K170489 · Skylit Medical · FTC · May 23, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170489
Device Name
Skylit Phototherapy System
Applicant
Skylit Medical
Product Code
FTC · General, Plastic Surgery
Decision Date
May 23, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Clarify Medical Phototherapy System is an Ultraviolet Light Emitting Medical Device. It is intended for use in localized phototheraputic treatment of dermatologic conditions such as psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and lenkoderma on all skintypes (I-VI).
Device Story
Handheld UV light-emitting diode (LED) device for localized phototherapy; delivers UVB light (300-320 nm). System includes handheld unit, mobile smartphone application (Android/iOS), and cloud-based physician portal. Physician prescribes treatment; patient uses handheld device at home or in clinic. Smartphone app serves as primary interface, providing treatment guidance, reminders, and dose adjustment based on physician protocol. Patient can capture photographs for comparison. Cloud portal links patient to physician for remote monitoring and data review. Device benefits patients by enabling home-based phototherapy with physician oversight, improving adherence and treatment management.
Clinical Evidence
Bench testing only. Performance testing verified optical emissions (300-320nm), electrical safety, battery performance, and biocompatibility (ISO 10993). System and software validation confirmed functionality of prescription storage, treatment reminders, auto-calculation of dose adjustments, and wireless communication. No clinical trial data presented.
Indicated for localized phototherapeutic treatment of psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis, and leukoderma in patients of all skin types (I-VI).
Regulatory Classification
Identification
An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W O66-G609 Silver Spring, MD 20993-0002
May 23, 2017
Skylit Medical Don Canal Vice President 401 North A Street Suite 775 San Diego, California 92101
Re: K170489
Trade/Device Name: Skylit Phototherapy System Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet Lamp For Dermatologic Disorders Regulatory Class: Class II Product Code: FTC Dated: February 22, 2017 Received: February 23, 2017
Dear Don Canal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{2}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely, Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170489
Device Name Clarify Medical Phototherapy System
#### Indications for Use (Describe)
The Clarify Medical Phototherapy System is an Ultraviolet Light Emitting Medical Device. It is intended for use in localized phototheraputic treatment of dermatologic conditions such as psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and lenkoderma on all skintypes (I-VI).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# "510(k) Summary" as required by section 807.92(c)
| Date Prepared: | May 22, 2017 |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer Name | Clarify Medical, Inc.<br>401 N. A Street, Suite 775<br>San Diego, CA 92101, USA<br>Telephone: (972) 955-7644 |
| Official Contact | Don Canal<br>Regulatory Affairs<br>401 N. A Street<br>San Diego, CA 92101, USA<br>Telephone:(972) 955-7644<br>Email: don@clarifymed.com |
# DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Clarify Phototherapy System |
|---------------------------------------------------|------------------------------------------------------------------------------------|
| Common Name | Light, Ultraviolet, Dermatological |
| Classification Name<br>Classification Regulations | Ultraviolet lamp for dermatologic disorders<br>21 CFR<br>878.4630<br>Class II |
| Product Code | FTC |
| Submission Type | 510(k) |
| Classification Panel<br>Premarket Review | General & Plastic Surgery<br>ODE/DSD/General Surgery Devices Branch One<br>(GSDB1) |
# Predicate Devices
The Clarify System is substantially equivalent to the other legally marketed phototherapy, UVB treatment devices. Specifically, the Clarify Phototherapy System is substantially equivalent to the Lerner Medical Inc. Levia Phototherapy System (K040062/K090097), Daavlin Distributing Co. DermaPal (K073587) and Allux, Inc. Resolve UVB Phototherapy System (K072035).
## Device Description
The Clarify Phototherapy System is a handheld medical device that produces UVB light for the purpose of phototherapy. It was developed and will be marketed by Skylit Medical, Inc. DBA Clarify Medical, Inc. and is referred to herein as "Clarify Medical" or "Clarify". The Clarify Medical Phototherapy System is an Ultraviolet Light Emitting Medical device.
It is indicated for use in localized phototherapeutic treatment of dermatologic conditions
{5}------------------------------------------------
such as psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and leukoderma on all skin types (I-VI). The system consists of the following components and accessories:
- Handheld UV light Emission device that delivers the UVB light 1.
