K172273 · Chongqing Peninsula Medical Technology Co., Ltd. · FTC · Jul 30, 2018 · General, Plastic Surgery
Device Facts
Record ID
K172273
Device Name
308nm Excimer System
Applicant
Chongqing Peninsula Medical Technology Co., Ltd.
Product Code
FTC · General, Plastic Surgery
Decision Date
Jul 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 308nm Excimer System is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only.
Device Story
308nm Excimer System is a dermatological light therapy device using a xenon-chloride (XeCl) gas mixture to generate monochromatic 308 nm UVB light. Operated by clinicians in a medical setting, the device delivers targeted ultraviolet radiation to skin lesions. The system features a touch screen control panel for setting treatment parameters, including fluence and pulse duration (1s to 40s). A handswitch controls light delivery from the applicator. The mechanism involves inducing T-cell apoptosis and stimulating melanin cell proliferation and differentiation in hair follicles. The device uses air circulation cooling and includes color-coded indicator lights to monitor state and function. Clinical benefit is derived from the therapeutic effect of UVB light on psoriasis and vitiligo lesions.
Clinical Evidence
Bench testing only. The device was evaluated for safety and performance according to IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), and IEC 60601-2-57 (non-laser light source equipment). No clinical data was provided.
Technological Characteristics
Xenon-chloride (XeCl) excimer lamp; monochromatic 308 nm (±3 nm) UVB output; air circulation cooling; touch screen interface; handswitch control; 16 cm² treatment area; max beam power 800 mW; max power density 50 mW/cm²; Class III beam; software compliant with IEC 62304.
Indications for Use
Indicated for the treatment of psoriasis and vitiligo on intact skin.
Regulatory Classification
Identification
An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. To the right of the square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
July 30, 2018
Chongqing Peninsula Medical Technology Co., Ltd. % Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road Guangzhou Science Park Guangzhou. 510000 Cn
Re: K172273
Trade/Device Name: 308nm Excimer System Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet Lamp For Dermatologic Disorders Regulatory Class: Class II Product Code: FTC Dated: June 30, 2018 Received: July 5, 2018
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -53
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172273
Device Name 308nm Excimer System
Indications for Use (Describe)
The 308nm Excimer System is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Chapter 5. 510(k) Summary
## 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
### 1. Submitter's Information
510(k) Owner's Name: Chongqing Peninsula Medical Technology Co., Ltd. Establishment Registration Number: Applying D-U-N-S Number: 545832254 Address: No.2-4, 2F Block I, Biological and Biochemical Park , 70-1&70-2# Ke Yuan 4th Road, Jiulongpo District, Chongqing, China Tel: +86-15012578550 Fax: +86-023-68611963 Contact Person: Zhang Sudi Email: zsd@peninsulalaser.com
### Application Correspondent:
Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: Suite 306, Kecheng Mansion, No.121 Science Road, Guangzhou Science Park, Guangzhou 510663. China Tel: +86-20-61099984 Email: regulatory@glomed-info.com
## 2. Subject Device Information
Trade Name: 308nm Excimer System Models: XECL-308C, XECL-308D Common Name: Powered Laser Surgical Instrument Classification name: Light, Ultraviolet, Dermatological Review Panel: Dermatology, Physical Medicine Product Code: FTC Regulation Class: II Regulation Number: 878.4630
## 3. Predicate Device Information
510(K) Number: K150752 Company Name: GME German Medical Engineering GmbH Address: Grimmstrasse 23, Bavaria, Germany 90491 Trade/Device Name: GME ExSys 308 Models: --Common Name: Ultraviolet Lamp for Dermatologic disorders
{5}------------------------------------------------
| Sponsor: | Chongqing Peninsula Medical Technology Co., Ltd. |
|-----------------|----------------------------------------------------|
| Subject Device: | 308nm Excimer System, Models: XECL-308C, XECL-308D |
| Document Name: | FDA 510(k) Submission Report |
Requlation Number: 878.4630 Requlatory Class: II Product Code: FTC
## 4. Device Description
308nm Excimer System is an ultratraviolet light system designed to be used in Dermatological practice for the treatment of psoriasis and vitiligo. The lamp is axenon-chloride excimer lamp which utilizes a XeCl gas mixture to generate specific ultraviolet light at wavelenath of 308 nm. It displays the treatment parameters. Users can set the device and fluence on the control touch screen and determine the machine's state and function under the help of colour-coded indicator light.
