. • The user can generate tortuosity calculations along a centerline. The TAVR Planning Assist feature offers the ability to automatically initialize visualizations and measurements, that users can edit. The calculation of these outputs utilizes artificial intelligence/machine learning methods.
Device Story
Non-invasive post-processing software for TAVR procedure planning; processes CT image data to provide 3D segmentation of heart structures/vessels. Uses deep learning algorithms to automatically initialize anatomical landmarks and aortic root measurements (annulus, coronary heights, LVOT, sinus of Valsalva); user reviews and edits all outputs before finalizing reports. Used in clinical settings by cardiologists/radiologists; assists in delivery path selection, C-arm angle determination, and device sizing. Benefits include standardized, efficient pre-procedural planning and improved visualization of anatomy/calcium. Results exported via DICOM/DICOM SR for clinical decision-making.
Clinical Evidence
Bench testing only. Retrospective study of 89 CT cases (diverse demographics, scanner vendors, calcification severity). Primary endpoints: measurement accuracy vs. expert manual ground truth (ICC, Bland-Altman bias). Results: ICCs generally >0.90 for most parameters (e.g., Annulus area 0.95, Aortic root angle 0.97). Leaflet visualization rated 'Good' or 'Excellent' in >93% of cases. No clinical data.
Technological Characteristics
SaMD; no hardware. Uses deep learning for segmentation and landmark detection. Connectivity: DICOM/DICOM SR. Software-based post-processing on CTA images. Operates on Vitrea platform. No specific material standards applicable as it is software-only.
Indications for Use
Indicated for patients identified as candidates for TAVR procedures. Enables assessment and measurement of the aortic valve and related ventricles to evaluate feasibility of trans-apical, iliac (transfemoral), or subclavian approaches for replacement or repair procedures. Intended for use by cardiologists, radiologists, and clinical specialists.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 22, 2026
Canon Medical Informatics, Inc.
Jay Vaishnav
Director, Regulatory Affairs
5850 Opus Parkway, Suite 300
Minnetonka, Minnesota 55343
Re: K260324
Trade/Device Name: Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: QIH
Dated: April 23, 2026
Received: April 23, 2026
Dear Jay Vaishnav:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260324 - Jay Vaishnav
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260324 - Jay Vaishnav
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

for
Jessica Lamb, Ph.D.
Assistant Director, Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260324 | ? |
| Please provide the device trade name(s). | | ? |
| Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning | | |
| Please provide your Indications for Use below. | | ? |
| Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning is a non-invasive post-processing application designed to assist medical professionals with the assessment of the aortic valve and in pre- operational planning and post-operative evaluation of transcatheter aortic valve replacement procedures. Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning includes general functionality such as: • The software processes CT (computed tomography) image data to provide 3D segmentation of heart structures and vessels relevant to approach planning. • The user can review the 2D and 3D images to select and plan the delivery path. • The user can determine C-arm angles for use during the procedure. • The user can verify and adjust the results of segmentation and cross-section measurements. • The software provides visualization techniques such as volume rendering, MIP, MPR and curved MPR. • The user can identify and edit contours and the centerline automatically or manually. • The user can generate a report with relevant approach planning data and measurements for device sizing. • The software can provide visualization of calcium. • The user can generate tortuosity calculations along a centerline. The TAVR Planning Assist feature offers the ability to automatically initialize visualizations and measurements, that users can edit. The calculation of these outputs utilizes artificial intelligence/machine learning methods. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Canon
K260324
Vitrea CT TAVR Planning 510(k) Submission
# 510(k) Summary - Vitrea CT TAVR Planning
# I. Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Canon Medical Informatics, Inc. |
| --- | --- |
| Applicant Address | 5850 Opus Parkway, Suite 300 Minnetonka MN 55343 United States of America |
| Applicant and Correspondent | Jay Vaishnav, PhD, RAC, FRAPS Director, Regulatory Affairs Phone 952-487-9530 E-mail Jay.Vaishnav@mi.medical.canon |
| Alternate Contacts | Vincent Swenson Senior Director, Quality and Regulatory Phone 952-487-9548 E-mail Vincent.Swenson@mi.medical.canon Gargeyi Pavuluri Senior Regulatory Affairs Specialist Phone 952-487-9875 E-mail Gargeyi.Pavuluri@mi.medical.canon Alisha Bouley Associate Regulatory Affairs Specialist Phone 952-487-9530 E-mail Alisha.Bouley@mi.medical.canon |