- 2. Mobile Software Application that runs on smartphone (Android and iOS) platforms. This Application is the primary user interface, and provides patient guidance when phototherapy is delivered and treatment reminders
- 3. Physician Portal Software is a cloud based software application that links the patient to the prescribing physician during home therapy.
## Federal (U.S.) law restricts this device to sale by or on the order of a physician.
The Clarify System delivers UV light between the wavelengths of 300 and 320 nm using Light Emitting Diodes (LEDs). The handheld device is powered by a rechargeable battery and is connected to a smartphone software application that enables the device and provides feedback during treatment. The smartphone also provides reminders on treatment days and provides a means for patient feedback and dose adjustment. The smartphone application also provides the means for the patient to take photographs for comparison purposes.
## Indications for Use Statement
The system is indicated for the treatment of psoriasis, vitiligo, atopic dermatitis (eczema). seborrheic dermatitis and leucoderma on all skin types (I-VI). The device can be used in the comfort of a patient's home or in a physician's office.
## Comparison to Predicate Devices
The Clarify System is substantially equivalent to the other legally marketed phototherapy, UVB treatment devices. Specifically, the Clarify Phototherapy System is substantially equivalent to the Lerner Medical Inc. Levia Phototherapy System (K040062/K090097). Daavlin Distributing Co. DermaPal (K073587) and Allux, Inc. Resolve UVB Phototherapy System (K072035). The detailed comparison between the Subject Device and the Predicate devices is provided in Table 1 below.
{6}------------------------------------------------
| | Subject Device | Predicate Device | Predicate Device | Predicate Device |
|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Clarify Medical<br>Phototherapy System<br>(Subject Device) | Lerner Medical<br>Devices, Inc.<br>Levia Phototherapy<br>System / LH-75T<br>Phototherapy<br>System<br>K040062 / K090097 | Daavlin Distributing<br>Co.<br>DermaPal K073587 | Allux<br>Resolve<br>UVB<br>Photothera<br>py System<br>K072035 |
| | Image: Clarify Medical Phototherapy System | Image: Levia Phototherapy System | Image: DermaPal K073587 | Image unavailable |
| Indications<br>for Use | The Clarify Medical<br>Phototherapy System is<br>an Ultraviolet Light<br>Emitting Medical<br>device. It is intended<br>for use in localized<br>phototherapeutic<br>treatment of<br>dermatologic<br>conditions such as<br>psoriasis, vitiligo,<br>atopic dermatitis<br>(eczema), seborrheic<br>dermatitis and<br>leukoderma on all skin<br>types (I-VI). | K040062: The Levia<br>Phototherapy System is<br>intended for use in<br>UVB phototherapy for<br>the treatment of<br>psoriasis, vitiligo,<br>atopic dermatitis<br>(eczema), seborrheic<br>dermatitis and<br>leucoderma. The Levia<br>Phototherapy System is<br>intended for use on all<br>skin types (I-VI).<br><br>K090097: The LH-75T<br>Phototherapy System is<br>intended for use, by or<br>under the direction of a<br>physician, in UVB<br>phototherapy for the<br>treatment of psoriasis,<br>vitiligo, atopic<br>dermatitis (eczema),<br>seborrheic dermatitis<br>and leucoderma. The<br>LH-75T Phototherapy<br>System is intended for<br>use on all skin types (I<br>-VI). | The DermaPal® is a<br>handheld ultraviolet<br>light emitting<br>fluorescent lamp. It is<br>intended for use, by or<br>under the direction of a<br>physician, for the<br>treatment of psoriasis,<br>vitiligo, and atopic<br>dermatitis (eczema) on<br>all skin types (I - VI). | The Resolve™ UVB<br>Phototherapy System is<br>an ultraviolet light<br>emitting medical device<br>for localized<br>phototherapeutic<br>treatment of<br>dermatologic conditions<br>such as psoriasis,<br>vitiligo, atopic<br>dermatitis (eczema),<br>and seborrheic<br>dermatitis. The<br>Resolve™ UVB<br>Phototherapy System is<br>intended to be used<br>with all Skin Types (I -<br>VI). |
| Product Codes<br>/ Regulation<br>Number | FTC / 21 CFR<br>878.4630 | FTC / 21 CFR<br>878.4630 | FTC / 21 CFR<br>878.4630 | FTC / 21 CFR<br>878.4630 |
| Treatment<br>area | $4.5 cm^2$ | $3 cm^2$ | 11.4 x 2.54 cm | 12.8 x 9.0 cm |
| | Subject Device | Predicate Device | Predicate Device | Predicate Device |
