K233209 · Shanghai United Imaging Healthcare Co., Ltd. · QIH · May 17, 2024 · Radiology
Device Facts
Record ID
K233209
Device Name
uOmnispace.CT
Applicant
Shanghai United Imaging Healthcare Co., Ltd.
Product Code
QIH · Radiology
Decision Date
May 17, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Spine labeling
Deep learning-based algorithm
Average score > 4 points
Average score 5 points
—
—
120 subjects
>1 (licensed physicians)
Rib labeling
Deep learning-based algorithm
Average score > 4 points
Average score 5 points
—
—
120 subjects
>1 (licensed physicians)
TAVR analysis
AI-based algorithm
Mean landmark error < 1 mm; Average score > 2
Mean landmark error 0.86 mm; Average score 3
—
—
60 subjects
2 (licensed physicians)
Indications for Use
uOmnispace.CT is a software for viewing, manipulating, evaluating and analyzing medical images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications: -The uOmnispace.CT Colon Analysis application is intended to provide the user a tool to enable easy visualization and efficient evaluation of CT volume data sets of the colon. -The uOmnispace.CT Dental application is intended to provide the user a tool to reconstruct panoramic and paraxial views of jaw. -The uOmnispace.CT Lung Density Analysis application is intended to segment pulmonary, lobes, and airway, providing the user quantitative parameters, structure information to evaluate the lung and airway. -The uOmnispace.CT Lung Nodule application is intended to provide the user a tool for the review and analysis of thoracic CT images, providing quantitative and characterizing information about nodules in the lung in a single study, or over the time course of several thoracic studies. -The uOmnispace.CT Vessel Analysis application is intended to provide a tool for viewing, and evaluating CT vascular images. -The uOmnispace.CT Brain Perfusion is intended to calculate the parameters such as: CBV, CBF, etc. in order to analyze functional blood flow information about a region of interest (ROI) in brain. -The uOmnispace.CT Heart application is intended to segment heart and extract coronary artery. It also provides analysis of vascular stenosis, plaque and heart function. -The uOmnispace.CT Calcium Scoring application is intended to identify calcifications and calculate the calcium soore. -The uOmnispace.CT Dynamic Analysis application is intended to support visualization of the CT datasets over time with the 3D/4D display modes. -The uOmnispace.CT Bone Structure Analysis application is intended to provide visualization and labels for the ribs and spine, and support batch function for intervertebral disk. -The uOmnispace. CT Liver Evaluation application is intended to processing and visualization for liver segmentation and vessel extraction. It also provides a tool for the user to perform liver separation and residual liver segments evaluation. -The uOmnispace. CT Dual Energy is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The u0mnispace.CT Dual Energy application is intended to provide information on the chemical composition of the scanned body materials and/or contrast agents. Additionally, it enables images to be generated at multiple energies within the available spectrum. -The uOmnispace.CT Cardiovascular Combined Analysis is an image analysis software package for evaluating contrast enhanced CT images. The CT Cardiovascular Combined Analysis is intended to analyze vascular and cardiac structures. It can be used in the qualitative and quantitative for the analysis of head-neck, abdomen, multi-body part combined, TAVR (Transcatheter Aortic Valve Replacement) CT data as input for the planning of cardiovascular procedures.
Device Story
uOmnispace.CT is a standalone post-processing software platform for CT image analysis; operates in healthcare institutions; used by clinicians for diagnostic support. Inputs: CT volume datasets. Processing: includes automated segmentation, labeling, and quantification algorithms (e.g., spine/rib labeling, TAVR analysis, perfusion parameter calculation). Outputs: reconstructed views, quantitative parameters, segmented structures, and analysis reports. Clinical impact: assists in visualization, characterization of nodules, vascular/cardiac assessment, and procedural planning (e.g., TAVR). Benefits: provides efficient, standardized quantitative analysis of complex CT data to aid clinical decision-making.
