K253630 · Medacta International S.A. · LPH · Aug 28, 2026 · Orthopedic
Device Facts
Record ID
K253630
Device Name
Mpact 3D Metal Foam
Applicant
Medacta International S.A.
Product Code
LPH · Orthopedic
Decision Date
Aug 28, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Mpact 3D Metal Foam is intended to be used in cementless hip replacement surgeries in combination with Medacta F-Cage and a Medacta cemented cup (Apricot or Versacem), or in combination with a Medacta cemented cup only. Mpact 3D Metal Foam, Medacta F-Cage and Apricot are indicated for use in the case of acetabular bony defect during total hip arthroplasty. Total hip arthroplasty is indicated in the following cases: - Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia. - Avascular necrosis of the femoral head when total hip arthroplasty has been chosen. - Acute traumatic fracture of the femoral head or neck when total hip arthroplasty has been chosen. - Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement, or total hip arthroplasty. When the Mpact 3D Metal Foam is used with the Versacem, is also recommended when dislocation prevention is the main driver for the prosthesis choice and the acetabular bone quality is damaged, as in the following cases: - Elderly patients or poor bone quality. - Revision for recurrent dislocation in elderly patients with poor bone quality. - Patients treated with radiotherapy or chemotherapy. Versacem is intended for cemented use only.
Device Story
Mpact 3D Metal Foam is an acetabular system extension for cementless hip replacement; includes 3D-printed Ti-6Al-4V cups, F-Cage reinforcement, and UHMWPE cemented cups (Apricot/Versacem). Used in OR by orthopedic surgeons for acetabular bony defects and total hip arthroplasty. Components provide structural support and joint reconstruction; F-Cage and Versacem combination specifically addresses dislocation risk in patients with compromised bone quality. Device is single-use and sterile. Clinical benefit derived from restoring joint function and stability in complex revision or primary arthroplasty cases.
Clinical Evidence
No clinical studies were conducted. Evidence is based on non-clinical bench testing, including fatigue testing (ASTM F3090), static tension (ASTM F1147), static/dynamic shear (ASTM F1044, ASTM F1160-14), Taber abrasion (ASTM F1978), ROM evaluation (ISO 21535), and impingement risk analysis. Biocompatibility, pyrogenicity, and shelf-life evaluations were also performed.
Technological Characteristics
Mpact 3D Metal Foam Cup: Ti-6Al-4V (ASTM F2924) via additive manufacturing. F-Cage: Titanium Grade 1 (ASTM F67). Apricot Cup: High Cross UHMWPE. System is cementless (cup) or cemented (liner/cage). Sterile, single-use. No software or electronics.
Indications for Use
Indicated for patients requiring total hip arthroplasty due to arthrosis, traumatic arthritis, rheumatoid polyarthritis, congenital hip dysplasia, avascular necrosis, acute traumatic fracture, or failed previous hip surgery. When used with Versacem, indicated for elderly patients, poor bone quality, revision for recurrent dislocation, or patients treated with radiotherapy/chemotherapy.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Trabecular metal acetabular revision shells (K050937)
Trabecular metal acetabular revision system cage (K061226)
{0}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
August 28, 2026
Medacta International S.A.
% Christopher Lussier
Senior Director, Quality and Regulatory
Medacta USA
6386 Global Dr.
Suite 101
Memphis, Tennessee 38141
Re: K253630
Trade/Device Name: Mpact 3D Metal Foam
Regulation Number: 21 CFR 888.3358
Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented
Prosthesis
Regulatory Class: Class II
Product Code: LPH
Dated: July 30, 2026
Received: July 30, 2026
Dear Christopher Lussier:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253630 - Christopher Lussier
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253630 - Christopher Lussier
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

For: Limin Sun, Ph.D.
Assistant Director
DHT6A: Division of Joint
Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
K253630 Page 1 of 1
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K253630
Device Name
Mpact 3D Metal Foam
# Indications for Use (Describe)
Mpact 3D Metal Foam is intended to be used in cementless hip replacement surgeries in combination with Medacta F-Cage and a Medacta cemented cup (Apricot or Versacem), or in combination with a Medacta cemented cup only. Mpact 3D Metal Foam, Medacta F-Cage and Apricot are indicated for use in the case of acetabular bony defect during total hip arthroplasty.
Total hip arthroplasty is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia.
- Avascular necrosis of the femoral head when total hip arthroplasty has been chosen.
- Acute traumatic fracture of the femoral head or neck when total hip arthroplasty has been chosen.
- Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement, or total hip arthroplasty.
When the Mpact 3D Metal Foam is used with the Versacem, is also recommended when dislocation prevention is the main driver for the prosthesis choice and the acetabular bone quality is damaged, as in the following cases:
- Elderly patients or poor bone quality.
- Revision for recurrent dislocation in elderly patients with poor bone quality.
- Patients treated with radiotherapy or chemotherapy.
Versacem is intended for cemented use only.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
K253630 Page 1 of 4
# 510(k) Summary
## I. Submitter
Medacta International SA
Strada Regina
6874 Castel San Pietro (CH)
Switzerland
Phone (+41) 91 696 60 60
Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA
Applicant Correspondent: Chris Lussier, Senior Director, Quality and Regulatory, Medacta USA
Date Prepared: November 19, 2025
Date Revised: August 27, 2026
## II. Device
| Device Proprietary Name: | Mpact 3D Metal Foam |
| --- | --- |
| Common or Usual Name: | Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented |
| Classification Name: | Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis |
| Primary Product Code | LPH |
| Regulation Number: | 21 CFR 888.3358 |
| Device Classification | II |
## III. Predicate Device
Substantial equivalence is claimed to the following predicate devices.