The mechanism of 308nm Excimer System as shown below:
Image /page/4/Picture/5 description: The image shows a diagram of the effects of 308nm light on skin cells. The first diagram shows lesion T cells attacking melanin cells in the follicle, which makes melanin cells less and less. The second diagram shows that 308nm Excimer systems induced the lesion T cells and stimulate the melanin cells proliferation and differentiation in the follicle. The image also states that 308nm Excimer light can make T cell apoptosis and stimulate the melanin cells proliferation and differentiation.
#### 5. Intended Use / Indications for Use
The 308nm Excimer System is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only.
#### 6. Test Summary
308nm Excimer System has been evaluated the safety and performance by lab bench testing as following:
- � IEC 60601-1:2005+A1:2012, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance,
- � IEC 60601-1-2:2007/(R)2012, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - collateral standard: Electromagnetic Compatibility
- � IEC 60601-2-57 (First Edition): 2011 for use in conjunction with IEC 60601-1:2005, Medical electrical equipment -- Part 2: Particular requirements for the basic safety and essential performance of nonlaser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- � IEC 62304: 2006 (First Edition), Medical device software, Software life- cycle processes.
## 7. Comparison to predicate device and conclusion
{6}------------------------------------------------
| Sponsor: | Chongqing Peninsula Medical Technology Co., Ltd. |
|-----------------|----------------------------------------------------|
| Subject Device: | 308nm Excimer System, Models: XECL-308C, XECL-308D |
| Document Name: | FDA 510(k) Submission Report |
The technological characteristics, features, specifications, materials, mode of operation, and intended use of 308nm Excimer System is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of Comparison | Subject Device | Predicate Device | Remark |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Company | Chongqing Peninsula M<br>edical Technology Co.,<br>Ltd. | GME German Medical<br>Engineering Gmbh | -- |
| Device Name and Model | 308nm Excimer System,<br>Models: XECL-308C,<br>XECL-308D | ExSys 308 System,<br>Models: -- | -- |
| Classification Name | Light, Ultraviolet,<br>Dermatological | Light, Ultraviolet,<br>Dermatological | SE |
| 510(k) Number | K172273 | K150752 | -- |
| Product code | FTC | FTC | SE |
| Intended Use & Indications<br>for Use | The 308nm Excimer<br>System is intended to be<br>used for the treatment of<br>psoriasis and vitiligo. It<br>is to be used on intact<br>skin only. | The GME 308nm<br>Excimer is intended to<br>be used for the<br>treatment of psoriasis<br>and vitiligo. It is to be<br>used on intact skin only | SE |
| Mode of<br>Operation | Continuous light source | Continuous light<br>source | SE |
| Wavelength | 308 nanometers (nm) ±3<br>nm | 308 nm+/-4nm | SE<br>Note 2 |
| Light Source | Xenon-Chlorine (XeCl)<br>excimer lamp produces<br>monochromatic UVB<br>light | XeCl excimer lamp<br>produces<br>monochromatic UVB<br>light | SE |
| Light Delivery | Light source in the<br>Applicator handpiece | Light source is in the<br>Applicator handpiece | SE |
| Cooling of light source | Air circulation cooling | Integrated air cooling | SE |
