K103721 · Medacta International · LPH · Mar 21, 2011 · Orthopedic
Device Facts
Record ID
K103721
Device Name
MPACT ACETABULAR SYSTEM
Applicant
Medacta International
Product Code
LPH · Orthopedic
Decision Date
Mar 21, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Mpact® Acetabular System is intended for cementless use in total hip arthroplasty and in primary or revision surgery. The patient should be skeletally mature. The patient's condition should be due to one or more of: - Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriactic arthritis, congenital hip dysplasia, or ankylosing spondylitis. - Avascular necrosis of the femoral head. - Acute traumatic fracture of the femoral head or neck. - Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
Device Story
Mpact® Acetabular System is a cementless total hip arthroplasty component set; includes titanium alloy (Ti-6Al-4V) hemispherical cups with TiGrowth porous coating and fixed UHMWPE or HighCross® highly crosslinked UHMWPE liners; includes optional titanium alloy cancellous bone screws. Used in clinical settings by orthopedic surgeons during primary or revision hip replacement. Device provides stable acetabular interface; liners accept various CoCrMo or ceramic ball heads. Benefits include restoration of joint function and pain relief for patients with degenerative or traumatic hip conditions. Performance verified via bench testing (push-out, lever-out, rotational stability, wear simulation).
Clinical Evidence
No clinical data provided. Evidence consists of bench testing including multi-axial hip joint simulation per ISO 14242-1 (3.4 kN peak load, 5 million cycles) to evaluate wear rates of HighCross® UHMWPE liners. Design verification testing included push-out, lever-out, rotational stability, coating validation, shell deformation resistance, range of motion, and bone screw testing. All tests met pre-defined acceptance criteria.
Technological Characteristics
Materials: Ti-6Al-4V (ASTM F136, ISO 5832-3) for cups/screws; pure titanium (ASTM F1580) for TiGrowth porous coating; UHMWPE (ISO 5834-2) and HighCross® HXUHMWPE for liners. Design: Hemispherical cup, modular fixed liner, optional bone screws. Sterilization: Sterile, single-use. Connectivity: None.
Indications for Use
Indicated for skeletally mature patients undergoing primary or revision total hip arthroplasty due to osteoarthritis, post-traumatic/rheumatoid/psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis, avascular necrosis, acute traumatic fracture, or failure of previous hip surgery (reconstruction, fixation, arthrodesis, hemiarthroplasty, or total hip replacement) with sufficient bone stock.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
DePuy Pinnacle® with Gription™ Acetabular Cups (K071784)
REFLECTION 3 Acetabular System, Smith & Nephew (K070756)
Apex Modular Heads, +10.5 mm offset, Omni Lifesciences (K101575)
Apex- LNK Poly Acetabular Liners and Apex Modular Head, 40 mm, Omni Lifesciences (K100555)
Versafitcup® Double Mobility, Medacta International (K083116)
Versafitcup® Double Mobility HighCross® UHMWPE Liners, Medacta International (K092265)
Medacta Bone Screws, Medacta International (K092265)
Medacta Total Hip Prosthesis, Medacta International (K072857)
Medacta Total Hip Prosthesis CoCrMo ball heads, Medacta International (K080885)
Submission Summary (Full Text)
{1}------------------------------------------------
K103721
p. 1/5
Image /page/0/Picture/2 description: The image shows the logo for Medacta International. The logo consists of the word "Medacta" in a sans-serif font, with two triangles above the "M". Below the word "Medacta" is the word "International" in a smaller font. To the right of the word "International" is a small square with a plus sign inside.
MAR 2 1 20il
## 510(k) Summary
| Manufacturer: | Medacta International SA |
|---------------|-----------------------------|
| | Strada Regina |
| | 6874 Castel San Pietro (CH) |
| | Switzerland |
| | Phone (+41) 91 696 60 60 |
| | FAX (+41) 91 696 60 66 |
| Contact Person: | Ms. Natalie J. Kennel<br>Consultant<br>NJK & Associates, Inc.<br>13721 Via Tres Vista<br>San Diego, CA 92129 USA<br>Phone: (858) 705-0350<br>Fax: (858) 764-9739<br>email: nkennel@njkconsulting.com |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
March 17, 2011 Date Prepared: :
# DEVICE INFORMATION
| Trade/Proprietary Name: | Mpact® Acetabular System<br>Medacta 40 mm Ball heads |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Total Hip Acetabular Components<br>Femoral ball heads |
| Classification Name: | 21 CFR 888.3358 Hip joint, femoral metal/<br>polymer/metal semi-<br>constrained porous-coated uncemented prosthesis<br>Device Product Code: LPH<br>21 CFR 888.3353 Hip joint metal/ceramic/polymer<br>semi-constrained cemented or nonporous uncemented<br>prosthesis<br>Device Product Code: LZO<br>Class II |
Mpact® Acetabular System 510(k)
.
