K210356 · Noris Medical , Ltd. · NHA · Feb 4, 2022 · Dental
Device Facts
Record ID
K210356
Device Name
Noris Medical Dental Implants System
Applicant
Noris Medical , Ltd.
Product Code
NHA · Dental
Decision Date
Feb 4, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K210356 · Feb 4, 2022
Noris Medical Dental Implants System
Noris Medical , Ltd.
Retrospective clinical case report (Noris Medical); Published retrospective cohort study (Armando L, et al., J. Clin. Med. 2021)
Retrospective clinical data were used to demonstrate the performance and success rates of the subject device's abutments (45°, 52°, 60°) compared to the predicate device, specifically addressing clinical concerns regarding the 60° abutment angulation.
Noris Medical Retrospective Clinical Case Report; Retrospective clinical case report; Follow-up/Duration: 6 to 48 months; Study Period: 2013-2020
33 patients requiring dental extraction and immediate replacement; Sample Size: 33 patients (88 Multi Unit abutments)
Not applicable for this study
Implant survival, stability, inflammation, soft tissue health
Indications for Use
Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Noris Medical Zygomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxillae.
Device Story
The Noris Medical Dental Implants System consists of endosseous dental implant abutments (Multi-Unit and Vari-Connect systems) used to support prosthetic devices. The system includes angled abutments (17°, 30°, 45°, 52°, 60°) and platform extenders to accommodate complex anatomical situations, such as tilted implants in atrophic maxillae. Components are manufactured from titanium alloy (ASTM F 136) and are anodized gold. The device is used by dental clinicians in surgical settings for one-stage or two-stage procedures. The abutments connect to dental implants to provide a stable interface for screw-retained or removable prostheses. By enabling the restoration of chewing function in edentulous or partially edentulous patients, the device improves patient quality of life. Clinical performance is supported by retrospective data demonstrating stability and success in prosthetic rehabilitation.
Clinical Evidence
Evidence includes a retrospective study of 33 patients (88 Multi-Unit abutments) with a follow-up period of 6-48 months. Primary endpoints included prosthetic success, implant stability, and absence of inflammation/complications. Results showed 100% survival rate for the Noris Medical system. Comparative real-world evidence from literature (Nobel Biocare Zygomatic system) was provided to support the safety of 60° angulated abutments, showing a 95.3% cumulative success rate at 2 years. Bench testing included comparative fatigue testing following a modified ISO 14801 approach.
Technological Characteristics
Materials: Titanium alloy (ASTM F 136). Abutments: Anodized gold. Design: Multi-Unit and Vari-Connect abutments with various angulations (17°, 30°, 45°, 52°, 60°) and gingival heights. Connection: Internal hex, 3.75 mm platform. Sterilization: Supplied non-sterile, steam sterilized before use. Connectivity: None (mechanical component).
Indications for Use
Indicated for edentulous or partially edentulous patients requiring replacement of missing teeth in either jaw, including the upper jaw arch for zygomatic implants, to support fixed or removable prosthetic devices and restore chewing function.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Noris Medical Ltd Simha Sibony QA/RA 8 Hataasia Nesher, 3688808 ISRAEL
Re: K210356
Trade/Device Name: Noris Medical Dental Implants System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: January 2, 2022 Received: January 5, 2022
Dear Simha Sibony:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K210356
Device Name
Noris Medical Dental Implants System
#### Indications for Use (Describe)
Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Noris Medical Zygomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxillae.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Noris Medical Dental Implant Solutions. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in gray. Below that, the text "Dental Implant Solutions" is also in gray.
