Zygoma Dental Implant system is intended to replace missing tooth/teeth by surgical installation in the zygoma region for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. The Dental implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. Zygomatic Dental Implants are intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxilla.
Device Story
Zygoma Dental Implant System consists of endosseous implants and abutments (AXA, Multi-unit) for surgical placement in the zygoma region; provides support for fixed or removable dental prostheses in patients with severely atrophic maxillae. Implants are CP Ti Grade 4 with SLA surface treatment; abutments are Ti-6Al-4V-ELI. System used by clinicians in surgical settings; supports immediate loading protocols when primary stability is achieved. Output is a stable foundation for prosthetic restoration; enables chewing function in patients with compromised maxillary anatomy. Device is MR Conditional.
Clinical Evidence
Clinical evidence derived from six studies involving ~250 patients and 677 zygomatic implants. Follow-up periods ranged from months to >7.5 years. Primary endpoints included implant survival and restoration of oral function. Survival rates reported between 93% and 100% (mostly >98%). Adverse events were consistent with known risks of zygomatic surgery (e.g., mucositis, peri-implantitis, soft tissue recession) and were manageable.
Technological Characteristics
Implants: CP Ti Grade 4 (ASTM F67), SLA surface. Abutments/Screws: Ti-6Al-4V-ELI (ASTM F136-13), anodized surface. Sterilization: Gamma irradiation (implants); steam sterilization by user (abutments/screws). MR Conditional. Fatigue testing performed per ISO 14801 (deviated).
Indications for Use
Indicated for patients with partially or fully edentulous maxilla requiring support for fixed or removable prosthetic devices in the zygoma region. Limited to ZAGA Class 4 cases. Requires multiple unit restorations in splinted applications using at least two implants.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Noris Medical Zygomatic Dental Implant System (K151909)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 18, 2026
MegaGen Implant Co., Ltd.
Hyo-Eun Lee
Official Correspondent
45, Secheon-ro 7-gil, Dasa-eup, Dalseong-gun
Daegu, 42921
REPUBLIC OF KOREA
Re: K251232
Trade/Device Name: MegaGen Zygoma Dental Implant System
Regulation Number: 21 CFR 872.3640
Regulation Name: Endosseous Dental Implant
Regulatory Class: Class II
Product Code: DZE, NHA
Dated: May 10, 2026
Received: May 14, 2026
Dear Hyo-Eun Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251232 - Hyo-Eun Lee
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K251232 - Hyo-Eun Lee
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ANDREW I. STEEN -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251232 |
| --- | --- |
| Please provide the device trade name(s). | |
MegaGen Zygoma Dental Implant System
Please provide your Indications for Use below.
Zygoma Dental Implant system is intended to replace missing tooth/teeth by surgical installation in the zygoma region for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. The Dental implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.
Zygomatic Dental Implants are intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxilla.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
MegaGen Zygoma Dental Implant System
{4}
K251232
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR Part 807.92.
Date: June 09, 2026
### 1. Applicant / Submitter
MegaGen Implant Co., Ltd.
45, Secheon-ro, 7-gil, Dasa-eup, Dalseong-gun,
Daegu, Republic of Korea
Tel: + 82-53-222-2828
### 2. Submission Correspondent
Hyo-Eun, Lee
MegaGen Implant Co., Ltd.