Section 5 - Page 1 of 5
{2}------------------------------------------------
K103721
Predicates:
K071784 DePuy Pinnacle® with Gription™ Acetabular Cups K001534 Pinnacle Acetabular System, DePuy Orthopedics K070756 REFLECTION 3 Acetabular System, Smith & Nephew K101575 Apex Modular Heads, +10.5 mm offset, Omni Lifesciences K100555 Apex- LNK Poly Acetabular Liners and Apex Modular Head, 40 mm, Omni Lifesciences K083116 Versafitcup® Double Mobility, Medacta International K092265 Versafitcup® Double Mobility HighCross® UHMWPE Liners, Medacta International K091069 Medacta Bone Screws, Medacta International K072857 Medacta Total Hip Prosthesis, Medacta International K080885 Medacta Total Hip Prosthesis CoCrMo ball heads, Medacta International
### Product Description:
The Mpact® Acetabular System components are designed to be used with the Medacta Total Hip Prosthesis System. The Medacta Total Hip Prosthesis system includes the Quadra S, H, R, and C Stems and CoCrMo and ceramic ball heads (K072857, K073337, K080885, K082792, and K083558). The AMIStem femoral stems also work with the Medacta Total Hip Prosthesis System (K093944). The Medacta Total Hip Prosthesis System is a total hip replacement system consisting of the femoral stem made of metal, a modular femoral head made of metal or ceramic, and acetabular components. The Mpact® Acetabular System acetabular components that are the subject of this 510(k) consist of a metal cup made of Titanium alloy and a fixed liner that is made of ultra-high molecular weight polyethylene (UHMWPE) or HighCross® highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE).
The metal acetabular component or shell is machined from titanium alloy (Ti-6Al-4V) conforming to ISO 5832-3 and ASTM F136. The outside of the metal component is a hemispherical design. The outer surface of the metallic cup has a porous coating called TiGrowth made of pure titanium conforming to ASTM F1580. The polyethylene liner is a fixed liner made of UHMWPE conforming to ISO 5834-2. The liner has a minimum thickness of at least 5.5 mm when made of standard UHMWPE and of least 4 mm when made of HXUHMWPE.
If supplemental bone screw fixation is needed, cancellous bone screws made of titanium alloy (Ti-6Al-4V) can be inserted through screw holes. Screw hole covers made of titanium alloy (Ti-6Al-4V) are also provided for the central hole.
Also included in this submission are additional Medacta ball heads made of CoCrMo in the 40 mm diameter to augment the Medacta ball heads previously cleared in K072857, K073337 and K080885.
The metal acetabular component is available in 11 sizes and is for use with
Mpact® Acetabular System 510(k)
Section 5 - Page 2 of 5
{3}------------------------------------------------
polyethylene liners that accept both CoCrMo with diameters of 22, 28, 32, and 36 mm and MectaCer BIOLOX® forte Ceramic ball heads with diameters of 28, 32, and 36 mm. They also work with the Medacta 40 mm CoCrMo ball heads. The fixed liners are available in both flat and hooded options in a range of sizes.
All the Mpact® Acetabular System components and Medacta 40 mm ball heads are supplied sterile in single-use individual packages.
### Indications for Use:
The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery.
The patient should be skeletally mature.
The patient's condition should be due to one or more of:
- Severely painful and/or disabled joint: as a result of osteoarthritis, post-. traumatic arthritis, rheumatoid arthritis or psoriatic arthritis, Congenital hip dysplasia, ankylosing spondylitis.
- Avascular necrosis of the femoral head. .
- Acute traumatic fracture of the femoral head or neck. t
- Failure of previous hip surgery: joint reconstruction, internal fixation, . arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
## Wear Claim
Medacta's Mpact® liners are made of HighCross™ highly crosslinked UHMWPE (ETO sterilized). The UHMWPE raw material, used for HighCross™ UHMWPE, is Chirulen 1020. The HighCross™ UHMWPE liners were gamma irradiated at 100 kGy and thermally stabilized at 150°C with controlled cooling, to optimize the crystallinity of the material. The liners were manufactured by turning compression molded UHMWPE bars and were finished products that followed all the manufacturing process flow including final sterilization.
The Mpact® liners tested were 4 mm thick, size Ø40/F. Testing was conducted under a multi-axial hip joint simulation with a standard walking gait cycle as specified by ISO 14242-1 with a 3.4 kN peak load for 5 million cycles using a 40 mm size M CoCrMo ball head and metal acetabular shell of size 54 mm. The lubricant was composed of calf serum with a 30 g/L protein content and deionized water. EDTA and Patricin were added to bind the calcium phosphate and to retard bacteria-induced degradation, respectively. The lubricant was filtered before doing the test and degraamon, replaced every 0.5 million cycles. The lubricant was replaced every 0.5 million cycles. Three liners were tested. Wear was measured at 0.5, 1, 2, 3, 4 and 5 million cycles.
