Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. Target Detachable Coils are indicated for endovascular embolization of: • Intracranial aneurysms • Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae • Arterial and venous embolizations in the peripheral vasculature
Device Story
Target Detachable Coils are platinum-tungsten alloy coils with multi-strand stretch-resistant cores; attached to stainless steel delivery wires. Used for endovascular embolization of intracranial aneurysms and other vascular abnormalities. Operated by physicians in clinical settings; requires use with Stryker Neurovascular InZone Detachment System (sold separately) and compatible microcatheters. Device introduced via microcatheter to occlude blood flow in target vessels. Provides mechanical obstruction to stabilize vascular lesions. Modifications include larger primary coil OD (0.017 inches) and updated introducer sheath dimensions/pigmentation.
Clinical Evidence
Bench testing only. No clinical data presented. Verification testing included dimensional analysis, visual inspection, durability (tensile strength), particulate release, friction testing, biocompatibility (MEM Elution, Hemolysis, USP <661>, FTIR), packaging integrity, and simulated use in tortuous models. All tests met criteria equivalent to predicate devices.
Technological Characteristics
Platinum-tungsten alloy coil; stainless steel delivery wire; high-density polyethylene introducer sheath. Stretch-resistant multi-strand core. Primary coil OD 0.017 inches. Compatible with microcatheters (max ID 0.48 mm/0.019 in). Sterilization process unchanged from predicate. No software or energy source.
Indications for Use
Indicated for endovascular embolization of intracranial aneurysms, neurovascular abnormalities (arteriovenous malformations, arteriovenous fistulae), and peripheral arterial/venous embolizations.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other to create a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 22, 2016
Stryker Neurovascular Ms. Kate Taylor Staff Regulatory Affairs Specialist 47900 Bayside Parkway Fremont, California 94538
Re: K161429
Trade/Device Name: Target Detachable Coils Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: May 20, 2016 Received: May 23, 2016
Dear Ms. Kate Taylor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161429
Device Name Target Detachable Coils
#### Indications for Use (Describe)
Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Target Detachable Coils are indicated for endovascular embolization of:
- · Intracranial aneurysms
- · Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
- · Arterial and venous embolizations in the peripheral vasculature
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is in all lowercase letters, and there is a registered trademark symbol to the right of the "r". The text is black and the background is white.
Neurovascular
# 510(k) Summary
| Summary Date: | May 20, 2016 |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name and Address: | Stryker Neurovascular<br>47900 Bayside Parkway<br>Fremont, CA. 94538 |
| Contact Person: | Kate Taylor<br>Staff Regulatory Affairs Specialist<br>Phone: 510 413-2175<br>Email: kate.taylor@stryker.com |
| Trade Name: | Target® Detachable Coils |
| Common Name: | Occlusion Coil, Vascular Occlusion Coil, Neurovascular Occlusion Coil |
| Classification Name: | Target Detachable Coils are classed as vascular and neurovascular<br>embolization devices under 21 CFR 870.3300 (KRD) and 21 CFR 882.5950<br>(HCG), respectively, and are Class II devices (special controls). |
The special control for the devices is FDA's guidance document, Class II
Special Controls Guidance Document: Vascular and Neurovascular
Embolization Devices (issued 29 Dec 2004).