# II. Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning |
| --- | --- |
| Common Name | TAVR Planning |
| Classification Name | Medical image management and processing system |
| Regulation Number | 21 CFR 892.2050 |
| Product Code(s) | Product Code: QIH |
# III. Predicate Device
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name | Product Code |
| --- | --- | --- |
| K122578 | Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning | LLZ |
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{5}
Canon
Vitrea CT TAVR Planning 510(k) Submission
# IV. Device Description Summary
21 CFR 807.92(a)(4)
## How the Device Functions
The Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning software is a non-invasive post-processing application designed to assist medical professionals with the assessment of the aortic valve and in pre-operational planning and post-operative evaluation of TAVR procedures. The software processes CT (computed tomography) image data to provide 3D segmentation of heart structures and vessels relevant to approach planning. It allows users to review 2D and 3D images to select and plan the delivery path, determine C-arm angles for use during the procedure, verify and adjust the results of segmentation and cross-section measurements, and generate reports with relevant approach planning data and measurements for device sizing.
## Scientific Concepts That Form the Basis of the Device
The TAVR Planning Assist feature¹ incorporates AI/ML methods for segmentation and landmark detection to establish the valve plane. The software uses deep learning algorithms for identifying anatomical landmarks. The segmentation module generates preliminary measurement results from the segmentations and landmarks. The device provides initial values of the aortic root measurements based on segmentation and landmarks, which the user can edit and review before finalizing a report.
## Proposed Conditions for Use
The Vitrea CT TAVR Planning software is intended for use on patients who have been identified as candidates for TAVR procedures. The software enables assessment and measurement of the aortic valve and related ventricles, providing techniques to assess the feasibility of a trans-apical, iliac (transfemoral), or subclavian approach to structures for replacement or repair procedures. The device is for use in a clinical setting and is intended for qualified users such as cardiologists, radiologists, and clinical specialists.
## Physical and Performance Characteristics of the Device
### Physical Characteristics
The device is Software as a Medical Device (SaMD) and has no hardware or material components.
### Performance Characteristics
Bench testing on 89 retrospective CT cases demonstrated that the device performs with high accuracy and strong agreement relative to expert manual measurements, as measured by intraclass correlation coefficients.² Bland-Altman analyses showed low measurement bias and variability comparable to human inter-reader performance. The device showed consistent performance across all subgroups—scanner type, slice thickness, kVp, calcification severity, demographics, and reconstruction kernels, and the leaflet visualization feature surpassed acceptance criteria with over 93% of cases rated "Good" or "Excellent" by
¹Canon Medical Informatics may market the TAVR Planning Assist feature as TAVR Assist+.
² LVOT measurement was validated at the default setting of 3 mm from the valve plane
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
Page 2 of 15
{6}
Canon
Vitrea CT TAVR Planning 510(k) Submission
expert readers. Overall, the software demonstrated robust, reliable, and clinically acceptable performance without introducing new safety or effectiveness concerns.
# How the Device Interacts with Other Devices and the Patient
The device is a software application that resides on the Vitrea platform (K172855). The device does not interact directly with patients; it performs postprocessing operations on CTA DICOM patient images. Results can be exported to image management, archival, or reporting systems that support DICOM standards for further review and interpretation. Results can also be saved in DICOM Structured Report (DICOM SR) format.
# V. Intended Use/Indications for Use
## 21 CFR 807.92(a)(5)
Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning is a non-invasive post-processing application designed to assist medical professionals with the assessment of the aortic valve and in pre-operational planning and post-operative evaluation of transcatheter aortic valve replacement procedures. Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning includes general functionality such as:
- The software processes CT (computed tomography) image data to provide 3D segmentation of heart structures and vessels relevant to approach planning.
- The user can review the 2D and 3D images to select and plan the delivery path.
- The user can determine C-arm angles for use during the procedure.
- The user can verify and adjust the results of segmentation and cross-section measurements.