# K210356
# 510(k) Summary
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the Noris Medical Dental Implants System
# 1. GENERAL INFORMATION
| Date Prepared: | February 4, 2022 |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade/Device Name: | Noris Medical Dental Implants System |
| Common Name: | Endosseous Dental Implant Abutment |
| Classification Name: | Abutment, Implant, Dental, Endosseous |
| Class: | II |
| Product Code: | NHA |
| CFR section: | 21 CFR§872.3630 |
| Device panel: | Dental Device Panel |
| Legally Marketed<br>Primary Predicate<br>Device: | K151909 - Noris Medical Ltd |
| Legally Marketed<br>Reference Devices: | K140440 - Noris Medical Ltd<br>K161598-Nobel Biocare A B |
| Submitter: | Noris Medical Ltd. 8 Hataasia street, Nesher 3688808, Israel |
| Contact 1: | Ms. Simha Sibony- Regulatory Affairs Consultant – GMRE<br>Ltd<br>RA/QA Consultant<br>M: +972 52-654-6625<br>T: +972 (73) 796-4477<br>F: +972 (4) 695-0991<br>E:simhasibony@gmail.com |
| Contact 2: | Mr. Udi Dailes - VP Executive<br>Noris Medical Ltd<br>8 Hataasia St. Nesher 3688808<br>ISRAEL<br>T: +972(54)2626719<br>F: +972(4)695-0991<br>E: udid@norismedical.com |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the words "NORIS Medical" in gray. Below "NORIS Medical" is the phrase "Dental Implant Solutions" also in gray. The logo is simple and modern, and the use of red and gray gives it a professional look.
# 2. DEVICE DESCRIPTION
# 2.1 NORIS MEDICAL MULTI UNIT AND VARI CONNECT ABUTMENTS
### Background:
Multi Unit, Vari Connect 52° and 60° 2mm height prosthetic system are subject of this current submission as a part of the Dental Implants System (K151909).
Multi Unit 17°, 30°, (K140440) with additional 4, 5 mm gingival height are subject of this current submission.
Vari Connect 17°, 30° (K140440) with additional 4, 5, 6 mm gingival heights are subject to this current submission.
Multi Unit 45°(K151909) with additional 4, 5 mm gingival height are subject of this current submission. Vari Connect 45°(K151909) with additional 4,5,6 mm gingival heights are subject to
Illustration of Gingival Height:
this current submission.
Multi Unit
Vari Connect
Image /page/4/Picture/11 description: The image shows a cross-sectional view of a threaded fitting. The fitting has an angled body with a threaded internal bore. The outer diameter of the fitting is labeled as 3.75, and the inner diameter is labeled as 2.42. The height of the fitting is also labeled.
Image /page/4/Picture/12 description: This image is a technical drawing of a part, showing its dimensions and features. The drawing includes measurements such as a diameter of 4.9, a height labeled as 'Height', and a platform with a diameter of 3.75. Additionally, the drawing specifies a thread size of M3 and a dimension of 2.1 on one side. The drawing provides a detailed view of the part's geometry and dimensions.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Noris Medical Dental Implant Solutions. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in gray. Below the text is the phrase "Dental Implant Solutions" in a smaller, lighter gray font.
# Vari Connect Extender
| Vari-Connect Platform Extender | | | | |
|----------------------------------|------------------------------------|------------------------------|------------------------------|----------------------------------------------|
| Platform Extender<br>Height (mm) | | | | |
| 2 | 3 | 4 | 5 | 6 |
| Image: Platform Extender 2mm | Image: Platform Extender 3mm | Image: Platform Extender 4mm | Image: Platform Extender 5mm | Image: Platform Extender 6mm |
| NM-T7282 | NM-T7283 | NM-T7284 | NM-T7285 | NM-T7286 |
| Image: Platform Extender Diagram | | | | |
| | Platform<br>Extender Height | | | Extender<br>Tightening<br>Torque<br>30-35Ncm |
| | Vari-Connect<br>Platform<br>Height | | | |
| | Total Platform<br>Height | | | |
By adding the extenders to Vari Connect 2mm height, the final heights of 4, 5,6,7,8 mm will be obtained.
The system also includes abutment fixation screws.
Noris Medical Multi Unit and Vari Connect 45°, 52°, 60° abutments are intended to be used only with Noris Medical Zygomatic Dental implant (K151909).
Noris Medical Multi Unit and Vari Connect 17°, 30° abutments are intended to be used only with Noris Medical Dental implants System(K140440).
### 2.2.1 Multi-Unit - screw retained reconstruction
The Multi-Unit system provides a solution for screw-retained prostheses even on complicated-to-restore implants (for example, multiple tilted implants). The Multi-Unit system comprises a full range of sizes for both the upper and lower jaws.
