K253844 · Mediso Medical Imaging Systems, Ltd. · KPS · Dec 30, 2025 · Radiology
Device Facts
Record ID
K253844
Device Name
AnyScan 3.0 NM Scanner Family
Applicant
Mediso Medical Imaging Systems, Ltd.
Product Code
KPS · Radiology
Decision Date
Dec 30, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The AnyScan 3.0 NM Scanner Family is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning or additional uses. SPECT: The SPECT subsystem is intended to provide projection and cross-sectional images through computer reconstruction of the data, representing radioisotope distribution in the patient body or in a specific organ using planar and tomographic scanning modes for isotopes with energies up to 588 keV. CT: CT component is intended to provide cross sectional images of the body by computer reconstruction of x-ray transmission data providing anatomical information. PET: The PET component is intended to provide cross- sectional images representing the distribution of tomographic scanning modes. SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images. The SPECT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (SPECT) and anatomical (CT) images. PET+CT: The PET and CT components used together acquire PET/CT images. The PET images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (PET) and anatomical (CT) images. The system maintains independent functionality of the SPECT, CT and PET components, allowing for single modality SPECT, CT and/ or PET diagnostic imaging. Software: The Nucline software is an acquisition, display and analysis package intended to aid the clinician to extract diagnostic information supported by image assessment tools, image enhancement features and image quantification of pathologies in images produced from SPECT, CT, PET and other imaging modalities. This CT system can be used for low dose lung cancer screening in high risk populations.
Device Story
AnyScan 3.0 NM Scanner Family is a multi-modality imaging system (SPECT, CT, PET) used in clinical settings by trained professionals. Device acquires physiological data via radioisotope tracers (SPECT/PET) and anatomical data via X-ray transmission (CT). Subsystems operate independently or in combination (SPECT/CT, PET/CT) for fused imaging; CT data provides attenuation correction and anatomical localization for nuclear medicine scans. Nucline software manages acquisition, reconstruction, and analysis; integrated InterView XP/FUSION and ClariCT software provide visualization, post-processing, and CT de-noising. Output consists of cross-sectional and projection images displayed on workstations. Clinicians use these images to assess metabolic function and anatomy, aiding in diagnosis, staging, and radiotherapy planning. Patient benefit includes precise lesion localization and improved diagnostic accuracy through fused physiological-anatomical imaging.
Clinical Evidence
Bench testing only. Performance validated against NEMA NU 1-2018 (SPECT) and NEMA NU 2-2018 (PET) standards. Metrics included intrinsic spatial resolution, linearity, energy resolution, flood field uniformity, count rate performance, system sensitivity, and co-registration accuracy. All results met predetermined acceptance criteria.
Technological Characteristics
Multi-modality scanner (SPECT, CT, PET). SPECT detectors: 3/8" (9.5mm) or 5/8" (15.9mm) crystals with 60 or 123 PMTs. CT: 16-slice, 60kW generator, 70cm bore. Energy source: X-ray and radioisotopes (up to 588 keV). Connectivity: DICOM-compliant. Software: Nucline (acquisition/analysis), ClariCT (de-noising), InterView (visualization). Safety: IEC 60601-1, IEC 60601-1-2, IEC 62366-1, ISO 10993-1, ISO 14971.
Indications for Use
Indicated for trained healthcare professionals to detect, localize, diagnose, stage, and restage lesions, tumors, and organ dysfunction in patients with cardiovascular disease, neurological disorders, or cancer. Includes low-dose lung cancer screening for high-risk populations.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 29, 2025
Mediso Medical Imaging Systems, Ltd.
% Prithul Bom
Most Responsible Person
Regulatory Technology Services, LLC
1000 Westgate Dr. Suite #510k
Saint Paul, Minnesota 55114
Re: K253844
Trade/Device Name: AnyScan 3.0 NM Scanner Family
Regulation Number: 21 CFR 892.1200
Regulation Name: Emission computed tomography system
Regulatory Class: Class II
Product Code: KPS JAK
Dated: December 1, 2025
Received: December 2, 2025
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253844 - Prithul Bom
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253844 - Prithul Bom
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D.
