K231845 · Foshan Safety Medical Equipment Co., Ltd. · EIA · Feb 28, 2024 · Dental
Device Facts
Record ID
K231845
Device Name
Dental unit Model: Mare
Applicant
Foshan Safety Medical Equipment Co., Ltd.
Product Code
EIA · Dental
Decision Date
Feb 28, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6640
Device Class
Class 1
Indications for Use
The Dental Unit is intended to supply power to and serve as a base for dental devices, and accessories. It is intended for use in the dental clinic /office environment and used by trained dentists and/or dental technicians and assistants. This product is attached with a dental chair.
Device Story
Dental unit (model: Mare) serves as a base for dental instruments; provides power, water, and compressed air. Components include dental chair, arm dentist unit, assistant's unit, LED oral lamp, cuspidor, and foot control. Operates via electrical, pneumatic, and hydraulic systems. Dentist uses control panel or foot switch to adjust chair position (motor-driven worm gear, gas spring support) and activate dental tools (handpieces, 3-way syringe, suction). LED lamp features non-contact infrared induction switch. Used in dental clinics by trained professionals. Output is mechanical movement of chair and delivery of air/water/power to connected ancillary devices. Benefits include stable patient positioning and integrated instrument support for dental procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), performance (ISO 7494-1, ISO 7494-2, ISO 9680, ISO 9168), and biocompatibility (ISO 10993 series).
Technological Characteristics
Materials: PU armrests, PVC leather, stainless steel syringe, TPU tubes. Energy: 110V/60Hz electrical, compressed air (200-400kPa). Features: LED oral lamp, hanging instrument tray, ISO 9168 connectors. Software: IEC 62304 compliant. Sterilization: Not provided sterile. Connectivity: Standalone unit with integrated control electronics.
Indications for Use
Indicated for use in dental clinics/offices by trained dentists, dental technicians, and assistants to provide power and a base for dental devices and accessories. Attached to a dental chair.
Regulatory Classification
Identification
A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 28, 2024
Foshan Safety Medical Equipment Co., Ltd % You Yijie Manager Qimmiq Medical Consulting Service Co., Ltd. RM.406, Building C, Run Science Park, No.18 Shenzhou Road, Huangpu Guangzhou, Guangdong 510663 CHINA
Re: K231845
Trade/Device Name: Dental Unit: model Mare Regulation Number: 21 CFR 872.6640 Regulation Name: Dental Operative Unit and Accessories Regulatory Class: Class I, reserved Product Code: EIA Dated: January 29, 2024 Received: January 29, 2024
Dear You Yijie:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K231845
Device Name Dental unit Model: Mare
#### Indications for Use (Describe)
The Dental Unit is intended to supply power to and serve as a base for dental devices, and accessories. It is intended for use in the dental clinic /office environment and used by trained dentists and/or dental technicians and assistants. This product is attached with a dental chair.
| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
Type of Use (Select one or both, as applicable)
{4}------------------------------------------------
# Section 5: 510(k) Summary K231845
# 1. Submitter's Information
#### Establishment Registration Information
Name: FOSHAN SAFETY MEDICAL EQUIPMENT CO.,LTD. Address: Floor 1,2,3, Building 6-1, lane1, Dianbian East Road, Nanyue, Shangan, Danzao Town, Nanhai, Foshan City, Guangdong Province, P.R. China
#### Contact Person of applicant
Name: May Xian Address: Floor 1,2,3, Building 6-1, lane1, Dianbian East Road, Nanyue, Shangan, Danzao Town, Nanhai, Foshan City, Guangdong Province, P.R. China TEL: +86 13668901542 FAX: No Email: x785286591@163.com
#### Contact Person of the Submission:
Name: Yijie You Address: RM.406, Building C, Run Science Park, No.18 Shenzhou Road, Huangpu district, Guangzhou city, Guangdong province, China TEL: +86 020-8224 5821 FAX: +86 020-8224 5821 Email: Jet.you@qimmiq-med.com
Date prepared: Feb. 27, 2024
## 2. Device Information
Trade Name: Dental unit Model: Mare Classification name: Dental Operative Unit & Accessories Review panel: Dental Product code: EIA Regulation Class: ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Regulation Number: 872.6640
## 3. Predicate Device Information
#### Primary predicate device:
Zhuhai Siger Medical Equipment Co., Ltd 510(k) submitter/holder:
{5}------------------------------------------------
| 510(K) Number: | K142206 |
|----------------------|-------------------------------------|
| Trade Name: | Dental Unit |
| Model: | U300 |
| Classification name: | Dental Operative Unit & Accessories |
| Review panel: | Dental |
| Product code: | EIA |
| Regulation Class: | I |
| Regulation Number: | 872.6640 |
# 4. Device description
The Dental Unit(model: Mare) is a dental treatment unit tested in accordance with IEC 80601-2-60. This product is used in dentistry only and may only be used by trained medical personnel and trained professional in the field of general dentistry.