### 2.2.2 Vari-Connect - screw retained or removable reconstruction
The Vari-Connect system presents a complete solution for removable prostheses on tilted implants. It provides all the required equipment for removable prostheses, both
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in gray. Below the text, it says "Dental Implant Solutions" also in gray.
on ball attachments and flat attachments, covering a wide range of possible situations. Complementary products: Vari Connect extenders are fixed to the Vari Connect abutment by the extender's thread.
| ANGLED MULTI-UNIT | | | | | ANGLED VARI-CONNECT | | |
|--------------------|---------------------------------|----------------------------------------------------------------------------------------|----------|----------|---------------------|-----------------------------------|----------|
| Angled Multi-Units | | | | | Angle Vari-Connect | | |
| | Multi-Unit Platform Height (mm) | | | | | Vari-Connect Platform Height (mm) | |
| Angle° | 2 | 3 | 4 | 5 | Angle° | 2 | 3 |
| 17° | NM-A7112 | NM-A7113 | NM-A7114 | NM-A7115 | 17° | NM-A7212 | NM-A7213 |
| 30° | NM-A7133 | NM-A7134 | NM-A7135 | NM-A7136 | 30° | NM-A7233 | NM-A7234 |
| 45° | NM-A7144 | NM-A7145 | NM-A7146 | NM-A7147 | 45° | NM-A7244 | |
| 52° | NM-A7152 | Included with all multi-unit bases<br><br>Abutment Cary-on handle<br>NM-S7101 NM-X7101 | | | 52° | NM-A7252 | |
| 60° | NM-A7160 | | | | 60° | NM-A7260 | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Noris Medical Dental Implant Solutions. The logo consists of a red abstract shape on the left, followed by the words "NORIS Medical" in gray. Below "NORIS Medical" is the text "Dental Implant Solutions" in a smaller font, also in gray. The logo is clean and modern, with a focus on the company's name and its specialization in dental implants.
# 3. MATERIALS AND PRODUCTION:
The implants and prosthetic components are manufactured from Titanium alloy complying with standard ASTM F 136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for surgical implant applications. Abutments are anodized gold, the same as the reference device K140440.
# 4. INTENDED USE
Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Noris Medical Zygomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxillae.
# 5. SUBSTANTIAL EQUIVALENCE CLAIMED TO PREDICATE DEVICES
Identification of the Legally Marketed Predicate Devices Used to Claim Substantial Equivalence:
Noris Medical Multi Unit and Vari Connect 52°,60° abutments subject to this submission are substantially equivalent to Primary predicate – Noris Medical Dental Implant System (K151909) in terms of intended use, design, and materials used.
Reference Device - Nobel Biocare K161598 Multi Unit 60° abutment connected to Zygomatic implant 50 mm-NobelZygoma 0° is equivalent to current submission in terms of performance testing. Reference Device Noris Medical Dental Implant System (K140440) is equivalent to current submission in terms of similar 3.75 mm Platform and Internal Hex connection for all abutments components connected to the same Noris Medical Dental implants system.
With regards to the intended use, material, design, characteristics and dimensions, the equivalence was determined through the points in following pages:
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Noris Medical Dental Implant Solutions. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in gray. Below the text is the phrase "Dental Implant Solutions" in a smaller, lighter gray font.
# Table 1: Comparison of subject and predicate abutments characteristics (K151909)
| | Nobel Biocare<br>Multi Unit 60°<br>abutment<br>Reference<br>Device | Noris Medical<br>Zygomatic Dental<br>Implant System-<br>Primary Predicate<br>Multi Unit/Variconnect | Noris Medical<br>Dental Implant<br>System-<br>Current<br>Submission<br>Multi<br>Unit/Variconnect<br>Connected to Noris<br>Zygomatic Dental<br>Implant system<br>K151909 only | Equivalence<br>Discussion |
|-----------------------|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| 510k | K161598 | K151909 | Current Submission | - |
| Indication for<br>use | [3] | [2] | [4] | Identical<br>Same indication for<br>use for subject and<br>primary predicate |
| Patient<br>Population | Edentulous or<br>partially<br>edentulous<br>individuals | Edentulous or partially<br>edentulous individuals | Edentulous or<br>partially edentulous<br>individuals | Identical |
| Material<br>abutments | Titanium alloy | Titanium alloy | Titanium alloy | Identical<br>Subject and<br>predicate devices<br>have the same<br>material of<br>construction. |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for NORIS Medical, a company that provides dental implant solutions. The logo consists of a red abstract shape on the left and the words "NORIS Medical" in gray on the right. Below the company name, the words "Dental Implant Solutions" are written in a smaller, lighter gray font. The logo is simple, modern, and professional.