O'hara -S
Date: 2025.12.30 09:34:22 -05'00"
For
Lora Weidner, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
Device Name
AnyScan 3.0 NM Scanner Family
Indications for Use (Describe)
The AnyScan 3.0 NM Scanner Family is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning or additional uses.
SPECT: The SPECT subsystem is intended to provide projection and cross-sectional images through computer reconstruction of the data, representing radioisotope distribution in the patient body or in a specific organ using planar and tomographic scanning modes for isotopes with energies up to 588 keV.
CT: CT component is intended to provide cross sectional images of the body by computer reconstruction of x-ray transmission data providing anatomical information.
PET: The PET component is intended to provide cross- sectional images representing the distribution of tomographic scanning modes.
SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images. The SPECT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (SPECT) and anatomical (CT) images.
PET+CT: The PET and CT components used together acquire PET/CT images. The PET images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (PET) and anatomical (CT) images.
The system maintains independent functionality of the SPECT, CT and PET components, allowing for single modality SPECT, CT and/ or PET diagnostic imaging.
Software: The Nucline software is an acquisition, display and analysis package intended to aid the clinician to extract diagnostic information supported by image assessment tools, image enhancement features and image quantification of pathologies in images produced from SPECT, CT, PET and other imaging modalities.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
{4}
# AnyScan 3.0 NM Scanner Family 510(k) Summary
## Submitter / Applicant
Mediso Medical Imaging Systems, Ltd.
Laborc utca 3.
Budapest HUNGARY H-1037
Phone: +36-1-3993030
Email: info@mediso.hu
Contact: Daniel CSER (daniel.cser@mediso.com)
Date: 4 December 2024
## Preparer/Consultant
510K FDA Inc.
156 East Granada Blvd.
Ormond Beach, FL 32176
Phone: 386-506-8711
Fax: (386) 675-4621
Primary Contact: Lee Strong, Regulatory Dept. Mgr (lee@510kfda.com)
Secondary Contacts: Claude Berthoin, CEO (claude@510kfda.com).
## Device Classification
Trade/Model Names: AnyScan 3.0 NM Scanner Family
Regulation Name: Emission computed tomography system
Regulation Number: 21 CFR 892.1200
Primary Product Code: KPS
Classification Name: System, tomography, computed, emission
Regulatory Class: II
510k Review Panel: Radiology
Regulation Name: Computed tomography x-ray system.
Regulation Number: 21 CFR 892.1750
Secondary Product Code: JAK
Classification Name: System, tomography, computed, emission
Regulatory Class: II
510k Review Panel: Radiology
Page 1 of 10
{5}
Page 2 of 10
AnyScan 3.0 NM Scanner Family
510(k) Summary
## Predicate Devices
| | Primary Predicate Device | Secondary Predicate Device |
| --- | --- | --- |
| Trade name: | Symbia 6.7 (Family) | Biograph Horizon (PET/CT) |
| Manufacturer: | Siemens Medical Solutions USA, Inc | Siemens Medical Solutions USA, Inc. |
| 510(k) Clearance: | K200474 | K193178 |
| Regulation Number: | 21 CFR 892.1200 | 21 CFR 892.1200 |
| Regulation Name: | Emission computed tomography system | Emission computed tomography system |
| Regulatory Class: | Class II | Class II |
| Product Code: | KPS, JAK | KPS, JAK |
This predicate has not been subject to a design-related recall.
## Device Description
The AnyScan 3.0 NM Scanner Family will enable clinicians to utilize the device to perform separate studies in SPECT-CT, PET-CT, SPECT, PET and multi-slice CT modalities.
