Retrospective clinical case study; Patient medical records/radiographic imaging
The sponsor provided retrospective clinical data from 18 patients (29 implants) to support the safety and performance of the Zygomatic Dental Implant System, specifically demonstrating a 96% success rate (osseointegration) at a minimum 6-month follow-up.
Retrospective clinical case study; Retrospective clinical case series; Follow-up/Duration: Minimum 6 months; Study Period: 2013-2014
Patients requiring dental extraction and immediate Zygomatic implant replacement; Sample Size: 18 patients (29 implants); Number of Sites: 1
Not applicable for this study
Osseointegration success rate
Indications for Use
Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Noris Medical Zygomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxillae.
Device Story
Zygomatic dental implant system; designed for atrophic maxilla cases. Implants feature tapered internal hex connection; sharp apical threads; lengths 30-57.5mm. System includes multi-unit and vari-connect prosthetic components (straight, 17°, 30°, 45°). Used by dental surgeons in clinical settings. Implants placed in upper jaw arch; support fixed or removable prostheses. Procedure involves one-stage or two-stage surgery; supports immediate loading if primary stability achieved. Output is structural support for dental prosthetics; aids patient chewing function. Benefits include restoration of oral function in patients with significant bone loss.
Clinical Evidence
Retrospective clinical study of 18 patients (29 Zygomatic implants) conducted in Israel (2013-2014). All patients underwent dental extraction and immediate implant replacement with immediate occlusal loading. Primary endpoint: osseointegration at minimum 6-month follow-up. Success rate reported as 96%. Evaluation methods included pre-op CT scans and pre/post-op panoramic radiographs.
Indicated for patients with partially or fully edentulous maxillae requiring dental implants in the upper jaw arch to support fixed or removable prosthetic devices and restore chewing function.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 8, 2016
Noris Medical Ltd. Ms. Simha Sibony VP Regulatory Science 8 Hataasia St. Nesher, 3688808 ISRAEL
Re: K151909
Trade/Device Name: Noris Medical Zygomatic Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: March 1, 2016 Received: March 9, 2016
Dear Ms. Sibony:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K151909
Device Name
Noris Medical Zygomatic Dental Implant System
Indications for Use (Describe)
Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Noris Medical Zygomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxillae.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017 See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K151909 510(k) Summary
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the Noris Medical Zygomatic Dental Implant System
### 1. GENERAL INFORMATION
| Date Prepared: | April 7, 2016 |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Noris Medical Zygomatic Dental Implant System |
| Classification Name: | Implant, Endosseous, Root-Form |
| Class: | II |
| Product Code: | DZE/NHA |
| CFR section: | 21 CFR§872.3640 |
| Device panel: | Dental |
| Legally Marketed<br>Predicate Device: | Primary Predicate: K093562-SOUTHERN IMPLANTS Inc.<br>K140440-Noris Medical Ltd. |
| Submitter: | Noris Medical Ltd. 8 Hataasia street, Nesher 3688808,<br>Israel |
| Contact 1: | Ms. Simha Sibony- VP Regulatory Science<br>Noris Medical Ltd<br>8 Hataasia St., Nesher 3688808 ISRAEL<br>M:+972 52-654-6625<br>T: +972 (73) 796-4477<br>F: +972 (4) 695-0991<br>E:simhas@norismedical.com |
| Contact 2: | Mr. Aharon Siev, President<br>Noris Medical Ltd<br>8 Hataasia St. Nesher 3688808<br>ISRAEL<br>T: +972(73)796-4477<br>F: +972(4)695-0991<br>E: asiev@norismedical.com |
## 2. DEVICE DESCRIPTION
- 2.1. BACKGROUND Noris Medical Dental Implants are tapered internal hex implants, designed to enable easy insertion while supporting excellent initial stability. The variable thread design enables self-tapping, thus providing
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in a bold, sans-serif font. Below the text "NORIS Medical" is the text "ENGINEERED FOR HEALTH" in a smaller, lighter font. The logo is clean and modern, and the use of red and gray gives it a professional look.
solutions for a variety of bone conditions. Noris Medical multi-design features offer a solution for immediate placement and immediate loading. The implantation procedure can be accomplished in a one-stage or twostage surgical operation for all implants type beside the Mono(Noris Medical cleared K140440) which is for one stage only. Packaging has been designed for quick identification and easy opening.
The scope of this submission is the following items:
### 2.2 NORIS MEDICAL ZYGOMATIC DENTAL IMPLANT
The Noris Medical Zygomatic Dental Implant is designed to provide a solution for cases of atrophic maxilla. The shape of the Noris Medical Zygomatic Dental Implant consists of sharp threads at the apical part. The platform of the Noris Medical Zygomatic Dental Implant is 3.75mm platform, with internal Hex connection. The Noris Medical Zygomatic Dental Implant is available in a large variety of lengths, from 30 mm to 57.5 mm.
