60 chest CTs (37 male, 23 female; age range 14-86)
>1 (clinical specialists)
Indications for Use
uOmnispace is a software solution intended to be used for viewing, manipulation, communication and storage of medical images. It allows processing and filming of multimodality DICOM images. It can be used as a stand-alone device or together with a variety of cleared and unmodified software options, and also support to plug in multi-vendor applications which meet interface requirements. uOmnispace is intended to be used by trained professionals, including but not limited to physicians and medical technicians. The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.
Device Story
uOmnispace is a software-only medical device for viewing, processing, and managing multimodality DICOM images (CT, MRI, PET, DR). It operates on off-the-shelf PCs in clinical environments, used by physicians and technicians. The device provides 2D/3D visualization, annotation, measurement, and automated segmentation tools. It supports multi-user access via a client-server architecture. The system includes a machine learning-based rib segmentation algorithm. Healthcare providers use the software to read images, create reports, and film studies. By automating segmentation and providing advanced visualization (e.g., volume rendering, virtual endoscopy), the device assists in radiological interpretation, potentially improving workflow efficiency and diagnostic accuracy.
Clinical Evidence
No clinical studies were required. Performance validation was conducted via bench testing of the ML-based rib segmentation algorithm on 60 chest CT datasets (37 male, 23 female; age range 14-86). The primary endpoint was the Dice similarity coefficient. The algorithm achieved an average Dice score of 0.855, exceeding the acceptance criterion of 0.8. Subgroup analysis confirmed consistent performance across age and gender.
Technological Characteristics
Software-only medical device; client-server architecture. Client OS: Windows 7/10; Server OS: Linux CentOS 7.7. Supports DICOM 3.0. Features: 2D/3D viewing, MPR, MIP, MinIP, CPR, SSD, volume rendering (HRR), and ML-based rib segmentation. Standards: NEMA PS 3.1-3.20 (DICOM), ISO 14971 (Risk Management), IEC 62304 (Software Lifecycle).
Indications for Use
Indicated for viewing, manipulation, communication, and storage of multimodality DICOM medical images (CT, MRI, PET, DR) by trained professionals (physicians, medical technicians). Not indicated for diagnostic display of digital mammography images in the U.S.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shanghai United Imaging Healthcare Co., Ltd. % Xin Gao RA Manager No. 2258 Chengbei Rd., Jiading District Jiading Industrial District Shanghai, Shanghai 201807 CHINA
### Re: K230039
Trade/Device Name: Medical Image Post-processing Software, Model: uOmnispace Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: June 20, 2023 Received: June 20, 2023
### Dear Xin Gao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K230039
Device Name
uOmnispace
Indications for Use (Describe)
uOmnispace is a software solution intended to be used for viewing, manipulation, communication and storage of medical images. It allows processing and filming of multimodality DICOM images.
It can be used as a stand-alone device or together with a variety of cleared and unmodified software options, and also support to plug in multi-vendor applications which meet interface requirements.
u Omnispace is intended to be used by trained professionals, including but not limited to physicians and medical technicians.
The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized "U" shape, which is also in a bold font. The color of the logo is a dark gray.
# 510 (k) SUMMARY
# 1. Date of Preparation: June 20, 2023
# 2. Sponsor Identification
Shanghai United Imaging Healthcare Co., Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China
Establishment Registration Number: 3011015597
Contact Person: Xin GAO Position: Regulatory Affairs Manager Tel: +86-021-67076888-5386 Fax: +86-021-67076889 Email: xin.gao@united-imaging.com
# 3. Identification of Proposed Device
Trade Name: Medical Image Post-processing Software Common Name: Medical image management and processing system Model(s): uOmnispace
Regulatory Information Classification Name: Medical image management and processing system Classification: II Product Code: QIH Regulation Number: 21 CFR 892.2050 Review Panel: Radiology
# 4. Identification of Predicate Device(s)
Predicate Device 510(k) Number: K191040 Device Name: syngo.via Classification: II Product Code: LLZ
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized "U" shape, which is dark gray. The logo is simple and modern.
Reference Device 510(k) Number: K183170 Device Name: uWS-CT Classification: II Product Code: LLZ
# 5. Device Description
uOmnispace is a software only medical device, the hardware itself is not seen as part of the medical device and therefore not in the scope of this product.
uOmnispace provides 2D and 3D viewing, annotation and measurement tools, manually and automatically segmentation tools (Rib extraction algorithm is based on Machine Learning) and film and report features to cover the radiological tasks reading images and reporting. uOmnispace supports DICOM formatted images and objects, CT, MRI, PET and DR multimodality are supported.
uOmnispace is a software medical device that allows multiple users to remotely access clinical applications from compatible computers on a network. The system allows processing and filming of multimodality DICOM images. This software is for use with off the-shelf PC computer technology that meets defined minimum specifications.
uOmnispace communicates with imaging systems of different modalities and medical information systems of the hospital using the DICOM3.0 standard.
