Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. Target Detachable Coils are indicated for endovascular embolization of: • Intracranial aneurysms • Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae • Arterial and venous embolizations in the peripheral vasculature
Device Story
Target Detachable Coils are platinum-tungsten alloy coils attached to stainless steel delivery wires; used for endovascular obstruction/occlusion of blood flow in neurovascular and peripheral vascular abnormalities. Operated by physicians in clinical settings; coils are delivered via microcatheters to target sites. Detachment is achieved using the InZone Detachment System (sold separately). The device is compatible with both transfemoral and transradial access routes. By occluding blood flow, the coils facilitate embolization of aneurysms and malformations, providing therapeutic benefit by reducing or eliminating flow into the target vascular abnormality.
Clinical Evidence
Bench testing only. Performance evaluated via tracking and deployment testing in clinically representative access models comparing radial vs. femoral access. Metrics included microcatheter friction, ease of deployment/retraction, and track force (advance/pullback). Results showed no noticeable or statistical differences between access routes.
Technological Characteristics
Materials: Platinum-tungsten alloy coil, stainless steel delivery wire. Form factor: Secondary coil diameters 1-24 mm, lengths 1-50 cm. Shapes: Helical, 360, 3D, Tetrahedral. Sterilization: Ethylene Oxide (EtO). Connectivity: Compatible with InZone Detachment System. No software/algorithm components.
Indications for Use
Indicated for endovascular embolization of intracranial aneurysms, neurovascular abnormalities (arteriovenous malformations, arteriovenous fistulae), and arterial/venous embolizations in peripheral vasculature.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 11, 2025
Stryker Neurovascular
Genae Tripp
Staff Regulatory Affairs Specialist
47900 Bayside Parkway
Fremont, California 94538-2588
Re: K252694
Trade/Device Name: Target Detachable Coils
Regulation Number: 21 CFR 882.5950
Regulation Name: Neurovascular Embolization Device
Regulatory Class: Class II
Product Code: HCG, KRD
Dated: August 26, 2025
Received: August 26, 2025
Dear Genae Tripp:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252694 - Genae Tripp
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252694 - Genae Tripp
Page 3
Sincerely,
SARA S. THOMPSON -S
Sara S. Thompson, D.V.M.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252694 | |
| Device Name Target Detachable Coils | |
| Indications for Use (Describe) Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. Target Detachable Coils are indicated for endovascular embolization of: • Intracranial aneurysms • Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae • Arterial and venous embolizations in the peripheral vasculature | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Stryker Neurovascular
510(k) Summary, K252694
Target Detachable Coils
Date Prepared: December 5, 2025
Submitter: Stryker Neurovascular
47900 Bayside Parkway
Fremont, CA 94538
Facility Registration #300853977
Contact: Genae Tripp
Staff Regulatory Affairs Specialist
Tel (805) 630-3378
E-mail: genae.tripp@stryker.com
Device Trade Name: Target Detachable Coils
Classification Name: Target Detachable Coils are vascular and neurovascular embolization devices under 21 CFR 870.3300 (KRD) and 21 CFR 882.5950 (HCG), respectively, and are Class II devices (special controls).
The special control for the devices is FDA’s guidance document, Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices (issued 29 Dec 2004).
Legally Marketed Predicate Device: K242243 (Cleared 29 Aug 2024)
Page 1 of 5
{5}
Stryker Neurovascular
510(k) Summary, K252694
Target Detachable Coils
# DEVICE DESCRIPTION
Stryker Neurovascular Target Detachable Coils are comprised of the following coil types:
| Target 360 Nano | Target Helical Nano | Target XL 360 Soft |
| --- | --- | --- |
| Target 360 Ultra | Target Helical Ultra | Target XL 360 Standard |
| Target 360 Soft | Target 3D | Target XL Helical |
| Target 360 Standard | Target XXL 360 | Target Tetra |
Please note that all Target Detachable Coils referenced in this submission are considered the Subject Device.