The AnyScan 3.0 NM Scanner Family includes the following products:
| AnyScan 3.0 NM Scanner Family | | |
| --- | --- | --- |
| Systems | Product Names | Detector Descriptions |
| SPECT | AnyScan DUO-Thera SPECT | XT-94/15.9 detector |
| | AnyScan DUO SPECT | UHP-60/9.5 detector |
| | AnyScan TRIO SPECT | |
| SPECT/CT | AnyScan DUO SPECT/CT | |
| | AnyScan TRIO SPECT/CT | |
| | AnyScan TRIO-IQMAX SPECT/CT | MAX-123/9.5 detector |
| | AnyScan TRIO-TheraMAX SPECT/CT | MAX-123/15.9 detector |
| SPECT/CT/PET | AnyScan DUO SPECT/CT/PET | UHP-60/9.5 detectors |
| | AnyScan TRIO SPECT/CT/PET | |
| | AnyScan TRIO-IQMAX SPECT/CT/PET | MAX-123/9.5 detector |
| | AnyScan TRIO-TheraMAX SPECT/CT/PET | MAX-123/15.9 detector |
| PET/CT | AnyScan PET/CT | PET and CT detectors |
{6}
Page 3 of 10
# AnyScan 3.0 NM Scanner Family 510(k) Summary
The partial product names 'TRIO' and 'DUO' only differentiate the number of built-in SPECT detectors.
The partial product names 'IQMAX' and 'TheraMAX' only differentiate the type of built-in SPECT detector. The SPECT gamma camera generates nuclear medicine images based on the uptake of radioisotope tracers in a patient's body, and supports integration with CT's anatomical detail for precise reference of the location of the metabolic activity.
The CT component produces cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.
The PET component images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.
The combination of SPECT, CT, and PET in a single device has several benefits. The SPECT subsystem images biochemical function while the CT subsystem images anatomy. The combination enables scans that not only indicate function, e.g., how active a tumor is, but precise localization, e.g., the precise location of that tumor in the body.
Combined SPECT and CT subsystems are intended for SPECT imaging enhanced with spatially registered CT image-based corrections, anatomical localization of tracer uptake and anatomical mapping. CT can be used to correct for the attenuation in SPECT acquisitions. Attenuation in SPECT is an unwanted side effect of the gamma rays scattering and being absorbed by tissue. This can lead to errors in the final image. The CT directly measures attenuation and can be used to create a 3D attenuation map of the patient which can be used to correct the SPECT images. The SPECT-CT scanner can be used to image and track how much dose was delivered to both the target and the surrounding tissue. The system maintains independent functionality of the CT and SPECT subsystems.
Combined PET and CT subsystems are intended for PET imaging enhanced with spatially registered CT image-based corrections, anatomical localization of tracer uptake and anatomical mapping. system maintains independent functionality of the CT and PET subsystems, allowing for single modality CT and/or PET diagnostic imaging.
A patient positioning light marker is generated by a low-power (Class II per IEC 60825-1) red laser.
Nucine software is installed on acquisition workstation to perform patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.
The systems also include display equipment, data storage devices, patient and equipment supports, software, and accessories.
InterView XP; InterView FUSION (K221984) and software is integrated for DICOM image visualization and post-processing.
ClariCT (K212074) software is integrated for DICOM CT de-noising.
## Indications for Use
The AnyScan 3.0 NM Scanner Family is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors,
{7}
Page 4 of 10
# AnyScan 3.0 NM Scanner Family 510(k) Summary
disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning or additional uses.
- The SPECT subsystem is intended to provide projection and cross-sectional images through computer reconstruction of the data, representing radioisotope distribution in the patient body or in a specific organ using planar and tomographic scanning modes for isotopes with energies up to 588 keV.
- CT component is intended to provide cross sectional images of the body by computer reconstruction of xray transmission data providing anatomical information.
- The PET component is intended to provide cross- sectional images representing the distribution of tomographic scanning modes.
- The SPECT and CT components used together acquire SPECT/CT images. The SPECT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (SPECT) and anatomical (CT) images.
- The PET and CT components used together acquire PET/CT images. The PET images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (PET) and anatomical (CT) images.
- The system maintains independent functionality of the SPECT, CT and PET components, allowing for single modality SPECT, CT and/ or PET diagnostic imaging.
- Software: The Nucline software is an acquisition, display and analysis package intended to aid the clinician to extract diagnostic information supported by image assessment tools, image enhancement features and image quantification of pathologies in images produced from SPECT, CT, PET and other imaging modalities.