45, Secheon-ro, 7-gil, Dasa-eup, Dalseong-gun,
Daegu, Republic of Korea
Tel: +82-53-222-3860 Fax: +82-53-289-3414
Email: ra7@imegagen.com
### 3. Device
- Trade Name: MegaGen Zygoma Dental Implant System
- Common Name: Endosseous Dental Implant
- Classification Name: Implant, Endosseous, Root-Form
- Classification Regulation: 872.3640
- Primary Product Code: DZE
- Secondary Product Code: NHA
### 4. Predicate Device
- **Primary Predicate Device:**
K151909 - Noris Medical Zygomatic Dental Implant System
- **Reference Device:**
K192651- ZAGA Zygomatic System
K232099 - Neodent Implant System - GM Zygomatic Implants System
K190491 - Blue Sky Bio Zygomatic Implant System
K212785 - GM Helix LG Neodent Implant System
K160119 - NobelSpeedy® Groovy
K241972- BLUEDIAMOND IMPLANT
K210356 - Noris Medical Dental Implant System
K242030 - MegaGen Dental Implant Abutment
K233450 - MegaGen Dental Implant Abutment - Scan Healing Abutment; Temporary Abutment; Temporary Cylinder; Comfort Cap; Healing Cap; Healing Cap Screw; Milling Abutment; EZ Post Abutment; Extra EZ Post Abutment; EZ Post Cylinder; ZrGEN Abutment; Multi-unit Abutment; Multi-unit Angled Abutment; AXA Abutment (Straight); AXA Abutment (Angled); Abutment Screw; Cylinder Screw; Crown Screw
K190958 - Neodent Implant System
K122231 - XPEED ANY RIDGE INTERNAL IMPLANT SYSTEM
{5}
K182448 - AnyRidge Octa 1 Implant System
K110955 - ANYRIDGE INTERNAL IMPLANT SYSTEM
K220562 - TiGEN Abutment, ZrGEN Abutment and Scan Healing Abutment
K150537 - MiNi Internal Implant System
K203554 - AnyOne External Implant System
## 5. Description
### • ZYGOMA IMPLANT (BD/AR/ AO ZYGOMA IMPLANT)
The ZYGOMA IMPLANT is intended to be implanted in the upper jaw arch to provide support for fixed or removable dental prostheses in partially edentulous or full arch prostheses. It is for immediate function when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. It is made of CP Ti Grade 4 with the surface treated by SLA method. The proposed MegaGen Zygoma Implants (BD, AR, AO models) are limited in use to ZAGA Class 4 only.
The dimensions of IMPLANT are follows:
| Product | Dimensions (Diameter x Length) |
| --- | --- |
| BD ZYGOMA IMPLANT | \( \emptyset \) 4.2 x 30, 32.5, 35, 37.5,40, 42.5, 45, 47.5, 50, 52.5, 55, 57.5 mm |
| AR ZYGOMA IMPLANT | \( \emptyset \) 4.2 x 30, 32.5, 35, 37.5,40, 42.5, 45, 47.5, 50, 52.5, 55, 57.5 mm |
| AO ZYGOMA IMPLANT | \( \emptyset \) 4.2 x 30, 32.5, 35, 37.5,40, 42.5, 45, 47.5, 50, 52.5, 55, 57.5 mm |
### • AXA Abutment
The AXA Abutment is angled and is two-piece type, the lower part is intended to be placed on the endosseous dental implant and the top part is connected Healing Cap, Impression Coping or Cylinders, such as Temporary Cylinder and EZ Post Cylinder with the Screw to fabricate temporary or final prosthesis. It is made of Ti-6Al-4V-ELI and offered in anodizing gold. The 45°, 52°, 60° angles and two types (Hex type and Octa type) are available for use with ZYGOMA IMPLANT. The AXA Abutment is supplied non-sterile, to be sterilized by the user according to the IFU and intended for single use.
The dimensions of AXA Abutment are follows:
| Product | Compatibility Implant | Dimensions (Diameter x Cuff) |
| --- | --- | --- |
| AXA Abutment | BD ZYGOMA IMPLANT | 45°: ∅ 5.0 x 2.0, 3.0, 4.0, 5.0 mm52°: ∅ 5.0 x 2.0mm60°: ∅ 5.0 x 2.0mm |
| | AR ZYGOMA IMPLANT | 45°: ∅ 5.0 x 2.0, 3.0, 4.0, 5.0 mm52°: ∅ 5.0 x 2.0mm60°: ∅ 5.0 x 2.0mm |
| | AO ZYGOMA IMPLANT | 45°: ∅ 5.0 x 2.0, 3.0, 4.0, 5.0 mm52°: ∅ 5.0 x 2.0mm60°: ∅ 5.0 x 2.0mm |
### • Multi-unit Abutment
The Multi-unit Abutment is angled and is two-piece type, the lower part is intended to be placed on the endosseous dental implant and the top part is connected Healing Cap, Impression Coping or Cylinders, such as Temporary Cylinder and CCM Cylinder with the Screw to fabricate temporary or final prosthesis. The Multi-unit Abutment is intended for both multiple unit restoration and single unit restoration. It is made of Ti-6Al-4V-ELI and offered in anodizing gold. The 45°, 52°, 60° angles and two types (Hex type and Octa type) are available for use with ZYGOMA IMPLANT. The Multi-unit Abutment is supplied non-sterile, to be sterilized by the user according to the IFU and intended for single use.
