K5 is intended to supply power to and serve as a base for dental devices and accessories. This product is intended for use by dentist and dental assistance to assist dental treatment of patients for adjusting the position of the patient seat, operating the instruments, and supplying water into the oral cavity through a water line in dental clinical environments.
Device Story
AC-powered dental operative unit; serves as base for dental instruments and accessories. Components: foot controller, headrest, backrest, seat, armrest, doctor table, unit, dental light, assistant table, doctor stool. Operation: electrically controlled chair movements; pneumatically controlled air/water supply for handpieces and 3-way syringes via foot pedal. Used in dental clinics by trained personnel. Provides power, water, and positioning support to facilitate dental procedures. Benefits: ergonomic patient positioning and integrated instrument control.
Clinical Evidence
No clinical data. Substantial equivalence established through bench testing, electrical safety, electromagnetic compatibility, and software verification/validation.
Technological Characteristics
AC-powered (100-240V). Materials: Polyvinyl Chloride (seat), Polyurethane (water/air lines), A16061 (suction/syringe), Stainless Steel 304 (warmer), C3604 (water block). Pneumatic control for air/water. Heating coil warmer (25-37°C). Complies with IEC 60601-1, IEC 60601-1-2, IEC 80601-2-60, IEC 60601-1-6, ISO 7494-1, ISO 7494-2, and ISO 16954 (waterline cleaning).
Indications for Use
Indicated for use by dentists and dental assistants to assist in dental treatment of patients in clinical environments, including patient seat positioning, instrument operation, and oral cavity water supply.
Regulatory Classification
Identification
A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 17, 2025
Osstem Implant Company., Ltd. Chair Business
% Leszczak Mateusz
Regulatory Affairs Manager
Hiossen Inc.
85 Ben Fairless Dr.
Fairless Hills, Pennsylvania 19030
Re: K251491
Trade/Device Name: K5 Cart, K5 Mount, K5 Swing
Regulation Number: 21 CFR 872.6640
Regulation Name: Dental Operative Unit And Accessories
Regulatory Class: Class I, reserved
Product Code: EIA, KLC
Dated: June 20, 2025
Received: June 20, 2025
Dear Leszczak Mateusz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251491 - Leszczak Mateusz
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251491 - Leszczak Mateusz
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
K5 Cart
Page 9 of 29
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251491 | ? |
| Please provide the device trade name(s). | | ? |
| K5 Cart, K5 Mount, K5 Swing | | |
| Please provide your Indications for Use below. | | ? |
| K5 is intended to supply power to and serve as a base for dental devices and accessories. This product is intended for use by dentist and dental assistance to assist dental treatment of patients for adjusting the position of the patient seat, operating the instruments, and supplying water into the oral cavity through a water line in dental clinical environments. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
OSSTEM
Osstem Implant Co., Ltd. Chair Business
192, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do, 15428, Republic of Korea
Tel: +82 2 2016 7000 Fax: +82 2 2016 7001 www.osstem.com
# 510(k) Summary
# K251491
Date: July 16, 2025
## 1. Company and Correspondent making the submission
- Submitter's Name : Osstem Implant Co., Ltd. Chair Business
- Address : 192, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do, 15428, Republic of Korea
- Contact : Ms. Jimin Hyun
- Phone : +82-70-4871-0191
- Correspondent's Name : Hiossen Inc.
- Address : 85 Ben Fairless Dr. Fairless Hills, PA 19030
- Contact : Mr. Mateusz Leszczak
- Phone : +1-201-266-0657
## 2. Proposed Device
- Trade or (Proprietary) Name : K5 Cart, K5 Mount, K5 Swing
- Classification Name : Dental operative unit and accessories
- Regulation Number : 21 CFR 872.6640
- Device Classification : Class I
- Classification Product Code : EIA, KLC
## 3. Predicated Device
- K5(K233805)
## 4. Description
The K5 Cart, K5 Mount and K5 Swing are designed for dental treatment and are intended for dental use only and are intended for use by trained medical personnel only.
These devices consist of a foot controller, headrest, backrest, seat, armrest, doctor table, unit, dental light, auxiliary table and doctor stool.
These are an AC-powered dental operative unit with accessories, intended to supply power to and serve as a base for other dental devices. It includes a treatment chair, dentist element, assistant element and a dental light as offering several additional options and electronically-controlled chair movements with software and water unit functions.