Dental Unit (model: Mare) consists of dental chair, arm dentist unit, assistant's unit, dental light, cuspidor, water unit, and multi-function foot control. It mainly relies on electricity, compressed air, water to achieve all functions. Various ancillary dental devices can be connected to the Dental Unit (model: Mare) which are attached by means of industry standard ISO connections. The ancillary dental devices include pneumatic handpieces, 3-way syringe, strong suction and weak suction vacuum instruments. The ancillary dental devices include 3-way syringe, strong suction and weak suction vacuum instruments are 510k clarence in the K142206. Based on Safety does not manufacture pneumatic handpieces, the pneumatic handpieces shall be purchased by the enduser. The recommended pneumatic handpieces are 510(k) clearance K170229 and K170236, which are held by GUANGDONG JINME MEDICAL TECHNOLOGY CO., LTD. None of the Dental Unit (model: Mare), parts or accessories are provided sterile.
[Note] The ancillary device, such as dental pneumatic handpieces is not included with this subject device. These optional device and accessories are not supplied by Safety and are installed by the end-users using the recommended installation method described in the user manual provided by the manufacturers of these ancillary device. Also, the strong suction tip and weak suction tip are not included with the subject device and not provided by the Safety.
### Principle of operation:
The dental chair is mainly composed of motors and simple mechanical equipment. During the entire operation process, the doctor can press keys on the control panel to meet their needs. The mechanical automatic controls the motor, which drives a worm gear device that lifts or tilts the seat as needed.
In order to make the whole process of the seat up and down activities stable, sitting and lying quiet, install the gas spring under the seat. It is a kind of spring that can save labor and take off. The gas spring is composed of a sleeve, piston and piston rod, etc., adding high pressure air or highpressure nitrogen in the sleeve, because of the differential pressure between the two areas of the piston, to make the piston and piston rod move, support people or heavy objects.
The design of handrail takes into account that it is convenient for patients to get on and off the dental chair. The right handrail is a movable handrail. When patients get on and off the dental chair, the right handrail can be rotated clockwise to the limit position.
{6}------------------------------------------------
Headrest of Dental Unit (model: Mare) is a kind of adjustable headrest. The Adjusting the headrest can make it fit more accurately with the shape of the patient's neck and back, meeting the needs of patients with different heights. To enable the headrest to reach various positions, the design separates the headrest and back, allowing for adjustment of the headrest at the upper end of the back. The headrest can be adjusted by moving the headrest slider up and down. Once adjusted to the correct height, it will automatically lock into place.
The treatment unit is composed of instrument tray, 3-way syringe, handpiece interface, strong suction and weak suction vacuum instruments, which mainly relies on water, air and electricity to achieve the treatment function.
The instrument tray of Dental Unit (model: Mare) is hanging instrument plate structure.
The 3-way syringe through the button type water/air valve to open or close water/ air. The 3-way syringes are taken and placed in line with the hanging instrument tray.