| Legally Marketed<br>Predicate Devices: | Primary Predicate<br>K153658 Target Detachable Coils (cleared 19 May 2016)<br>Reference Predicates<br>K123377 Target Detachable Coils (cleared 30 November 2012) |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | K093142 Target Detachable Coils (cleared 04 Feb 2010)<br>K102672 Target Detachable Coils (cleared 15 October 2010)<br>K112385 Target Detachable Coils (cleared 15 September 2011)<br>K113412 Target Detachable Coils (cleared 13 December 2012) |
| Device Description: | Stryker Neurovascular <b>Target Detachable Coils</b> are comprised of the following coil types:<br>Target 360 Nano |
{4}------------------------------------------------
| Device Description:<br>(continued) | All Target Coils are stretch resistant coils. Target Coils incorporate a length<br>of multi-strand material through the center of the coil designed to help resist<br>stretching. Target Coils are designed for use with the Stryker Neurovascular<br>InZone® Detachment System (sold separately). |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Each Target Coil type consists of a platinum-tungsten alloy coil attached to<br>a stainless steel delivery wire. For Target Coils in the 360-shape, the distal<br>end of the main coil is formed such that there is a smaller distal loop at the<br>end of the main coil to facilitate placement of the coil. The diameter of the<br>distal loop is 75% that of the rest of the main coil loops. |
| | The Stryker Neurovascular InZone Detachment System is intended for use<br>with all Stryker Neurovascular Detachable Coils in the embolization of<br>intracranial aneurysms and other vascular malformations of the neuro and<br>peripheral vasculature. |
| Comparison to the<br>Predicate Device: | Target Detachable Coils<br>This 510(k) requests clearance of 8 new Target Detachable Coil UPNs –<br>Target XXL 360 Detachable Coils. The change which differentiates the<br>Target XXL 360 Coils from the rest of the Target Detachable Coil product<br>family is a new, larger primary coil OD (0.017 inches), as compared to<br>current range of Target Coils primary coil ODs (0.010 - 0.014 inches).<br>The larger primary coil OD of the Target XXL 360 Detachable Coils<br>necessitates minor dimensional increase to the ID of the introducer sheath,<br>which is a component of the packaging hoop and facilitates introduction<br>into the microcatheter. The new introducer sheath is made of the same high<br>density polyethylene material as the current introducer sheath, but uses a<br>different pigment to color the sheath to provide differentiation from other<br>Target Coils. The larger primary OD requires that the Target XXL Coil be<br>used with the larger of the microcatheters recommended for all other Target<br>Coils: maximum internal diameter 0.48 mm [0.019 in].<br>The new UPNs utilize the same design and materials of existing Target<br>Coils, and the same manufacturing, packaging and sterilization processes.<br>The modified Stryker Neurovascular Target Detachable Coils have the same<br>intended use/indications for use as the predicate Target<br>Detachable Coils.<br>The modifications to the Target Detachable Coils do not alter the intended<br>use, indications for use, or the fundamental scientific technology of the<br>predicate devices.<br>Risk assessment of the modifications in the form of design and use failure<br>modes and effects analysis (design and use FMEAs) has been conducted in<br>accordance with EN ISO 14971:2012. Stryker Neurovascular has<br>determined the modifications to the predicate devices raise no new<br>questions of safety or effectiveness. |
| | Verification testing has demonstrated the modified Target Detachable Coils<br>are substantially equivalent to the predicate Target Detachable Coils. |
{5}------------------------------------------------
### Verification Testing:
Accessories:
Verification testing of the modified Target Detachable Coil consisted of the following:
| Test | Test Method Summary/Purpose | Results |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Dimensions | Coil Secondary Diameter and 2D<br>Loop Secondary Diameter are<br>measured. | Met the same criteria as<br>the predicate device. |
| Visual | Visual confirmation that the entire coil<br>is contained within the introducer<br>sheath. | Met the same criteria as<br>the predicate device. |
| Durability | The coil is visually inspected for<br>damage and Main Junction Tensile<br>Strength is tested after simulated<br>deployment/retraction in a tortuous<br>model. | Met the same criteria as<br>the predicate device. |
| Particulates | Particulate release due to delivery of<br>the coil is measured. | Met the same criteria as<br>the predicate device. |
| Friction | Frictional force through an introducer<br>sheath and a compatible<br>microcatheter is measured. | Met the same criteria as<br>the predicate device. |
| Biocompatibility | • MEM Elution Cytotoxicity/Part 5<br>• Hemolysis Direct/Extract<br>Contact/Part 4<br>• USP Physicochemical USP <661><br>• FTIR/Part 18<br>• Natural Rubber Latex | Met the same criteria as<br>the predicate device. |
| Packaging | Assess the ability of the packaging<br>system to protect the finished device. | Met the same criteria as<br>the predicate device. |
| Simulated Use | • Coil conformability<br>• Coil durability during repositioning<br>• Microcatheter compatibility<br>• Coil friction during delivery through<br>the microcatheter<br>• Introducer sheath friction<br>acceptability<br>• Removal of coil from the packaging<br>hoop without damaging the device | Met the same criteria as<br>the predicate device. |
Target Detachable Coils are not packaged with any accessories.
| Intended Use /<br>Indications for Use | Intended Use/Indications for use |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.<br><br>Target Detachable Coils are indicated for endovascular embolization of:<br>• Intracranial aneurysms<br>• Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae<br>• Arterial and venous embolizations in the peripheral vasculature |
| Conclusion: | Because the subject modifications do not alter the intended use or indications for use of the predicate devices, or the fundamental scientific technology of the predicate devices; and because risk assessment of the modifications and successful verification testing raise no new questions of safety and effectiveness, Stryker Neurovascular has determined the modified Target Detachable Coils to be substantially equivalent to the predicate devices. |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.