The dimensions of Multi-unit Abutment are follows:
| Product | Compatibility Implant | Dimensions (Diameter x Cuff height) |
| --- | --- | --- |
| Multi-unit Abutment | BD ZYGOMA IMPLANT | 45°: ∅ 4.8 x 2.0, 3.0, 4.0, 5.0 mm52°: ∅ 4.8 x 2.0mm60°: ∅ 4.8 x 2.0mm |
| | AR ZYGOMA IMPLANT | 45°: ∅ 4.8 x 2.0, 3.0, 4.0, 5.0 mm52°: ∅ 4.8 x 2.0mm60°: ∅ 4.8 x 2.0mm |
| | AO ZYGOMA IMPLANT | 45°: ∅ 4.8 x 2.0, 3.0, 4.0, 5.0 mm52°: ∅ 4.8 x 2.0mm |
{6}
60': ∅ 4.8 x 2.0mm
# - **Abutment Screw**
The Abutment Screw is used for securing the abutment to the endosseous implant. It offers Hex type to select the right direction for connection. It is made of Ti-6Al-4V-EL. The Abutment Screw is supplied non-sterile, to be sterilized by the user according to the IFU and intended for single use.
The dimensions of Abutment Screw are follows:
| Product | System | Dimensions (Diameter x Length) |
| --- | --- | --- |
| Abutement Screw | BD ZYGOMA IMPLANT | ∅ 1.9 x 2, 3, 4, 5, 6 mm |
| | AR ZYGOMA IMPLANT | ∅ 1.95 x 2, 3, 4, 5, 6 mm |
| | AO ZYGOMA IMPLANT | ∅ 1.95 x 2, 3, 4, 5, 6 mm |
### 6. Indication for use
Zygoma Dental Implant system is intended to replace missing tooth/teeth by surgical installation in the zygoma region for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. The Dental implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.
Zygomaic Dental Implants is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxilla.
{7}
## 7. Basis for Substantial Equivalence
The MegaGen Zygoma Dental Implant System is substantially equivalent to the predicate device and reference devices in terms of indication for use, technical characteristic and function. They are made of the same material and have similar design, and it is not affecting substantial equivalence. Based on the technological characteristic comparison tables below and test results provided in this submission, we conclude that the subject device is substantially equivalent to the predicate device.
### Indication for Use Statement
| | 510K | Indication for Use Statement |
| --- | --- | --- |
| **Subject Device** | K251232 | Zygoma Dental Implant system is intended to replace missing tooth/teeth by surgical installation in the zygoma region for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. The Dental implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. Zygomatic Dental Implants is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxilla. |
| **Predicate Device** | K151909 | Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Noris Medical Zygomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxillae. |
| **Reference Device** | K192651 | Southern Implants ZAGA Zygomatic System implants are intended to be implanted in the upper jaw arch to provide support for fixed dental prostheses in patients with partially or fully edentulous maxillae. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. |
| | K241972 | The BLUEDIAMOND IMPLANT is intended to be surgically placed in the maxillary or mandibular molar areas for the purpose providing prosthetic support for dental restorations (Crown, bridges, and overdentures) in partially or fully edentulous individuals. It is used to restore a patient's chewing function in the following situations and with the clinical protocols: - Delayed loading - Immediate loading when good primary stability is achieved and with appropriate occlusal loading. For the BLUEDIAMOND IMPLANTS with a Thread Length of 5mm, it is indicated for fixed or removable reconstruction in situations of moderate to severely atrophic jawbone and with adequate bone quality that allows primary stability after implant insertion, where a longer implant cannot be placed due to limited vertical bone height. The recommended healing time before loading is between 10 to 12 weeks. It is specifically recommended for: - Fixed partial dentures/splinted units (one implant per unit) - Pontic cases in combination with at least one longer implant - Fully edentulous cases with at least one 5 mm Short Implant in combination with 2 longer implants in the anterior region and at least four total implants |
| | K212785 | Blue Sky Bio Long Implant System is intended for surgical placement in the bone of the upper jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. Implants may be used with single-stage or two-stage procedures, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Blue Sky Bio Long implants can be placed bicortically in cases of reduced bone density. Blue Sky Bio Long implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. Blue Sky Bio Long Implant System with a 45° angulation are indicated for surgical installation in the pterygoid region only, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. |
| | K190958 | The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Neodent GM Helix LG implants can be placed bicortically in cases of reduced bone density. The Neodent GM Helix LG implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. |