| | 45,60 | 45 | 45,52,60 | SE |
|-----------------------------------|------------------------------------------|-----------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Abutments° | | | | the angulation is<br>equivalent to the<br>reference device |
| Abutment<br>Gingival<br>Height-mm | 45°: 6mm,<br>8mm, 10mm<br>60°: 6mm,8mm | 2 mm | Multi Unit 45°:<br>3,4,5 mm<br>Variconnect 45°<br>extender:<br>3,4,5,6 mm<br>52°,60°: 2 mm | SE: The subject<br>device is equivalent<br>to reference device.<br>Dynamic fatigue<br>test of the subject<br>device shows<br>equivalent<br>performance in<br>comparison to<br>predicate device<br>See Section 6.3 |
| Screw<br>material | Titanium alloy | Titanium alloy | Titanium alloy | Identical |
| Compatible<br>implant<br>abutment | External Hex<br>Regular<br>Platform (RP) | Internal Hex<br>3.75mm platform | Internal Hex<br>3.75mm platform | Identical to<br>predicate device |
| Sterility | Supplied sterile<br>Single use | Supplied non-sterile.<br>Single use<br>Steam sterilized before<br>use | Supplied non-sterile.<br>Single use<br>Steam sterilized<br>before use | Identical to primary<br>predicate device |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for Noris Medical. The logo consists of a red abstract shape on the left and the words "NORIS Medical" in gray on the right. Below the company name, the words "Dental Implant Solutions" are written in a smaller font, also in gray.
# Table 2: Comparison of subject and predicate abutments characteristics (K140440)
| | Noris Medical<br>Dental Implants<br>system- Reference<br>Device | Noris Medical<br>Dental Implants System-<br>Current Submission<br>Multi Unit/Variconnect<br>Connected to Noris Medical<br>Dental implants K140440<br>only | Equivalence<br>Discussion:<br>Identical/SE |
|------------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k | K140440 | Current Submission | - |
| Intended use | [1] | [4] | Identical<br>Same indication for<br>use for subject and<br>primary predicate |
| Patient<br>Population | Edentulous or partially<br>edentulous individuals | Edentulous or partially<br>edentulous individuals | Identical |
| Material<br>abutments | Titanium alloy | Titanium alloy | Identical<br>Subject and predicate<br>devices have the<br>same material of<br>construction |
| Abutments° | 0,15,17,25,30 | Multi Unit/Variconnect 17,30 | Identical<br>the angulation is<br>equivalent to the<br>primary predicate<br>device |
| Abutment<br>Gingival Height-<br>mm | Multi Unit 17,30°:<br>2,3 mm<br>Variconnect 17,30°:<br>2mm | Multi Unit 17,30°:<br>4,5 mm<br>Variconnect 17°,30° extender:<br>2,3,4,5,6 mm | SE : The subject<br>device has additional<br>gingival Height.<br>Worst cases<br>calculations for Multi<br>unit with higher<br>angulations show<br>equivalent<br>performance in<br>comparison to their<br>primary predicate<br>device<br>See Section 6.3 |
| Screw material | Titanium alloy | Titanium alloy | Identical |
| Compatible<br>implant<br>abutment | Internal Hex<br>3.75mm platform | Internal Hex<br>3.75mm platform | Identical<br>Subject and primary<br>predicate devices are<br>compatible with the<br>same devices. |
| Sterility | Supplied non-sterile.<br>Single use<br>Steam sterilized before<br>use | Supplied non-sterile.<br>Single use<br>Steam sterilized before use | Identical to primary<br>predicate device |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image is the logo for Noris Medical Dental Implant Solutions. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in gray. Below the text is the phrase "Dental Implant Solutions" also in gray.
[1] Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in gray. Below the company name, the text "Dental Implant Solutions" is written in a smaller font, also in gray.
Mono implants are specifically indicated for replacing maxillary lateral incisors and mandibular central and lateral incisors. They are used for immediate, non-occlusal provisionalization in singletooth restorations. Multiple-unit restorations should be splinted together and may be used immediately when clinically appropriate.
[2] Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Noris Medical Zyqomatic Dental Implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices with partially or fully edentulous maxillae.
