K243130 · Guangdong Yadeng Medical Apparatus Co., Ltd., · EIA · Jun 27, 2025 · Dental
Device Facts
Record ID
K243130
Device Name
Integral Dental Unit
Applicant
Guangdong Yadeng Medical Apparatus Co., Ltd.,
Product Code
EIA · Dental
Decision Date
Jun 27, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6640
Device Class
Class 1
Indications for Use
The Integral Dental Unit is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants.
Device Story
Integral Dental Unit (Model YD-A4) serves as a stationary base for dental instruments; provides power, water, and air supply to handpieces and accessories. System includes an electronically-controlled patient chair, dentist element, assistant element, and dental light. Operated by dental professionals in clinical settings via foot switch and integrated control unit. Device facilitates patient positioning and instrument management during dental procedures. Does not include handpieces or diagnostic tools; these are provided by the user. Benefits include ergonomic patient positioning and centralized utility management for dental operations.
Clinical Evidence
No clinical data. Bench testing only. Device evaluated for electrical safety (IEC 60601-1, IEC 80601-2-60), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993-1), and performance (ISO 7494-1/2).
Technological Characteristics
Stationary dental unit; hydraulic electromotor chair lift; pneumatically controlled valves for air/water flow. Power: 115/230 Vac, 50/60Hz. Connectivity: ISO 9168 compliant connectors. IPX4 rated foot switch. Software: IEC 62304 compliant. Materials: Not specified. Sterilization: Not specified (standard clinical cleaning/disinfection).
Indications for Use
Indicated for use by trained dental professionals, dentists, and dental assistants in a dental clinic or hospital environment to provide a base and power supply for dental diagnostic and treatment procedures.
Regulatory Classification
Identification
A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 27, 2025
Guangdong Yadeng Medical Apparatus Co., Ltd.,
% Jett Lee
Official Applicant
Guangdong Jianda Medical Technology Co., Ltd.
906 Room
Longxiang Garden, Tianhe district
GuangZhou, Guangdong 510000
CHINA
Re: K243130
Trade/Device Name: Integral Dental Unit
Regulation Number: 21 CFR 872.6640
Regulation Name: Dental Operative Unit And Accessories
Regulatory Class: Class I, reserved
Product Code: EIA, KLC
Dated: May 26, 2025
Received: May 27, 2025
Dear Jett Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243130 - Jett Lee
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243130 - Jett Lee
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243130
Device Name
Integral Dental Unit
Indications for Use (Describe)
The Integral Dental Unit is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
June 23, 2025
# K243130 - 510(k) Summary
## 1. Submitter's Information
### Sponsor
- Company Name: GUANGDONG YADENG MEDICAL APPARATUS CO., LTD.
- Address: Factory 2, No.12, Fenggong Road, Nansha Industrial Zone, Danzao Town, Nanhai District, Foshan, Guangdong, 528216, P.R. China
- Phone: +86-15916558838
- Fax: +86-0757-86603813
- Contact Person (including title): Weng changming
- E-mail: 541405168@qq.com
### Application Correspondent
- Company: Guangdong Jianda Medical Technology Co., Ltd
- Address: 906 Room, Longxiang Garden, Tianhe district, Guangzhou, China
- Contact Person: Jett Lee
- Tel: 86-13512755282
- Email: jianda-lee@foxmail.com
## 2. Subject Device Information
- Type of 510(k) submission: Traditional
- Common Name: Unit, Operative Dental
- Trade Name: Integral Dental Unit
- Model: YD-A4
- Classification Name: Dental Operative Unit and Accessories
- Review Panel: Dental
- Product Code: EIA
- Regulation Number: 872.6640
- Regulation Class: I
## 3. Predicate Device Information
| | Predicate Device |
| --- | --- |
| Sponsor | Osstem Implant Co., Ltd. |
| Device Name | K3 |
| Model | / |
| --- | --- |
| 510(k) Number | K183347 |
{5}
2
| Product Code | EIA |
| --- | --- |
| Regulation Number | 872.6640 |
| Regulation Class | I |
# 4. Device Description
The Integral Dental Unit (Model: YD-A4) is a dental operative unit specially designed and provided for a qualified dentist to be used in a professional clinic or hospital facility to carry out dental procedures. Integral Dental Unit is intended to be used in a professional environment for dental diagnosis, treatment, or operation.
It includes a treatment chair, dentist element, assistant element and a dental light as offering several additional options and electronically-controlled chair movements with software and water/air unit functions.
The dental unit consists of an electronically operated dental chair and integrated control unit control for electricity, water and air supply to handpieces or some other dental instrument. The device is equipped with an instrument tray, pipes for water supply and tube air supply, a mouth lamp, a saliva aspirator, a spittoon, a three-way syringe, a film viewer, a foot switch and chair with driving motors and armrest.
The dental chair is intended to be used with dental hand pieces, cure light, ultrasonic scaler, camera system or other doctor stool, which is not provided by the manufacturer. The user will select the dental instruments and accessories with FDA clearance by themselves. So, the device in the submission does not include these parts and accessories. The connector standard type complies with ISO 9168.
Basic parameters/use conditions/power supply specifications is as follows:
| ◆ Noise | <70 dBA |
| --- | --- |
| ◆ Base box | Power supply: 115/230 Vac, 50/60Hz, single-phase 3core, protective grounding.