| | K160119 | NobelSpeedy® Groovy implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting tooth replacements to restore patient |
{8}
| | esthetics and chewing function. NobelSpeedy® Groovy implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. This can be achieved by a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique. Implants allow also for bi- cortical anchorage in cases of reduced bone density NobelSpeedy® Groovy implants 20, 22, 25 mm when placed in the maxilla are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. |
| --- | --- |
| K210356 | Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Noris Medical Zygomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxillae. |
| K233450 | The MegaGen Dental Implant Abutment is intended to be surgically placed in the maxillary or mandibular areas for the purpose providing prosthetic support for dental restorations (Crown, bridges, and overdentures) in partially or fully edentulous individuals. It is used to restore a patients chewing function. All digitally designed abutments for use with ZrGEN Abutment are intended to be sent to a MegaGen validated milling center for manufacture. |
| **Substantial Equivalence Discussion** | |
| **Discussion** The indications for use statement of the subject device is nearly identical to the primary predicate, except for minor changes in wording that do not affect the intended use. Both indicated for splinted restorations utilizing a minimum of two implants to support fixed or removable prosthetic in the maxilla and may be loaded immediately when good primary stability is achieved. | |
{9}
# MegaGen Zygoma Dental Implant
| | Subject Device | Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 510k | K251232 | K151909 | K192651 | K241972 | K212785 | K190958 | K160119 |
| Device Name | MegaGen Zygoma Dental Implant | Noris Medical Zygomatic Dental Implant | Southern Implants ZAGA Zygomatic System implants | BLUEDIAMOND IMPLANT | Blue Sky Bio Dental Implant System - BIOILONG Implants | Neodent Implant System - GM Helix LG Implants | NobelSpeedy® Groovy |
| Manufacturer | MegaGen Implant Co., Ltd. | Noris Medical Ltd. | Southern Implants (Pty) Ltd | MegaGen Implant Co., Ltd. | Blue Sky Bio, LLC | JJGC Industria e Comercio de Materiais Dentarios SA | Nobel Biocare AB |
| Patient Population | Edentulous or partially edentulous individuals | Edentulous or partially edentulous individuals | Unknown in 510k | Edentulous or partially edentulous individuals | Unknown in 510k | Unknown in 510k | Unknown in 510k |
| Material | CP Ti Grade 4 (ASTM F67) | Titanium alloy | Unalloyed titanium (ASTM F67) Grade 4, and UTS ≥ 900MPa (cold-worked) | CP Ti Grade 4 (ASTM F67) | Ti-6Al-4V ELI (ASTM F136) | CP Ti Grade 4 (ASTM F67) | CP Titanium |
| Implant Diameter (Ø, mm) | 4.2 | 4.2 | 4.3 | Normal Thread: 4.0, 4.4, 4.7 Deep Thread: 4.4, 4.8, 5.1 | Platform: 3.5, NP (3.5) Implant Body: 3.7, 4.3, 5.0 | 3.75, 4.0 | 4.0 |
| Length (mm) | 30, 32.5, 35, 37.5, 40, 42.5, 45, 47.5, 50, 52.5, 55, 57.5 | 30, 35, 37.5, 40, 42.5, 45, 47.5, 50, 52.5, 55, 57.5 | 30, 32.5, 35, 37.5, 40, 42.5, 45, 47.5, 50, 52.5 | - Normal Thread Ø 4.0: 7.0, 9.0 Ø 4.4: 5.0, 7.0, 9.0 Ø 4.7: 5.0, 7.0, 9.0 - Deep Thread Ø 4.4: 7.0, 9.0 Ø 4.8: 5.0, 7.0, 9.0 Ø 5.1: 5.0, 7.0, 9.0 | 20.0, 22.5, 25.0 | 20.0, 22.5, 25.0 | 20.0, 22.5, 25.0 |
| Abutments* | 45, 52, 60 | 45, 52*, 60* (*K210356) | 55 | 30 | Straight, 12, 17, 24, 30 | Straight, 17, 30, 45* (*K190718) | 45* (*K161598) |
| Surface Treatment | Sand-blasted, Large grit, Acid-etched (S.L.A) Machined collar | RBM (Resorbable Blast Media) | Grit-blasted | Sand-blasted, Large grit, Acid-etched (S.L.A) Machined collar | Grit blasted and acid etched | Machined, acid-etched NeoPoros surface. | Ti Unite |
| Thread Design | Tapered Screw type | Tapered Screw type | Unknown in 510k | Tapered Screw type | Unknown in 510k | Unknown in 510k | Unknown in 510k |
| Sterilization | Gamma irradiation | Gamma irradiation | Gamma irradiation | Gamma irradiation | Gamma irradiation | Gamma irradiation | Gamma irradiation |
| Clinical procedure | Immediate loading or for loading after a conventional healing period | Immediate loading or for loading after a conventional healing period | Immediate loading or for loading after a conventional healing period | Immediate loading or for loading after a conventional healing period | Immediate loading or for loading after a conventional healing period | Immediate loading or for loading after a conventional healing period | Immediate loading or for loading after a conventional healing period |
Substantial Equivalence Discussion
# Discussion
The proposed subject device and Predicate/reference devices have common in all items in the comparison chart.