[3] Nobel Zygoma implants are endosseous implants intended to be surgically placed in the bone of the upper iaw arch to provide support for the prosthetic devices, such artificial teeth, in order to restore patient esthetics and chewing function. The NobelZygoma Implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
[4] Noris Medical Dental Implants System is intended to replace missing tooth/teeth in either jaw for supporting prosthetic devices that may aid in restoring the patient's chewing function. The procedure can be accomplished in a one-stage or two-stage surgical operation. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading,
Evidence of equivalence has been demonstrated through:
* The Noris Medical Dental Implants System intended use and indications for use are similar to previously cleared FDA cleared predicate and reference devices.
* The technical characteristics of the Noris Medical Dental implants System are similar to those of the predicate and reference devices.
* Clinical and non-clinical performance testing results of the Noris Medical Dental implants System are similar to those of the predicate and reference devices.
Therefore, the Noris Medical Dental Implants System is substantially equivalent to the predicate device in terms of intended use, materials used, and technological characteristic
### 6. NON-CLINICAL TEST
6.1 Biocompatibility tests were conducted in compliance with the following:
AAMI ANSI ISO 10993-5: 2009 (R) 2014 - Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
#### 6.2 Sterilization
Sterilization validation for subject devices was conducted according to ISO 17665-1, ISO 17665-2, and ANSI/AAMI ST79.
### 6.3 Mechanical tests
Comparative fatigue testing was performed on both zygomatic predicate systems as well as the subject devices, following a modified ISO 14801 approach to reflect the extra-maxillary surgical techniques of the subject and predicate systems, utilizing worst-case constructs for each system
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in gray. Below the text, it says "Dental Implant Solutions" in a smaller, lighter gray font.
# 7. CLINICAL DATA
Real-world evidence of Nobel Biocare Zygomatic implant system cases were provided from the literature to address the clinical concerns regarding the worst case angulation 60° Multi Unit abutment subject to this submission to legally marketed Nobel Biocare Multi Unit 60°
Nobel Biocare Multi Unit (predicate device) presented in Ref. 1 below is provided as real world evidence data: A retrospective study including a total of 44 completely edentulous maxillary rehabilitations (77 Multi unit abutments) were performed.
The zygomatic implants used in this study were NobelZygoma 0 with a TiUnite surface (Nobel Biocare AB, Göteborg, Sweden), and typically emerged between the first premolar and the first molar on the residual crest of the ridge, near its center).The study abutments were 45 and 60 Multi-unit abutments External Hex RP6 mm of height. The surgical protocol followed previous indications for zygomatic implants inserted through the extramaxillary surgical technique.
This implant is the predicate device presented in this current submission and the extramaxillary surgical technique is the same as Noris Medical surgical technique for Zygomatic implants.
### Outcome Measures
Outcome measures were evaluated at implant surgery and at 2 years post-surgery.
The primary outcome measures were prosthetic success, abutment success, abutment success and complications. modified plaque index (mPLI), modified bleeding index (mBI), mucosal seal efficacy evaluation (MSEE) >4 mm, and Zygomatic implants classification level (ZICL). Prosthetic success was judged in terms of function, being considered a failure if need to be replaced by a new prosthesis. Two patients (4.5%) were lost to follow-up. No prosthesis was lost; one patient lost one zygomatic implant; two angulated abutments of 60 degrees needed to be replaced in one patient due to an aesthetic complaint; rendering a cumulative success rate at 2-years of 95.3% and 95.9% using patient and implant/abutment as unit of analysis, respectively.
### Table 3: Summary of clinical cases of Nobel Biocare Zygomatic/ Multi Unit system
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for NORIS Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in gray. Below the text "NORIS Medical" is the text "Dental Implant Solutions" in a smaller, lighter gray font. The logo is clean and modern, and the use of red and gray gives it a professional look.
| (a) | | Cumulative Proportion Success at the Time | | | | |
|------------------|-------------------------------------------|-------------------------------------------|-------------------|---------------------------|--------------------------|-------------------------------|
| Time (Months) | Status<br>(0 = Success;<br>1 = Failure *) | Estimate | Std. Error | N of Cumulative<br>Events | N of Patients at<br>Risk | |
| 0 | 0 | | | 0 | 44 | |
| 7 | 1 | 0.977 | 0.022 | 1 | 43 | |
| 9 | 0 | | | 1 | 42 | |
| 10 | 0 | | | 1 | 41 | |
| 12 | 0 | | | 1 | 41 | |
| 13 | 0 | | | 1 | 40 | |
| 24 | 0 | | | 1 | 40 | |
| (b) | | | | | | |
| Duration | Total | Failed | Lost to Follow-Up | Censored | Survival Rate % | Cumulative<br>Survival Rate % |
| Placement—1 year | 77 | 1 | 1 | 5 | 98.7% | 98.7% |
| 1 year—2 years | 70 | 0 | 2 | 2 | 100% | 98.7% |
| 2 years—3 years | 52 | 0 | 0 | 25 | 100% | 98.7% |
* Failure was defined as the first implant to fail in one patient.
