K091498 · Vital Images, Inc. · LLZ · Jun 5, 2009 · Radiology
Device Facts
Record ID
K091498
Device Name
VITREA ENDOVASCULAR STENT PLANNING SOFTWARE
Applicant
Vital Images, Inc.
Product Code
LLZ · Radiology
Decision Date
Jun 5, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
Vitrea® is a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea is not meant for primary image interpretation in mammography. In addition, Vitrea has the following additional indication: The Endovascular Stent Planning Software application is intended for use with CT (computed tomography) images to assist medical professionals in the analysis, treatment and follow-up of aortic vascular disorders that may require a stent procedure. The software provides 3D segmentation of the aorta and initializes stent measurements based on a template provided by the stent manufacturers. The user can review the 2D and 3D images, verify and correct the results of the segmentation and initialization, and generate a report with the stent measurements.
Device Story
Software module for Vitrea system; processes CT images for aortic vascular disorder management. Inputs: CT image data. Processing: automated 3D segmentation of aorta and thrombus; initialization of measurements (diameters, lengths, angles, volumes) using manufacturer-specific stent-graft templates. Outputs: 2D/3D visualizations, stent-graft sizing measurements, and clinical reports. Used in clinical settings by physicians/radiologists for pre-treatment planning and post-treatment follow-up. Workflow includes side-by-side comparative review; user verifies and edits automated segmentation and measurement results. Benefits: facilitates accurate stent sizing and procedural planning.
Clinical Evidence
Bench testing only. Software verification and validation testing confirmed that features function according to requirements and do not negatively impact existing Vitrea system functionality.
Technological Characteristics
Radiological image processing software module; operates on Vitrea platform. Features automated 3D segmentation and measurement initialization based on stent-graft templates. Connectivity: integrates with CT imaging systems. Software-based analysis.
Indications for Use
Indicated for medical professionals to assist in the analysis, treatment, and follow-up of patients with aortic vascular disorders requiring stent procedures using CT images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Vitrea®, Version 4.0 Medical Image Processing Software (K071331)
{0}------------------------------------------------
K091498
Page 1 of 2
JUN - 5 2009
## 2. 510(K) SUMMARY AS REQUIRED BY 21 CFR 807.92
510(k) Number: K
Date Prepared: May 5, 2009
| SUBMITTER INFORMATION | | |
|-----------------------------|----------------------------------------------------------|-------------------------------|
| Submitter: | Vital Images, Inc. | Contact Person |
| Street Address: | 5850 Opus Parkway Suite 300<br>Minnetonka, MN 55343-4414 | Darin Busch |
| Establishment registration: | 2134213 | Manager, Regulatory Affairs |
| | | Phone: (952) 487-9680 |
| | | Fax: (952) 487-9510 |
| | | Email: dbusch@vitalimages.com |
## DEVICE INFORMATION
| Trade Name | Endovascular Stent Planning Software |
|----------------------------|----------------------------------------|
| Common Name | Radiological Image Processing Software |
| Classification Name | System, Image Processing, Radiological |
| Regulation /Product Code | 21 CFR 892.2050 |
| Product Code | LLZ |
| Regulatory Classification: | Class II |
| Device Panel: | Radiology |
The Endovascular Stent Planning Software is substantially equivalent to the following predicate devices.
| Predicate Device | Manufacturer | FDA 510(k) |
|-----------------------------------------------------------|-------------------------------|------------|
| Vitrea®, Version 4.0 Medical Image<br>Processing Software | Vital Images, Inc | K071331 |
| Vitrea2, version 2.1 | Vital Images, Inc | K002519 |
| Preview® Treatment Planning<br>Software | MEDICAL MEDIA SYSTEMS,<br>INC | K040852 |
| GE AVA II/Vessel Xpress<br>(AVA II/Vessel Xpress) | GE HealthCare<br>Waukesha, WI | K060779 |
{1}------------------------------------------------
KC91498
Page 2 of 2
### 2.1. Device Description
The Endovascular Stent Planning Software, an application module for the Vitrea software, enables visualization and measurements of the aortic vessel for evaluation, treatment and follow up for aortic vascular disorders that may require a stent procedure. It has the following features:
- Automated segmentation of the aortic vessel, including thrombus .
- Custom and manufacturer-recommended stent-graft worksheets and report . templates for stent sizing
- Automatic initialization of endovascular measurements (diameters, lengths, . angles, volumes) based on the stent template selection, with easy centerline and contour editing tools
- Workflows for pre-treatment (stent, surgery) planning and post-treatment . follow up, including a side-by-side comparative review layout
### 2.2. Intended Use/Indications for Use
Vitrea® is a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea is not meant for primary image interpretation in mammography. In addition, Vitrea has the following additional indication:
The Endovascular Stent Planning Software application is intended for use with CT (computed tomography) images to assist medical professionals in the analysis, treatment and follow-up of aortic vascular disorders that may require a stent procedure. The software provides 3D segmentation of the aorta and initializes stent measurements based on a template provided by the stent manufacturers. The user can review the 2D and 3D images, verify and correct the results of the segmentation and initialization, and generate a report with the stent measurements.
#### 2.3. Summary of Testing
The software was designed, developed and tested according to written procedures. Software testing was completed to insure the Endovascular Stent Planning Software features function according to the requirements and interact without impact to existing functionality. The test results support a determination of substantial equivalence.
#### 2.4. Conclusions
Vitrea Endovascular Stent Planning software has similar intended use as the predicate devices and essentially identical technological characteristics. Minor technological differences do not raise any new questions regarding safety or effectiveness of the device. The Vitrea Endovascular Stent Planning Software application performs as intended, and presents with no unacceptable risks to the intended patient population or end user. Vitrea Endovascular Stent Planning Software is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of four stylized human figures in profile, arranged in a stacked formation. The figures are depicted in a simple, abstract manner.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## IIIN - 5 2009
Vital Images, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street, NW BUFFALO MN 55313
Re: K091498
Trade/Device Name: Vitrea® Endovascular Stent Planning Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class:_II Product Code; LLZ Dated: May 19, 2009
Received: May 20, 2009
#### Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology). | (240)-276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Samme M. Marre
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 1. INDICATIONS FOR USE STATEMENT
# Indications for Use
510(k) Number (if known): K09/498
Device Name: Vitrea® Endovascular Stent Planning Software
Indications for Use:
Vitrea® is a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea is not meant for primary image interpretation in mammography. Vitrea has the following additional indication:
The Endovascular Stent Planning Software application is intended for use with CT (computed tomography) images to assist medical professionals in the analysis, treatment and follow-up of aortic vascular disorders that may require a stent procedure. The software provides 3D segmentation of the aorta and initializes stent measurements based on a template provided by the stent manufacturers. The user can review the 2D and 3D images, verify and correct the results of the segmentation and initialization, and generate a report with the stent measurements.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Helene
(Division'Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.