On the basis of the discussion above, it is concluded that the subject device is substantially equivalent to the predicate and reference devices.
{10}
AXA Abutment – Angled Type
| | Subject Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- |
| 510k | K251232 | K151909 | K210356 | K233450 |
| Device Name | AXA Abutment | Multi-unit abutment | Multi-unit abutment | AXA Abutment(Angled) |
| Manufacturer | MegaGen Implant Co., Ltd. | Noris Medical Ltd. | Noris Medical Ltd. | MegaGen Implant Co., Ltd. |
| Diameter (Ø, mm) | 5.0 | Unknown in 510k | Unknown in 510k | 4.0, 5.0 |
| Gingival Height (mm) | 45°: 2.0, 3.0, 4.0, 5.0 52°, 60°: 2.0 | 45°: 2.0* (*refer to K210356) | 45°: 3.0, 4.0, 5.0 52°, 60°: 2.0 | 3.8, 5.8, 7.8 3.8, 5.8, 7.8 3.74, 5.74, 7.74 |
| Angulation (°) | 45, 52, 60 | 45 | 45, 52, 60 | 20, 30 |
| Connection Interface | Internal Hex, Internal Non-Hex, Internal Octa, Internal Non-Octa | Unknown | Internal Hex 3.75mm platform | Internal Hex, Internal Non-Hex, Internal Octa, Internal Non-Octa |
| Material | Ti-6Al-4V ELI (ASTM F136-13) | titanium alloy Ti 6Al 4V ELI | Titanium alloy | Ti-6Al-4V ELI (ASTM F136-13) |
| Surface Treatment | Machined, Anodizing | Anodizing | Anodizing | Machined, Anodizing |
| Single Use | Yes | Yes | Yes | Yes |
| Sterilization | Supplied non-sterile. Single use Steam sterilized before use | Unknown in 510K | Supplied non-sterile. Single use Steam sterilized before use | Non-sterile |
| Additional Post Component | EZ Post Cylinder | Unknown in 510K | Unknown in 510K | EZ Post Cylinder |
| Compatible Implant System | MegaGen Zygoma Dental Implant System | Noris Medical Dental implants System | Noris Medical Dental implants System | BLUEDIAMOND IMPLANT System XPEED AnyRidge internal Implant System AnyOne Internal Implant System |
| Substantial Equivalence Discussion | | | | |
| Discussion The proposed subject device and Predicate/reference devices have common in all items in the comparison chart. And the fatigue test was performed as a representative of the worst-case model with angle. On the basis of the discussion above, it is concluded that the subject device is substantially equivalent to the predicate and reference devices. | | | | |
{11}
# Multi-unit Abutment
| | Subject Device | Predicate Device | Predicate Device | Reference Device 1 |
| --- | --- | --- | --- | --- |
| 510k | K251232 | K151909 | K210356 | K233450 |
| Device Name | Multi-unit Abutment | Multi-unit abutment | Multi-unit abutment | Multi-unit Angled Abutment |
| Manufacturer | MegaGen Implant Co., Ltd. | Noris Medical Ltd. | Noris Medical Ltd. | MegaGen Implant Co., Ltd. |
| Diameter (Ø, mm) | 4.8 | Unknown in 510k | Unknown | 4.8 |
| Gingival Height (mm) | 45°: 2.0, 3.0, 4.0, 5.0 52°, 60°: 2.0 | 45°: 2.0* (*refer to K210356) | 45°: 3.0, 4.0, 5.0 52°, 60°: 2.0 | 5.3, 6.3, 5.24, 6.24 |
| Angulation(°) | 45, 52, 60 | 45 | 45, 52, 60 | 17, 30 |
| Connection Interface | Internal Octa, Internal Non-Octa, Internal Hex, Internal Non-Hex, | Unknown | Internal Hex 3.75mm platform | Internal Octa, Internal Non-Octa, Internal Hex, Internal Non-Hex, |
| Material | Ti-6Al-4V ELI(ASTM F136-13) | titanium alloy Ti 6Al 4V ELI | Titanium alloy | Ti-6Al-4V ELI(ASTM F136-13) |
| Surface Treatment | Anodizing | Anodizing | Anodizing | Anodizing |
| Single Use | Yes | Yes | Yes | Yes |
| Sterilization | Supplied non-sterile. Single use Steam sterilized before use | Unknown in 510K | Supplied non-sterile. Single use Steam sterilized before use | Non-sterile |