- The software provides visualization techniques such as volume rendering, MIP, MPR and curved MPR.
- The user can identify and edit contours and the centerline automatically or manually. The user can generate a report with relevant approach planning data and measurements for device sizing.
- The software can provide visualization of calcium.
- The user can generate tortuosity calculations along a centerline.
The TAVR Planning Assist feature offers the ability to automatically initialize visualizations and measurements, that users can edit. The calculation of these outputs utilizes artificial intelligence/machine learning methods.
# VI. Indications for Use Comparison
## 21 CFR 807.92(a)(5)
The following text has been added to the IFU, as compared to the predicate:
"The TAVR Planning Assist feature offers the ability to automatically initialize visualizations and measurements, that users can edit. The calculation of these outputs utilizes artificial intelligence/machine learning methods."
The additional text reflects the addition of automatic initialization of device outputs, that does not change the Intended Use of the subject device from the predicate device. Users retain the ability to edit any automatically initialized results, and performance testing demonstrates that the automatically initialized measurements maintain agreement with expert ground-truth and predicate measurements.
# VII. Technological Comparison
## 21 CFR 807.92(a)(6)
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{7}
Canon
Vitrea CT TAVR Planning 510(k) Submission
The subject and predicate devices are intended for use with CTA DICOM images to support TAVR planning by measuring aortic root anatomy and vascular access dimensions. The subject device uses deep learning methods to perform some of the segmentation and landmark identification required to initialize measurements, whereas the predicate device uses traditional segmentation and pattern identification algorithms and automatically initializes fewer outputs. The reference device automates some of the measurements using AI/ML. All automatically initialized outputs from the subject device are user reviewable and editable. All algorithms and functionality associated with this feature are imported unchanged from Canon's previously cleared Endovascular Stent Planning Software (cleared as K091498, updated K143079). Regression testing was conducted to validate that the functionality integrated correctly with the rest of the device.
Table 1 summarizes the key feature/technological differences and similarities between the predicate device and the proposed device:
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{8}
Canon
Vitrea CT TAVR Planning 510(k) Submission
Table 1: Substantial Equivalence Comparison
| Characteristic | Subject Device Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: QIH | Predicate Device (K122578) Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: LLZ | Analysis of Similarities and Differences |
| --- | --- | --- | --- |
| Intended Use | Allows users to make measurements that assist qualified medical professionals in either planning TAVR procedures or in evaluating results post-operatively. | Same | Same |
| Indications for Use | Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning is a non-invasive post-processing application designed to assist medical professionals with the assessment of the aortic valve and in pre-operational planning and post-operative evaluation of transcatheter aortic valve replacement procedures. Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning includes general functionality such as • The software processes CT (computed tomography) image data to provide 3D segmentation of heart structures and vessels relevant to approach planning. • The user can review the 2D and 3D images to select and plan the delivery path. • The user can determine C-arm angles for use during the procedure. • The user can verify and adjust the results of segmentation and cross-section measurements. | Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning is a non-invasive post-processing application designed to assist medical professionals with the assessment of the aortic valve and in pre-operational planning and post-operative evaluation of transcatheter aortic valve replacement procedures. Vitrea CT Transcatheter Aortic Valve Replacement (TAVR) Planning includes general functionality such as • The software processes CT (computed tomography) image data to provide 3D segmentation of heart structures and vessels relevant to approach planning. • The user can review the 2D and 3D images to select and plan the delivery path. • The user can determine C-arm angles for use during the procedure. • The user can verify and adjust the results of segmentation and cross-section measurements. | The following text has been added to the IFU: “The TAVR Planning Assist feature offers the ability to automatically initialize visualizations and measurements, that users can edit. The calculation of these outputs utilizes artificial intelligence/machine learning methods.” The additional text reflects the addition of automatic initialization of device outputs, that does not change the Intended Use of the subject device from the predicate device. Users retain the ability to edit any automatically initialized results, and performance testing demonstrates that the |
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{9}
Canon
Vitrea CT TAVR Planning 510(k) Submission