### Citation:
Ref. 1 : Angulated Abutments for Full-Arch Rehabilitation of Extremely Atrophic Maxillae : Short-Term Outcome of a Retrospective Cohort - Armando L, Miguel de Araújo Nobre, Ana Ferro, Carlos Moura Guede, Ricardo Almeida and Mariana Nunes , J. Clin. Med. 2021, 10, 3600
Noris Medical Multi Unit (subject device) clinical cases report was submitted.
This Noris Medical retrospective study included 33 patients (19 females and 14 males) with 88 Multi Unit abutment. The patients were on average 61 (36–86) years old. Patients required dental extraction and immediate replacement.
The implants were placed during the years 2013 to 2020. Patients were treated with Multi Unit abutments.
Panoramic radiographs and CT were taken Pre-op and Panoramic radiographs immediately after implantation and after at least 6 months follow up.
In this report 33 patients (88 Multi unit abutments) showed success relative to stability and no micro movements, no bleeding or soft tissue downgrowth, no pain was observed. Table 4 below provides a summary of the cases demonstrating that Noris Medical Zygomatic implant/MU abutment survival was 100 %( No failures). To evaluate the performance and local effects of implantation, 10 patients representing 14 clinical cases of Noris Medical Zygomatic implants system were provided to address concerns regarding the worst case angulation 60° Multi Unit abutment subject to this submission. Of the 33 patients noted, 10 used the 60° abutment.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Noris Medical. The logo consists of a red abstract shape on the left, followed by the text "NORIS Medical" in gray. Below the company name, the text "Dental Implant Solutions" is written in a smaller font size, also in gray.
r
# Table 4: Summary of clinical cases of Noris Medical Zygomatic/Multi Unit System
| Case<br>no. | Gender | Age<br>[yea<br>rs] | Position | Catalog<br>number | Follow-<br>up period<br>[month] | Angle<br>of<br>Multi<br>Unit | Status<br>(0=Suc<br>cess;<br>1=Failu<br>re) |
|-------------|--------|--------------------|--------------|-------------------|---------------------------------|------------------------------|---------------------------------------------|
| 1 | F | 79 | 4, 2, 13 | NM-A7144 | 8 | 45° | 0 |
| 2 | F | 53 | 4, 2, 13, 15 | NM-A7144 | 9 | 45° | 0 |
| 3 | F | 52 | 2, 13, 15 | NM-A7144 | 11 | 45° | 0 |
| 4 | F | 65 | 4, 2, 13, 15 | NM-A7144 | 13 | 45° | 0 |
| 5 | M | 52 | 2, 13, 15 | NM-A7144 | 8 | 45° | 0 |
| 6 | F | 56 | 4, 2, 13, 15 | NM-A7144 | 17 | 45° | 0 |
| 7 | F | 64 | 4, 2, 13, 15 | NM-A7144 | 7 | 45° | 0 |
| 8 | M | 71 | 4, 13 | NM-A7144 | 11 | 45° | 0 |
| 9 | M | 58 | 4, 2, 13, 15 | NM-A7144 | 10 | 45° | 0 |
| 10 | M | 51 | 2, 13, 15 | NM-A7144 | 14 | 45° | 0 |
| 11 | M | 68 | 4, 2, 15 | NM-A7144 | 13 | 45° | 0 |
| 12 | F | 55 | 2, 13, 15 | NM-A7144 | 8 | 45° | 0 |
| 13 | M | 64 | 4, 2, 13, 15 | NM-A7144 | 11 | 45° | 0 |
| 14 | M | 80 | 4, 2, 13, 15 | NM-A7144 | 10 | 45° | 0 |
| 15 | F | 60 | 4, 2, 13, 15 | NM-A7144 | 10 | 45° | 0 |
| 16 | F | 36 | 2, 13 | NM-A7144 | 6 | 45° | 0 |
| 17 | F | 52 | 2, 15 | NM-A7144 | 7 | 45° | 0 |
| 18 | F | 43 | 15 | NM-A7144 | 7 | 45° | 0 |
| 19 | M | 43 | 2 | NM-A7144 | 7 | 45° | 0 |
| 20 | F | 58 | 15 | NM-A7144 | 6 | 45° | 0 |
| 21 | M | 45 | 15 | NM-A7144 | 8 | 45° | 0 |
| 22 | M | 66 | 4 | NM-A7144 | 11 | 45° | 0 |
| 23 | F | 67 | 6 | NM-A7152 | 13 | 52° | 0 |
| 24 | F | 61 | 3, 14 | NM-A7145 | 47 | 45° | 0 |
| 24 | F | 61 | 6 | NM-A7152 | 47 | 52° | 0 |