There are three handpiece interfaces on the arm of the instrument table, and it is the four-hole handpiece interfaces which meet ISO 9168. The high or low output of water/air flow from the handpiece hose is determined by the water pressure valve, diaphragm valve, and foot valve.
Foot switch is composed of main gas control valve, cooling water control valve, chip blowing valve, foot control switch, flushing, water supply switch. The main gas control valve, cooling water control valve, and chip blowing valve can switch and adjust the size of compressed air through the presstype air control valve. This enables the pressure water valve and diaphragm valve on the instrument panel to be controlled, providing necessary working conditions for handpieces and other dental devices.
Foot control switch is through four microswitches to open and close the control signal to the program controller, so as to achieve the seat lifting and pitching function. Foot control switch four microswitches are responsible for the control of the seat's rise, fall, tilt, four directions of electrical signals.
Flushing and water supply switch is through two metal button switches to send the solenoid valve electrical signals to the program controller to achieve the flushing, water supply solenoid valve open and close.
The dental unit (model: Mare) adopts integral cuspidor, mainly composed of flush nozzle, water supply nozzle and cuspidor bowl.
Flushing nozzle through the main control key, auxiliary control key, foot key switch control flushing solenoid valve switch open phlegm water to achieve flushing cuspidor bowl. By rotating the flushing nozzle direction to adjust the flushing phlegm flow direction, it is convenient to wash the stains in different directions of the cuspidor bowl.
Water supply nozzle through the main control key, auxiliary control key, foot key switch control water supply solenoid valve switch open mouthwash to realize the filling of mouthwash. Cuspidor bowl are made of ceramic. The cuspidor bowl is connected to a drainpipe through a drainage interface to discharge mouthwash and phlegm.
The dental unit (model: Mare) adopts LED oral lamp. The LED oral lamp is mainly composed of
{7}------------------------------------------------
LED lamp bead, voltage regulator circuit board, induction switch, input power supply.
Through controlling the input power supply to realize the opening and closing of the LED oral lamp. AC 24V power input to the oral lamp voltage control circuit board, voltage control circuit board through the rectifier, voltage control circuit for LED lamp bead to provide stable voltage and current.
The induction switch consists of an infrared emitter and an infrared receiver. By passing through the induction switch, the output power of the voltage regulator circuit can be turned on or off, achieving non-contact control over the opening and closing of the oral lamp.
# 5. Indications for Use
The Dental Unit is intended to supply power to and serve as a base for dental devices, and accessories. It is intended for use in the dental clinic /office environment and used by trained dentists and/or dental technicians and assistants. This product is attached with a dental chair.
{8}------------------------------------------------
# 6. Summary of technological characteristics of device compared to the predicate device (K142206)
| SE<br>Comparisons | | Subject device<br>(Dental Unit,<br>model: Mare) | Predicate device<br>(Dental Unit, Model: U300) | Discussion of<br>difference |
|----------------------------------------------|--------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| 510K Number | | / | K142206 | / |
| Classification | | 21 CFR 872.6640 | 21 CFR 872.6640 | Same |
| Product Code | | EIA | EIA | Same |
| FDA Class | | / | / | Same |
| Indications for Use | | The Dental Unit is intended to supply power to and<br>serve as a base for dental devices, and accessories. It<br>is intended for use in the dental clinic /office<br>environment and used by trained dentists and/or dental<br>technicians and assistants. This product is attached<br>with a dental chair. | The Dental Unit is intended to supply power to and<br>serve as a base for dental devices, and accessories.<br>It is intended for use in the dental clinic /office<br>environment and used by trained dentists and/or<br>dental technicians and assistants. This product is<br>attached with a dental chair. | Same |
| Model | | Mare | U300 | / |
