308nm UV Phototherapy System (UV-K); 308nm UV Phototherapy System (UV-X); 308nm UV Phototherapy System (UV-Y); 308nm UV Phototherapy System (UV-Z)
K244022 · Boston Aesthetics, Inc. · FTC · Mar 13, 2025 · General, Plastic Surgery
Device Facts
Record ID
K244022
Device Name
308nm UV Phototherapy System (UV-K); 308nm UV Phototherapy System (UV-X); 308nm UV Phototherapy System (UV-Y); 308nm UV Phototherapy System (UV-Z)
Applicant
Boston Aesthetics, Inc.
Product Code
FTC · General, Plastic Surgery
Decision Date
Mar 13, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 308nm UV Phototherapy System is intended to be used for the treatment of psoriasis, vitiligo, seborrheic dermatitis, atopic dermatitis, and leukoderma. It is used on intact skin only.
Device Story
Portable 308nm UV phototherapy system (models UV-K, UV-X, UV-Y, UV-Z) for localized dermatologic treatment. Device utilizes LED light board to emit narrowband UVB light (308nm ±2nm). Operated by physicians or patients in hospitals, clinics, or home environments. System includes host unit, power adapter, and battery (for UV-X, UV-Z). User controls treatment duration via hardware interface; output is delivered directly to affected skin areas. Provides targeted phototherapy to manage skin disorders; intended to improve patient outcomes through controlled UV exposure. Safety features include adherence to IEC 60601 series standards for electrical and photobiological safety.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via electrical safety (IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-57, 60601-2-83), photobiological safety (IEC 62471), and biocompatibility (ISO 10993-5, 10993-10, 10993-23).
Indicated for the treatment of psoriasis, vitiligo, seborrheic dermatitis, atopic dermatitis, and leukoderma in patients requiring ultraviolet radiation therapy. For use on intact skin only.
Regulatory Classification
Identification
An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date March 13, 2025. The month is March, the day is the 13th, and the year is 2025. The date is written in a clear, legible font. The text is black on a white background.
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
Boston Aesthetics INC Cao Hongmei General Manager 1521 Concord Pike Suite 201 Wilmington, Delaware 19803
Re: K244022
Trade/Device Name: 308nm UV Phototherapy System (UV-K); 308nm UV Phototherapy System (UV-X); 308nm UV Phototherapy System (UV-Y); 308nm UV Phototherapy System (UV-Z) Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet Lamp For Dermatologic Disorders Regulatory Class: Class II Product Code: FTC Dated: December 26, 2024 Received: December 27, 2024
Dear Cao Hongmei:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by YAN FU -S
YAN FU-S Date: 2025.03.13 19:17:31
-04'00'
for Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K244022
Device Name
308mm UV Phototherapy System (UV-K): 308mm UV Phototherapy System (UV-X): 308nm UV Phototherapy System (UV-Y); 308nm UV Phototherapy System (UV-Z)
#### Indications for Use (Describe)
The 308nm UV Phototherapy System is intended to be used for the treatment of psorrasis, vitiligo, seborrheic dermatitis, atopic dermatitis, and leukoderma. It is used on intact skin only.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# K244022 - 510(k) Summary
This summary of 510(K) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR §807.92.
#### 1. Administrative Information
#### Submission Date
December 26, 2024
#### Submission Correspondent
Name: Boston Aesthetics INC Address: 1521 Concord Pike Suite 201 Wilmington DE 19803 Tel: +001 949-792-8168 E-mail: bsnaesthetics@gmail.com Contact: Ms. Hongmei Cao
### 2. Device Information
| Type of 510(k) Submission: | Traditional |
|-------------------------------|---------------------------------------------|
| Device Name: | 308nm UV Phototherapy System |
| Model: | UV-K; UV-X; UV-Y; UV-Z |
| Device Classification: | Light, Ultraviolet, Dermatological |
| Regulation Description: | Ultraviolet lamp for dermatologic disorders |
| Regulation Medical Specialty: | General & Plastic Surgery |
| Regulation Number: | 878.4630 |
| Product Code: | FTC |
| Device Class: | 2 |
### 3. Predicate Device
| 510(K) Number: | K172273 |
|-------------------------------|--------------------------------------------------|
| Device name: | 308nm Excimer System |
| Device model: | XECL-308C |
| Manufacturer: | Chongqing Peninsula Medical Technology Co., Ltd. |
| Classification Name: | Ultraviolet lamp for dermatologic disorders |
| Regulation Medical Specialty: | General&Plastic Surgery |
| Product code: | FTC |
| Regulation number: | 878.4630 |
{6}------------------------------------------------
### 4. Reference Device
| Items | Reference 1 | Reference 2 |
|-------------------------------|---------------------------------------------|-----------------------------------------------------------|
| 510(K) Number: | K170489 | K191571 |
| Manufacturer: | Clarify Medical, Inc. | Xuzhou Yongkang Electronic<br>Science Technology Co., Ltd |
| Device name: | Clarify Phototherapy System | UV Radiation Treatment System |
| Model: | N/A | YK-6000B |
| Classification Name: | Ultraviolet lamp for dermatologic disorders | |
| Regulation Medical Specialty: | General&Plastic Surgery | |
| Product code: | FTC | |
| Regulation number: | 878.4630 | |
#### 5. Device Description
The 308nm UV Phototherapy System is available in four models, UV-K , UV-Y and UV-Z. It is a portable medical device, which consists of LED light board, heat dissipation module, control circuit, drive circuit, battery (applicable to UV-X, UV-Z), and power adapter. It is a therapeutic product under the direction of a physician for individuals who require ultraviolet radiation for diagnosed skin disorders, which can be used in hospital, clinics and households.