| Characteristic | Subject Device Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: QIH | Predicate Device (K122578) Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: LLZ | Analysis of Similarities and Differences |
| --- | --- | --- | --- |
| | ·The software provides visualization techniques such as volume rendering, MIP, MPR and curved MPR. ·The user can identify and edit contours and the centerline automatically or manually. The user can generate a report with relevant approach planning data and measurements for device sizing. ·The software can provide visualization of calcium. ·The user can generate tortuosity calculations along a centerline. The TAVR Planning Assist feature offers the ability to automatically initialize visualizations and measurements, that users can edit. The calculation of these outputs utilizes artificial intelligence/machine learning methods. | ·The software provides visualization techniques such as volume rendering, MIP, MPR and curved MPR. ·The user can identify and edit contours and the centerline automatically or manually. The user can generate a report with relevant approach planning data and measurements for device sizing. ·The software can provide visualization of calcium. ·The user can generate tortuosity calculations along a centerline. | automatically initialized measurements maintain agreement with expert ground-truth and predicate measurements. |
| Target Population | Patients indicated for a TAVR procedure | Same | Same |
| Intended Users | Cardiologists, radiologists and clinical specialists | Same | Same |
| Anatomical Sites | Aortic valve and ventricles | Same | Same |
| Use Environment | Clinical environment | Same | Same |
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{10}
Canon
Vitrea CT TAVR Planning 510(k) Submission
| Characteristic | Subject Device Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: QIH | Predicate Device (K122578) Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: LLZ | Analysis of Similarities and Differences |
| --- | --- | --- | --- |
| Human Factors | The software is designed for use on a radiology workstation | Same | Same |
| Use of Contrast | The software requires contrast enhanced CTA images | Same | Same |
| DICOM Standard Compliance | The software processes DICOM-compliant CT image data | Same | Same |
| Image Display and Visualization | | | |
| Provides an option to display vascular anatomy in stretched MPR, orthogonal, multi-dimensional, curved MPR, oblique, MIP, MinIP, and 3D views. | Yes | Yes | Same |
| Allows visualization of calcium | Yes | Yes | Same |
| Image Processing, including Segmentation and Measurement | | | |
| Measurements of aortic structures | The software uses AI/ML algorithms to automatically initialize the following measurements (the user can edit the initialized value): • Aortic annulus contour measurements (area, perimeter, min/max/effective diameters) • Annulus to Left Main coronary artery (LM) height. • Annulus to Right Coronary Artery (RCA) height. • Aortic root coronal angle. | Users can manually make the following measurements: • Aortic annulus contour measurements (area, perimeter, min/max/effective diameters) • Annulus to Left Main coronary artery (LM) height • Annulus to Right Coronary Artery (RCA) height • Aortic root coronal angle • Left ventricular outflow tract (LVOT) contour measurements (area, perimeter, min/max/effective diameters) | The subject device initializes some of the measurements that a user had to make manually in the predicate. The initialization uses locked AI/ML algorithms in segmentation and landmark detection. The predicate device allowed the user to make and label any measurement, but the subject device explicitly adds the max |
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{11}
Canon
Vitrea CT TAVR Planning 510(k) Submission
| Characteristic | Subject Device Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: QIH | Predicate Device (K122578) Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: LLZ | Analysis of Similarities and Differences |
| --- | --- | --- | --- |
| | • Left ventricular outflow tract (LVOT) contour measurements (area, perimeter, min/max/effective diameters). • The sinus of Valsalva cusp to commissure distance measurements Users can manually make the following measurements: • Sinotubular junction (STJ) contour measurements (area, perimeter, min/max/effective diameters) • Sinotubular Junction (STJ) height • Sinus of Valsalva contour measurements (area, perimeter, min/max/effective diameters) • Sinus of Valsalva height • Max ascending aorta contour measurements (area, perimeter, min/max/effective diameters) Users can also make any other manual measurement they choose, and label it appropriately. | • Sinotubular junction (STJ) contour measurements (area, perimeter, min/max/effective diameters) • Sinotubular Junction (STJ) height • Sinus of Valsalva contour measurements (area, perimeter, min/max/effective diameters) • Sinus of Valsalva height • Sinus of Valsalva cusp to commissure distances Users can also make any other manual measurement they choose, and label it appropriately. | ascending aorta contour measurements to the list of available manual measurements. The user has the ability to edit all initialized outputs and must manually approve all output before reporting. |