- This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
# Comparison of Technological Characteristics with The Predicate Device
AnyScan 3.0 NM Scanner Family is similar in materials, energy source, or technological characteristics when compared to the predicate devices. Both the subject device and predicate devices are SPECT, CT, PET scanners that support various visualization and evaluation tools. The intended use and fundamental scientific technology are similar to the predicate devices; therefore, Mediso believes that they are substantially equivalent to the predicate devices.
# Performance data
The following performance data were provided in support of the substantial equivalence determination.
## 1.1 SPECT Performance
AnyScan 3.0 conforms to the NEMA NU 1-2018 international standard, which describes the performance of SPECT devices. 99mTc isotope measurements unless other isotope specified.
The predetermined acceptance values between the subject and predicate device are similar. There is no impact on safety and effectiveness.
All performance testing of the subject device met the predetermined acceptance values.
{8}
AnyScan 3.0 NM Scanner Family
510(k) Summary
1.1.1 Comparison of SPECT Performance with 3/8” (9.5 mm) detector
| SPECT Performance Criteria | Acceptance values | | |
| --- | --- | --- | --- |
| | Subject device
AnyScan 3.0 NM Scanner Family with TRIO-IQMAX MAX-123/9.5 detector (3/8” crystal, 123 pcs PMT) | Subject device
AnyScan 3.0 NM Scanner Family with UHP-60/9.5 detector (3/8” crystal, 60 pcs PMT) | Predicate device
Symbia 6.7
3/8” detector (3/8” crystal, 59 pcs PMT) |
| Intrinsic Spatial Resolution | | | |
| FWHM in CFOV | ≤2.6 mm | ≤3.4 mm | ≤3.8 mm |
| FWHM in UFOV | ≤2.7 mm | ≤3.5 mm | ≤3.9 mm |
| FWTM in CFOV | ≤5.1 mm | ≤6.6 mm | ≤7.5 mm |
| FWTM in UFOV | ≤5.3 mm | ≤6.8 mm | ≤7.7 mm |
| Intrinsic Spatial Linearity | | | |
| Differential in CFOV | ≤0.10 mm | ≤0.17 mm | ≤0.2 mm |
| Differential in UFOV | ≤0.10 mm | ≤0.18 mm | ≤0.2 mm |
| Absolute in CFOV | ≤0.20 mm | ≤0.30 mm | ≤0.4 mm |
| Absolute in UFOV | ≤0.22 mm | ≤0.36 mm | ≤0.7 mm |
| Intrinsic Energy Resolution | | | |
| FWHM in CFOV | ≤9.3% | ≤9.5% | ≤9.9% |
| Intrinsic Flood Field Uniformity (Uncorrected) | | | |
| Differential in CFOV | ≤1.8 % | ≤2.0 % | ≤2.5 % |
| Differential in UFOV | ≤1.9 % | ≤2.2 % | ≤2.7 % |
| Integral in CFOV | ≤2.2 % | ≤2.5 % | ≤2.9 % |
| Integral in UFOV | ≤2.6 % | ≤3.3 % | ≤3.7 % |
| Multiple Window Spatial Registration | ≤0.6 mm | ≤0.6 mm | ≤0.6 mm |
| Intrinsic Count Rate Performance in Air | | | |
| Maximum Count Rate @20% window | ≥600 kcps | ≥600 kcps | ≥310 kcps |
| System Spatial Resolution Without Scatter (at 10 cm) | LEHR-HS | | LEHR |
| FWHM in CFOV | ≤7.3 mm | ≤7.4 mm | ≤7.5 mm |
| FWTM in CFOV | ≤13.2 mm | ≤13.4 mm | ≤13.6 mm |
| System Spatial Resolution with Scatter (at 10 cm) | LEHR-HS | | LEHR |
| FWHM in CFOV | ≤8.0 mm | ≤8.1 mm | ≤8.3 mm |
| FWTM in CFOV | ≤18.5 mm | ≤19.1 mm | ≤18.6 mm |
| System Planar Sensitivity (at 10 cm) | LEHR-HS | | LEHR |
| Absolute | 220 cpm/μCi | 220 cpm/μCi | 202 cpm/μCi |
| System volume sensitivity | LEHR-HS | | LEHR |
