GME ExSys 308

K150752 · Gme (German Medical Engineering) GmbH · FTC · Jul 31, 2015 · General, Plastic Surgery

Device Facts

Record IDK150752
Device NameGME ExSys 308
ApplicantGme (German Medical Engineering) GmbH
Product CodeFTC · General, Plastic Surgery
Decision DateJul 31, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The GME ExSys 308 System is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only.

Device Story

ExSys 308 is an ultraviolet light system for dermatological practice; treats psoriasis and vitiligo. System comprises base unit with touch screen controls and applicator handpiece containing XeCl excimer lamp. Physician operates device in clinical setting. UVB light (308 nm) emitted from applicator penetrates skin; modulates immune system; stimulates melanocytes to produce melanin. Dosage controlled via menu-driven interface; energy density (J/cm²) and pulse duration settings. Device includes distance spacer (treatment tip) for patient contact. Benefits include targeted phototherapy for skin conditions. Output affects clinical decision-making by providing controlled UV exposure for therapeutic effect.

Clinical Evidence

Bench testing only. Testing included illumination homogeneity validation (<+/-20% deviation), treatment tip transmission analysis, and electrical safety/EMC testing per IEC 60601-2-57, EN 60601-1, and EN 60601-1-2.

Technological Characteristics

XeCl excimer lamp light source; 308 nm +/- 4 nm wavelength; 50 mW/cm2 max power density; 875 mW max beam power. Integrated air cooling. Type B applied part. Distance spacer made of aluminum and PMMA. Touch screen control panel. Electrical requirements 100V-240V. Internal automatic power calibration. Sterilization via disinfection of applicator between patients.

Indications for Use

Indicated for the treatment of psoriasis and vitiligo on intact skin.