Power input: 380 VA
Water filter hole diameter: <90μm
Air filter hole diameter: <25μm |
| ◆ Saliva ejector | Weak saliva:
— vacuum degree >15 kPa;
— water pumping rate>80 mL/min
Strong saliva:
— vacuum degree >25 kPa;
— water pumping rate>1000 mL/min |
| --- | --- |
| ◆ Instrument tray | Rotating angle: >270°
Up-down moving range: >440mm
Max. Load: <5 Kg |
| ◆ Film viewer | Power supply: 24 V ac Inside power supply
Power input: 80 VA |
{6}
3
| ◆Operating light | Power supply: 12 V ac Inside power supply
Power input: 50 VA
Illuminance: 5 000~20 000 Lx
Radiant heat: <200 W/m² @ max. illuminance
Color rendering index: >85 Ra |
| --- | --- |
| ◆Foot switch | Tripping force: >10N and <30N
Degree of protection against harmful ingress of water: IPX4(foot switch)
Service life: >25 000 repeats |
| ◆Dental chair | Power supply: 24 V dc Inside power supply
Loading capacity: 1323N (about 135 Kg)
Loading capacity of headrest: 300N (about 30Kg)
Moving range of headrest: 120mm
Range of backrest when going backwards: 90°~170°
Seat cushion's maximum height away from ground: 730mm
Seat cushion's minimum height away from ground: 450mm |
| Attachment
--- Amalgam separation device | It has a medical device product registration card
Attachment parameters are reflected in its operating instructions |
| ◆Work space | L: ≥3 000 mm; W: ≥2 000 mm; H: ≥2 500 mm |
| ◆Environment for operation | Temperature: +5°C to +40°C
Relative humidity: 30% - 80%
Atmospheric pressure: 86kPa ~ 106kPa |
# 5. Intended Use / Indications for Use
The Integral Dental Unit is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants.
# 6. Test Summary
The Integral Dental Unit was evaluated for conformance to recognized international standards. The following is a list of these evaluations and tests that were found to be in conformance:
{7}
4
- Electrical safety test
IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
IEC 80601-2-60:2019, Medical electrical equipment Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment
- Electromagnetic compatibility test
IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
- Biocompatibility test
ISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- Software verification and validation test
FDA “Guidance for Premarket Submissions and for Software Contained in Medical Devices”
- Software Life Cycle Processes
IEC 62304:2006+AMD1:2015 Medical Device Software
- Performance Testing
ISO 7494-1:2018, Dentistry – Stationary dental units and dental patient chairs, Part 1: General requirements
ISO 7494-2:2018, Dentistry – Stationary dental units and dental patient chairs, Part 2: Air, water, suction and wastewater systems
7. Comparison to predicate device and conclusion
The subject device Integral Dental Unit is substantially equivalent to the predicated device based on intended use, design, specifications and performance. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
Information for predicate device was obtained from publicly available sources. A technical comparison to the predicate is provided below.
| Elements of Comparison | Subject Device | Predicate Device I | Verdict |
| --- | --- | --- | --- |
| 510(k) Number | K243130 | K183347 | -- |
| Device Name | Integral Dental Unit (Model: YD-A4) | K3 | -- |
| Product Code | EIA | EIA | Same |
| Regulation Number | 21 CFR 872.6640 | 21 CFR 872.6640 | Same |
| Regulation Class | I | I | Same |
| Prescription | Prescription use | Prescription use | Same |
{8}
5
| Intended Use | The Integral Dental Unit is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants. | K3 is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants. | Same |
| --- | --- | --- | --- |
| Power & Utility Supply | AC 230V, 50/60Hz, compressed air and water | AC 100-120/220-240V, 50/60Hz, compressed air and water | Similar Note 1 |
| Main Components | Ground box, foot controller, patient chair, weak suction system, strong suction system, Triple function handpiece(hot),cuspidor: water flues& | Chair, Unit, Table, Seat, Stool, Monitor Arm*, Hanaro Console* (Note: K3 Cart* model applied ONLY) | Similar Note 1 |
| Elements of Comparison | Subject Device | Predicate Device I | Verdict |
| --- | --- | --- | --- |
| | supply, cabinet box, operation lamp, film viewer, instrument control system, instrument tray, Triple function handpiece(cold), cup collector, glass plate | | |
| Syringe | 3-way syringe | 3-way syringe | Same |
| Control of water and air | Uses pneumatically controlled vales to water control the flow of air and water. On/off and intensity controlled by foot pedal. | Uses pneumatically controlled vales to water control the flow of air and water. On/off and intensity controlled by foot pedal. | Same |
| Air Pressure | 0.6MPa~0.80 MPa | 500kPa(min)/750kPa(max) | Similar Note 1 |
| Water Pressure | 0.20 MPa ~0.40 MPa | 250kPa(min)/600 kPa(max) | Similar Note 1 |
| Water System | City water supply | City water supply | Same |
| Patient Load | Max. 135kg | Max. 135kg | Same |
| Chair Height | Max. 730±10mm, Min. 450±10mm | Max. 795±10mm, Min. 365±10mm | Similar Note 1 |
| Lift Motor | Hydraulic electromotor | Hydraulic electromotor | Same |
| Electrical Safety | Complied with IEC 60601-1 | Complied with IEC 60601-1 | Same |
| Electromagnetic compatibility | Complied with IEC 60601-1-2 | Complied with IEC 60601-1-2 | Same |
## Summary for clinical test
Clinical performance testing was not performed for this device.
## Conclusion
Based on the above performance as documented in this application, Integral Dental Unit is substantially equivalent to the predicate device, K3 (K183347). Thus, the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.