Average gravimetric and volumetric wear rates were 6.28 mg +/- 1.68 mg per million Avelage grannon of - 1.8 mm3 per million cycles, respectively, for the Mpact® oycles and 0.78 HMWPE liners. The average total wear for 5 million cycles was 31.54 mg for the 40 mm inner diameter liners.
{4}------------------------------------------------
The in vitro hip wear simulator tests have not been shown to quantitatively predict clinical wear performance. Also, a reduction in total polyethylene wear volume or wear rate alone may not result in an improved clinical outcome as wear particle size and morphology are also critical factors in the evaluation of the potential for wear mediated osteolysis and associated aseptic implant loosening. Particle size and morphology were not evaluated as part of this wear claim.
#### Comparison to Predicate Devices
The Mpact® Acetabular System family is similar to the DePuy Pinnacle® with Gription™ Acetabular Cups and the Smith & Nephew REFLECTION 3 Acetabular system in indications for use, design, materials, coating, and technological characteristics. The bone screws are similar to the Medacta Bone Screws in materials and design. The standard UHMWPE liners are made of the same material and process as the Versafitcup® Double Mobility standard UHMWPE liners. The HighCross® highly crosslinked UHMWPE liners of the Mpact® Acetabular System is the same HighCross® highly crosslinked UHMWPE of which the Versafitcup® Double Mobility HighCross® UHMWPE liners are made. The Medacta 40 mm ball heads are similar to Apex Modular heads in indications for use, material, size and options. The Medacta 40 mm ball heads have the same indications for use, material, design, and performance characteristics as the previously cleared CoCrMo ball heads in the Medacta Total Hip Prosthesis System.
## Performance Testing
No performance standards applicable to this device have been adopted under Section 514 of the Food, Drug and Cosmetic Act. Performance testing of the Mpact® Acetabular System was conducted in accordance with various international standards and FDA guidance documents.
The Mpact® Acetabular System and Medacta 40 mm ball heads were tested as part of design verification to written protocols with pre-defined acceptance criteria. The Mpact® testing was conducted on the worst case component size and option/design. The design verification testing included push out, lever out and rotational stability of the modular connection of the fixed liner to the metal shell, coating validation, metal shell deformation resistance during impaction, range of motion, wear, and bone screw testing. The Medacta 40 mm ball heads testing were included in the worst case testing of the previously cleared ball heads. The testing met all acceptance criteria and verifies that the performance of the Mpact® Acetabular System and Medacta 40 mm ball heads are substantially equivalent to the predicate devices.
#### Conclusion:
The data and information provided in this submission support the conclusion that the Mpact® Acetabular System and Medacta 40 mm ball head are substantially equivalent to its main' predicates devices: DePuy Pinnacle® with Gription® eqarabular Cups and Smith & Nephew REFLECTION 3 Acetabular System with respect to indications for use and technological characteristics. The Mpact® Acetabular System is also substantially equivalent to the Medacta Bone Screws and Versafitcup® Double Mobility and Versafitcup® Double Mobility HighCross® I crounted.po to the chnological characteristics. The Medacta 40 mm ball heads are
Section 5 - Page 4 of 5
{5}------------------------------------------------
K103721
p. 5/5
substantially equivalent to the Apex Modular heads and the Medacta Total Hip Prosthesis CoCrMo ball heads with respect to indications for use and technological characteristics.
> . .
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle or bird in flight, with its wings and body represented by curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Medacta International SA % NJK & Associates, Inc. Ms. Natalie J. Kennel 13721 Via Tres Vista San Diego, California 92129
MAR 2 1 201
Re: K103721
Trade/Device Name: Mpact® Acetabular System
Regulation Number: 21 CFR 888.3358
Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis
Regulatory Class: II Product Code: LPH, LZO Dated: December 14, 2010 Received: December 21, 2010
Dear Ms. Kennel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{7}------------------------------------------------
Page 2 - Ms. Natalie J. Kennel
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
for D. Holmquist
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known):
Device Name: Mpact® Acetabular System
Indications for Use:
The Mpact® Acetabular System is intended for cementless use in total hip arthroplasty and in primary or revision surgery. The patient should be skeletally mature. The patient's condition should be due to one or more of:
- Severely painful and/or disabled joint: as a result of osteoarthritis, post-. traumatic arthritis, rheumatoid arthritis or psoriactic arthritis, congenital hip dysplasia, or ankylosing spondylitis.
- . Avascular necrosis of the femoral head.
- Acute traumatic fracture of the femoral head or neck. .
- . Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
Prescription Use x Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for n. melkerson
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
Page _ of _ l
510(k) Number K103721
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.