Primary Predicate device:
- Trabecular metal acetabular revision shells, K050937, Zimmer Trabecular
Additional Predicate devices:
- Trabecular metal acetabular revision system cage, K061226, Zimmer Trabecular
- Trabecular metal revision shell liners, K051516, Zimmer Trabecular
- REDAPT Porous Acetabular Shell and Cemented Liner, K150790, SMITH & NEPHEW, INC.
- Mpact Extension, K230011, Medacta International SA
- Mpact acetabular system, K103721, Medacta International SA
- Mpact extension, K122641, Medacta International SA
Reference devices:
- Mpact 3D Metal Implants– DMLS Technology, K202568, Medacta International SA
{5}
K253630 Page 2 of 4
- GMK 3D Metal Tibial Tray, K221850, Medacta International SA
- M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System, K171595, Medacta International SA
- Medacta Anatomic Shoulder Prosthesis K170910, Medacta International SA
#### IV. Device Description
The Mpact 3D Metal Foam is an extension of Medacta Mpact 3D Metal acetabular system portfolio intended to be used in hip replacement surgeries. The Mpact 3D Metal Foam includes implantable devices, provided individually packed, sterile and single-use. Specifically, the devices subject of this submission are:
- Mpact 3D Metal Foam Cup, available in 18 sizes, from ø46 mm to ø80 mm (in 2 mm increments) and made of Ti-6Al-4V according to ASTM F2924 by using additive manufacturing.
- Apricot Cup, an High Cross UHMWPE cemented cup available from size 36 mm to size 64 mm (in 2 mm increments) with internal diameters varying with sizes.
- F-Cage, a reinforcement cage made of Titanium Grade 1 according to ASTM F67 and available in 2 versions (Long and Short) and 10 sizes, 5 left and 5 right, from ø46 mm to ø66 mm.
#### V. Indications for Use
Mpact 3D Metal Foam is intended to be used in cementless hip replacement surgeries in combination with Medacta F-Cage and a Medacta cemented cup (Apricot or Versacem), or in combination with a Medacta cemented cup only. Mpact 3D Metal Foam, Medacta F-Cage and Apricot are indicated for use in the case of acetabular bony defect during total hip arthroplasty.
Total hip arthroplasty is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia.
- Avascular necrosis of the femoral head when total hip arthroplasty has been chosen.
- Acute traumatic fracture of the femoral head or neck when total hip arthroplasty has been chosen.
- Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement, or total hip arthroplasty.
When the Mpact 3D Metal Foam is used with the Versacem, is also recommended when dislocation prevention is the main driver for the prosthesis choice and the acetabular bone quality is damaged, as in the following cases:
- Elderly patients or poor bone quality.
- Revision for recurrent dislocation in elderly patients with poor bone quality.
- Patients treated with radiotherapy or chemotherapy.
Versacem is intended for cemented use only.
#### VI. Comparison of Technological Characteristics
The subject and predicate devices (K050937, K051516, K061226) are substantially equivalent with respect to the following characteristics:
- Design, except for the Mpact 3D Metal Foam Cup;
- Fixation;
- Materials, except for the Mpact 3D Metal Foam Cup;
- Biocompatibility;
- Device usage; and
{6}
K253630 Page 3 of 4
- Packaging.
The subject devices differ from the predicate devices (K050937, K051516, K061226) with respect to:
- Range of product;
- Mpact 3D Metal Foam Cup design; and
- Mpact 3D Metal Foam Cup materials.
# Discussion
The subject Mpact 3D Metal Foam Cup and Apricot Cup range of sizes, slightly different respect to the one of the predicate devices (K050937, K051516), has been designed to cover a wider population, but it has no impact on safety and effectiveness as demonstrated by fatigue testing.
The different diameters of the subject F-Cage and predicate devices (K061226) do not arise any new issue of safety and effectiveness since both devices cover the same range of product being compatible with the same group of acetabular shells.
The different Mpact 3D Metal Foam Cup design and feature for impactor/alignment with respect to the predicate devices (K050937) do not arise any new issue with respect to safety and effectiveness since it is shared with the reference devices (K202568).
The different material of the subject Mpact 3D Metal Foam Cup and the predicate devices (K050937) does not introduce new questions of safety and effectiveness since the subject devices' material and manufacturing process are shared with the reference devices (K221850, K202568).
Finally, subject devices shelf-life and sterilization is shared with Medacta already cleared devices.
The comparison of technological characteristics and performance data provided within the submission supports the substantial equivalence of the subject devices respect to the predicate devices.
# VII. Performance Data
Based on the risk analysis, testing activities were conducted to written protocols. The following validations and tests are provided in support of the substantial equivalence determination:
# Non-Clinical Studies
# o DESIGN VALIDATION
- Mpact 3D Metal Foam and Medacta F-Cage, Design validation
# o PERFORMANCE TESTING
- Mpact 3D Metal Foam Acetabular Cup Fatigue Test according to ASTM F3090
- Static tension test according to ASTM F1147
- Static shear test according to ASTM F1044
- Dynamic shear according ASTM F1160-14
- Taber abrasion test according to ASTM F1978
- Acetabular Cage Fatigue Test – F-Cage
- Apricot Full PE HC Cemented Cup - ROM Evaluation according to ISO 21535
- Apricot Full PE HC – Evaluation of impingement risk – Rationale
{7}
K253630 Page 4 of 4
○ PYROGENICITY
- Bacterial endotoxin test (LAL test) according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>)
- Pyrogen test according to USP chapter <151> for pyrogenicity determination
- The subject devices are not labeled as non-pyrogenic or pyrogen free.
○ BIOCOMPATIBILITY assessment
○ SHELF-LIFE evaluation
Clinical Studies:
- No clinical studies were conducted.
### VIII. Conclusion
The information provided above supports that the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.