| Environment of Use | | Dental clinic /office environment | Dental clinic /office environment | Same |
| | Operating<br>Light | LED | Halogen | Different<br>(Discussion is<br>indicated in<br>D1) |
| Features | Connection<br>Joint | Comply with ISO9168 | Comply with ISO9168 | Same |
| | Water<br>Heating | No | Yes | Different<br>(Discussion is<br>indicated in<br>D2) |
| Operation Method | | Control Panel / Assistant Control Panel | Control Panel / Assistant Control Panel | Same |
| | | / Foot Controller | / Foot Controller | |
| Power Supply | | 110V | 110V | Same |
| Frequency | | 60Hz | 50/60Hz | (Discussion is<br>indicated in<br>D3) |
| | Power (with dental chair) | 1100VA | 900VA | Different<br>(Discussion is<br>indicated in<br>D3) |
| Pressure of Air Supply | | 200kPa-400kPa | 200kPa-400kPa | Same |
| Pressure of Air Supply | | ≥550 kPa | ≥550 kPa | Same |
| | Loading<br>Capacity | 150kg | 135kg | Different<br>(Discussion is<br>indicated in<br>D4) |
| | Movement<br>Range<br>(Chair) | 420mm-820mm | 420mm-820mm | Same |
| Dental<br>Chair | Range of<br>angular<br>movement<br>for the<br>backrest | -5°-85° | 0°-80° | Different<br>(Discussion is<br>indicated in<br>D5) |
| | Movement<br>Range<br>(Backrest) | 130mm | 200mm | Different<br>(Discussion is<br>indicated in<br>D6) |
| Accessories can be<br>attached to the device | | Handpiece/ Syringe | Handpiece / Scaler / Curing Light /<br>Syringe | Same |
| | | | | (included by<br>Predicate<br>device) |
| Rate of | Suction | ≥ 1L/min | ≥ 1L/min | Same |
| Water<br>Suction | Silva<br>Ejector | >750mL/min | > 750mL/min | Same |
| Performance Standards | | Comply with ISO7494-1 and ISO7494-2 | Comply with ISO7494-1, ISO7494-2<br>and ISO6875 | Same<br>(ISO7494-<br>1:2018 replace<br>the ISO<br>6875:2011) |
| Electrical Safety | | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Same |
| EMC | | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same |
| Patient Contact Material | | Armrest: PU<br>Leather for patient chair: PVC<br>Syringe: Stainless steel<br>Tube: TPU | Armrest: PU<br>Leather for patient chair: PVC<br>Syringe: Stainless steel<br>Tubes: TPU | Same |
{9}------------------------------------------------
{10}------------------------------------------------
The discussion of differences exist between the subject and predicate devices is listed in following:
- D : The subject device and predicate device have allamp. The subject device adapted the LED oral lamp, and the predicate device used the Halogen. But the LED oral lamp of subject device has been validated for t per SO 9680, and the subject device has demonstrated electrical safety by passing IEC 0001 - 1, IEC 6061-1-2 and IEC 80601-2-60 test. Therefore will not affect the safety and effectiveness of the subject device.
- D2: The water heating is used to generate warn water for use during dental treatment procedure, but the biofilm in the waterline. The subject device lacks water heating to reduce. The waterline biofilm treatment of subject device was performed as recommended by the manufacturer and verified according to ISO 16954, and the subject device has demonstrated electrical safety by passing IEC 6061-1-2 and IEC 80601-2-60 test. Therefore, the difference will not affectiveness of the subject device compared to the predicate device.
- D3: The subject device has demonstrated election leC 60601-1. IEC 6061-1-2 and IEC 80601-2-60 test. Therefore, will not affect the safety and effectiveness of the subject device compared to the predicate device.
- D4: The subject device and predicate devices have differences in dental capacity. According to the ISO 7494-1.2018, the patient chair shall support the patient mass shall be al least 150kg. The subject device has demonstrated meet the requirement by passing ISO 7494-1.2018 test and subject device has
{11}------------------------------------------------
demonstrated electrical safety by passing IEC 6061-1-2 and IEC 80601-2-60 test. Therefore, the difference will not affect the safety and effectiveness of the subject device compared to the predicate device.
- DE: The subject device and predicate devices in range of angular movement for the backrest. The subject device has demonstrated meet the requirement by passing ISO 7494-1:2018 test and subject device has demonstrated electrical safety by passing IEC 6061-1-2 and IEC 8001-2-60 test. Therefore, the difference will not affectiveness of the subject device compared to the predicate device.
