MallyaG Injection Pen Adapter (MallyaG)

K251500 · Biocorp Production · QOG · Jun 9, 2025 · General Hospital

Device Facts

Record IDK251500
Device NameMallyaG Injection Pen Adapter (MallyaG)
ApplicantBiocorp Production
Product CodeQOG · General Hospital
Decision DateJun 9, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5860
Device ClassClass 2

Indications for Use

The MallyaG Injection Pen Adapter is indicated for the capture and wireless transmission of dosing information of compatible reusable and disposable pen injectors. The Novo Nordisk PDS290 Injection pens compatible for growth disorder treatment are: - Sogroya (somapacitan-beco) - Norditropin FlexPro

Device Story

MallyaG is a smart sensor adapter that attaches to the button of compatible Novo Nordisk PDS290 pen injectors. It monitors the rotation of the pen's dosing mechanism to detect and record the dose increment dialed by the user. The device differentiates between priming and actual dosing. Data, including dose amount, time, and date, are transmitted via Bluetooth Low Energy (BLE) to a mobile application. Used by patients or caregivers in home or clinical settings. The output allows users to track injection history, supporting adherence and clinical decision-making. The device is rechargeable and intended for single-patient, multi-use over a 3-year lifespan.

Clinical Evidence

Bench testing only. Performance verified per ISO 11608-1; 99% of recorded doses match dialed doses within +/- 1 increment. Biocompatibility assessed per ISO 10993-1. Electrical safety and EMC tested per IEC 60601-1 and IEC 60601-1-11. Human factors validation performed to confirm usability.

Technological Characteristics

Smart sensor adapter; 25.8 mm diameter, 40.1 mm height. Sensing via mechanical rotation tracking of pen dosing button. Connectivity: Bluetooth Low Energy (BLE). Power: Rechargeable, non-replaceable battery. Biocompatibility: Permanent contact with intact skin (ISO 10993-1). Electrical safety: IEC 60601-1, IEC 60601-1-11. Software: Embedded firmware for dose calculation and transmission.