Clinical Evidence
No clinical studies were required. Performance was established via bench testing and software verification/validation. ML/AI algorithms (spine labeling, rib labeling, TAVR analysis) were validated on datasets (n=60-120) using ground truth comparisons and subjective scoring by licensed physicians. Spine/rib labeling achieved average scores >4/5. TAVR analysis showed mean landmark error of 0.86 mm (<1 mm threshold) and average physician scores >2/3.
Technological Characteristics
Standalone software; DICOM-compliant; utilizes deep learning-based algorithms for spine/rib labeling and TAVR analysis. Complies with NEMA PS 3.1-3.20, ISO 14971, and IEC 62304 standards. No hardware components; no biocompatibility or electrical safety requirements.
Indications for Use
Indicated for healthcare professionals to view, manipulate, evaluate, and analyze CT medical images for diagnostic support. Applications include colon, dental, lung density/nodule, vessel, brain perfusion, heart, calcium scoring, dynamic analysis, bone structure, liver, dual energy, and cardiovascular combined analysis (including TAVR planning).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shanghai United Imaging Healthcare Co., Ltd. Xin Gao Official Correspondent No.2258 Chengbei Rd. Jiading District Shanghai, Shanghai 201807 China
## Re: K233209
Trade/Device Name: uOmnispace.CT Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: April 12, 2024 Received: April 12, 2024
May 17, 2024
## Dear Xin Gao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, PhD Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233209
Device Name uOmnispace.CT
### Indications for Use (Describe)
uOmnispace. CT is a software for viewing, manipulating, evaluating and analyzing medical images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additions: -The uOmnispace. CT Colon Analysis application is intended to provide the user a tool to enable easy visualization and efficient evaluation of CT volume data sets of the colon. -The uOmnispace. CT Dental application is intended to provide the user a tool to reconstruct panoramic and paraxial views of jaw. -The uOmnispace. CT Lung Density Analysis application is intended to segment pulmonary, lobes, and airway, providing the user quantitative parameters, structure information to evaluate the lung and airway. -The uOmnispace.CT Lung Nodule application is intended to provide the user a tool for the review and analysis of thoracic CT images, providing quantitative and characterizing information about nodules in the lung in a single study, or over the time course of several thoracic studies. -The uOmnispace.CT Vessel Analysis application is intended to provide a tool for viewing, and evaluating CT vascular images. -The uOmnispace. CT Brain Perfusion is intended to calculate the parameters such as: CBV, CBF, etc. in order to analyze functional blood flow information about a region of interest (ROI) in brain. -The uOmnispace.CT Heart application is intended to segment heart and extract coronary artery. It also provides analysis of vascular stenosis, plaque and heart function. -The uOmnispace. CT Calcium Scoring application is intended to identify calcifications and calculate the calcium soore. -The uOmnispace. CT Dynamic Analysis application is intended to support visualization of the CT datasets over time with the 3D/4D display modes. -The uOmnispace.CT Bone Structure Analysis application is intended to provide visualization and labels for the ribs and spine, and support batch function for intervertebral disk.
-The uOmnispace. CT Liver Evaluation application is intended to processing and visualization for liver segmentation and vessel extraction. It also provides a tool for the user to perform liver separation and residual liver segments evaluation.
-The uOmnispace. CT Dual Energy is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The u0mnispace.CT Dual Energy application is intended to provide information on the chemical composition of the scanned body materials and/or contrast agents. Additionally, it enables images to be generated at multiple energies within the available spectrum.
-The uOmnispace.CT Cardiovascular Combined Analysis is an image analysis software package for evaluating contrast enhanced CT images. The CT Cardiovascular Combined Analysis is intended to analyze vascular and cardiac structures. It can be used in the qualitative and quantitative for the analysis of head-neck, abdomen, multi-body part combined, TAVR (Transcatheter Aortic Valve Replacement) CT data as input for the planning of cardiovascular procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is dark gray. The logo is simple and modern.