Zygomatic Implant sizes and dimensions:
Image /page/4/Picture/6 description: The image shows a dental implant. The implant has a diameter of 4.2 mm and lengths of 30, 35, 37.5, 40, 42.5, 45, 47.5, 50, 52.5, 55, and 57.5 mm. The implant is made of titanium alloy Ti 6Al 4V ELI.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in a gray sans-serif font. Below the text is the tagline "ENGINEERED FOR HEALTH" in a smaller, lighter gray font.
### 2.3 PROSTHETIC COMPONENTS FOR NORIS MEDICAL ZYGOMATIC IMPLANT
Background: Straight, 17°, 30° Multi unit and Vari connect prosthetic system are already cleared under previous Noris Medical 510k submission -K140440
Multi unit and Vari connect 45° prosthetic system are subject of this current submission as a part of the Zyqomatic Dental Implant System.
#### Multi-Unit - screw retained reconstruction
The Multi-Unit system provides a solution for screw-retained prostheses even on complicated-to-restore implants (for example, multiple tilted implants). The Multi-Unit system comprises a full range of sizes for both the upper and lower jaws. Straight, 17°, 30°, 45° components, in a variety of heights, connects to a wide range of complementary products.
Image /page/5/Picture/6 description: The image shows a close-up of a brass-colored plumbing fitting. The fitting has a hexagonal base and a cylindrical top, with a joint in between that allows for angular adjustment. The fitting appears to be new and unused, with a clean and shiny surface.
Angulated 45° - Length 4 mm Made of titanium alloy Ti 6AI 4V ELI
### Vari-Connect - screw retained or removable reconstruction
The Vari-Connect system presents a complete solution for removable prostheses on tilted implants. It provides all the required equipment for removable prostheses, both on ball attachments and flat attachments, covering a wide range of possible situations. Straight, 17°, 30°, 45° components are available. Complementary products are fixed to the adaptors by the adaptor thread.
Image /page/5/Picture/10 description: The image shows a close-up of a brass-colored fitting. The fitting has a 90-degree angle and appears to be made of metal. The top part of the fitting has a threaded section, while the bottom part has a hexagonal shape. The fitting is likely used for connecting pipes or hoses in plumbing or other fluid transfer systems.
Angulated 45° - Length 4 mm Made of titanium alloy Ti 6AI 4V ELI
### Materials and Production:
The implants and prosthetic components are manufactured from Titanium alloy complying with standard ASTM F 136 Standard Specification for
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in red and gray. Below the text, there is a tagline that reads "ENGINEERED FOR HEALTH" in gray.
Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for surgical implant applications.
Noris Medical employs the RBM (Resorbable Blast Media) surface treatment technology.
The product is packed using materials known in the industry to be appropriate for medical device packaging and is provided with a minimum sterility assurance level of 10 ° validated in compliance with ANSI/AAMI/ISO 11137-1 Sterilization of healthcare products -- Radiation -- Part 1 Requirements for development, validation and routine control of a sterilization process for medical devices.
### 3. INTENDED USE
Noris Medical Dental Implants System is intended to replace missing tooth/ teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a onestage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Noris Medical Zygomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxillae.
## 4. SUBSTANTIAL EQUIVALENCE CLAIMED TO PREDICATE DEVICES
Identification of the Legally Marketed Predicate Devices Used to Claim Substantial Equivalence:
Noris Medical Zygomatic Dental Implant System is substantially equivalent to Primary predicate -Southern Implants Zygomatic Implant (K093562) in terms of intended use, design, materials used and performance testing. As well as to secondary predicate - Noris Medical Dental Implant System(K140440) in terms of Internal Hex connection, packaging and sterilization.
With regards to the intended use, material, design, characteristics and dimensions, the equivalence was determined through the points in following table:
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red square with a white diagonal line running through it, followed by the words "NORIS Medical" in red and gray. Below the words "NORIS Medical" is the phrase "ENGINEERED FOR HEALTH" in gray.
| Device Name | Southern Implants<br>Zygomatic Implant<br>(Primary Predicate) | Noris Medical<br>Dental Implants<br>system | Noris Medical<br>Zygomatic Dental<br>Implant System<br>(Current Submission) |
|-----------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| 510k | K093562 | K140440 | K151909 |
| Intended use | [1] | [2] | [3] |
| Patient<br>Population | Edentulous or partially<br>edentulous individuals | Edentulous or partially<br>edentulous individuals | Edentulous or partially<br>edentulous individuals |
| Material | Titanium Grade 4 | Titanium alloy | Titanium alloy |
| Diameter(mm) | 4.05 | Tuff Implant<br>3.3,3.75,4.2,5,6 | 4.2 |
| Length(mm) | 35, 40, 42.5, 45, 47.5,<br>50, 52.5, 55 | Tuff Implant Series:<br>6,8,10,11.5,13,16 | 35,37.5,40,42.5,45,47.<br>5, 50,52.5,55,57.5 |
| Abutments° | 55 | 0,15,17,25,30 | 45 |
| Surface treatment | Machined,<br>Oxidized surface | Sand blasted<br>Acid-etched | RBM(Resorbable Blast<br>Media) |
| Thread Design | Screw type | Tapered screw type | Tapered screw type |
| Sterility | Gamma irradiation | Gamma irradiation | Gamma irradiation |
| Clinical procedure | Immediate loading or<br>for loading after a<br>conventional healing<br>period | Immediate loading or<br>for loading after a<br>conventional healing<br>period | Immediate loading or<br>for loading after a<br>conventional healing<br>period |
[1] The Zygomatic implant is intended to be implanted in the upper jaw arch to provide support for fixed or removable dental prostheses in partially edentulous or full arch prostheses. It further adds the option for immediate function when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. This implant is not intended, nor should it be used, in conjunction with an angled abutment.
