K250555 · Biocorp Production · QOG · Mar 27, 2025 · General Hospital
Device Facts
Record ID
K250555
Device Name
MallyaD injection pen adapter (MallyaD)
Applicant
Biocorp Production
Product Code
QOG · General Hospital
Decision Date
Mar 27, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The MallyaD Injection Pen Adapter is indicated for the capture and wireless transmission of compatible reusable and disposable pen injectors. The Novo Nordisk PDS290 Injection pens compatible for diabetes treatment are: - insulin degludec molecule (Tresiba U-100 FlexTouch and Tresiba U-200 FlexTouch) - insulin aspart molecule (Fiasp FlexTouch) - insulin degludec and liraglutide molecules (Xultophy)
Device Story
MallyaD is a smart sensor adapter that attaches to the button of compatible Novo Nordisk PDS290 injection pens. It captures dosing information by sensing the rotation of the pen's dosing mechanism during dose setting. The device transmits the dialed dose increment, date, and time of injection via Bluetooth Low Energy (BLE) to a mobile application. It is intended for use by patients or caregivers in home or clinical environments. The device provides visual feedback via an LED. It does not control or impact drug delivery. By automating the logging of injection data, it assists patients in tracking their diabetes treatment adherence. The device is rechargeable via USB and has a 3-year lifetime.
Clinical Evidence
Bench testing only. Performance verified per ISO 11608-1 guidelines. Testing confirmed 99% dose recording accuracy (within +/- 1 increment). Additional testing included biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601-1/2/11), and cybersecurity/software validation. Human factors validation confirmed usability of the device form factor.
Technological Characteristics
Smart sensor adapter; 25.8mm diameter, 40.1mm height. Sensing via mechanical rotation of pen dosing button. Connectivity: Bluetooth Low Energy (BLE). Power: Non-replaceable rechargeable battery. Biocompatibility: ISO 10993-1 (permanent contact with intact skin). Standards: ISO 11608-1 (dose accuracy), IEC 60601-1/2/11 (electrical/EMC).
Indications for Use
Indicated for capture and wireless transmission of dosing information from compatible reusable and disposable pen injectors for patients with diabetes using Novo Nordisk PDS290 platform pens (Tresiba U-100/U-200 FlexTouch, Fiasp FlexTouch, Xultophy).
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{1}------------------------------------------------
March 27, 2025
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
BioCorp Production Alexia Garin QARA Director ZI de Lavaur La Bechade Issoire, 63500 France
Re: K250555
Trade/Device Name: MallyaD injection pen adapter (MallyaD) Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QOG Dated: February 18, 2025 Received: February 25, 2025
Dear Alexia Garin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}------------------------------------------------
Sincerely,
# Shruti N. Mistry -S
Shruti Mistry Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K250555
Device Name MallyaD injection pen adapter (MallyaD)
Indications for Use (Describe)
The MallyaD Injection Pen Adapter is indicated for the capture and wireless transmission of compatible reusable and disposable pen injectors.
The Novo Nordisk PDS290 Injection pens compatible for diabetes treatment are:
- insulin degludec molecule (Tresiba U-100 FlexTouch and Tresiba U-200 FlexTouch)
- insulin aspart molecule (Fiasp FlexTouch)
- insulin degludec and liraglutide molecules (Xultophy)
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Biocorp, a Novo Nordisk company. The logo is in a dark gray color and features the word "BIOCORP" in a bold, sans-serif font. The two O's in the word are interlocked, creating a visual link. Below the word "BIOCORP" is the text "a novo nordisk company" in a smaller font size.
MALLYAP
## 510(k) Summary - K250555
| 1 SUBMITTER | |
|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Biocorp Production<br>ZI Lavaur La Béchade<br>63500 Issoire - France |
| Prepared By:<br>Date Prepared: | Alexia Garin<br>March 26, 2025 |
| 2 DEVICE | |
| Trade/Proprietary Name:<br>Common/Usual Name:<br>Classification Name:<br>Product Code:<br>Class: | MallyaD Injection Pen Adapter (MallyaD)<br>Injection Data Capture Device<br>21 CFR §880.5860 Piston Syringe<br>QOG<br>II |
### 3 PREDICATE DEVICE
K231820, Biocorp Production, SoloSmart Injection Pen Adapter (SoloSmart®).
### 4 SUBSTANTIAL EQUIVALENCE
The Mallyaº Injection Pen Adapter device is substantially equivalent to the SoloSmart® (K231820). The device has the same indication for use, intended use and same fundamental scientific technology.