| | Clarify Medical<br>Phototherapy System<br>(Subject Device) | Lerner Medical<br>Devices, Inc.<br>Levia Phototherapy<br>System / LH-75T<br>Phototherapy<br>System<br>K040062 / K090097 | Daavlin Distributing<br>Co.<br>DermaPal K073587 | Allux<br>Resolve<br>UVB<br>Photothera<br>py System<br>K072035 |
| UVB light<br>source | LED UV lamps | Arc Lamp | Fluorescent UV lamps | LED UV lamps |
| User<br>Interface | Touch screen on<br>Mobile Device and<br>Start button on | Control Box with<br>cables to handheld<br>device | Control Box with<br>cables to handheld<br>device | Control Box with<br>cables to handheld<br>device |
| Max Power<br>Output | 3-15 mW/cm2 | 90-100 mW/ cm2 | 3.9-6.8 mW/ cm2 | 1 mW/cm2 |
| UV light<br>Wavelength | 300-320 nm | 300-320 nm | 300-320 nm | 300-320nm |
| Mode of<br>operation | Hand piece used with<br>wireless connection<br>to mobile phone | Hand piece<br>connected to console<br>with touch screen<br>control panel to<br>control treatment | Hand piece used in<br>standalone treatment<br>mode | Patient attached<br>unit connected<br>to a power<br>supply used in<br>standalone<br>treatment mode |
| | Physician<br>provides dosing<br>instructions and<br>physician can<br>adjust time/dose<br>based on | Physician provides<br>dosing instructions<br>and patient can<br>adjust time/dose<br>based on treatment<br>outcome | Physician provides<br>dosing instructions<br>and patient can<br>adjust time/dose<br>based on treatment<br>outcome | Unknown |
| Increase UVB<br>treatment Dose<br>after each<br>treatment | Yes, this<br>capability is built<br>into the SW in<br>accordance with<br>the prescribed<br>protocol. | Yes, this capability<br>is built into the SW<br>in accordance with<br>the prescribed<br>protocol. | Instructions are<br>provided in the<br>DFU. Not built into<br>the the SW, the<br>patient must<br>manually calculate<br>and enter the dose. | Unknown |
| Delivery<br>method | Spot treatment window | Spot treatment attachment<br>(LiteSpot™), Quartz fiber-<br>optic brush attachment<br>(LiteBrush™) | Spot treatment wand with<br>integrated comb | Spot treatment window |
| Power<br>Source | Rechargeable battery | AC outlet | AC outlet | AC outlet |
| Prescription<br>stored on<br>Device | Yes - Cloud server | Yes - USB | No | No |
| Scalp<br>Attachment | No | Yes | Yes | No |
| Use<br>Environme | Home / Physician's<br>office | Home / Physician's<br>office | Home / Physician's<br>office | Home / Physician's<br>office |
| | Subject Device | Predicate Device | Predicate Device | Predicate Device |
| | Clarify Medical<br>Phototherapy System<br>(Subject Device) | Lerner Medical<br>Devices, Inc.<br>Levia Phototherapy<br>System / LH-75T<br>Phototherapy<br>System<br>K040062 / K090097 | Daavlin Distributing<br>Co.<br>DermaPal K073587 | Allux<br>Resolve<br>UVB<br>Photothera<br>py System<br>K072035 |
| IPX -<br>Rating/water<br>Resistance | Not water resistant | Not water resistant | Not water resistant | Not water resistant |
| Communications<br>between | Bluetooth wireless | Wired connection | Wired connection | Wired connection |
| Records<br>retention<br>and remote<br>physician<br>review of<br>results. | Yes. Treatment data is<br>stored and uploaded to<br>Server for remote<br>review | Yes. Treatment data is<br>stored locally on the<br>device. no remote<br>review of data is<br>possible | No record retention<br>capability | No record retention<br>capability |
| Treatment<br>day/Time<br>reminder<br>set by User | Yes | No | No | No |
| Timer to<br>control<br>treatment<br>duration | Yes - stored in the<br>Software, based on<br>Prescription | Yes - entered by the<br>user for each treatment | Yes - entered by the<br>user for each treatment | Yes - entered by the<br>user for each treatment |
| Target<br>Population | Patients and clinicians | Patients and<br>clinicians | Patients and<br>clinicians | Patients and<br>clinicians |
# Table 1 – Predicate Comparison Table
{7}------------------------------------------------
{8}------------------------------------------------
# Summary of Performance Data – Bench
The determination of substantial equivalence is based on an assessment of non-clinical performance data. Table 2 below contains a summary of the performance testing that was submitted, referenced, or relied on in the premarket notification submission to determine substantial equivalence and how their results support a determination of substantial equivalence.