The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.
## 6. Indications for use
uOmnispace is a software solution intended to be used for viewing, manipulation, communication and storage of medical images.
It can be used as a stand-alone device or together with a variety of cleared and unmodified software options, and also support to plug in multi-vendor applications which meet interface requirements.
uOmnispace is intended to be used by trained professionals, including but not limited to physicians and medical technicians.
The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is displayed in a bold, sans-serif font. To the right of the text is a stylized "U" symbol, which is dark gray with a white vertical line running through the center. The logo appears to be clean and modern.
# 7. Summary of Technological Characteristics
The technology characteristics of uOmnispace, reflected in this 510(k) submission is substantially equivalent to those of the predicate device.
The following tables compare the technology and intended use of uOmnispace when compared to the predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is divided vertically by a white line. The logo is simple and modern in design.
| Item | Proposed Device | Predicate Device | Reference Device | Comparison |
|---------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | uOmnispace | syngo.via | uWS-CT | |
| | | (K191040) | (K183170) | |
| Device Classification<br>Name | Medical image<br>management and<br>processing system | Medical image<br>management and<br>processing system | / | Same |
| Product Code | LLZ, QIH | LLZ | / | Same |
| Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 | / | Same |
| Device Class | II | II | / | Same |
| Classification Panel | Radiology | Radiology | / | Same |
| Intended Use | uOmnispace is a software<br>solution intended to be<br>used for viewing,<br>manipulation,<br>communication and storage<br>of medical images. It<br>allows processing and<br>filming of multimodality<br>DICOM images.<br>It can be used as a stand-<br>alone device or together<br>with a variety of cleared<br>and unmodified software<br>options, and also support | syngo.via is a software<br>solution intended to be<br>used for viewing,<br>manipulation,<br>communication, and<br>storage of medical<br>images.<br>It can be used as a stand-<br>alone device or together<br>with a variety of cleared<br>and unmodified<br>syngo based software<br>options. | / | Substantial Equivalence<br>This difference between the predicate<br>device and the proposed device is about<br>description, it doesn't impact the safety<br>and effectiveness of the subject device. |
| | to plug in multi-vendor<br>applications which meet<br>interface requirements.<br>uOmnispace is intended to<br>be used by trained<br>professionals, including but<br>not limited to physicians<br>and medical technicians.<br>The system is not intended<br>for the displaying of digital<br>mammography images for<br>diagnosis in the U.S. | syngo.via supports<br>interpretation and<br>evaluation of<br>examinations within<br>healthcare institutions,<br>For example, in<br>Radiology, Nuclear<br>Medicine and Cardiology<br>environments.<br>The system is not<br>intended for the<br>displaying of digital<br>mammography images<br>for diagnosis in the U.S. | | |
| Client-Server<br>Architecture and Multi-<br>User Access | Yes<br>Based on client-server<br>architecture and supports<br>multi-user access. | Yes<br>Based on client-server<br>architecture and supports<br>multi-user access. | / Same | |
| Image communication<br>and storage | Yes<br>Communicate and store<br>medical images based on<br>standard communication<br>protocol DICOM. | Yes<br>Communicate and store<br>medical images based on<br>standard communication<br>protocol DICOM. | / Same | |
| Hardware /OS | Yes<br>Client:<br>Microsoft Windows 7 or<br>Microsoft | Yes<br>Client:<br>Microsoft Windows 7 SP1<br>or Microsoft | / Functional Substantially Equivalent<br>The client and server operating systems of<br>the proposed device are Microsoft Windows<br>and Linux respectively, while those of the | |
| | Windows 10 or compatible<br>versions<br>Server:<br>Linux Core CentOS7.7 or<br>compatible versions | Windows 8.1 or Microsoft<br>Windows 10<br>Server:<br>Microsoft Windows Server<br>2008 R2 or<br>Microsoft Windows Server<br>2012 R2. | | predicate device are Microsoft Windows and<br>Microsoft Windows Server respectively.<br>This difference doesn't impact the safety<br>and effectiveness of the subject device. |
| | | or Microsoft Server 2016 | | |
| Workflow control | Yes<br>Pre-processing: Auto<br>process images before<br>loading into post-processing<br>applications.<br>The client workflow<br>supports both single monitor<br>and dual monitors. | Yes<br>Workflows support the<br>user in preparing images<br>for examination.<br>Supports various monitor<br>setups workflow. | / | Functional Substantially Equivalent<br>Pre-processing is equivalent to part of the<br>"preparing images for examination", the<br>predicate device also assigns the<br>examination to an application. This<br>difference doesn't impact the safety and<br>effectiveness of the subject device. |