| Compatible Implant System | MegaGen Zygoma Dental Implant System | Noris Medical Dental implants System | Noris Medical Dental implants System | BLUEDIAMOND IMPLANT System XPEED AnyRidge internal Implant System AnyOne Internal Implant System |
| Patient Population | Edentulous or partially edentulous individuals | Edentulous or partially edentulous individuals | Edentulous or partially edentulous individuals | Edentulous or partially edentulous individuals |
| **Substantial Equivalence Discussion** | | | | |
| **Discussion** The proposed subject device and Predicate/reference devices have common in all items in the comparison chart. And the fatigue test was performed as a representative of the worst-case model with angle. On the basis of the discussion above, it is concluded that the subject device is substantially equivalent to the predicate and reference devices. | | | | |
{12}
Abutment Screw
| | Subject Device | Reference Device 1 |
| --- | --- | --- |
| 510k | K251232 | K233450 |
| Device Name | Abutment Screw | Abutment Screw |
| Manufacturer | MegaGen Implant Co., Ltd. | MegaGen Implant Co., Ltd. |
| Diameter (Ø, mm) | 1.9, 1.95 | 2.0, 2.1, 2.2 |
| Total Length (mm) | 7.2~12.3 | 7.9 ~ 12.7 |
| Material | Ti-6Al-4V ELI (ASTM F136-13) | Ti-6Al-4V ELI (ASTM F136-13) |
| Surface Treatment | Machined | Machined, Anodizing |
| Single Use | Yes | Yes |
| Sterilization | Non-sterile | Non-sterile |
| Compatible Implant System | BLUEDIAMOND IMPLANT System XPEED AnyRidge internal Implant System AnyOne Internal Implant System | BLUEDIAMOND IMPLANT System XPEED AnyRidge internal Implant System AnyOne Internal Implant System |
Substantial Equivalence Discussion
# Similarities
The subject device has the same characteristic for the followings compared to the reference devices.
- Indication for use, Design, Material, Surface Treatment, Single Use, Sterilization and Compatible Implant System
# Differences
The subject device has the different characteristic for the followings compared to the reference device.
- Diameter
The diameters of the subject device are smaller than the reference devices, but the variety of the size can be possible to operate more precise treatment to meet each patient's condition. Therefore, it does not cause a matter in substantial equivalence.
- Total Length
The Total Length of the subject device is slightly different with the reference devices but it does not cause a matter in substantial equivalence since the size differences are very minor.
# Discussion
The proposed Abutment Screw and Predicate device have common in all the terms in the comparison chart except Diameter and Total length. These differences are explained not affecting on device's fundamental functions and safety. On the basis of the discussion above, it is concluded that the subject device is substantially equivalent to the Predicate device.
{13}
### 8. Summary of Non-Clinical Testing
The non-clinical testing data which are submitted, referenced, or relied on in this submission support demonstrating substantial equivalence.
#### Biocompatibility
The biocompatibility evaluation has been performed in accordance with International Standard ISO 10993-1 and ISO 10993-5. The subject device has same material composition, manufacturing process and patient contacting parts as our previously cleared device, XPEED AnyRidge Internal System (K122231), BLUEDIAMOND Implant System (K182448) and AnyRidge Internal Implant System (K110955).