{5}
OSSTEM
Osstem Implant Co., Ltd. Chair Business
192, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do, 15428, Republic of Korea
Tel: +82 2 2016 7000 Fax: +82 2 2016 7001 www.osstem.com
# 5. Indication for use
K5 is intended to supply power to and serve as a base for dental devices and accessories. This product is intended for use by dentist and dental assistance to assist dental treatment of patients for adjusting the position of the patient seat, operating the instruments, and supplying water into the oral cavity through a water line in dental clinical environments.
# 6. Comparison of the Indications for Use
Predicated Device: K5 is intended to supply power to and serve as a base for dental devices and accessories.
This product includes a dental chair. The dental treatment unit is intended for use in the dental clinic environment and is used by trained dentists and/or dental assistants.
The intended use of the predicated device and the subject device are the same, but more detailed information has been added on how the subject device is used to assist in dental treatment, so there is no effect on substantial equivalence.
# 7. Summary of Technological Characteristics with predicated device and proposed device
The K5 Cart, K5 Mount, K5 Swing functions in a manner similar to and is intended for the same use to the predicate device. Primarily, K5 Cart, K5 Mount, K5 Swing are substantially equivalent to the K5 (K233805) marketed by OSSTEM IMPLANT Co., Ltd. Chair Business and have some different technological characteristics and slightly different external design. However, these differences do not raise new concerns of substantial equivalence as the performance data and testing of the K5 Cart, K5 Mount, K5 Swing demonstrate that the devices are deemed to be substantially equivalent as described in a following comparison table:
| Description | Proposed Devices | Predicate Device | Remark |
| --- | --- | --- | --- |
| Indications for Use | The K5 is intended to supply power to and serve as a base for dental devices and accessories. This product is intended for use by dentist and dental assistance to assist dental treatment of patients for adjusting the position of the patient seat, operating the instruments, and supplying water into the oral cavity through a water line in dental clinical environments. | K5 is intended to supply power to and serve as a base for dental devices and accessories. This product includes a dental chair. The dental treatment unit is intended for use in the dental clinic environment and is used by trained dentists and/or dental assistants. | Different |
| Product Name | K5 Cart, K5 Mount, K5 Swing | K5 | Same |
| 510(k) No. | Proposed | K233805 | - |
| Manufacturer | Osstem Implant Co., Ltd. Chair business | Osstem Implant Co., Ltd. Chair business | Same |
| Product Code | EIA, KLC | EIA, KLC | Same |
| Model Type | Mount type, Cart type, Swing type | Mount type, Cart type | Different |
| Unit Type | Single type | Single type | Same |
| Power & Utility Supply | AC 100-120/220-240V, 50/60Hz, compressed air and water | AC 100-120/220-240V, 50/60Hz, compressed air and water | Same |
{6}
OSSTEM
Osstem Implant Co., Ltd. Chair Business
192, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do, 15428, Republic of Korea
Tel: +82 2 2016 7000 Fax: +82 2 2016 7001 www.osstem.com
| Doctor Table | - 3-way syringe
- Control of water supply, scaler vibration power, table height, patient position, System Power, Film viewer, Patient Chair Positioning, Light On/Off, Handpiece function, Timer, Mode Selection, LED Display | - : 3-way syringe
- Control of water supply, scaler vibration power, table height, patient position, System Power, Film viewer, Patient Chair Positioning, Light On/Off, Handpiece function, Timer, Mode Selection, LED Display | Same |
| --- | --- | --- | --- |
| Assist Table | - 3-way syringe, HVE and SE suction
- Patient chair positioning, Dental Light on/off, Control of water | - 3-way syringe, HVE and SE suction
- Patient chair positioning, Dental Light on/off, Control of water | Same |
| Main Components | Foot controller, Headrest, Backrest, Seat, Armrest, Doctor Table, Unit, Dental Light, Assistant Table, Doctor Stool, Monitor Arm, Hanaro Console, Joystick Control | Chair, Unit, Table, Seat, Stool, Monitor Arm, Hanaro Console | Same |
| Syringe | 3-way syringe | 3-way syringe | Same |
| Control of water and air | Uses pneumatically controlled vales to water control the flow of air and water. On/off and intensity controlled by foot pedal. | Uses pneumatically controlled vales to water control the flow of air and water. On/off and intensity controlled by foot pedal. | Same |