| Patient Administration | Yes<br>Patient administration<br>displays the patient data and<br>offers the function of<br>searching, sorting, and<br>editing of patient data and<br>image preview. | Yes<br>With simplified search<br>functionality, clearer<br>structure of search results,<br>unlimited search results,<br>periodic updates of search<br>results, image preview and<br>flexible floating patient<br>browser window. | / | Functional Substantially Equivalent<br>The proposed device does not support<br>flexible floating patient browser window.<br>This difference doesn't impact the safety<br>and effectiveness of the subject device. |
| Review 2D | Yes | / | Yes | Same |
| | | | | |
| | 2D image viewing, Textual<br>and graphical annotations,<br>distance, angle, ROI, image<br>addition and subtraction,<br>image filter are supported. | 2D image viewing, Textual<br>and graphical<br>annotations, distance,<br>angle, ROI, image addition<br>and subtraction, image<br>filter are supported. | | |
| Review 3D | Yes<br>3D image viewing, Save 3D<br>images in batches.<br>Segmentation functions such<br>as CT bone removal tools,<br>CT/MR tissue growing tools,<br>VOI, cut. | Yes<br>3D image viewing.<br>Save 3D images in<br>batches.<br>Segmentation functions<br>such as CT bone removal<br>tools, CT tissue growing<br>tools, VOI, cut. | Same | |
| Review 3D algorithm | Yes<br>Volume rendering (VR) with<br>Hyper Realistic Rendering<br>(HRR), Multi-Planar<br>Reconstruction (MPR),<br>Maximum Intensity<br>Projection (MIP), Minimum<br>Intensity Projection (MinIP),<br>Curved Planar Reformation<br>(CPR), Surface-shaded<br>Display (SSD).<br>Automatic body bone<br>removal algorithm | Yes<br>Volume rendering (VR),<br>Multi-Planar<br>Reconstruction (MPR),<br>Maximum Intensity<br>Projection (MIP),<br>Minimum Intensity<br>Projection (MinIP),<br>Curved Planar<br>Reformation (CPR),<br>Surface-shaded Display<br>(SSD).<br>Automatic body bone<br>removal algorithm | Functional Substantially Equivalent<br>HRR is an extension to the standard VR<br>rendering algorithm to visualize the<br>photorealistic images.<br>This difference between the proposed device<br>and the reference device doesn't impact the<br>safety and effectiveness of the subject device<br>as the necessary measures taken for the<br>safety and effectiveness of the proposed<br>device. | |
| | Automatic head bone<br>removal algorithm | | Automatic head bone<br>removal algorithm | |
| Rib segmentation | Yes<br>Machine learning based<br>algorithm | / | Yes<br>Threshold based<br>interactive algorithm | Functional Substantially Equivalent<br>Both are used for rib segmentation, the<br>reference device requires interaction,<br>while the proposed device does not.<br>This difference between the proposed device<br>and the reference device doesn't impact the<br>safety and effectiveness of the subject device<br>as the necessary measures taken for the<br>safety and effectiveness of the proposed<br>device |
| Inner view | Yes<br>3D virtual endoscopy view<br>and extract the centerline of<br>vessel, airway and colon.<br>Imaging algorithms:<br>CT colon Inner View<br>algorithm<br>CT vessel Inner View<br>algorithm<br>CT lung trachea Inner<br>View algorithm<br>MR vessel Inner View<br>algorithm | / | Yes<br>3D virtual endoscopy view<br>and extract the centerline<br>of vessel, airway and<br>colon.<br>Imaging algorithms:<br>CT colon Inner View<br>algorithm, CT vessel Inner<br>View algorithm<br>CT lung trachea Inner<br>View algorithm | Functional Substantially Equivalent<br>The MR vessel Inner View algorithm is<br>same with CT vessel Inner View algorithm,<br>the difference of input modality does not<br>affects the output.<br>This difference between the proposed device<br>and the reference device doesn't impact the<br>safety and effectiveness of the subject device<br>as the necessary measures taken for the<br>safety and effectiveness of the proposed<br>device. |
| Filming | Yes | / | Yes | Same |
| | | | | |
| Report | Support to print image and selection of printer.<br>Yes<br>Support to create reports, text editing and image inserting, customize report template, exporting and printing of report. | / | Support to print image and selection of printer.<br>Yes<br>Support to create reports, text editing and image inserting, customize report template, exporting and printing of report. | Same |
| Archiving | Yes<br>Import of DICOM images from configured network nodes (modalities, medical imaging process software, PACS, etc) or local and network drives or DVD/CD.<br>Export (archive) DICOM images to network nodes, or local and network drives or DVD/CD. | Yes<br>Import of DICOM data from network nodes or external media, and of DICOM-compliant or non DICOM compliant data from external media and Windows file system.<br>Export to CD/DVD, Windows file system, or other DICOM nodes. | / | Functional Substantially Equivalent<br>The proposed product does not support to import non DICOM compliant data from external media and Windows file system.<br>This difference between the proposed device and the reference device doesn't impact the safety and effectiveness of the subject device as the necessary measures taken for the safety and effectiveness of the proposed device |
Table 1 Substantial equivalent discussion for basic functions
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED IMAGING" are in bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized "U" shape, which is dark blue. The logo is simple and modern.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other and are in a bold, sans-serif font. To the right of the words is a stylized "U" symbol. The symbol is a dark gray color and has a vertical line running through the center of the "U".