### 510 (k) SUMMARY K233209
- 1. Date of Preparation: May 16, 2024
# 2. Sponsor Identification
Shanghai United Imaging Healthcare Co.,Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China
Establishment Registration Number: 3011015597
Contact Person: Xin GAO Position: Regulatory Affairs Specialist Tel: +86-021-67076888-5386 Fax: +86-021-67076889 Email: xin.gao@united-imaging.com
# 3. Identification of Proposed Device
Trade Name: uOmnispace.CT Common Name: Medical image management and processing system Model: uOmnispace.CT
## Regulatory Information Classification Name: Medical image management and processing system Classification: II
Product Code: QIH Regulation Number: 21 CFR 892.2050 Review Panel: Radiology
# 4. Identification of Predicate Device(s)
Predicate Device 510(k) Number: K183170 Common Name: Medical image management and processing system Model: uWS-CT
## Regulatory Information
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is dark gray. The logo is simple and modern in design.
Classification Name: Medical image management and processing system Classification: II Product Code: LLZ Regulation Number: 21 CFR 892.2050 Review Panel: Radiology
Reference Device#1 510(k) Number: K230162 Device Name: uCT 760 with uWS-CT-Dual Energy Analysis, uCT 780 with uWS-CT-Dual Energy Analysis
Reference Device#2 510(k) Number: K230039 Device Name: uOmnispace
Reference Device#3 510(k) Number: K170221 Device Name: syngo.CT Cardiac Planning
Reference Device#4 510(k) Number: K133643 Device Name: syngo.CT Liver Analysis
Reference Device#5 510(k) Number: K182130 Device Name: iSchema View's RAPID
# 5. Device Description
The uOmnispace.CT is a post-processing software based on the uOmnispace platform for viewing, manipulating, evaluating and analyzing medical images, can run alone or with other advanced commercially cleared applications.
uOmnispace.CT contains the following applications:
- uOmnispace.CT Calcium Scoring ●
- uOmnispace.CT Lung Nodule ●
- uOmnispace.CT Colon Analysis
- . uOmnispace.CT Lung Density Analysis
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is dark gray. The logo is simple and modern in design.
- uOmnispace.CT Dental Application ●
- . uOmnispace.CT Bone Structure Analysis
- uOmnispace.CT Dual Energy
- . uOmnispace.CT Vessel Analysis
- uOmnispace.CT Heart
- . uOmnispace.CT Brain Perfusion
- uOmnispace.CT Dynamic Analysis
- . uOmnispace.CT Liver Evaluation
- . uOmnispace.CT Cardiovascular Combined Analysis
# 6. Indications for use
uOmnispace.CT is a software for viewing, manipulating, evaluating and analyzing medical images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications:
- . The uOmnispace.CT Colon Analysis application is intended to provide the user a tool to enable easy visualization and efficient evaluation of CT volume data sets of the colon.
- . The uOmnispace.CT Dental application is intended to provide the user a tool to reconstruct panoramic and paraxial views of jaw.
- . The uOmnispace.CT Lung Density Analysis application is intended to segment pulmonary, lobes, and airway, providing the user quantitative parameters, structure information to evaluate the lung and airway.
- . The uOmnispace.CT Lung Nodule application is intended to provide the user a tool for the review and analysis of thoracic CT images, providing quantitative and characterizing information about nodules in the lung in a single study, or over the time course of several thoracic studies.
- . The uOmnispace.CT Vessel Analysis application is intended to provide a tool for viewing, manipulating, and evaluating CT vascular images.
- . The uOmnispace.CT Brain Perfusion application is intended to calculate the parameters such as: CBV, CBF, etc. in order to analyze functional blood flow information about a region of interest (ROI) in brain.
- . The uOmnispace.CT Heart application is intended to segment heart and extract coronary artery. It also provides analysis of vascular stenosis, plaque and heart function.
- . The uOmnispace.CT Calcium Scoring application is intended to identify calcifications and calculate the calcium score.