- D6: The subject device and predicate devices in backrest novement range. The subject device has demonstrated meet the requirement by passing ISO 7494-1:2018 test and subject device has demonstrated electrical safety by passing IEC 6061-1-2 and IEC 8001-2-60 test. Therefore, the difference will not affect the safety and effectiveness of the subject device compared to the predicate device.
{12}------------------------------------------------
# 7. Discussion of Non-Clinical Tests Performed for Safety and
## effectiveness are as follows
The recognized consensus standards for safety of medical electrical equipment: IEC 60601-1, IEC 80601-2-60 for safety, IEC 60601-1-2 for electromagnetic compatibility, ISO 7494-1:2018, ISO 7494-2:2015, ISO 9168:2021, and ISO 9680:2021for performance, IEC 62304 for software verification and ISO 16954:2015 for waterline biofilm treatment are complied. See below table for details:
| Standards | Standards Name |
|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 Edition 3.2 2020-<br>08 | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance |
| IEC 60601-1-2 Edition 4.0<br>2014-02 | Medical electrical equipment - Part 1-2: General requirements for<br>basic safety and essential performance - Collateral Standard:<br>Electromagnetic disturbances - Requirements and tests |
| IEC 80601-2-60 Edition 2.0<br>2019-06 | Medical electrical equipment - Part 2-60: Particular requirements<br>for the basic safety and essential performance of dental<br>equipment |
| ISO 7494-1 Third edition<br>2018-06 | Dentistry - Stationary dental units and dental patient chairs - Part<br>1: General requirements |
| ISO 7494-2 Second edition<br>2015-04-01 | Dentistry - Dental units - Part 2: Air, water, suction and waste<br>water systems |
| ISO 9680:2021 | Dentistry. Operating lights |
| ISO 17665-1 First edition<br>2006-08-15 | Sterilization of health care products - Moist heat - Part 1:<br>Requirements for the development, validation and routine control<br>of a sterilization process for medical devices |
| ISO 16954:2015 | Dentistry. Test methods for dental unit waterline biofilm treatment |
| AAMI TIR 12:2020 | Designing, testing, and labeling medical devices intended<br>for processing by health care facilities: A guide for<br>device manufacturers |
| AAMI TIR 30:2011/(R)2016 | A compendium of processes, materials, test methods,<br>and acceptance criteria for cleaning reusable medical device |
| ANSI/AAMI ST79 :2017<br>& 2020 Amendments A1,<br>A2, A3, A4 | Comprehensive guide to steam sterilization and<br>sterility assurance in health care facilities |
| ISO 9168 Third edition 2009-<br>07-15 | Dentistry - Hose connectors for air driven dental handpieces |
| ISO 10993-5 Third edition<br>2009-06-01 | Biological evaluation of medical devices - Part 5: Tests for in<br>vitro cytotoxicity |
| ISO 10993-10 Third Edition<br>2010-08-01 | Biological evaluation of medical devices - Part 10: Tests for<br>irritation and skin sensitization |
| ISO 10993-10 Fourth edition | Biological evaluation of medical devices - Part 10: Tests for skin |
| ISO 10993-23 First edition<br>2021-01 | Biological evaluation of medical devices - Part 23: Tests for<br>irritation |
| IEC 62304 Edition 1.1 2015-<br>06 CONSOLIDATED<br>VERSION | Medical device software - Software life cycle processes |
| ASTM D4169-22 | Standard Practice for Performance Testing of Shipping<br>Containers and Systems |
{13}------------------------------------------------
Software verification and validation was performed for the subject device in accordance with Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices -Guidance for Industry and FDA Staff, May 2005.
Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff MARCH 2015
# 8. Discussion of Clinical Accuracy Testing Performed
There was no clinical testing performed.
# 9. Conclusions
Based on performance testing, comparison and analysis, the subject device Dental Unit (model: Mare) is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.