Indications for Use

Indicated for capture and wireless transmission of dosing data from compatible Novo Nordisk PDS290 pen injectors (Sogroya, Norditropin FlexPro) used for growth disorder treatment. Intended for use by patients or caregivers.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION June 9, 2025 Biocorp Production Alexia Garin QARA Director Zi Lavaur -La Béchade Issoire, 63500 France Re: K251500 Trade/Device Name: Mallya® Injection Pen Adapter (MallyaG) Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QOG Dated: May 14, 2025 Received: May 15, 2025 Dear Alexia Garin: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {2} K251500 - Alexia Garin Page 2 Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {3} K251500 - Alexia Garin Page 3 Sincerely, # Shruti N. Mistry -S Shruti Mistry Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251500 | | | Device Name MallyaG Injection Pen Adapter (MallyaG) | | | Indications for Use (Describe) The MallyaG Injection Pen Adapter is indicated for the capture and wireless transmission of dosing information of compatible reusable and disposable pen injectors. The Novo Nordisk PDS290 Injection pens compatible for growth disorder treatment are: - Sogroya (somapacitan-beco) - Norditropin FlexPro | | | Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {5} BIOCORP a novo nordisk company MALLYA® 510(k) Summary - K251500 # 1 SUBMITTER Applicant: Biocorp Production ZI Lavaur La Béchade 63500 Issoire – France Prepared By: Alexia Garin Date Prepared: June 09, 2025 # 2 DEVICE Trade/Proprietary Name: Mallya® Injection Pen Adapter (Mallya®) Common/Usual Name: Injection Data Capture Device Classification Name: 21 CFR §880.5860 Piston Syringe Product Code: QOG Class: II 510(k) number: K251500 # 3 PREDICATE DEVICE K250555, Biocorp Production, Mallya® Injection Pen Adapter (Mallya®). # 4 SUBSTANTIAL EQUIVALENCE The Mallya® Injection Pen Adapter device is substantially equivalent to the Mallya® (K250555). The device has the same indication for use and same scientific technology. # 4.1 DEVICE DESCRIPTION The Mallya® Injection Pen Adapter is a smart sensor composed of a button to be assembled onto a pen injector by covering the injection pen button. It is designed to be mounted on specific Novo Nordisk® PDS290 platform injection pens. A USB cable, necessary to charge the Mallya® device, is also provided in the package. # 4.2 TECHNOLOGICAL CHARACTERISTICS The Mallya® has the same technological characteristics as the predicate device to capture and transmit dosing information. The information is captured through the rotation of the dosing button of the pen into a value indicating the dose increment injected. Both the Mallya® and the predicate device use same Low Energy Bluetooth (BLE) communication technology and same protocol to pair and transmit the information to a mobile device with a connectable app. BIOCORP Page 1 of 4 {6} BIOCORP a novo nordisk company MALLYA # 4.3 INDICATIONS FOR USE The MallyaG Injection Pen Adapter is indicated for the capture and wireless transmission of dosing information of compatible reusable and disposable pen injectors. The Novo Nordisk PDS290 Injection pens compatible for growth disorder treatment are: - Sogroya (somapacitan-beco) - Norditropin FlexPro # 4.4 DEVICE PHYSICAL DESCRIPTION The device is made of a cap to be assembled onto a pen injector by covering the injection pen button. The overall device size is 25.8 mm diameter and 40.1 mm height. A USB cable, necessary to charge the Mallya® device, is also provided in the package. # 4.5 SUBSTANTIAL EQUIVALENCE DISCUSSION | Attribute | Subject Device | Predicate device | | --- | --- | --- | | Device Name/Brandname | Mallya Injection Pen Adapter Mallya^{G} | Mallya Injection Pen Adapter Mallya^{D} | | | Data Capture and Transmission Technology | | 510(k) Number | N/A | K250555 | | Attribute | Subject Device | Predicate device | | Device Classification-Product Code | Class 2 – QOG | Class 2 – QOG | | Indication for Use | The Mallya® Injection Pen Adapter is indicated for the capture and wireless transmission of dosing information of compatible reusable and disposable pen injectors. The Novo Nordisk PDS290 Injection pens compatible for growth disorder treatment are: - Sogroya (somapacitan-beco) - Norditropin FlexPro | The Mallya® Injection Pen Adapter is indicated for the capture and wireless transmission of dosing information of compatible reusable and disposable pen injectors. The Novo Nordisk PDS290 Injection pens compatible for diabetes treatment are: - insulin degludec molecule (Tresiba U-100 FlexTouch and Tresiba U-200 FlexTouch) - insulin aspart molecule (Fiasp FlexTouch) - insulin degludec and liraglutide molecules (Xultophy) | | Same indication for use baseline | | | | Same pen platform compatibility (Novo Nordisk PDS290) | BIOCORP Page 2 of 4 {7} BIOCORP a novo nordisk company MALLYA | Attribute | Subject Device | Predicate device | | | --- | --- | --- | --- | | Single patient Use | Yes | Yes | Same | | Reusable Device | Yes | Yes | Same | | Intended Use | Intended to be used by patients in the same use environment as their compatible injection pen. | Intended to be used by patients in the same use environment as their compatible injection pen. | Same | | Prescription use | No | No | Same | | User Group | User of Compatible Injection Pens; includes patients or patient relatives/caregiver | User of Compatible Injection Pens; includes patients or patient relatives/caregiver | Same | | User Feedback | Electronic – LED (light) | Electronic – LED (light) | Same light indication | | Wireless Connectivity | Bluetooth Low Energy (BLE) | Bluetooth Low Energy (BLE) | Same connectivity BLE protocol | | Control or impact Drug delivery | No | No | Same | | Fluid Pathway Contact | None | None | Same | | Software controlled | Yes | Yes | Same | | Dose Recorded | Calculated based on dose increment set | Calculated based on dose increment set | Same | | Information Transmitted | Dialed dose increment, and time and date of injection | Dialed dose increment, and time and date of injection | | | Mechanism for Recording dose dialed | Sense dose increment dialed through movement and rotation of dosing mechanism during dose setting | Sense dose increment dialed through movement and rotation of dosing mechanism during dose setting | | | Differentiates Prime vs. Dose | Yes | Yes | Same | | Battery | Non-Replaceable; Rechargeable | Non-Replaceable; Rechargeable | Same | | Lifetime | Up to 3 years of use from manufacturing date | Up to 3 years of use from manufacturing date | Same | # 5 SIGNIFICANT PERFORMANCE CHARACTERISTICS ## 5.1 PERFORMANCE CLAIMS Mallya® claims are based on the reported dose accuracy: - Device accuracy: 99% of the recorded doses match the dialed doses with a margin of error of +/- 1 increment. BIOCORP Page 3 of 4 {8} BIOCORP a novo nordisk company MALLYA ## 5.2 SUPPORTING PERFORMANCE DATA Data from the following testing identified, as appropriate to assess the impact of the modifications, were generated verifying the design of the modified device: - Bench testing on performance, using ISO 11608-1 as a guide. - Dose Prime Differentiation. - Biocompatibility – FDA guidance; permanent contact with intact skin based on ISO10993-1. - Electrical safety, EMC and Radiocommunication – per FDA Guidelines based on IEC 60601-1 and appropriate collateral requirements from IEC 60601-2 and IEC 60601-1-11. - Cybersecurity Testing and Software Verification and Validation per FDA Guidelines. ## 5.3 HUMAN FACTOR VALIDATION Biocorp performed a human factor impact assessment for performance and safety validation to confirm the usability of the device and associated instructions for use per FDA human factor guidance. ## 6 CONCLUSION Biocorp as concluded that the evidence documented during design control activities supports that the Mallya® Injection Pen Adapter (Mallya®), is substantially equivalent to the predicate for specific indications and technological characteristics of the predicate device with regards to the data recording and transmission. ## 7 CONTACT PERSON Alexia Garin, pms@biocorp.fr +33 473 557050 BIOCORP Page 4 of 4
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...