| Vascular access measurements | Algorithms and functionality associated with vascular access features have been replaced with the version from Canon's previously cleared Endovascular Stent Planning Software (cleared as K091498, updated K143079). | The software initializes the following measurements (the user can edit the initialized value, and AI/ML is not used in the initialization): • Left & right iliac maximum tortuosity Users can manually make the following measurements: | The predicate device initialized one vascular access measurement. The subject device replaces the vascular access algorithm with the version from Canon's previously cleared Endovascular |
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{12}
Canon
Vitrea CT TAVR Planning 510(k) Submission
| Characteristic | Subject Device Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: QIH | Predicate Device (K122578) Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: LLZ | Analysis of Similarities and Differences |
| --- | --- | --- | --- |
| | The software initializes the following measurements (the user can edit the initialized value): • Left & right iliac maximum tortuosity • Left & right femoral iliac minimum diameter • Left & right external iliac minimum diameter • Left & right common iliac minimum diameter • Minimum abdominal aorta diameter The initialization does not use AI/ML. | • Left & right femoral iliac minimum diameter • Left & right external iliac minimum diameter • Left & right common iliac minimum diameter • Minimum abdominal aorta diameter | Stent Planning Software (cleared as K091498, updated K143079), that initializes all the vascular access measurements. For both devices, automatic initialization does not use AI/ML. For both devices, the user has the ability to edit, and is required to review, all initialized values. Integration testing was performed to ensure that the integration of previously cleared algorithms did not impact the safety and effectiveness of the device. |
| Segmentation and display of aortic structures | Semi-automated; adds segmentation of aortic leaflets for qualitative visualization purposes only | Semi-automated | Subject device adds visualization of leaflets. The visualization was evaluated by expert readers and does not impact the safety and effectiveness of the device as compared to the predicate. |
| Detection of annulus plane | Semi-automated | Same | Same |
| Detection of anatomical landmarks | Semi-automated detection of • right coronary cusp landmark, • left coronary cusp landmark, • non-coronary cusp landmark, | Manual identification of the same landmarks | The subject device uses AI/ML methods to automatically detect several landmarks that had to be identified manually in the |
| | non-coronary and non-lateral landmarks | | in the last 12 months. |
| Data collection and analysis | Semi-automated data collection and analysis of the same landmarks | Data collection and analysis of the same landmarks | Subject device uses the same landmarks as in the previous 12 months. |
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{13}
Canon
Vitrea CT TAVR Planning 510(k) Submission
| Characteristic | Subject Device Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: QIH | Predicate Device (K122578) Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: LLZ | Analysis of Similarities and Differences |
| --- | --- | --- | --- |
| | • right coronary ostium and left coronary ostium. | | predicate. The user can edit and must review any automatic output before reporting. Reference device K233209 also uses AI/ML methods for automatic TAVR planning functionality. |
| Detection of relevant vessel centerlines and calculation of diameter, area, and tortuosity | Semi-automated | Same | Same |
| Contour detection | Semi-automated contour detection with diameter, area, and distance measurements of annulus plane, as well as distance to each coronary ostia from the annulus plane | Same | Same |
| Use of AI/ML methods | Uses AI/ML for segmentation and landmark detection. | Does not use AI/ML | The predicate device does not use any AI/ML. The TAVR planning feature in Reference device K233209 uses AI/ML. The use of AI/ML in automatic initialization of device outputs does not change the Intended Use of the subject device from that of the predicate device. |
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{14}
Canon
Vitrea CT TAVR Planning 510(k) Submission
| Characteristic | Subject Device Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: QIH | Predicate Device (K122578) Vitrea CT TAVR Planning 21 CFR 892.2050 Product Code: LLZ | Analysis of Similarities and Differences |
| --- | --- | --- | --- |
| | | | Users retain the ability to edit any automatically initialized results, and performance testing demonstrates that the automatically initialized measurements maintain agreement with expert ground-truth and predicate measurements. |
| Reporting | | | |
| The software provides tools to generate reports to distribute findings. | Yes | Yes | Same |
| DICOM SR | Yes | No | Users can export reports in a Structured Report format, if they have a license. |
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{15}
Canon
Vitrea CT TAVR Planning 510(k) Submission
In summary, the main difference between the subject and predicate device is that the subject device adds AI/ML based automatic initialization of aortic root measurements, whereas the predicate device only allowed the user to make these measurements manually. These differences do not raise new questions of safety or effectiveness. The subject device's algorithms have been validated through performance testing, meeting pre-specified acceptance criteria and demonstrating accuracy and reliability.
