K142206 · Zhuhai Siger Medical Equipment Co., Ltd. · EIA · Mar 13, 2015 · Dental
Device Facts
Record ID
K142206
Device Name
Dental Unit
Applicant
Zhuhai Siger Medical Equipment Co., Ltd.
Product Code
EIA · Dental
Decision Date
Mar 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6640
Device Class
Class 1
Indications for Use
The Dental Unit is intended to supply power to and serve as a base for dental devices; and accessories. It is intended for use in the dental clinic /office environment and used by trained dentists and/or dental technicians and assistants. This product is attached with a dental chair.
Device Story
Dental unit (models U300, U500) serves as base/power supply for dental instruments; includes patient chair, instrument tray, operation light, assistant position, and pedal switch. Used in dental clinics by dentists/technicians. Operates via control panels or foot controller to manage chair positioning and instrument power. Provides stable platform and utility connections (water/air) for handpieces, scalers, curing lights, and syringes. Benefits patient by providing ergonomic, stable environment for dental procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance verified via testing to IEC 60601-1, IEC 60601-1-2, ISO 7494-1/2, ISO 6875, and ISO 10993-5/10.
Technological Characteristics
Dental unit with integrated chair. Materials: PU (armrest), PVC (upholstery), stainless steel (syringe), TPU (tubes). Power: 110V, 50/60Hz. Connectivity: ISO9168 compliant joints. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 7494-1/2 (dental units), ISO 6875 (chair), ISO 10993 (biocompatibility).
Indications for Use
Indicated for use in dental clinic/office environments by trained dentists, dental technicians, and assistants to provide power and a base for dental devices and accessories.
Regulatory Classification
Identification
A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, suggesting a sense of community and care. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 13, 2015
Zhuhai Siger Medical Equipment Co., Ltd. c/o Ms. Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 120-119 Shanghai, 200120 CHINA
Re: K142206
Trade/Device Name: Dental Unit Regulation Number: 21 CFR 872.6640 Regulation Name: Dental operative unit and accessories Regulatory Class: I Product Code: EIA Dated: January 21, 2015 Received: February 12, 2015
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K142206
Device Name Dental Unit
Indications for Use (Describe)
The Dental Unit is intended to supply power to and serve as a base for dental devices; and accessories. It is intended for use in the dental clinic /office environment and used by trained dentists and/or dental technicians and assistants. This product is attached with a dental chair.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
#### K142206
# 510(k) Summary
This 510(k) Summary of 510(k) information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K142206
- 1. Date of Submission: 01/20/2015
#### 2. Sponsor Identification
Zhuhai Siger Medical Equipment Co., Ltd Building 2, No. 1 Chuangxin Yi Road, Tangjiawan Town, Zhuhai City, Guangdong, China
Establishment Registration Number: Not yet registered
Contact Person: Huisheng Wang Position: General Manager Tel: 86 13802677946 Fax: 86 756 3881011 Email: wanghs@siger.cn
#### 3. Submission Correspondent
Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Proposed Device Identification
Proposed Device Name: Dental Unit Proposed Device Model: U300, U500 Proposed Device Common Name: operative dental unit
Regulatory Information: Classification Name: Unit, Operative Dental Classification: I Product Code: EIA Regulation Number: 21 CFR 872.6640 Review Panel: Dental
Intended Use Statement:
to The Dental Unit is intended to supply power to and serve as a base for dental devices; and accessories. It is intended for use in the dental clinic /office environment and used by trained dentists and/or dental technicians and assistants. This product is attached with a dental chair."
#### 5. Predicate Device Identification
510(k) Number: K130410 Product Name: Dental Unit with Chair S2310 Manufacturer: Xianyang North West Medical Instrument (Group) Co., Ltd
### 6. Device Description
The proposed devices Dental Unit are well equipped dental unit, which are intended to supply power to and serve as a base for dental devices and accessories. They are intended for use in the dental clinic/office environment and used by trained dentists and/or dental technicians and assistants. The products are attached with dental chair.
