ExpressCore Biopsy Device

K243886 · Inrad, Inc. · KNW · Apr 29, 2025 · Gastroenterology, Urology

Device Facts

Record IDK243886
Device NameExpressCore Biopsy Device
ApplicantInrad, Inc.
Product CodeKNW · Gastroenterology, Urology
Decision DateApr 29, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1075
Device ClassClass 2

Indications for Use

The ExpressCore Full Core Biopsy Device is intended for use in obtaining biopsies from soft tissues.

Device Story

ExpressCore is a sterile, single-use, handheld biopsy device for soft tissue sampling. It features a plastic handle and three stainless steel needle components. The device operates via a manual mechanism allowing for variable throw distances (10-30mm) and multiple firing modes (semi or automatic) to suit physician preference and procedural needs. It is available in various gauge sizes (12ga to 18ga) and lengths (10cm to 25cm). The device is used by clinicians to capture full core biopsy samples. The output is a physical tissue core sample used for diagnostic evaluation. The device is intended for clinical use.

Clinical Evidence

Bench testing only. Verification data included simulated use, usability, biopsy sample collection, tensile testing of cannula and stylet components, representative tissue sampling, biocompatibility, sterilization, packaging, and shelf life.

Technological Characteristics

Materials: 304 stainless steel (needles), ABS/Polycarbonate (handle). Mechanism: Manual, variable throw (10-30mm), semi/automatic firing modes. Connectivity: None. Sterilization: Ethylene Oxide. Single-use, disposable.

Indications for Use

Indicated for obtaining biopsies from soft tissues in patients requiring percutaneous tissue core biopsy.