The new functionality is similar to that in reference device K233209 (uOmnispace.CT, United Imaging), which also makes AI/ML enabled semi-automated measurements of the aortic root.
## VIII. Non-Clinical and/or Clinical Tests, Summary and Conclusions
The device was trained and tuned using diverse retrospective CT data representing a wide range of patient anatomies, acquisition protocols, scanner manufacturers, and disease presentations relevant to TAVR planning. The training set consisted of 218 cases.
The validation dataset for the aortic root measurements included 89 independent TAVR CT planning cases spanning multiple U.S. and international institutions, with diversity in demographics (age, sex), imaging characteristics (scanner vendor, scanner model, slice thickness, kVp), and calcification severity. Scanner representation included Canon, GE, Philips, and Siemens systems).
## Characteristics of the dataset follow:
### Demographic distribution (Gender, Age, Ethnicity)
#### Gender
- Female: 30
- Male: 42
- Unknown: 17
#### Age
- 50–59 years: 2
- 60–69 years: 8
- 70–79 years: 15
- 80–89 years: 20
- 90–99 years: 5
- Unknown: 39
#### Ethnicity
Structured race and ethnicity data were not available in the anonymized dataset. Geographic and site diversity were used as a proxy for racial and ethnic diversity, based on U.S. Census data for the contributing regions. The sites used included a mix of academic, community, and federal hospitals spanning four major US geographic regions, as well as data from outside the US.
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{16}
Canon
Vitrea CT TAVR Planning 510(k) Submission
# Clinical Subgroups and Confounders
The validation dataset was intentionally designed to include these clinical subgroups and confounders, with subgroup performance analyses performed:
- Sex: Male, Female, Unknown
Age: 50-79 years, $\geq 80$ years, Unknown
- Valve Calcification Severity: None (n=18), Mild/Moderate (n=35), Severe (n=36)
The validation dataset also includes challenging and edge cases related to anatomy and imaging conditions:
- LVOT calcification
Horizontal aorta (high coronal angle)
- Reduced IV contrast
Low kVp imaging (<120 kVp)
# Exclusions
- Cases with severe motion artifacts
Poor contrast or non-diagnostic phases
# Equipment and Protocols for Image Acquisition
# CT Equipment
- Scanners from Canon, Siemens, GE, and Philips were used.
# Imaging Protocol Parameters
- Slice thickness: $\leq 1.0$ mm (subgroups $\leq 0.5$ mm, $>0.5 - 1.0$ mm).
Tube energy: 100-120 kVp.
Cardiac phases: systolic (20–30%) and diastolic (~80%).
# Truthing Process
- Ground truth measurements were determined by two independent, board-certified cardiac CT readers.
- The FDA-cleared predicate device was used to manually generate these measurements.
Each reader was unaware of the other's findings and did not see the automated results.
- For every measurement, the ground truth was established by averaging the values from both readers.
- Agreement between the readers was documented to ensure consistency with the TAVR literature previously published.
Validation truthers were completely independent of training truthers. The validation dataset was retrospective and distinct from any training activities. The device performance was measured by comparing device outputs with ground truth created manually by expert physicians. The overall device performance is summarized in Table 2.