- . The uOmnispace.CT Dynamic Analysis application is intended to support visualization of the CT datasets over time with the 3D/4D display modes.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for United Imaging. The logo consists of the word "UNITED" stacked on top of the word "IMAGING", both in a bold, sans-serif font. To the right of the text is a stylized "U" shape, also in a bold, sans-serif font, with a vertical line running through the center.
- The uOmnispace.CT Bone Structure Analysis application is intended to provide visualization and labels for the ribs and spine, and support batch function for intervertebral disk.
- . The uOmnispace.CT Liver Evaluation application is intended to provide processing and visualization for liver segmentation and vessel extraction. It also provides a tool for the user to perform liver separation and residual liver segments evaluation.
- . The uOmnispace.CT Dual Energy is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The uOmnispace.CT Dual Energy application is intended to provide information on the chemical composition of the scanned body materials and/or contrast agents. Additionally, it enables images to be generated at multiple energies within the available spectrum.
- . The uOmnispace.CT Cardiovascular Combined Analysis is an image analysis software package for evaluating contrast enhanced CT images. The CT Cardiovascular Combined Analysis is intended to analyze vascular and cardiac structures. It can be used in the qualitative and quantitative for the analysis of headneck, abdomen, multi-body part combined, TAVR (Transcatheter Aortic Valve Replacement) CT data as input for the planning of cardiovascular procedures.
### Summary of Technological Characteristics 7.
The technology characteristics of uOmnispace.CT, reflected in this 510(k) submission are substantially equivalent to those of the predicate devices.
The following tables compare the main features, principles of operation, fundamental scientific technology and intended use of uOmnispace.CT when compared to the predicate devices.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized icon that resembles the letter "U" with a vertical line running through the center, creating a negative space "I" shape. The icon is a solid color, providing a visual contrast to the white background.
| Item | Proposed Device<br>uOmnispace.CT | Predicate Device<br>uWS-CT(K183170) | Remark |
|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General | | | |
| Intended Use | uOmnispace.CT is a software for viewing,<br>manipulating, evaluating and analyzing medical<br>images. It supports interpretation and evaluation of<br>examinations within healthcare institutions. It has<br>the following additional indications:<br>The uOmnispace.CT Colon Analysis<br>application is intended to provide the user a tool to<br>enable easy visualization and efficient evaluation of<br>CT volume data sets of the colon. The uOmnispace.CT Dental application is<br>intended to provide the user a tool to reconstruct<br>panoramic and paraxial views of jaw. The uOmnispace.CT Lung Density Analysis<br>application is intended to segment pulmonary, lobes,<br>and airway, providing the user quantitative<br>parameters, structure information to evaluate the<br>lung and airway. The uOmnispace.CT Lung Nodule<br>application is intended to provide the user a tool for<br>the review and analysis of thoracic CT images,<br>providing quantitative and characterizing<br>information about nodules in the lung in a single<br>study, or over the time course of several thoracic<br>studies. The uOmnispace.CT Vessel Analysis<br>application is intended to provide a tool for viewing,<br>manipulating, and evaluating CT vascular images. The uOmnispace.CT Brain Perfusion<br>application is intended to calculate the parameters | uWS-CT is a software solution intended to be used<br>for viewing, manipulation, communication, and<br>storage of medical images. It supports interpretation<br>and evaluation of examinations within healthcare<br>institutions. It has the following additional<br>indications:<br>The CT Colon Analysis application is<br>intended to provide the user a tool to enable easy<br>visualization and efficient evaluation of CT volume<br>data sets of the colon. The CT Dental application is intended to<br>provide the user a tool to reconstruct panoramic and<br>paraxial views of jaw. The CT Lung Density application is<br>intended to provide the user a number of density<br>parameters and structure information for evaluating<br>tomogram scans of the lung. The CT Lung Nodule application is<br>intended to provide the user a tool for the review and<br>analysis of thoracic CT images, providing<br>quantitative and characterizing information about<br>nodules in the lung in a single study, or over the<br>time course of several thoracic studies. The CT Vessel Analysis application is<br>intended to provide a tool for viewing, manipulating,<br>and evaluating CT vascular images. The CT Brain Perfusion application is<br>intended to calculate the parameters such as: CBV,<br>CBF, etc. in order to analyze functional blood flow | Substantial Equivalence.