K200474 · Siemens Medical Solutions USA, Inc. · KPS · Mar 20, 2020 · Radiology
Device Facts
Record ID
K200474
Device Name
Symbia 6.7
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
KPS · Radiology
Decision Date
Mar 20, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1200
Device Class
Class 2
Indications for Use
The Siemens Symbia series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures. SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques; Planar imaging, and tomographic imaging for isotopes with energies up to 588 keV. CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data (anatomy) from either the same axial plane taken at different angles or spiral planes take at different angles. SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (SPECT) and anatomical (CT) images. Software: the syngo MI Applications software is a display and analysis package intended to aid the clinician in the assessment and quantification of pathologies in images produced from SPECT, PET, CT and other imaging modalities. The following statement applies only to Siemens Symbia T16, and Symbia Intevo Bold systems: This CT system can be used for low dose lung cancer screening in high risk populations.
Device Story
Symbia 6.7 is a hybrid SPECT/CT imaging system; combines SPECT gamma camera for physiological radionuclide distribution imaging with X-ray CT for anatomical localization. Inputs: radionuclide emissions (SPECT) and X-ray transmission data (CT). Operation: system reconstructs cross-sectional images; SPECT and CT data can be fused/registered for attenuation correction and anatomical mapping. Used in clinical settings by trained healthcare professionals (radiologists, nuclear medicine physicians). Output: cross-sectional images and quantitative analysis via syngo MI Applications software. Clinicians use outputs to diagnose, stage, and plan radiotherapy or interventional procedures. Benefits: improved diagnostic accuracy through combined anatomical/physiological data; enables low-dose lung cancer screening in specific models.
Clinical Evidence
Bench testing only. Performance testing conducted per NEMA NU-1:2018 standards for SPECT and existing CT validation. Quantitative accuracy for isotopes (Tc99m, I123, In111, Lu177, I131) demonstrated errors ≤10% in phantom studies. No clinical data required for this modification.
Technological Characteristics
Hybrid SPECT/CT system. SPECT: variable angle gamma camera, NEMA NU-1:2018 compliant. CT: spiral X-ray reconstruction. Software: syngo MI Applications VB20A/VB22A, Windows 10. Connectivity: DICOM (NEMA PS3.1-3.20). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-44, NEMA XR 25/28. Sterilization: N/A (imaging equipment).
Indications for Use
Indicated for patients requiring detection, localization, diagnosis, staging, or restaging of lesions, tumors, disease, and organ function evaluation (e.g., cardiovascular, neurological, cancer). CT component indicated for low-dose lung cancer screening in high-risk populations for specific models (T16, Intevo 16, Intevo Bold).
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a bold, sans-serif font. Below that, the word "ADMINISTRATION" is written in a smaller font size.
Siemens Medical Solutions USA, Inc. % Ms. Cynthia Busch Regulatory Affairs Specialist 2501 N Barrington Road HOFFMAN ESTATES IL 60192
March 20, 2020
## Re: K200474
Trade/Device Name: Symbia 6.7 Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: Class II Product Code: KPS, JAK Dated: February 24, 2020 Received: February 26, 2020
Dear Ms. Busch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K200474
Device Name Symbia 6.7
#### Indications for Use (Describe)
The Siemens Symbia series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques; Planar imaging, and tomographic imaging for isotopes with energies up to 588 keV.
CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data (anatomy) from either the same axial plane taken at different angles or spiral planes take at different angles.
SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (SPECT) and anatomical (CT) images.
Software: the syngo MI Applications software is a display and analysis package intended to aid the clinician in the assessment and quantification of pathologies in images produced from SPECT, PET, CT and other imaging modalities.