## 4.1 DEVICE DESCRIPTION
The concerned Mallya Injection Pen Adapter, also called Mallyaº, is a smart sensor composed of a button to be assembled onto a pen injector by covering the injection pen button. It is designed to be mounted on specific Novo Nordisk® PDS290 platform injection pens. A USB cable, necessary to charge the Mallya® device is also provided in the package.
## 4.2 TECHNOLOGICAL CHARACTERISTICS
The Mallya® has the same technological characteristics as the predicate device to capture and transmit dosing information. The information is captured through the rotation of the dosing button of the pen into a value indicating the dose increment injected. Both the Mallyaº and the predicate device use same Low Energy Bluetooth (BLE) communication technology and same protocol to pair and transmit the information to a mobile device with a compatible app.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for BICORP, a Novo Nordisk company. The logo is in a dark gray color and features the word "BICORP" in a bold, sans-serif font. The two O's in the word are interlocked. Below the word "BICORP" is the text "a novo nordisk company" in a smaller, lighter font.
## 4.3 INDICATIONS FOR USE
The Mallyaº is indicated for the capture and wireless transmission of dosing information of compatible reusable and disposable pen injectors.
The Novo Nordisk PDS290 Injection pens compatible for diabetes treatment are:
- insulin degludec molecule (Tresiba U-100 FlexTouch and Tresiba U-200 FlexTouch)
- insulin aspart molecule (Fiasp FlexTouch)
- insulin degludec and liraglutide molecules (Xultophy)
## 4.4 DEVICE PHYSICAL DESCRIPTION
The device is made of a cap to be assembled onto a pen injector by covering the injection pen button. The overall device size is 25.8 mm diameter and 40.1 mm height. A USB cable, necessary to charge the Mallya® device is also provided in the package.
| Attribute | Subject Device | Predicate device | | | | | | | | | |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name/<br>Brandname | MallyaD Injection Pen Adapter<br>MallyaD | SoloSmart Injection Pen Adapter<br>SoloSmart | | | | | | | | | |
| | | Data Capture and Transmission Technology | | | | | | | | | |
| 510(k) Number | K250555 | K231820 | | | | | | | | | |
| Attribute | Subject Device | Predicate device | Discussion/<br>Comments | | | | | | | | |
| Classification-<br>Product Code | Class 2 - QOG | Class 2 - QOG | Same | | | | | | | | |
| Indication for<br>Use | The MallyaD Injection Pen Adapter is<br>indicated for the capture and wireless<br>transmission of dosing information of<br>compatible reusable and disposable<br>pen injectors.<br><br>The Novo Nordisk PDS290 Injection<br>pens compatible for diabetes<br>treatment are:<br>- insulin degludec molecule (Tresiba<br>U-100 FlexTouch and Tresiba U-200<br>FlexTouch)<br>- insulin aspart molecule (Fiasp<br>FlexTouch)<br>- insulin degludec and liraglutide<br>molecules (Xultophy) | The SoloSmart Injection Pen Adapter is<br>indicated for the capture and wireless<br>transmission of dosing information<br>from compatible reusable and<br>disposable pen injectors.<br><br>The following Injection pens are<br>compatible:<br>DEVICE MODEL<br>NAME INSULIN<br>BRAND NAME MOLECULE NAME MOLECULE<br>CONCENTRATION SoloSmart®<br>designed for<br>SoloStar®<br>SANOFI injection<br>pen Lantus<br>Toujeo<br>Admelog<br>Apidra<br>Soliqua 100/33 GLARGINE<br>GLARGINE<br>LISPRO<br>GLULISINE<br>GLARGINE AND<br>LIXISENATIDE 100 IU/mL<br>300 IU/mL<br>100 IU/mL<br>100 IU/mL<br>100 IU/mL<br>+33 mcg/mL | | | | | | | | | Similar indication<br>for use,<br>compatibility with<br>different<br>manufacturer<br>injection pen for<br>same therapeutic<br>area. Verification<br>tests conducted<br>with compatible<br>pen. |
| Type of pens | The following Novo Nordisk PDS290<br>pen platform are compatible:<br>Tresiba® U-100 FlexTouch® and<br>Tresiba® U-200 FlexTouch®, Fiasp®<br>FlexTouch®,Xultophy® | The following Sanofi SoloStar® Injection<br>pens are compatible: Lantus<br>Toujeo Toujeo Max Aprida<br>Soliqua 100/33 | Compatible with a<br>different pen<br>platform: Novo<br>Nordisk PDS290.<br>Verification tests<br>conducted with the<br>compatible pen. | | | | | | | | |
## 4.5 SUBSTANTIAL EQUIVALENCE DISCUSSION
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Biocorp, a Novo Nordisk company. The logo is in a dark gray color and features the word "BIOCORP" in a bold, sans-serif font. The two O's in the word are linked together, creating a visual element that resembles a chain or connection. Below the word "BIOCORP" is the text "a novo nordisk company" in a smaller font.