The device is mainly composed of a host and a power adapter. The LED light board of the device host contains light emitting diodes (LEDs), which create narrowband UVB light centered at a wavelength 308nm(±2nm) for the purpose of phototherapy. They are intended to be use in localized phototherapeutic treatment of dermatologic conditions(excluding eyes), such as psoriasis, vitiligo, seborrheic dermatitis, atopic dermatitis, and leukoderma. It is for the partial treatment and is intended to be used on intact skin only.
The materials of the parts of the 308nm UV Phototherapy System that come into contact with the human body are mainly: PC, ABS.
## 5. Indication for Use
The 308nm UV Phototherapy System is intended to be used for the treatment of psoriasis, vitiligo, seborrheic dermatitis, atopic dermatitis, and leukoderma. It is used on intact skin only.
{7}------------------------------------------------
## 6. Comparison with predicate device
A comparison of the intended use and technological characteristics of the subject device is provided in the table below:
| Items | Subject Device (K######) | Predicate Device (K172273) | Reference Device (K170489) | Reference Device (K191571) | Comparison |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Device name | 308nm UV Phototherapy System | 308nm Excimer System | Clarify Medical Phototherapy<br>System | UV Radiation Treatment<br>System | / |
| Device model | UV-K, UV-X, UV-Y, UV-Z | XECL-308E | N/A | YK-6000A | / |
| Device picture | Image: 308nm UV Phototherapy System | Image: 308nm Excimer System | Image: Clarify Medical Phototherapy System | Image: UV Radiation Treatment System | / |
| Type of Use | Prescription Use | Prescription Use | Prescription Use | Prescription Use | Same |
| Use environment | Hospitals, clinics and households | Hospitals, clinics and<br>households | Home / Physician's office | Hospitals, clinics and<br>households | Same |
| Indications for use | The 308nm UV Phototherapy System is<br>intended to be used for the treatment of<br>psoriasis, vitiligo, seborrheic dermatitis,<br>atopic dermatitis, and leukoderma. It is<br>used on intact skin only. | The 308nm Excimer System is<br>intended to be used for the<br>treatment of psoriasis and<br>vitiligo. It is to be used on intact<br>skin only. | The Clarify Medical<br>Phototherapy System is an<br>Ultraviolet Light Emitting<br>Medical Device. It is intended<br>for use in localized<br>phototheraputic treatment of<br>dermatologic conditions such<br>as psoriasis, vitiligo, atopic<br>dermatitis (eczema), seborrheic<br>dermatitis and lenkoderma on<br>all skintypes (I-VI). | The UV Radiation Treatment<br>System models YK-6000A, YK-<br>6000A-T, YK-6000B, and YK-<br>6000B-T, are intended for use<br>under the direction of a<br>physician, for the treatment of<br>vitiligo, psoriasis, and eczema.<br>The devices are intended for<br>treatment of these conditions<br>on Fitzpatrick Skin Types (I-VI). | Similar &<br>Comparable |
| Mode of operation | Handheld | Handheld | Handheld | Handheld | Same |
| Re-usable | Yes | Yes | Yes | Yes | Same |
| Light source | LED UV lamps | Xenon-Chlorine (XeCI) excimer<br>lamp | LED UV lamps | LED UV lamps | Same |
| UV spectral | UVB | UVB | UVB | UVA&UVB | Same |
| UV light<br>wavelength | 308nm±2nm | 308nm±2nm | 300nm-320nm | UVA: 320nm~400nm<br>UVB: 300nm~320nm | Same |
| Max Power Output | UV-K&UV-X: 30 mW/cm²<br>UV-Y&UV-Z: 50 mW/cm² | 50mW/cm² | 3-15mW/cm² | 10mW/cm² | Comparable |
| Items | Subject Device (K######) | Predicate Device (K172273) | Reference Device (K170489) | Reference Device (K191571) | Comparison |
| Maximum<br>treatment dose | ≤ 5J/cm2 | 2J/cm2 | Not Publicly Available | Not Publicly Available | Comparable |
| Action area | Partial treatment excluding eyes | Partial treatment excluding<br>eyes | Not Publicly Available | Partial treatment excluding<br>eyes | Same |
| Treatment area | 3cm2 ~ 3.41cm² | 16cm2 | 4.5cm2 | Max 15cm² | Comparable |
| Treatment time | UV-K&UV-X: 1~140s<br>UV-Y&UV-Z: 1~100s | 1~40s | Not Publicly Available | 5~20 min | Comparable |
| Display Type | OLED | Not Publicly Available | Not Publicly Available | LED | Same |