<br>The proposed device<br>includes more<br>applications, which are<br>discussed in the following<br>sections, than the predicate<br>device. This difference<br>will not impact the safety<br>and effectiveness of the<br>device. |
| Item | Proposed Device<br>uOmnispace.CT | Predicate Device<br>uWS-CT(K183170) | Remark |
| | such as: CBV, CBF, etc. in order to analyze<br>functional blood flow information about a region of<br>interest (ROI) in brain. | information about a region of interest (ROI) in the<br>brain. | |
| | • The uOmnispace.CT Heart application is<br>intended to segment heart and extract coronary<br>artery. It also provides analysis of vascular stenosis,<br>plaque and heart function. | • The CT Heart application is intended to<br>segment heart and extract coronary artery. It also<br>provides analysis of vascular stenosis, plaque and<br>heart function. | |
| | • The uOmnispace.CT Calcium Scoring<br>application is intended to identify calcifications and<br>calculate the calcium score. | • The CT Calcium Scoring application is<br>intended to identify calcifications and calculate the<br>calcium score. | |
| | • The uOmnispace.CT Dynamic Analysis<br>application is intended to support visualization of the<br>CT datasets over time with the 3D/4D display<br>modes. | • The CT Dynamic Analysis application is<br>intended to support visualization of the CT datasets<br>over time with the 3D/4D display modes. | |
| | • The uOmnispace.CT Bone Structure<br>Analysis application is intended to provide<br>visualization and labels for the ribs and spine, and<br>support batch function for intervertebral disk. | • The CT Bone Structure Analysis application<br>is intended to provide visualization and labels for the<br>ribs and spine, and support batch function for<br>intervertebral disk. | |
| | • The uOmnispace.CT Liver Evaluation<br>application is intended to provide processing and<br>visualization for liver segmentation and vessel<br>extraction. It also provides a tool for the user to<br>perform liver separation and residual liver segments<br>evaluation. | • The CT Liver Evaluation application is<br>intended to provide processing and visualization for<br>liver segmentation and vessel extraction. It also<br>provides a tool for the user to perform liver<br>separation and residual liver segments evaluation. | |
| | • The uOmnispace.CT Dual Energy Analysis<br>is a post-processing software package that accepts<br>UIH CT images acquired using different tube<br>voltages and/or tube currents of the same anatomical<br>location. The uOmnispace.CT Dual Energy<br>application is intended to provide information on the<br>chemical composition of the scanned body materials<br>and/or contrast agents. Additionally it enables | | |
| Item | Proposed Device<br>uOmnispace.CT | Predicate Device<br>uWS-CT(K183170) | Remark |
| | images to be generated at multiple energies within<br>the available spectrum.<br>• The uOmnispace.CT Cardiovascular<br>Combined Analysis is an image analysis software<br>package for evaluating contrast enhanced CT<br>images. The CT Cardiovascular Combined Analysis<br>is intended to analyze vascular and cardiac<br>structures.It can be used in the qualitative and<br>quantitative for the analysis of head-neck, abdomen,<br>multi-body part combined, TAVR (Transcatheter<br>Aortic Valve Replacement) CT data as input for the<br>planning of cardiovascular procedures. | | |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for United Imaging. The text "UNITED IMAGING" is in bold, sans-serif font. To the right of the text is a stylized "U" shape, which is dark blue with a white horizontal line through the middle. The logo is simple and modern.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the word "UNITED" stacked on top of the word "IMAGING" in a bold, sans-serif font. To the right of the text is a stylized, dark blue icon that resembles the letter "U" with a horizontal line through the middle, creating a shape that also resembles the letter "I".