MALLYAº
| Attribute | Subject Device | Predicate device | |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Single patient<br>Use | Yes | Yes | Same |
| Reusable<br>Device | Yes | Yes | Same |
| Intended Use | Intended to be used by patients in the<br>same use environment as their<br>compatible injection pen. | Intended to be used by patients in the<br>same use environment as their<br>compatible injection pen. | Same |
| Prescription<br>use | No | No | Same |
| User Group | User of Compatible Injection Pens;<br>includes patients or patient<br>relatives/caregiver | User of Compatible Injection Pens;<br>includes patients or patient<br>relatives/caregiver | |
| User Feedback | Electronic - LED (light) | Electronic - LED (light) | Same type of<br>coloured light<br>indication adapted<br>to pen mechanical<br>behaviour.<br>Software<br>verification tests<br>conducted with the<br>compatible pen. |
| Wireless<br>Connectivity | Bluetooth Low Energy (BLE) | Bluetooth Low Energy (BLE) | Same connectivity<br>BLE protocol |
| Control or<br>impact Drug<br>delivery | No | No | Same |
| Fluid Pathway<br>Contact | None | None | Same |
| Software<br>controlled | Yes | Yes | Same |
| Dose Recorded | Calculated based on dose increment<br>set | Calculated based on dose delivered | Change from<br>predicate adapted<br>to pen mechanical<br>behaviour. |
| Information<br>Transmitted | Dialed dose increment, and time and<br>date of injection | Delivered dose, and time and date of<br>injection | Dose recording<br>accuracy<br>verification tests<br>conducted with the<br>compatible pen. |
| Mechanism for<br>Recording dose<br>dialed | Sense dose increment dialed through<br>movement and rotation of dosing<br>mechanism during dose setting | Sense dose increment delivered<br>through rotation of dosing mechanism<br>during dose administration | |
| Differentiates<br>Prime vs. Dose | Yes | Yes | Same |
| Battery | Non-Replaceable; Rechargeable | Non-Replaceable; Rechargeable | Same |
| Lifetime | Up to 3 years of use from<br>manufacturing date | Up to 3 years of use from<br>manufacturing date | Same |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for BICORP, a Novo Nordisk company. The logo is in a dark gray color and features the word "BICORP" in a bold, sans-serif font. The two O's in the word are designed to look like interlocking links of a chain. Below the word "BICORP" is the text "a novo nordisk company" in a smaller, sans-serif font.
.
MALLYAP
## 5 SIGNIFICANT PERFORMANCE CHARACTERISTICS
### 5.1 PERFORMANCE CLAIMS
Mallyaº claims are based on the reported dose accuracy:
- Device accuracy: 99% of the recorded doses match the dialed doses with a margin of error of +/- 1 increment.
### 5.2 SUPPORTING PERFORMANCE DATA
Data on the following testing were generated verifying the design of the device:
- Bench testing on performance, using ISO 11608-1 as a guide.
- Dose accuracy of pen is not affected by Mallya® ●
- . Recording and transmission Accuracy
- . Dose Prime Differentiation
- . Complete Dose Notification
- Incomplete Injection Identification
- Physical Interaction with pen, including force to mount, maintain and remove Mallya®
- Biocompatibility –FDA guidance; permanent contact with intact skin based on ● ISO10993-1.
- Lifetime The product met test criteria to verify it will operate/function before the ● expiration date of 3 years including storage and Use Lifetime.
- Electrical safety, EMC and Radiocommunication per FDA Guidelines based on ● IEC 60601-1 and appropriate collateral requirements from IEC 60601-2 and IEC 60601-1-11
- Cybersecurity Testing and Software Verification and Validation per FDA ● Guidelines
### 5.3 HUMAN FACTOR VALIDATION
Biocorp performed a HF impact assessment for performance and safety validation to confirm the usability of the device form factor and associated instructions for use per FDA guidance.
#### 6 CONCLUSION
Biocorp as concluded that the evidence documented during design control activities supports that the Mallya® Injection Pen Adapter (Mallyaº), is substantially equivalent to the predicate for specific indications and technological characteristics of the predicate device with regards to the data recording and transmission.
#### 7 CONTACT PERSON
Alexia Garin, pms@biocorp.fr +33 473 557050
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.