| Treatment Area | 16 cm² (4 x 4 cm) | 50x35 mm² | SE<br>Note 2 |
| Maximum Beam Power | 800 mW | 875 mW | SE<br>Note 2 |
| Maximum Beam Power<br>Density | 50 mW/cm2 | 50 mW/cm2 | SE |
| Beam Class | III | III | SE |
| Pulse Duration | 1s to 40 s | 1 sec - 40 sec | SE |
| Controls | Handswitch | Footswitch or<br>handswitch | SE<br>Note 1 |
| Power Supply | Adapter Input: 100~240<br>Vac, 0.8 A, 50/60 Hz | Adaptor Input:<br>100~240 Vac, 50/60 | SE<br>Note 1 |
| Elements of Comparison | Subject Device | Predicate Device | Remark |
| | Main unit input: 48 Vdc,<br>2.94 A, 135 VA max | Hz, 0.1 A<br>Main Unit Input: 5 Vdc,<br>1 A | |
| Power Calibration<br>Method | Internal, automatic | Internal, automatic | SE |
| MED Dose<br>Determination | Menu driven | Menu driven | SE |
| Sterilization<br>Aspects | Optical head reducer<br>mask is disinfected<br>between patients. | Applicator is<br>disinfected<br>between patients. | SE |
| Dosage Controls | Dosage (or energy<br>density<br>J/cm2), pulse duration | Dosage (or energy<br>density<br>J/cm2), pulse duration | SE |
| Display | Touch Screen Control<br>Panel | Touch Screen Control<br>Panel | SE |
| Patient Leakage Current | Complied with IEC<br>60601-1 and IEC 60601-<br>2-57 | Comply with IEC<br>60601-1 and IEC<br>60601-2-57 | SE |
| Weight | < 2.6 kg | 12 kg | SE<br>Note 1 |
| Dimensions<br>(H x W x D) | 26 cm x 24 cm x 27 cm<br>(For lamp)<br>5 cm x 9.5 cm x 18 cm<br>(For Adapter) | 25 cm x 30 cm x 30 cm | SE<br>Note 1 |
| Operating Environment | Temperature: 15~35°C,<br>Humidity: ≤ 80%RH,<br>Atmospheric Pressure:<br>86 ~ 106 kPa | Temperature: 5~45°C,<br>Humidity: 20~65% RH | SE<br>Note 1 |
| Storage Environment | Temperature: -20~45°C,<br>Humidity: 10-85%RH<br>Atmospheric Pressure:<br>50 ~ 106 kPa | Temperature: 0~45°C,<br>Humidity: 10~90% RH,<br>Electrode Pad:<br>10~20°C | SE<br>Note 1 |
| Electrical Safety | Comply with IEC 60601-<br>1 and IEC 60601-2-57 | Comply with IEC<br>60601-1 and IEC<br>60601-2-57 | SE |
| EMC | Comply with IEC 60601-<br>1-2 | Comply with IEC<br>60601-1-2 | SE |
{7}------------------------------------------------
## Comparison in Detail(s):
#### Note 1:
Although the "Power Supply", "Weight", "Dimension", "Operating Environment" and "Storage Environment" are a little different from the predicate devices, they all comply with IEC 60601-1 and IEC 60601-2-57 requirements. So the differences will not raise any safety or effectiveness issue.
{8}------------------------------------------------
| Sponsor: | Chongqing Peninsula Medical Technology Co., Ltd. |
|-----------------|----------------------------------------------------|
| Subject Device: | 308nm Excimer System, Models: XECL-308C, XECL-308D |
| Document Name: | FDA 510(k) Submission Report |
### Note 2:
Although some output specifications "Wavelength", "Treatment area", "Output Intensity", and "Maximum Beam Power" are a little different from the predicate devices, they all comply with IEC 6060 1-1, IEC 60601-1-2, IEC 60601-2-57 requirement. So the differences of function specification will not raise any safety or effectiveness issue.
#### Finial Conclusion:
The subject device "308nm Excimer System" is Substantial Equivalent to the predicate devices.
#### 8. Date of the summary prepared: June 30, 2018
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.