Regulatory Classification

Identification

An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines representing hair or clothing. The profiles are arranged in a way that suggests unity and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 GME German Medical Engineering GmbH Mike Johnson M.D. Philosopher's River LLC P.O. Box 106 Willow Creek, Montana 59760 July 31, 2015 Re: K150752 Trade/Device Name: GME ExSys 308 Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet lamp for dermatologic disorders Regulatory Class: Class II Product Code: FTC Dated: June 30, 2015 Received: July 2, 2015 Dear Dr. Johnson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {2}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Joshua C. Nipper -S Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K150752 Device Name GME ExSys 308 Indications for Use (Describe) The GME ExSys 308 System is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # Section 5 ## 510(k) Summary This 510(k) Summary is submitted in accordance with the requirements of 21 CFR 807.87 and 807.92. Summary preparation date 07-31-2015 [21 CFR 807.92(a)(1)]. #### Applicant Name and Address [21 CFR 807.92(a)(1)] A. GME German Medical Engineering GmbH. Grimmstrasse 23 Bavaria, Germany 90491 Tel: +49 9131 934159 10 Fax: +49 9131 934159 99 #### B. Contact Information Philosopher's River llc P O Box 106 Willow Creek, MT 59760 Tel: 406-209-3039 Fax: 406 2093039 Contact person: Mike Johnson M.D. mike(@philosophersriver.com {5}------------------------------------------------ #### C. Device Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] Trade Name: ExSys 308 System Device Common Name: Ultraviolet Light for Dermatology Classification Name: Light, Ultraviolet, Dermatological 21 CFR 878.4630 Product Code: FTC Device Classification: Class II #### D. Predicate Devices [21 CFR 807.92(a)(3)] The ExSys 308 System uses similar technology and physical output characteristics as the following predicate devices: K090762 LEDA from Quantel Derma GmbH #### E. Device Description [21 CFR 807.92(a)(4)] The ExSys 308 is an ultratraviolet light system designed to be used in Dermatological practice for the treatment of psoriasis and vitiligo. The ExSys 308 System consists of a base unit (touch screen, mains switch, key switch, Emergency Stop button, speaker), which controls an applicator unit. The applicator unit contains the UV light source. The model number is ExSys 308. The ExSys 308 System includes the following accessories: Power cord, optional foot switch, laser protective goggles, applicator holder, and set of treatment tips. The treatment tip (a distance spacer) is the only patient contacting part of the system. There are no single use parts in the ExSys 308 System. - F. Principle of Operation: The UVB light penetrates the skin and modulates the immune system and stimulates melanocytes to produce melanin. {6}------------------------------------------------ #### Device Specifications and Comparison to Predicates [21 CFR 807.92(a)(6)] G. The ExSys 308 is compared to the predicate, the Quantel Medical 308 Dermatological Excimer System. The indications for use and intended use for the ExSys 308 are identical to the Quantel Medical 308 Dermatological Excimer System. Below is a comparison table. | Characteristic | GME ExSys 308 System | Quantel 308 Dermatological<br>Excimer System | |---------------------------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------| | | "ExSys 308" | "Quantel 308" | | Applicable<br>510(k)s<br>Panel/ | NA | K090762 | | Product Code/<br>Regulation<br>Number | FTC<br>21 CFR 878.4630 | FTC<br>21 CFR 878.4630 | | Indications for<br>Use Statement | The GME ExSys 308 System is<br>intended to be used for the<br>treatment of psoriasis and<br>vitiligo. | The Quantel 308 Excimer System is<br>intended to be used for the treatment of<br>psoriasis and vitiligo. | | Mode of<br>Operation | Continuous light source | Continuous light source | | Wavelength | 308 nm+/-4nm | 308 nm+/-4nm | | Light Source | XeCl excimer lamp produces<br>monochromatic UVB light | XeCl excimer lamp produces<br>monochromatic UVB light | | Light Delivery | Light source is in the<br>Applicator handpiece | Light source is in the Applicator<br>handpiece | | Cooling of Light<br>Source | Integrated air cooling | Integrated air cooling | | Treatment Area<br>Size | 50x35mm2 | 40x40mm2 | | Maximum Beam<br>Power | 875mW | 800mW | | Maximum Beam<br>Power Density | 50 mW/cm2 | 50 mW/cm2 | {7}------------------------------------------------ | Beam Class | III | III | |--------------------------------|-----------------------------------------------------|----------------------------------------------------------------------------| | Pulse Duration | 1 - 40 seconds | 1-120 seconds | | Applied Part<br>Safety Class | Type B | Type B | | Patient<br>Contacting Parts | Distance Spacer made from<br>aluminum and PMMA | Distance tips made of several materials:<br>Quartz glass, aluminum, rubber | | Controls | Footswitch or handswitch | Handswitch | | Electrical<br>Requirements | 100V - 240V | 100V - 240V | | Power<br>Calibration<br>Method | Internal, automatic | Internal, automatic | | Sterilization<br>Aspects | Applicator is disinfected<br>between patients. | Applicator is disinfected between<br>patients. | | MED Dose<br>Determination | Menu driven | Menu driven | | Duty Cycle | 100% | 100% | | Dosage Controls | Dosage (or energy density<br>J/cm²), pulse duration | Dosage (or energy density J/cm²), pulse<br>duration | | Display | Touch Screen Control Panel | Touch Screen Control Panel | #### H. Indications for Use [21 CFR 807.92(a)(5)] The GME ExSys 308 System is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only. {8}------------------------------------------------ #### I. Performance Data [21 CFR 807.92(b)(2)] Noncinical testing of performance and safety was directed by IEC 60601-2-57: Ed. 1.0, without deviation from the standard. A report of this testing was submitted. In addition, validation testing of illumination homogeneity and Treatment Tip transmission were performed. The submitted report demonstrated a less than +/-20% deviation from illumination homogeneity and no measurable unwanted transmission around the Treatment Tip. Third party testing for EN 60601-1 and EN 60601-1-2 was performed to provide further evidence of safety. This preclinical testing demonstrates that in terms of performance and safety, the ExSys 308 System is equivalent to the predicate device, K090762 LEDA from Quantel Derma GmbH. #### J. Conclusion [21 CFR 807.92(b)(3)] The GME ExSys 308 was found to be substantially equivalent to the predicate device, the 308 Dermatological Excimer System from Quantel Medical, in terms of technology, function and intended use. The indications for use are identical to the previously cleared device (K073066) Quantel Medical 308 Dermatological Excimer System. We believe that there are no new questions of safety or efficacy raised by the introduction of the GME ExSys 308 System. ર-ર
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...