[2] NORIS Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Mono implants are specifically indicated for replacing maxillary lateral incisors and mandibular central and lateral incisors. They are used for immediate, non-occlusal provisionalization in single-tooth restorations. Multiple-unit restorations should be splinted together and may be used immediately when clinically appropriate.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the words "NORIS Medical" in a bold, sans-serif font. Below the company name, the tagline "ENGINEERED FOR HEALTH" is written in a smaller, lighter font.
[3] NORIS Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Noris Medical Zyqomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially of fully edentulous maxillae.
Evidence of equivalence has been demonstrated through the following: * The Noris Medical Zygomatic Dental Implant System's intended use and indications for use were previously cleared by FDA for the predicate devices.
* The technical characteristics of the Noris Medical Dental implant System are similar to those of the predicate devices.
* Nonclinical testing results for the Noris Medical Zygomatic Dental implants System are similar to those of the predicate devices.
Therefore, the Noris Medical Zygomatic Dental Implant System is substantially equivalent to the predicate devices in terms of intended use, materials used, and technological characteristics
## 4. NON-CLINICAL TESTING
The Noris Medical Zygomatic Dental Implant is manufactured from medical grade Titanium alloy per ASTM F136.
SEM and Surface analysis (EDS) after RBM process demonstrated the morphology and cleanliness of the final product.
Sterilization validation tests for the Noris Medical Dental Implants System were conducted in compliance with the following:
- ANSI/AAMI/ISO 11137-1:06 and EN ISO 11137-2:12 and ANSI/AAMI ST79:2010 & ● A1:2010.
Biocompatibility tests were conducted in compliance with the following:
- AAMI ANSI ISO 10993-5: 2009 (R)2014 Biological evaluation of medical devices ● Part 5: Tests for in vitro cytotoxicity
- AAMI ANSI ISO 10993-11: 2006 (R)2010 Biological evaluation of medical devices ● Part 11: Tests for systemic toxicity
Accelerated aging per ASTM-F-1980:07 has been applied on the final packaging followed by validating durability to peel, dye and burst tests conditions, in order to substantiate 5 years shelf life.
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in a dark gray sans-serif font. Below the company name, the text "ENGINEERED FOR HEALTH" is written in a smaller, lighter gray font.
Static and dynamic compression performance testing was conducted per ISO 14801: 07- Dentistry-Implants-Dynamic fatique test for Endosseous Dental implants. The worst case scenario was chosen based on the FDA guideline "Class II Special Controls Guidance Document: Root form for Endosseous dental implants and Endosseous dental Implant Abutments".
Comparative testing was conducted according to a modified ISO 14801 set-up to account for the nominal bone level for a zygomatic dental implant design
The results of the testing indicate that the Noris Medical Zygomatic Dental Implant is substantial equivalent to the predicate devices cited in this submission.
## 5. CLINICAL EVALUATION
Because the subject device design represented a new technology, featuring a different threading pattern from its primary predicate, clinical testing was requested. The sponsor provided a series of retrospective, randomized clinical case studies to support the proposed device. The device was implanted in 18 patients (a total of 29 Zygomatic implants). All patients required dental extraction and immediate subject device implant replacement. The location for Zygomatic implant was usually in the upper second premolar.
The implants were placed during the years 2013 to 2014 in one clinical site in Israel, and in all 29 cases, immediate occlusal loading was employed. Patients were evaluated at a minimum of 6 months post-surgery, with the success endpoint identified as osseointegration following implant placement. Computed Tomography (CT) was taken Pre-op. Panoramic radiographs were taken Pre-op, immediately after implantation and after at least 6 month follow up.
In the clinical study report, the sponsor presented long-term results of several patients implanted with the subject device. A success rate of 96% was obtained after at least 6 months follow up. Those results correspond to the CSR (Cumulative Success Rate) reported worldwide by the sponsor for Zygomatic dental implants.
## 6. CONCLUSION
The results of the testing conducted on the Noris Medical Zygomatic Dental Implant System demonstrated that the system is substantially equivalent to the predicate devices in terms of functional, mechanical properties, indications for use and material.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.