| 24 | F | 61 | 11 | NM-A7160 | 47 | 60° | 0 |
| 25 | F | 86 | 4, 2,13,15 | NM-A7145 | 11 | 45° | 0 |
| 25 | F | 86 | 7 | NM-A7160 | 11 | 60° | 0 |
| 26 | M | 43 | 3, 14 | NM-A7152 | 48 | 52° | 0 |
| 26 | M | 43 | 6, 11 | NM-A7160 | 48 | 60° | 0 |
| 27 | M | 72 | 3 | NM-A7160 | 13 | 60° | 0 |
| 28 | F | 68 | 6, 11 | NM-A7160 | 12 | 60° | 0 |
| 29 | F | 67 | 4 | NM-A7160 | 13 | 60° | 0 |
| 30 | M | 78 | 6, 11 | NM-A7160 | 15 | 60° | 0 |
| 31 | M | 81 | 6, 4 | NM-A7160 | 12 | 60° | 0 |
| 32 | F | 81 | 11 | NM-A7160 | 14 | 60° | 0 |
| 33 | F | 79 | 6 | NM-A7160 | 14 | 60° | 0 |
| 33 | F | 79 | 4, 11,14 | NM-A7144 | 14 | 45° | 0 |
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for Noris Medical Dental Implant Solutions. The logo consists of a red abstract shape on the left, followed by the words "NORIS Medical" in gray. Below that, the words "Dental Implant Solutions" are also written in gray. The logo is simple and modern, and it conveys a sense of professionalism and expertise.
### Conclusion:
The Nobel Biocare MU cases presented in this report (real world evidence data -see ref. 1 above) demonstrated substantially equivalent success of the predicate Device All the Noris Medical MU cases presented in the submitted cases report demonstrate substantial equivalence results compared to Nobel Biocare predicate device in term of status of success.
Table 5 below details the Number of different Multi Unit angles that is included in the Noris Medical cases report and in the Nobel Biocare real world evidence clinical data.
Table 5: No. of MU abutment Cases- Noris Medical vs Nobel Biocare
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for Noris Medical Dental Implant Solutions. The logo consists of a red abstract shape on the left, followed by the words "NORIS Medical" in gray. Below the words "NORIS Medical" is the phrase "Dental Implant Solutions" in a smaller, lighter gray font. The logo is clean and modern, with a focus on the company's name and its area of expertise.
| Multi-unit Angle<br>(degrees) | No. of abutments used –<br>Noris Medical | No. of abutments<br>used –Nobel Biocare |
|-------------------------------|------------------------------------------|-----------------------------------------|
| 45 | 70 | 31 |
| 52 | 4 | - |
| 60 | 14 | 46 |
| Total: | 88 | 77 |
No adverse events norinflammation were reported and the follow up showed 100% success of the Noris Medical MU abutment by attaining the end point of continuous stability of implant with no inflammation after at least 6 months.
Survival Rate of the 60° Zygomatic implant/abutment system Noris Medical was 100% vs Nobel Biocare was 93.5% within 6 months to 48 months Follow up and 24 months Follow up respectively.
# 8. CONCLUSION
The data presented in this submission demonstrates that the subject devices are substantially equivalent with respect to performance and intended use. The proposed devices perform as well as the legally marketed primary and reference devices. The subject device, the primary predicate device, and the reference device have the same intended use, have similar technological characteristics. The subject device and the primary predicate, are made of identical materials, are packaged in the same materials, and are sterilized using the same methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Furthermore, the subject devices do not pose any new or increased risks as compared to the legally marketed predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.