The proposed devices include two models. U300 and U500. Both of the two models mainly consist of instruments, instrument arm tray, cabinet group, operation light arm, assistant position, pedal switch, and patient chair.
#### 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
{5}------------------------------------------------
- IEC 60601-1: 2005 +COOR.1 (2006) + CORR.2 (2007), Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2: 2007, Medical Electrical Equipment Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements And Tests
- ISO7494-1: 2004, Dentistry Dental units Part 1: General requirements and test methods. ISO7494-2: 2003, Dentistry - Dental units - Part 2: Water and air supply.
- ISO 6875: 2011, Dentistry -- Patient chair
- ISO 10993-5: 2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity.
- ISO 10993-10: 2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
## 8. Substantially Equivalent (SE) Conclusion
The following table compares the DEVICE to the predicate device with respect to intended use, technological characteristics and principles of operation, etc.
| Item | Proposed Devices | Predicate Device | |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Product Code | EIA | EIA | |
| Regulation No. | 21 CFR 872.6640 | 21 CFR 872.6640 | |
| Class | 1 | 1 | |
| Intended Use | The Dental Unit is intended to supply<br>power to and serve as a base for dental<br>devices; and accessories. It is intended<br>for use in the dental clinic /office<br>environment and used by trained<br>dentists and/or dental technicians and<br>assistants. This product is attached with<br>a dental chair. | The Dental Unit with Chair is intended<br>to supply power to and serve as a base<br>for dental devices and accessories. It is<br>intended for use in the dental<br>clinic/office environment and used by<br>trained dentists and/or dental<br>technicians and assistants. This product<br>is attached with a dental chair. | |
| Features | Operating<br>Light | Halogen | LED |
| | Connection<br>Joint | Comply with ISO9168 | Comply with ISO9168 |
| | Water<br>Heating | Yes | No |
| Operation Method | Control Panel / Assistant Control Panel<br>/ Foot Controller | Control Panel / Assistant Control Panel<br>/ Foot Controller | |
| Power Supply | 110V | 110V | |
| Frequency | 50/60Hz | 50/60Hz | |
#### Table 3-1 Comparison of Technology Characteristics
{6}------------------------------------------------
| Power (with dental chair) | | | 900VA | 400VA |
|----------------------------------------------|---------------------------------|---------------------------------|-----------------------------------------------------------------------------------------|----------------------------------------------------------|
| Pressure of Water Supply | | | 200 kPa ~ 400 kPa | 0.2MPa-0.4MPa |
| Pressure of Air Supply | | | ≥550 kPa | 0.55MPa |
| Dental<br>Chair | Loading<br>Capacity | | 135Kg | 200kg |
| | Movement<br>Range<br>(Chair) | | 420-820 MM | 390mm-740mm |
| | Movement<br>Range<br>(Backrest) | | 0°~80° | 1°~70° |
| | | Movement<br>Range<br>(Headrest) | 200MM | 150mm |
| Accessories can be<br>attached to the device | | | Handpiece / Scaler / Curing Light /<br>Syringe | Handpiece / Scaler / Curing Light /<br>Three-way-Syringe |
| Performance Standards | | | Comply with ISO7494-1, ISO7494-2<br>and ISO6875 | Comply with ISO7494-1, ISO7494-2<br>and ISO6875 |
| Rate of Water<br>Suction | Suction | | ≥ 1L/min | 1 L/min |
| | Silva<br>Ejector | | > 750mL/min | 750ml/min |
| Electrical Safety | | | Comply with IEC 60601-1 | Comply with IEC 60601-1 |
| EMC | | | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
| Patient Contact Material | | | Armrest: PU<br>Leather for patient chair: PVC<br>Syringe: Stainless steel Tubes:<br>TPU | PU, PVC, Polyamide, ABS, Stainless<br>steel |
The proposed devices, Dental Unit U300 and U500, are determined to be Substantially Equivalent (SE) to the predicate device, Dental Unit with Chair S2310 (K130410), with respect to intended use, technological characteristics and principles of operation.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.