The following statement applies only to Siemens Symbia T16, and Symbia Intevo Bold systems: This CT system can be used for low dose lung cancer screening in high risk populations.* * As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung
Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
K200474 as required by 21 CFR Part 807.87(h) and 21 CFR Part 807.92(c)
## Identification of the Submitter
| Submitter: | Cynthia Busch<br>Regulatory Technical Specialist<br>Siemens Medical Solutions USA, Inc.<br>2501 N. Barrington Road<br>Hoffman Estates, IL 60192<br>USA |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name / Address of Manufacturer: | Siemens Medical Solutions USA, Inc<br>Molecular Imaging<br>2501 N. Barrington Road<br>Hoffman Estates, IL 60192<br>USA |
| Telephone Number: | (847) 643-6818 |
| Date of Submission: | February 24, 2020 |
| Identification of the product | |
| Device Proprietary Name: | Symbia 6.7 |
| Common Name: | Single-photon emission computed tomography (SPECT)<br>system |
| Classification Name: | Emission Computed Tomography per 21 CFR 892.1200 |
| Product Code: | KPS (primary)<br>JAK (secondary) |
| Classification Panel: | Radiology |
| Class: | II |
{4}------------------------------------------------
#### Marketed Devices to which Equivalence is claimed
## Primary Predicate:
| Device | Manufacturer | 510(k) Number | Device Classification<br>Name | Regulation |
|------------|---------------------------------------|---------------|---------------------------------|-----------------|
| Symbia 6.5 | Siemens Medical<br>Solutions USA, Inc | K162337 | Emission Computed<br>Tomography | 21 CFR 892.1200 |
#### Secondary Predicate:
| Device | Manufacturer | 510(k) Number | Device Classification<br>Name | Regulation |
|-----------------------|---------------------------------------|---------------|---------------------------------|-----------------|
| Symbia<br>Intevo Bold | Siemens Medical<br>Solutions USA, Inc | K162483 | Emission Computed<br>Tomography | 21 CFR 892.1200 |
## Reference Predicate Devices:
| Device | Manufacturer | 510(k) Number | Device Classification<br>Name | Regulation |
|----------------------------------------------------------------------------|---------------------------------------|---------------|----------------------------------------|--------------------|
| SOMATOM<br>Scope/Scope<br>Power, and<br>SOMATOM<br>Emotion 16<br>with VC50 | Siemens Medical<br>Solutions USA, Inc | K183548 | Computed<br>Tomography<br>X-Ray System | 21 CFR<br>892.1750 |
| Device | Manufacturer | 510(k) Number | Device<br>Classification Name | Requlation |
|------------|---------------------------------------|---------------|---------------------------------|-----------------|
| Symbia 5.0 | Siemens Medical<br>Solutions USA, Inc | K131634 | Emission Computed<br>Tomography | 21 CFR 892.1200 |
{5}------------------------------------------------
## Device Description:
The Siemens Symbia systems consist of Sinqle Photon Emission Computed Tomography (SPECT) scanners and integrated hybrid X-Ray Computed Tomography (CT) and SPECT scanners. The SPECT subsystem images and measures the distribution of radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and integrates CT's anatomical detail for precise reference of the location of the metabolic activity. The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The system can be used as an integrated SPECT and CT modality while also enabling independent functionality of SPECT and CT as standalone diagnostic imaging devices.
| Symbia 6.7 Family | | |
|--------------------|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SPECT only systems | Symbia E Single | variable angle, single detector gamma camera |
| | Symbia E Dual | variable angle, dual detector gamma camera |
| | Symbia S | variable angle dual detector SPECT system |
| | Symbia Evo | variable angle dual detector SPECT system |
| | Symbia Evo Excel | variable angle dual detector SPECT system |
| SPECT/CT Systems | Symbia T | SPECT/CT system with non-diagnostic CT support for only attenuation correction and anatomical localization |
| | Symbia T2, T6, T16 | a variable angle dual detector SPECT with a 2, 6, or 16-slice spiral CT to give the system full functionality for all SPECT-only, xSPECT, or stand-alone CT diagnostic applications |
| | Symbia Intevo Excel | SPECT/CT system with non-diagnostic CT support for only attenuation correction and anatomical localization |
| | Symbia Intevo 2, Intevo 6, Intevo 16, Intevo Bold | variable angle dual detector SPECT and 2, 6, or 16-slice spiral CT to give the system full functionality for all SPECT-only, xSPECT, or stand-alone CT diagnostic applications |
All systems implement a new software version syngo MI Applications VB20A.
Figure 1: Symbia models
{6}------------------------------------------------
## Modifications in Symbia 6.7 include:
#### a) Software: Modifications to xSPECT features and other software updates:
- update to organ processing features: ●
- a. qastric retention addition:
- b. renal processing improvement.
- Addition of SPECT Dose Reporting
- . xSPECT Quantification: expansion of commercially available xSPECT Quant (Symbia 5.0 K131634) to support Absolute Quantification of additional isotope I-131.
- upgrade to Windows® 10 Operating System for syngo MI Applications VB22A including:
- a. updated version of synqo software; VI20
- b. Somaris 5 VC50 (K183548), 3rd party software applications; Corridor 4DM, Cedars, Scenium, and TrueD are updated for Windows 10 compatibility.