| Element of Comparison | Method for Safety<br>Assessment | Test Result Meet<br>Design Input<br>Specifications and<br>Demonstrate<br>Equivalence |
|---------------------------------------|---------------------------------|--------------------------------------------------------------------------------------|
| UBV light Source, Max Power<br>Output | Performance Testing | Yes |
# Table 2 – Performance Test Summary
{9}------------------------------------------------
| Element of Comparison | Method for Safety<br>Assessment | Test Result Meet<br>Design Input<br>Specifications and<br>Demonstrate<br>Equivalence |
|---------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Power Source | Electrical Safety Testing,<br>Battery Performance<br>Testing | Yes |
| Prescription Stored on the device | System Validation | Yes |
| Treatment Reminders | Software Validation | Yes |
| Auto Calculation of treatment<br>adjustment | System Validation | Yes |
| Skin Contact Materials | Performance testing | Yes |
| Wireless communications between<br>handheld and console | System Validation | Yes |
| Records Retention | System Validation | Yes |
| Remote Physician Review of the<br>results | System Validation | Yes |
# Performance Test Conclusion
Based on technical comparison, the Subject device has been shown to be equivalent to the predicate devices.
# Standards and Guidance Documents
# Optical emissions testing:
Clarify has verified that optical emissions within the 300-320nm range is equivalent to that of the predicate devices. With the LEDs switched on, the optical emissions spectrum is measured at the treatment surface. The biologically weighted emissions curve is then computed using the International Commission on Illumination accepted erythema curve. Clarify has completed verification testing to demonstrate that the weighted peak of emissions is within the established therapeutic range (300-320nm). The optical power at the treatment surface is also measured to ensure it meets the specified requirements. The device has been evaluated for safety and performance in accordance with IEC 60601-2-57 Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.
# Electrical safety:
Clarify has completed testing to the applicable sections of IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, and IEC 60601-1-4 Medical electrical equipment – Part 1-4: General requirements for safety - Collateral Standard: Programmable electrical medical systems.
{10}------------------------------------------------
# Software and wireless safety:
Due to the software component and wireless functionality of the device. Clarify will review and comply with the following FDA guidance documents: Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (Issued 5/11/2005), Guidance for Industry and FDA Staff: Mobile Medical Applications (Issued 9/25/2013), and Guidance for Industry and FDA Staff: Radio Frequency Wireless Technology in Medical Devices (Issued 8/13/2013). The following standard has been considered when evaluating the software: IEC 62304: Medical device software - Software life cycle processes.
# Electromagnetic compatibility (EMC):
Since the system includes electronics, EMC has been assessed according to IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances.
## Biocompatibility:
The distal end of the Clarify Phototherapy System will have limited contact duration (< 24 hours) with skin and is therefore considered a "surface device" according to Annex A of ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. Based on the device type and contact duration, FDA Blue Book Memorandum #G95-1 Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," recommends the following biocompatibility tests be performed: cytotoxicity, sensitization, and irritation. The Clarify Device distal end material has been tested to ISO-10993.
# Human factors:
Clarify medical has completed numerous formative studies and software validation of the Clarify System. Clarify medical will not market the device until compliance with IEC 62366: Medical devices – Application of usability engineering to medical devices is verified by an independent third party.
-------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.