| Air Pressure | 550(min), 600kPa(max) | 550(min), 750kPa(max) | Different |
| Water Pressure | 250(min), 600kPa(max) | 250(min), 600kPa(max) | Same |
| Water System | City water supply | City water supply | Same |
| Water Sanitation System | Distilled water container added | Distilled water container added | Same |
| Cleaning | Waterline cleaning according to ISO 16954
Waterline: Routine(Daily), Biofilm treatment(2 week)
Water Flushing (each patient) | Waterline cleaning according to ISO 16954
Waterline: Routine(Daily), Biofilm treatment(2 week)
Water Flushing (each patient) | Same |
| Warmer | Heating Method : Heating Coil
Water Temperature : 25°C~37 °C
Warmer Volume : 360ml
Warmer Power : 300W
Warmer input voltage : AC48V | Heating Method : Heating Coil
Water Temperature : 25°C~37 °C
Warmer Volume : 360ml
Warmer Power : 300W
Warmer input voltage : AC48V | Same |
| Suction | HVE (High volume evacuator)
SE | HVE (High volume evacuator)
SE | Same |
| Patient Load | Max. 150kg | Max. 150kg | Same |
| Chair Height | Max. 850±30mm, Min. 450±30mm | Max. 840±30mm, Min. 440±30mm | Different |
| Back Rest | 0°±5° to 70°±5° | 0°±5° to 70°±5° | Same |
| Head Rest | -90° to 75° | -90° to 70° | Different |
| Patient contacting components | Seat: Polyvinyl Chloride
Waterline: Polyurethane resins
Airline: Polyurethane resins
Suction: A16061
3-way syringe: A16061
Warmer: Stainless Steel 304
Water block: C3604 | Seat: Polyvinyl Chloride
Waterline: Polyurethane resins
Airline: Polyurethane resins
Suction: A16061
3-way syringe: A16061
Warmer: Stainless Steel 304
Water block: C3604 | Same |
| Principle of Operation | The chair is operated, the rising S/W is activated and the chair is electrically operated.
The handpiece is operated, pressing the foot control pedal opens the air supply valve to rotate the handpiece turbine.
The 3-way syringe is operated by air | The chair is operated, the rising S/W is activated and the chair is electrically operated.
The handpiece is operated, pressing the foot control pedal opens the air supply valve to rotate the handpiece turbine.
The 3-way syringe is operated by air | Same |
- 3/5 -
{7}
OSSTEM
Osstem Implant Co., Ltd. Chair Business
192, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do, 15428, Republic of Korea
Tel: +82 2 2016 7000 Fax: +82 2 2016 7001 www.osstem.com
| | pressure or electronic circuit S/W | pressure or electronic circuit S/W | |
| --- | --- | --- | --- |
| Lift Motor | Electromotor | Electromotor | Same |
| Dental Light | Available | Available | Same |
| Foot Control | Standard Wireless | Standard Wireless | Same |
| Electrical Safety | Complied with IEC 60601-1 | Complied with IEC 60601-1 | Same |
| Electromagnetic compatibility | Complied with IEC 60601-1-2 | Complied with IEC 60601-1-2 | Same |
[Table 1] Comparison between K5 Cart, K5 Mount, K5 Swing and K233805
- 4/5 -
{8}
OSSTEM
Osstem Implant Co., Ltd. Chair Business
192, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do, 15428, Republic of Korea
Tel: +82 2 2016 7000 Fax: +82 2 2016 7001 www.osstem.com
## 8. Non-clinical Test Data
### Electrical Safety and Electromagnetic compatibility
K5 Cart, K5 Mount, K5 Swing have been verified according to IEC 60601-1, IEC 60601-1-2, IEC 80601-2-60, IEC 60601-1-6 and thus have electrical safety.
Since these standards are verified according to the same standards as the predicated device, their electrical safety is equivalent to that of the predicate device.
### Performance test
K5 Cart, K5 Mount, K5 Swing have been verified according to ISO 7494-1, ISO 7494-2, and thus the performance of the product has been demonstrated.
These standards have been verified according to the same standards as predicate devices, so the performance items are equivalent to predicate devices.
### Software and System Verification and Validation
The proposed device is identical to the original device in all major software functions, except for the user interface.
The updated software was verified in accordance with IEC 62304 and the FDA Guidance Content of Premarket Submissions for Device Software Functions, so the original device and the submitted device are substantially equivalent, although they have different software versions.
## 9. Clinical Test Data
Clinical test data is not needed to characterize its performance and establish substantial equivalence.
## 10. Conclusion
Based upon the above mentioned data and comparison table, the K5 Cart, K5 Mount, K5 Swing are substantially equivalent to the predicate device as described herein.
- 5/5 -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.