## b) Hardware:
Updated computer models for obsolescence
## Intended Use:
The Symbia Intevo Excel, Symbia Intevo series and Symbia T series are radiological imaqinq systems that combine a single photon emission computed tomography (SPECT) camera system for nuclear medicine images, and a computed tomography (CT) camera system for x-ray images. The Symbia E series, Symbia Evo, and Symbia Evo Excel systems are SPECT camera systems.
The SPECT system is intended to detect the location and distribution of qamma ray radionuclides in the body and produce cross-sectional images through computer reconstruction of the data, and the CT system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. The nuclear medicine images and the x-ray images may be reqistered and displayed in a fused format (overlaid in the same orientation) for the anatomical localization of the nuclear medicine data (that is, distribution of radiopharmaceuticals).
The SPECT and CT portions of the system may be used independently or in combination, and may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories. The SPECT and CT images may be transferred to other systems for radiation therapy planning or additional uses.
{7}------------------------------------------------
#### Indications for Use:
The Siemens Symbia series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
SPECT: To detect or imaqe the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV
CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different anqles.
SPECT+CT: Perform CT scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.
Software: The MI Applications software is a display and analysis package intended to aid the clinician in the assessment and quantification of pathologies taken from SPECT, PET, CT and other imaging modalities.
The following statement applies only to Siemens Symbia T16, Symbia Intevo 16, and Symbia Intevo Bold systems:
This CT system can be used for low dose lung cancer screening in high risk populations. *
* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lunq Screening Trial (N Enql J Med 2011; 365:395-409) and subsequent literature, for further information.
#### Technological Characteristics:
Symbia 6.7 systems are based on the commercially available Symbia 6.5 (K162337) and Symbia Intevo Bold (K162483) . The software updates are based on the same fundamental technology of the xSPECT quantification in Symbia SPECT/CT predicate components. SPECT detector, existing collimators, and CT performance specifications do not change between the commercially available Symbia 6.5 systems and Symbia 6.7 systems.
{8}------------------------------------------------
## Performance Testing:
Performance testing for the CT subsystem was included in the original premarket notification for the CT subsystems and there have been no changes affecting this testing.
Symbia 6.7 is designed in accordance with the 60601-1 series including all relevant collateral standards general and specific (see standards table in Safety and Effectiveness section). Collimator performance testinq is conducted according to NEMA NU-1:2018. All Performance testing met the predetermined acceptance values.
Fiqure 2 below Depicts Detector Specifications which are valid for Symbia 6.7 and published in the predicate device Symbia 6.5 data sheet. Specifications have not changed from the predicate device.
| Detector Specifications | 3/8" |
|--------------------------------------------------------------|----------|
| Intrinsic Spatial Resolution | |
| FWHM in CFOV | ≤3.8 mm |
| FWHM in UFOV | ≤3.9 mm |
| FWTM in CFOV | ≤7.5 mm |
| FWTM in UFOV | ≤7.7 mm |
| Intrinsic Spatial Linearity | |
| Differential in CFOV | ≤0.2 mm |
| Differential in UFOV | ≤0.2 mm |
| Absolute in CFOV | ≤0.4 mm |
| Absolute in UFOV | ≤0.7 mm |
| Intrinsic Energy Resolution | |
| FWHM in CFOV | ≤9.9% |
| Intrinsic Flood Field Uniformity (Uncorrected) | |
| Differential in CFOV | ≤2.5% |
| Differential in UFOV | ≤2.7% |
| Integral in CFOV | ≤2.9% |
| Integral in UFOV | ≤3.7% |
| Multiple Window Spatial Registration | ≤0.6 mm |
| Intrinsic Count Rate Performance in Air | |
| Maximum Count Rate | 310 kcps |
| Intrinsic Spatial Resolution at 75 kcps | |
| FWHM in UFOV | ≤4.1 mm |
| FWTM in UFOV | ≤7.8 mm |
| Intrinsic Flood Field Uniformity at 75 kcps | |
| Differential in CFOV | ≤2.5% |
| Differential in UFOV | ≤2.7% |
| Integral in CFOV | ≤2.9% |
| Integral in UFOV | ≤3.7% |
| System Spatial Resolution Without Scatter (LEHR<br>at 10 cm) | |
{9}------------------------------------------------
| FWHM in CFOV | ≤7.5 mm |
|----------------------------------------------------------------------------------------------|-------------------------------|
| FWTM in CFOV | ≤13.6 mm |
| System Spatial Resolution With Scatter (LEHR at 10 cm) | |
| FWHM in CFOV | ≤8.3 mm |
| FWTM in CFOV | ≤18.6 mm |
| System Planar Sensitivity (LEHR at 10 cm) | |
| Absolute | 202 cpm/µCi |
| System Planar Sensitivity (ME at 10 cm) | |
| Absolute 111In | 430 cpm/µCi |
| Average Volume Sensitivity per axial centimeter | |
| LEHR Tc99m | 12,000<br>(cts/sec)/(MBq/cm²) |
| Detector-Detector Sensitivity Variation (LEHR, 99mTc) | ≤5.0% |
| Whole-Body System Spatial Resolution Without Scatter at 10 cm/min Scan Speed (LEHR at 10 cm) | |
| FWHM Perpendicular | ≤7.5 mm |
| FWHM Parallel | ≤7.9 mm |
| FWTM Perpendicular | ≤14.0 mm |
| FWTM Parallel | ≤14.2 mm |
Figure 2 Detector Specifications
#### Image Quality
Figure 3 below shows NEMA performance testing results demonstrate accurate quantification in phantoms for all the isotopes and collimators tested. NEMA and high count performance results meet specifications.