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized "U" shape that is divided into three sections by two vertical white lines. The logo is simple and modern, and the use of bold font and geometric shapes gives it a strong visual impact.
| Application | Function name | Proposed device<br>uOmnispace.CT | Predicate device:<br>uWS-CT(K183170) | Remark |
|----------------------------------------|------------------------------------|----------------------------------|--------------------------------------|--------|
| uOmnispace.CT Lung<br>Density Analysis | Lung Segmentation | Yes | Yes | Same |
| | Lung Density Analysis | Yes | Yes | Same |
| | Lung Contour Editing | Yes | Yes | Same |
| | Pulmonary lobes Segmentation | Yes | Yes | Same |
| | Airway Segmentation | Yes | Yes | Same |
| | Airway Tree Extraction and Editing | Yes | Yes | Same |
| | Airway Contour Editing | Yes | Yes | Same |
| | Statistical Analysis | Yes | Yes | Same |
| | Save, Report, Print | Yes | Yes | Same |
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is dark blue with a white vertical line running through the center. The logo is simple and modern in design.
| Application | Function Name | Proposed Device<br>uOmnisapce.CT | Predicate<br>Device:<br>uWS-CT<br>(K183170) | Reference Device#5<br>iSchema View's RAPID<br>(K182130) | Remark |
|----------------------------------|-----------------------------------|----------------------------------|---------------------------------------------|---------------------------------------------------------|------------------------------------------------|
| uOmnispace.CT<br>Brain Perfusion | Motion Correction | Yes | Yes | / | Same |
| | Arterial Detection | Yes | Yes | / | Same |
| | Parameter Map<br>Calculation | Yes | Yes | / | Same |
| | Time-density curve<br>analysis | Yes | Yes | / | Same |
| | Tmax | Yes | / | Yes | Functional Substantial<br>Equivalent<br>Note 1 |
| | Ischemic penumbra<br>analysis | Yes | / | Yes | Functional Substantial<br>Equivalent<br>Note 2 |
| | Symmetric ROI and<br>ROI Template | Yes | Yes | / | Same |
| | Save, Report, Print | Yes | Yes | / | Same |
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized letter "U" that is also in a bold font. The color of the logo is a dark gray.
| Application | Function name | Proposed device<br>uOmnispace.CT | Predicate device:<br>uWS-CT<br>(K183170) | Reference<br>device#2<br>UOmnispace<br>(K230039) | Remark |
|-------------------------|-----------------------------|----------------------------------|------------------------------------------|--------------------------------------------------|------------------------------------------------|
| uOmnispace.<br>CT Heart | Multi-Phase Loading | Yes | Yes | / | Same |
| | Hyper Realistic Rendering | Yes | / | Yes | Same |
| | Heart Chamber Segmentation | Yes | Yes | / | Same |
| | Coronary Artery Extraction | Yes | Yes | / | Same |
| | Editing Tools | Yes | Yes | / | Same |
| | Centerline Extraction | Yes | Yes | / | Same |
| | Stenosis Analysis | Yes | Yes | / | Same |
| | Plaque Analysis | Yes | Yes | / | Same |
| | Cardiac function Assessment | Yes | Yes | / | Functional Substantial<br>Equivalent<br>Note 3 |
| | Save, Report, Print | Yes | Yes | / | Same |
| Application | Function name | Proposed device uOmnispace.CT | Predicate device: uWS-CT(K183170) | Remark |
|----------------------------------|----------------------------|-------------------------------|-----------------------------------|--------------------------------------------------|
| uOmnispace.CT<br>Calcium Scoring | Calcium sites segmentation | Yes | Yes | Same |
| | Calculate Calcium score | Yes | Yes | Functional Substantially<br>Equivalent<br>Note 4 |
| | Save, Report, Print | Yes | Yes | Same |
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is displayed in bold, sans-serif font. To the right of the text is a stylized icon that resembles a shield or a letter "U" with a vertical line running through the center. The logo is simple and modern in design.