| Sensitivity per axial centimeter (VSAC)
(Dual-Head/Triple-Head system) | ≥ -/36.0 (kcts/sec)/(MBq/cm²) | ≥ 24.0/36.0 (kcts/sec)/(MBq/cm²) | 12.0/No triple-head system is available (kcts/sec)/(MBq/cm²) |
| Detector-Detector Sensitivity Variation | ≤4.0% | ≤5.0% | ≤5.0% |
Page 5 of 10
{9}
Page 6 of 10
# AnyScan 3.0 NM Scanner Family
510(k) Summary
| SPECT Performance Criteria | Acceptance values | | |
| --- | --- | --- | --- |
| | Subject device
AnyScan 3.0 NM
Scanner Family with
TRIO-IQMAX MAX-
123/9.5 detector
(3/8” crystal,
123 pcs PMT) | Subject device
AnyScan 3.0 NM
Scanner Family
with UHP-60/9.5
detector
(3/8” crystal, 60 pcs
PMT) | Predicate device
Symbia 6.7
3/8” detector
(3/8” crystal,
59 pcs PMT) |
| Reconstructed spatial
resolution with scatter at FBP/
Iterative reconstruction SW | LEHR-HS | | LEHR |
| Center | ≤9.6/4.5 mm | ≤9.9/4.6 mm | ≤10.7/5.8 mm |
| Radial | ≤10.2/4.3 mm | ≤10.2/4.5 mm | ≤10.9/5.0 mm |
| Tangential | ≤6.8/3.1 mm | ≤7.1/3.4 mm | ≤7.9/4.1 mm |
{10}
AnyScan 3.0 NM Scanner Family
510(k) Summary
1.1.2 Comparison of SPECT Performance with 5/8" (15.9 mm) detector
| SPECT Performance Criteria | Acceptance values | |
| --- | --- | --- |
| | Subject device
AnyScan 3.0 NM Scanner Family with TRIO
TheraMAX
MAX-123/15.9 mm detector (5/8" crystal thickness, 123 pcs PMT) | Predicate device
Symbia 6.7 5/8" detector (5/8" crystal thickness, 59 pcs PMT) |
| Intrinsic Spatial Resolution | | |
| FWHM in CFOV | ≤3.3 mm | ≤4.5 mm |
| FWHM in UFOV | ≤3.4 mm | ≤4.6 mm |
| FWTM in CFOV | ≤6.5 mm | ≤8.7 mm |
| FWTM in UFOV | ≤6.6 mm | ≤8.9 mm |
| Intrinsic Spatial Linearity | | |
| Differential in CFOV | ≤0.10 mm | ≤0.2 mm |
| Differential in UFOV | ≤0.10 mm | ≤0.2 mm |
| Absolute in CFOV | ≤0.20 mm | ≤0.5 mm |
| Absolute in UFOV | ≤0.22 mm | ≤1.0 mm |
| Intrinsic Energy Resolution | | |
| FWHM in CFOV | ≤9.3% | ≤9.9% |
| Intrinsic Flood Field Uniformity (Uncorrected) | | |
| Differential in CFOV | ≤1.8 % | ≤2.5 % |
| Differential in UFOV | ≤1.9 % | ≤2.7 % |
| Integral in CFOV | ≤2.2 % | ≤2.9 % |
| Integral in UFOV | ≤2.6 % | ≤3.7 % |
| Multiple Window Spatial Registration | ≤0.6 mm | ≤1.0 mm |
| Intrinsic Count Rate Performance in Air | | |
| Maximum Count Rate @20% window | ≥600 kcps | ≥460 kcps |
| Intrinsic Spatial Resolution at 75 kcps | | |
| FWHM in UFOV | ≤3.4 mm | ≤4.6 mm |
| FWTM in UFOV | ≤6.6 mm | ≤8.9 mm |
| System Spatial Resolution Without Scatter (at 10 cm) | LEHR-HS | LEHR |
| FWHM in CFOV | ≤7.4 mm | ≤7.8 mm |
| FWTM in CFOV | ≤13.4 mm | ≤14.9 mm |
| System Spatial Resolution With Scatter (at 10 cm) | LEHR-HS | LEHR |
| FWHM in CFOV | ≤8.1 mm | ≤8.9 mm |
| FWTM in CFOV | ≤19.1 mm | ≤19.5 mm |
| System Planar Sensitivity (at 10 cm) | LEHR-HS | LEHR |
| Absolute | 244 cpm/μCi | 225 cpm/μCi |
| Reconstructed spatial resolution with scatter at FBP | LEHR-HS | LEHR |
| Center | ≤9.9 mm | ≤11.5 mm |
| Radial | ≤10.2 mm | ≤12.0 mm |
| Tangential | ≤7.1 mm | ≤8.8 mm |
Page 7 of 10
{11}
AnyScan 3.0 NM Scanner Family
510(k) Summary
1.2 CT performance
The performance values between the subject and predicate device are similar. There is no impact on safety and effectiveness.