#### Pyrogen and Endotoxin Test
The subject device will not be labeled as “non-pyrogenic”, and the endotoxin testing will be conducted on every batch for the subject device with the testing limit of below 20 EU/device in accordance with the USP 43 <85> and <161>.
#### Sterilization validation and Shelf life
The ZYGOMA IMPLANTs are supplied in sterile state. Sterilization validating testing has been performed in accordance with ISO 11137 to verify the sterility assurance level (10-6). The tests to validate the shelf life of the device through the proposed shelf life were conducted using the accelerated aging method in accordance to ASTM F1980 and the test results validated 5 years shelf life.
The AXA Abutment, Multi-unit Abutment and Abutment Screw are supplied in non-sterile state. Sterilization validating testing for steam sterilization by the user has been performed in accordance with ISO 17665-1 and ISO 17665-2 to verify the sterility assurance level (10⁻⁶). Validation Testing was conducted on a worst-case test article from our previously cleared device, K220562.
#### Modified Surface Treatment
The surface treatment evaluation has been performed in accordance with ‘Section 11 of Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments – Guidance for Industry and FDA Staff’.
The ZYGOMA IMPLANT has same surface and manufacturing process with the previously our cleared devices of BLUEDIAMOND IMPLANT (K241972) for the surface treatment of S.L.A.
The AXA Abutment, Multi-unit Abutment and Abutment Screw have the same anodized surface treatment and manufacturing process as our previously cleared device, MiNi Internal Implant System (K150537), AnyOne External Implant System (K203554) and BLUEDIAMOND IMPLANT System (K182448).
#### Performance test
Mechanical performance testing was conducted to evaluate the subject devices in accordance with the Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments.
Fatigue testing was performed using a deviated ISO 14801 test method(only potting the threaded apical end with no supporting alveolar bone) to reflect the intended clinical use of the ZYGOMA IMPLANT under an extramaxillary approach in use to ZAGA Class 4 only. All test results met the pre-established acceptance criteria, supporting substantial equivalence to the predicate devices.
#### MR Compatibility
The MR compatibility in additional simulation was performed to assess the risk of exposing patients who have implantable medical devices according to FDA’s guidance “Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment” and applicable ASTM standards (ASTM F2052-21, ASTM F2213-17, ASTM F2119-07, ASTM F2182-19e2).
Based on the MR testing, the subject devices are classified as MR Conditional. Patients implanted with the subject devices may be safely scanned under the specific MR conditions established and consistent with those of the predicate devices, confirming substantial equivalence with respect to MR safety.
{14}
## 9. Summary of Clinical Data
Clinical evidence supporting the safety and effectiveness of the Subject Device (MegaGen ZYGOMA Implant System) was derived from six clinical studies involving patients with severely atrophic maxillae, maxillary defects, and other complex maxillofacial conditions requiring implant-supported rehabilitation.
Across the six studies, approximately 250 patients received a total of 677 zygomatic implants, with follow-up periods ranging from several months to more than 7.5 years. The study populations included patients with severe maxillary resorption, edentulous maxillae, maxillary defects, and complex reconstructive conditions requiring zygomatic implant-supported rehabilitation.
The reviewed studies consistently demonstrated successful implant placement, restoration of oral function, and predictable prosthetic rehabilitation. Implant survival rates ranged from approximately 93% to 100%, with most studies reporting survival rates near or above 98%, supporting the long-term effectiveness of zygomatic implant treatment. Reported adverse events were infrequent and consistent with the known risks associated with zygomatic implant surgery. Observed complications included soft tissue recession, sinus-related findings, mucositis, cutaneous fistula, peri-implantitis, oroantral communication, temporary paresthesia, and a limited number of implant failures. These events were generally manageable through routine clinical intervention, and no previously unrecognized safety concerns or unexpected device-related adverse events were identified. Overall, the clinical evidence demonstrates that zygomatic implant systems provide a safe and effective treatment option for rehabilitation of severely compromised maxillary anatomy and supports the substantial equivalence of the Subject Device to the predicate devices.
## 10. Conclusions
Based on the information provided in this premarket notification, We, MegaGen Implant Co., Ltd. conclude that the subject device is substantially equivalent to the predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.