Table 2: Measurement Accuracy (Retrospective validation; n=89)
| Measurement | Intra-Class Correlation (ICC) | Acceptance criterion | 95% CI | Bland-Altman Bias (%) |
| --- | --- | --- | --- | --- |
| Systolic | 0.80 | 0.75 | 0.72–0.80 | 0.80 |
| Diastolic | 0.80 | 0.75 | 0.72–0.80 | 0.80 |
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{17}
Canon
Vitrea CT TAVR Planning 510(k) Submission
| Annulus area (mm²) | 0.95 | >0.90 | [0.92, 0.96] | -4.1% |
| --- | --- | --- | --- | --- |
| Annulus perimeter (mm) | 0.94 | >0.90 | [0.91, 0.96] | -2.5% |
| Derived diameter (mm) | 0.95 | >0.90 | [0.92, 0.97] | -1.9% |
| Min diameter (mm) | 0.91 | >0.85 | [0.87, 0.94] | -1.9% |
| Max diameter (mm) | 0.93 | >0.85 | [0.90, 0.95] | -2.8% |
| LM coronary height (mm) | 0.91 | >0.80 | [0.87, 0.94] | 3.0% |
| RCA coronary height (mm) | 0.94 | >0.80 | [0.91, 0.96] | -2.4% |
| Aortic root angle (°) | 0.97 | >0.90 | [0.95, 0.98] | 1.0% |
| LVOT area (mm²) | 0.92 | >0.90 | [0.89, 0.95] | -5.4% |
| LVOT perimeter (mm) | 0.91 | >0.90 | [0.86, 0.94] | -3.0% |
| LVOT derived diameter (mm) | 0.93 | >0.90 | [0.90, 0.95] | -2.4% |
| LVOT min diameter (mm) | 0.92 | >0.90 | [0.87, 0.94] | -2.4% |
| LVOT max diameter (mm) | 0.86 | >0.90 | [0.80, 0.91] | -3.6% |
| Sinus of Valsalva R (mm) | 0.96 | >0.90 | [0.93, 0.97] | -1.0% |
| Sinus of Valsalva L (mm) | 0.96 | >0.90 | [0.95, 0.98] | -1.0% |
| Sinus of Valsalva NC (mm) | 0.97 | >0.90 | [0.95, 0.98] | 0.1% |
All primary measurements met their pre-specified acceptance criteria except for the LVOT maximum diameter, which demonstrated a slightly lower ICC; however, the associated absolute differences and Bland-Altman limits of agreement were small, within clinically acceptable margins, and not expected to affect annular sizing or overall clinical use.
Comprehensive subgroup analyses across CT scanner vendors, slice thickness, tube energy, reconstruction kernels, calcification severity, patient age, sex, and geographic distribution showed consistent performance. A lower ICC observed in a female-only subgroup was determined to be due to a single case with severe motion artifact and a very narrow measurement range, conditions under which ICC is not a reliable indicator of accuracy; absolute error and Bland-Altman results for the subgroup were consistent with the overall dataset. These isolated deviations were attributed to limitations on the quality of the input image rather than systematic algorithm behavior and do not impact device safety or effectiveness.
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
{18}
Canon
Vitrea CT TAVR Planning 510(k) Submission
The optional aortic leaflet-coloring feature was evaluated separately using a multi-reader Likert-scale quality assessment, with over 93% of cases rated Good or Excellent, exceeding acceptance criteria. Overall, the bench testing demonstrated that the device performs as intended, with measurement accuracy comparable to expert clinical assessment, supporting substantial equivalence to the predicate device.
## Conclusion on Non-Clinical Testing
The non-clinical testing of the CT TAVR Planning Assist demonstrated that the device delivers accurate and reliable measurements for key anatomical parameters relevant to TAVR planning. Performance validation using retrospective CT datasets confirmed strong agreement between the device outputs and clinician-generated ground truth measurements, with high intraclass correlation coefficients (ICC) across all measured parameters. The results indicate minimal bias and narrow confidence intervals, supporting the device's precision and consistency. Collectively, these findings confirm that the device's AI-assisted workflow is robust and suitable for its intended use in pre-procedural planning, meeting the established performance criteria without introducing new questions of safety or effectiveness.
## IX. Substantial Equivalence Analysis Conclusion
The results of the non-clinical and performance validation testing demonstrate that the subject device is as safe and effective as the predicate device for its intended use. The TAVR planning measurements demonstrate robust agreement to clinician-generated ground truth, with no evidence of clinically significant bias or performance variability across subgroups. The aortic leaflet segmentation delivered consistently high visualization quality. These findings support the substantial equivalence of the device to the predicate devices, wot no new questions of safety or effectiveness raised by its introduction. Therefore, the CT TAVR Planning Assist device meets the requirements for market clearance under the 510(k) pathway.
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | mi.medical.canon
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.