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" symbol, which is dark blue. The logo is simple and modern.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is formed by two vertical bars connected by a horizontal bar in the middle. The logo is simple and modern in design.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized symbol that resembles a shield or a stylized letter "U" with a horizontal line through the middle. The color scheme is a dark, muted blue-green.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized symbol that resembles a shield or a stylized letter "U" with a horizontal line through the middle. The color of the text and symbol is a dark teal or blue-gray.
#### 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility
Not Applicable to the proposed device, because the device is stand-alone software.
### Electrical Safety and Electromagnetic Compatibility (EMC)
Not Applicable to the proposed device, because the device is stand-alone software.
### Software Verification and Validation
Software verification and validation testing was provided to demonstrate safety and efficacy of the proposed device. This includes a hazard analysis, and the potential hazards have been classified as a moderate level of concern (LOC). Those documentations include:
- · Software Description
- Device Hazard Analysis
- Software Requirements Specification
- Software Architecture Design Chart
- · Software Development Environment Description
- Software Verification and Validation
- · Cybersecurity Documents
# Animal Study
No animal study was required.
# Clinical Studies
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in bold, dark blue font. To the right of the words is a stylized symbol, also in dark blue, that resembles a shield or a stylized letter "U" with a vertical white line running through the center.
No clinical study was required.
# Performance Verification
To validate the uOmnispace software from a clinical perspective, the ML-based rib segmentation algorithm contained in the product underwent a scientific evaluation. The results of clinical data-based software validation for the subject device demonstrated equivalent performance in comparison to the reference device.
The performance testing for ML-based rib segmentation algorithm was performed on 60 subjects (data shown in Table 8-2) during the product development.
#### Acceptance Criteria .
The validation type and acceptance criteria is shown in the Table 8-1 below:
Table 8-1. Validation type and acceptance criteria
| Validation Type | Acceptance Criteria |
|-----------------|-----------------------------------------------------|
| Average DICE | The average dice of testing data is higher than 0.8 |
- Testing Data Information .
Table 8-2. Testing data information
| Information of data | 60 chest CTs |
|---------------------|------------------------------------------------|
| Sex | Male 37<br>Female 23 |
| Age | [14, 35] : 5<br>[36, 69] : 41<br>[70, 86] : 14 |
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized symbol that resembles the letter "U" with a horizontal line through the middle. The logo is simple and modern in design.
#### Performance Testing Summary: .
The average dice on testing data set is 0.855, which is higher than 0.8. Meanwhile, the subgroup analysis shows that (Table 8-3) the performance of algorithm are consist in different subgroups.
| Age | DICE |
|---------|-------|
| [14,35] | 0.848 |
| [36,69] | 0.856 |
| [70,86] | 0.856 |
| Gender | DICE |
| Female | 0.856 |
| Male | 0.855 |
Toblo 0 2 Subaraum norfam
#### Standard Annotation Process .
For ground truth annotations, all ground truth are annotators. A threshold based interactive tool is used to generate initial rib mask, then annotators will refine the first round annotation, they will check each other's annotation. At last, a senior clinical specialist will check and modify annotations to make sure the ground truth correct.
#### Testing & Training Data Independence .
The training data used for the training of the post-processing algorithm is independent of the digorithm
# Other Standards and Guidance
- NEMA PS 3.1 3.20 Digital Imaging and Communications in Medicine (DICOM) Set (2016). ●
- ISO 14971 Medical devices Application of risk management to medical devices (Edition 2.0, corrected version, 2007). ●
- IEC 62304 Medical device software Software life cycle processes (Edition 1.1, 2015). ●
#### Summary
{15}------------------------------------------------
Image /page/15/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized "U" symbol, which is made up of two vertical lines and a horizontal line. The logo is simple and modern, and the colors are muted.
The features described in this premarket submission are supported with the results of the testing mentioned above; the uOmnispace was found to have a safety and effectiveness profile that is similar to the predicate device and reference devices.
# 9. Substantially Equivalent Conclusion
The proposed device is equivalent to the predicate device with regard to safety and efficacy. This conclusion is based upon a comparison of intended use, technological characteristics, performance specification, device hazards, as well as verification and validation results.
In summary, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.