| Summary of Quantitative accuracy specifications* | |
|--------------------------------------------------|------|
| Quantitative error Tc99m LEHR/LPHR | ≤10% |
| Quantitative error I123 LPHR/MELP | ≤10% |
| Quantitative error In111 MELP | ≤10% |
| Quantitative error Lu177 MELP | ≤10% |
| Quantitative error Lu177 MELP at 310kcps** | ≤10% |
| Quantitative error I131 HE ** | ≤10% |
Figure 3 Quantitative Accuracy Specifications
* In phantoms for objects with negligible partial volume effect
- ** incident count rate
The quantitative error for all isotopes with the collimators is smaller or equal to 10%, and met the predefined acceptance criteria.
{10}------------------------------------------------
SPECT detector, existing collimators, quantification supported with additional isotopes and CT performance specifications do not change the fundamental technology between the commercially available Symbia 6.5 systems and Symbia 6.7 systems. There are no changes in the system design that could impact the SPECT performance specifications, and the commercially available NEMA NU1:2018 specifications.
Performance testing for the CT subsystem was included in the original premarket notification for the CT subsystems and there have been no modifications from the original FDA clearance via CT 510k Somaris 5 VC50 (K183548) that affect device performance.
{11}------------------------------------------------
## Safety and Effectiveness:
The device labelinq contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product.
Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards such as IEC 60601-1 series and 21 CFR 1020.30 and 21 CFR 1020.33 to minimize electrical, mechanical and radiation hazards.
Siemens claims compliance with the following product standards for the Symbia systems:
- IEC 60601-1: 2005+ A1:2012 ●
- IEC 60601-1-2: 2014 ●
- IEC 60601-1-3: 2013
- IEC 60601-1-6:2010 +A1:2013 ●
- IEC 60601-2-28:2010 ●
- IEC 60601-2-44: 2009 + A1:2012
- IEC 61223-2-6:2006 ●
- . IEC 61223-3-5:2004
- IEC 62366-1:2015 ●
- IEC 62304:2015
- NEMA XR 25: 2010 ●
- NEMA XR 28: 2013
- . NEMA PS3.1-3.20: 2016
Symbia 6.7 conforms to applicable FDA recognized and international IEC, ISO and NEMA standards with reqards to performance and safety as required by the respective SPECT FDA Guidance Documents. SPECT detector and CT performance is conducted according to NEMA NU1:2018, and the performance does not change from the predicate device.
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Cybersecurity features have been assessed according to FDA Guidance issued October 2, 2014 'Content of Premarket Submissions for Management of Cybersecurity in Medical Devices'.
The Symbia systems' software has specific cybersecurity controls to prevent unauthorized access, modifications, misuse or denial of use. Additionally, controls are enabled to prevent the unauthorized use of information that is stored, accessed or transferred between the Symbia systems and external devices.
Verification and validation of Siemens systems is performed in accordance with documented procedures, design and code reviews, test plans and specifications. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured durinq component integration, software validation and system testing.
{12}------------------------------------------------
#### Substantial Equivalence:
Symbia 6.7 has the same intended use and utilizes the same fundamental scientific technology as the predicate device. There have been no changes implemented in the modifications of the Symbia systems that impact the fundamental technology or the indications for use. Siemens considers Symbia 6.7 to be as safe, as effective, and with performance substantially equivalent to the commercially available predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.