| Application | Function name | Proposed device<br>uOmnispace.CT | Predicate device:<br>uWS-CT(K183170) | Remark |
|------------------------------------------|-----------------------------------|----------------------------------|---------------------------------------|---------------------------------------------|
| uOmnispace.CT Bone<br>Structure Analysis | Labeling Ribs | Yes | Yes | Functional Substantial Equivalent<br>Note 5 |
| uOmnispace.CT Bone<br>Structure Analysis | Labeling Spine | Yes | Yes | Functional Substantial Equivalent<br>Note 6 |
| uOmnispace.CT Bone<br>Structure Analysis | Batch | Yes | Yes | Same |
| uOmnispace.CT Bone<br>Structure Analysis | Save, Report, Print | Yes | Yes | Same |
| Application | Function name | Proposed device<br>uOmnispace.CT | Predicate device:<br>uWS-CT (K183170) | Remark |
| uOmnispace.CT<br>Dynamic Analysis | Motion correction | Yes | Yes | Same |
| uOmnispace.CT<br>Dynamic Analysis | Multiple phase viewing | Yes | Yes | Same |
| uOmnispace.CT<br>Dynamic Analysis | Bone Removal | Yes | Yes | Same |
| uOmnispace.CT<br>Dynamic Analysis | Data Loading and 3D/4D<br>Display | Yes | Yes | Same |
| uOmnispace.CT<br>Dynamic Analysis | Artery and Vein Display | Yes | Yes | Same |
| uOmnispace.CT<br>Dynamic Analysis | Save, Report, Print | Yes | Yes | Same |
{15}------------------------------------------------
Image /page/15/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold, sans-serif font. The color of the logo is a dark teal.
| Application | Function name | Proposed device<br>uOmnispace.CT | Predicate device<br>uWS-CT<br>(K183170) | Reference device#3<br>uOmnispace<br>(K230039) | Reference<br>device#4<br>syngo.CT<br>Liver Analysis<br>(K133643) | Remark |
|-----------------------------------|----------------------------------------------------------|----------------------------------|-----------------------------------------|-----------------------------------------------|------------------------------------------------------------------|------------------------------------------------|
| uOmnispace.CT<br>Liver Evaluation | Phase Selection | Yes | Yes | / | / | Same |
| | Liver Segmentation | Yes | Yes | / | / | Same |
| | Lesion Segmentation | Yes | Yes | / | / | Same |
| | Rib Segmentation<br>and<br>manual correction | Yes | / | Yes | / | Same |
| | Vessel Extraction | Yes | Yes | / | / | Same |
| | Vascular Editing | Yes | Yes | / | / | Same |
| | Liver Segments | Yes | Yes | / | / | Same |
| | Virtual Planning | Yes | Yes | / | / | Same |
| | RFA | Yes | Yes | / | / | Same |
| | Vascular Territories<br>Computation and<br>visualization | Yes | / | / | Yes | Functional Substantial<br>Equivalent<br>Note 7 |
| | Measurement | Yes | Yes | / | / | Same |
| | Save, Report, Print | Yes | Yes | / | / | Same |
{16}------------------------------------------------
Image /page/16/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED IMAGING" are in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized logo that resembles a shield or a letter 'U' with a vertical line running through the center, creating a negative space 'I' shape.
| Application | Function name | Proposed device<br>uOmnispace.CT | Reference Device#1:<br>uCT 760 with uWS-CT-Dual<br>Energy Analysis, uCT 780<br>with uWS-CT-Dual Energy<br>Analysis (K230162) | Remark |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| uOmnispace.CT<br>Dual<br>Energy | Mono Energetic Image | Yes | Yes | Same |
| | Mixed Enhanced Image | Yes | Yes | Same |
| | CNR(Contrast Noise Ratio) Image | Yes | Yes | Same |
| | Base Material Images: Including<br>Water-Iodine, Water-Calcium,<br>Calcium-Iodine, Uric acid-Calcium,<br>Water-HAP, Liver-Fat Base<br>Material Pair image. | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.