All performance testing of the subject device met the performance criteria.
| CT Performance Criteria | Subject Device
AnyScan 3.0 NM Scanner Family | Reference Device
Symbia Intevo Bold (K162483) |
| --- | --- | --- |
| Acquired/Reconstructed CT slice/rot. | 16/32 | 16/32 |
| Min. rotational time (sec) | 0.5 | 0.5 |
| Min. full (360*) scan time (sec) | 0.5 | 0.5 |
| Min. partial (240*) scan time (sec) | 0.33 | 0.33 |
| Max. z-coverage (mm) | 20 | 19.2 |
| Min. slice thickness (mm) | 0.625 | 0.6 |
| Tube min. current (mA) | 10 | 25 |
| Min. current-exposure time @360*
(mAs) | 5 | 12.5 |
| Min. current-exposure time @240*
(mAs) | 3.3 | 8.25 |
| Tube max. current (mA) | 440 | 345 |
| X-Ray generator power (kW) | 60 | 50 |
| Tube anode heat storage (MHU) | 7.5 | 5 |
| Tube min.-max. current (mA) | 10-440 | 25-345 |
| High Power CT Imaging | 80kV: 440mA
100kV: 440mA
120kV: 440mA
140kV: 380mA | 80, 110, 130 kV
25 - 345 mA |
| Single breath-hold im. (400 mm)
time(sec) | 5.5 | 5.2 |
| High-contrast res. @ 2% MTF
(lpm/cm) (±10%) | 15.9 | 15.8 |
| Helical & continuous scan length
(mm) | 1920 | 1860 |
| High Voltage KVpp range | 80, 100, 120, 140 | 80, 110, 130 |
| Bore size (cm) | 70 | 70 |
| Generator Output | 53.2 kW | 50kW |
| # Elements | 24 x 896 | 24 x 1472 |
| Topogram (Length x Scan Field) | 100-2000 x 500mm | 128 -1530 x
500 mm |
| LC Resolution (20cm Catphan /
CTDIvol 32cm) | 3.0 mm/ 2.9HU / 13.60 mGy | 5 mm/3HU/12.88 mGy |
| CTDI(Dose/100 mAs) /
16cm Phant. (typical head)
32cm Phant. (typical body) | Typical head: 120keV
Center: 14.0 mGy
Surface: 14.2 mGy
Typical body: 120keV
Center: 4.7 mGy
Surface: 8.4 mGy | Typical head: 130 kV
A: 18.7 mGy
B: 19.6 mGy
Typical body: 130 kV
A: 5.8 mGy
B: 10.9 mGy |
| Max. HC Resolution (2% MTF wire
in plastic) | 15.16 lp/cm | 15.15 lp/cm |
| Max. scan range // Max. Scantime | 202 cm // 120 s | 157 cm // 100 s |
Page 8 of 10
{12}
Page 9 of 10
# AnyScan 3.0 NM Scanner Family 510(k) Summary
## 1.3 PET performance
AnyScan 3.0 conforms to NEMA NU 2-2018 international standard, which describes the performance of PET devices. 18F isotope measurements unless otherwise indicated.