Target Detachable Coils are stretch resistant, electrolytically detachable coils consisting of a platinum-tungsten alloy coil attached to a stainless steel delivery wire.
Target Detachable Coils are specifically designed for use with Stryker Neurovascular’s InZone® Detachment System (sold separately).
Target Detachable Coils are compatible with Stryker Neurovascular 2-tip marker microcatheters; refer to Instructions for Use (IFU) for the compatible microcatheter sizes.
# COMPARISON TO THE PREDICATE DEVICE: TARGET DETACHABLE COILS
This 510(k) requests clearance for the revised Instructions for Use to support the use of Target Detachable Coils via transradial access in addition to the existing use via transfemoral access.
A comparison of the subject device with the predicate device is summarized in Table 1 below.
Page 2 of 5
{6}
Stryker Neurovascular
510(k) Summary, K252694
Target Detachable Coils
Table 1. Substantial Equivalence Comparison
| Characteristic | Predicate Device K242243 | Subject Device |
| --- | --- | --- |
| Manufacturer | Stryker | Same |
| Trade Name | Target Detachable Coil | Same |
| Device Type | Vascular Embolization Device; Neurovascular Embolization Device | Same |
| Classification | 870.3300, Class 2 | Same |
| Regulation (21 CFR) | 882.5950, Class 2 | Same |
| Product Code: | KRD, HCG | Same |
| Intended Use/Indication for Use | Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Target Detachable Coils are indicated for endovascular embolization of:
• Intracranial aneurysms
• Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
• Arterial and venous embolization’s in the peripheral vasculature | Same |
| How Supplied | Single Use/Sterile | Same |
| Method of Sterilization | Ethylene Oxide (EtO) Gas | Same |
| Coil Material | Platinum/Tungsten alloy | Same |
| Delivery Wire Material | Stainless steel | Same |
| Features | | |
| Secondary Coil Outer Diameter | 1 mm – 24 mm | Same |
| Coil Length | 1 cm – 50 cm | Same |
| Coil Secondary Shape Types | Helical, 360, 3D, Tetrahedral | Same |
| Power Supply Compatibility | InZone® Detachment System | Same |
| Packaging Configuration and Materials | Pouch: Tyvek/Film pouch
Carton: Chipboard carton
Dispenser coil (Hoop):
HDPE tubing and clips | Same |
Page 3 of 5
{7}
Stryker Neurovascular
510(k) Premarket Notification
Target Detachable Coils
# ACCESSORIES
Target Detachable Coils are not packaged with any accessories.
# INTENDED USE / INDICATIONS FOR USE:
Intended Use:
Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Indications for Use:
Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Target Detachable Coils are indicated for endovascular embolization of:
- Intracranial aneurysms
- Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
- Arterial and venous embolizations in the peripheral vasculature
# BIOCOMPATIBILITY
The biocompatibility profile of the products remains unchanged, and previous biocompatibility assessments remain applicable.
# PERFORMANCE TESTING
## Bench Testing
Non-clinical bench testing was performed to assess the tracking and deployment performance of Target Detachable Coils through radial access, and to compare these results with femoral access using clinically representative access models.
| Parameter | Results |
| --- | --- |
| Microcatheter Friction | No noticeable difference between radial access and femoral access |
| Ease of Deployment | No noticeable difference between radial access and femoral access |
| Ease of Retraction | No noticeable difference between radial access and femoral access |
| Any noticeable difference in device behavior | No noticeable difference between radial access and femoral access |
| Track Force Testing – Advance Force | No statistical difference between radial access and femoral access |
| Track Force Testing – Pullback Force | No statistical difference between radial access and femoral access |
Page 4 of 5
{8}
Stryker Neurovascular
510(k) Premarket Notification
Target Detachable Coils
CONCLUSION
Because the proposed changes to the subject device do not alter the intended use or indications for use of the predicate device, nor its fundamental scientific technology, and the risk assessment of these changes raises no new questions regarding safety and effectiveness, Stryker Neurovascular has determined that the proposed change to the subject device is substantially equivalent to the predicate device.
Page 5 of 5
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.