| User Interface | Hardware interface | Hardware interface | Touch screen on mobile device<br>and start button on host body | Hardware interface | Same |
| Power source | UV-K & UV-Y: Adapter<br>UV-X & UV-Z: Adapter + lithium battery | Adapter | Rechargeable battery | AC outlet | Same |
| Operating<br>environment | Ambient temperature: 5℃~35℃<br>Relative humidity: ≤ 80%<br>Atmospheric pressure:<br>780hPa~1060hPa | Ambient temperature:<br>15℃~35℃<br>Relative humidity: ≤ 80%<br>Atmospheric pressure:<br>860hPa~1060hPa | Not Publicly Available | Not Publicly Available | Comparable |
| Transport/Storage<br>environment | Ambient temperature: -20℃~50℃<br>Relative humidity: ≤ 93%<br>Atmospheric pressure:<br>700hPa~1060hPa | Ambient temperature:<br>-20℃~45℃<br>Relative humidity: 10-85%RH<br>Atmospheric pressure:<br>500hPa~1060hPa | Not Publicly Available | Not Publicly Available | Comparable |
| Safety &<br>Performance | IEC60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 62471<br>IEC 60601-2-57<br>IEC 60601-2-83 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-57<br>IEC 62471 | IEC60601-1<br>IEC 60601-1-2<br>IEC 60601-1-4<br>IEC 60601-2-57<br>IEC 62471 | IEC60601-1<br>IEC 60601-1-2<br>IEC 60601-2-57<br>IEC 62471 | Same |
| Patient-Contact<br>materia | ABS; PC | ABS; PC | ABS | ABS | Same |
{8}------------------------------------------------
The subject device and the predicate device have intended use. Although the subject device have several different technological characteristics as noted in the differences do not raise different questions of safety and effectiveness. Therefore, we consider that the subject device is as safe and effective as the predicate device.
{9}------------------------------------------------
## 7. Summary of Non-Clinical performance Testing
#### 7.1. Electromagnetic Compatibility and Electrical Safety Test
The subject device has passed safety testing in according to following standards.
- AAMI/IEC 60601-1:2005/(R)2012 and A1:2012,C1:2019/(R)2012 and A2:2010/(R)2012 Medical 1) electrical equipment - Part 1: General requirements for basic safety and essential performance.
- 2) ANSI AAMI IEC 60601-1-2: 2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances-Requirements and tests.
- 3) IEC 60601-1-11 Edition 2.0 2015-01 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- 4) IEC 60601-2-57 Edition 1.0 2011-01 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- 5) IEC 60601-2-83 Edition 1.0 2019-05 Medical electrical equipment -Part 2-83: Particular requirementsfor the basic safety and essential performance of home lighttherapy equipment
- 6) IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems
#### 7.2. Biocompatibility Test
Biocompatibility testing was conducted in accordance with the 2020 FDA guidance "Use of International Standard ISO 10993, Biological Evaluation of Medical Device Part 1: Evaluation and Testing." The testing includes:
- 1) Cytotoxicity per ISO 10993-5: 2009 Biological evaluation of medical devices - Part 5: Tests for in vitro Cytotoxicity.
- Sensitization per ISO 10993-10: 2021 Biological evaluation of medical devices Part 10: Tests for 2) skin sensitization
- 3) Irritation per ISO 10993-23: 2021 Biological evaluation of medical devices - Part 23: Tests for irritation
The user- contacting materials were shown to be non-cytotoxic, non-irritating and non-sensitizing.
#### 7.3. Software
Software documentation of the subject device was provided in accordance with the FDA guidance Document- "Content of Premarket Submissions for Device Software Functions: Guidance for Industry and Food and Drug Administration Staff", which was issued in 06/14/2023 to support a device's Basic Documentation Level.
## 8. Conclusion
The subject device and the predicate device have the same intended use and any difference in the technological characteristics does not raise any new issues or concerns of safety or effectiveness. The results of the testing described above demonstrate that the subject device is as safe and effective as the predicate device (K172273) and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.