The performance values between the subject and predicate device are similar. There is no impact on safety and effectiveness.
All performance testing of the subject device met the predetermined acceptance values.
| PET Performance Criteria | Acceptance values | |
| --- | --- | --- |
| | Subject device
AnyScan 3.0 NM Scanner Family | Predicate device
Biograph Horizon PET/CT (K193178) |
| Resolution – Full size | | |
| Transverse Resolution FWHM @ 1 cm | <5.2 mm | ≤4.7 mm |
| Transverse Resolution FWHM @ 10 cm | <5.5 mm | ≤5.5 mm |
| Transverse Resolution FWHM @ 20 cm | <6.4 mm | ≤7.6 mm |
| Axial Resolution FWHM @ 1 cm | <4.9 mm | ≤5.0 mm |
| Axial Resolution FWHM @ 10 cm | <5.0 mm | ≤7.0 mm |
| Axial Resolution FWHM @ 20 cm | <6.2 mm | ≤11.3 mm |
| Sensitivity / Scatter / Bias | | |
| Sensitivity @435 keV LLD | 5.0 cps/kBq | ≥5.8 cps/kBq |
| Scatter Fraction at peak NECR | ≤44% | ≤40% |
| Mean bias (%) at peak NEC | ≤+/- 6% | ≤+/- 6% |
| Image Quality (4 to 1) - (% Contrast / Background Variability) | | |
| 10mm sphere | ≥15% / <9% | ≥10% / ≤10% |
| 13mm sphere | ≥30% / <8% | ≥25% / ≤10% |
| 17mm sphere | ≥35% / <7% | ≥40% / ≤10% |
| 22mm sphere | ≥50% / <7% | ≥55% / ≤10% |
| 28mm sphere | ≥70% / <5% | ≥60% / ≤10% |
| 37mm sphere | ≥85% / <5% | ≥65% / ≤10% |
| Co-Registration Accuracy | | |
| Max Error | ≤4 mm | ≤5 mm |
## 1.4 Safety and effectiveness
Safety and effectiveness of the AnyScan 3.0 NM Scanner Family device was further documented by the following:
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Regarding Biocompatibility, the patient-contacting components are evaluated according to ISO 10993-1.
Regarding the Software, the software verification report is according to IEC 62304.
Regarding Risk Management, the risk management report was completed according to ISO 14971.
{13}
Page 10 of 10
# AnyScan 3.0 NM Scanner Family 510(k) Summary
AnyScan 3.0 NM Scanner Family conforms to applicable FDA recognized and international IEC, ISO and NEMA standards with regards to performance and safety as required by the respective SPECT FDA Guidance Documents. SPECT detector and CT performance is conducted according to NEMA NU1-2018, and PET detector performance is conducted according to NEMA NU2-2018. The performance is similar to the predicate devices.
Regarding Electromagnetic Compatibility, the tests were completed according to IEC 60601-1-2. The device passed these tests to the IEC standard.
Regarding Electrical, Mechanical and Thermal Safety, the tests were completed according to IEC 60601-1 and IEC 62366-1. The device passed these tests to the IEC standards.
Regarding Performance Testing, the performance tests were completed for SPECT according to NEMA NU 1-2018 and for PET according to NEMA NU 2-2018. The device passed these tests to the NEMA standard.
## Conclusions
A comparison of the subject to a legally marketed predicate was performed to demonstrate substantial equivalence. Results of that comparison revealed the following:
1. The subject and predicate devices have similar indications for use and intended use.
2. The subject and predicate devices have similar technological characteristics, with only minor differences that do not raise different questions of safety and effectiveness.
3. Tests based on consensus industry standards confirmed the performance capabilities of the subject device.
The comparison analysis and performance test results demonstrate the subject device is at least as safe and effective as the legally marketed predicate. Furthermore, this finding of substantial equivalence warrants clearance for marketing activities.
In conclusion, AnyScan 3.0 NM Scanner Family is substantially equivalent to the legally marketed Symbia 6.7 and Biograph Horizon, and thus warrants clearance from FDA for marketing activities in the United States.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.