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION April 29, 2025 INRAD, Inc Heidi Halverson RA/QA Manager 4375 Donker Court SE Kentwood, Michigan 49512 Re: K243886 Trade/Device Name: ExpressCore Biopsy Device Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: March 28, 2025 Received: March 28, 2025 Dear Heidi Halverson: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2} K243886 - Heidi Halverson Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {3} K243886 - Heidi Halverson Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jessica Carr -S Jessica Carr, PhD Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K243886 | | | Device Name ExpressCore Full Core Biopsy Device | | | Indications for Use (Describe) The ExpressCore Full Core Biopsy Device is intended for use in obtaining biopsies from soft tissues. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {5} ExpressCore Full Core Biopsy Device 510(k) Submission K243886 510(k) Summary I. SUBMITTER Sponsor/Manufacturer: INRAD, Inc. 4375 Donker Court SE, Kentwood, MI 49512, USA Establishment Registration Number: 1835568 Contact Person: Heidi Halverson RA/QA Manager INRAD, Inc. 4375 Donker Court SE, Kentwood, MI 49512, USA Phone: (616) 301-7800 Fax: (616) 301-7799 Email: hhalverson@inradinc.com Date Prepared: 12/10/2024 II. DEVICE Device Name(s): ExpressCore Full Core Biopsy Device ExpressCore Full Core Biopsy Device w/ HiLiter® Common or Usual Name: Biopsy Instrument Classification Name: Gastroenterology-urology Biopsy Instrument (21 CFR 876.1075) Regulatory Class: II Product Code: KNW III. PREDICATE DEVICE The INRAD, Inc. ExpressCore Full Core Biopsy Device [subject device] is substantially equivalent (SE) to the Sponsor's own predicate device: Revolution Full Core Biopsy Device, K090611 This predicate device has not been subject to a design related recall. Reference Devices: EliteCore Full Core Biopsy Device, K112945 AccuCore Biopsy Device, K000612 {6} The INRAD, Inc. ExpressCore Full Core Biopsy Device [subject device] is substantially equivalent (SE) to the INRAD, Inc. Revolution Full Core Biopsy Device (K090611) [predicate device] in terms of Indications for Use / Intended Use to obtain biopsies from soft tissues. Substantial equivalency (SE) of the subject device has also been based on substantially equivalent design, functionality, and performance characteristics as the predicate device. ## IV. DEVICE DESCRIPTION The ExpressCore Device is a sterile, single patient use, hand held full core biopsy device comprised of a plastic handle and three stainless steel needle components that work in sequence to capture a full core biopsy sample. The device is capable of infinitely variable throw distance between 10 and 30mm and is equipped with multiple firing modes to accommodate procedural requirements and physician preference. Multiple device gauge size and needle option combinations have been developed from 12ga x 10cm to 18ga x 25cm. ## V. INTENDED USE / INDICATIONS FOR USE The Intended Use and the Indications for Use are the same for the ExpressCore Full Core Biopsy Device [subject device] and the INRAD, Inc. Revolution Full Core Biopsy Device (K090611) [predicate device] with the exception of specific tissue listings being omitted for ExpressCore. [subject device]: The ExpressCore Full Core Biopsy Device is intended for use in obtaining biopsies from soft tissues. [predicate device]: The Revolution Full Core Biopsy Device is intended for use in obtaining biopsies from soft tissues such as breast, liver, kidney, prostate, spleen, lymph nodes, and various soft tissue tumors. ## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The INRAD, Inc. ExpressCore Full Core Biopsy Device [subject device] is substantially equivalent (SE) to the INRAD, Inc. Revolution Full Core Biopsy Device (K090611) [predicate device] based on the same functional and performance characteristics of the subject device when compared to the predicate device. The difference between the subject device's and predicate device's gauge size, length options and packaging do not raise concerns of safety and effectiveness. To further analyze risks associated with the varying gauge sizes, length options and packaging, additional reference devices; EliteCore and AccuCore by INRAD have been included for reference. The technological characteristics of design, function, and materials present in both the subject device and the predicate device has been evaluated. The comparisons supporting a determination of substantial equivalency (SE) are provided below. {7} | Comparison Criteria | Subject Device ExpressCore Biopsy Device | Predicate Device Revolution Biopsy Device (K090611) | Reference Device EliteCore Biopsy Device (K112945) | Reference Device AccuCore Biopsy Device (K000612) | | --- | --- | --- | --- | --- | | Manufacturer | INRAD, Inc. | Same | Same | Same | | Device Class | Class II | Same | Same | Same | | Device Classification Name | Biopsy Instrument | Same | Same | Same | | Product Code | KNW | Same | Same | Same | | Regulation Number | 21 CFR §876.1075 | Same | Same | Same | | Indications for Use | This device is intended for use in obtaining biopsies from soft tissues. | This device is intended for use in obtaining biopsies from soft tissues such as breast, liver, kidney, prostate, spleen, lymph nodes, and various soft tissue tumors. | This device is intended for use in obtaining biopsies from soft tissues such as breast, liver, kidney, prostate, spleen, lymph nodes, and various soft tissue tumors. | This device is intended to be used as an automated instrument for percutaneous tissue core biopsy of several human organs including, but not limited to, prostate, kidney, liver, spleen, lung, breast, lymph nodes and different tumors. | {8} | Comparison Criteria | Subject Device ExpressCore Biopsy Device | Predicate Device Revolution Biopsy Device (K090611) | Reference Device EliteCore Biopsy Device (K112945) | Reference Device AccuCore Biopsy Device (K000612) | | --- | --- | --- | --- | --- | | Handle Composition (non-patient contact) | ABS, Polycarbonate | ABS, TPE, Polyester, Acetal | ABS, TPE, Polyester, Acetal | Polycarbonate | | Needle Composition (patient contact) | 304 Stainless Steel | 304 Stainless Steel 17-7 Stainless Steel | 304 Stainless Steel 17-7 Stainless Steel | Same | | Needle Processing | • Straightening • Cut to length • Grit blast • Tip geometry • Electropolish • Depth marks • Clean and passivate • Finger bend • Inspect and pack | Same | Same | Same + Notch grind (stylet) - Finger bend | | Needle Gauge Size | 12ga, 14ga, 16ga, 18ga | 14ga | 18ga | Same | | Needle Lengths | 10cm, 15cm, 20cm, 25cm | 10cm, 15cm | Same | Same | | Needle Echogenicity | HiLiter Stylet (select product codes) | Same | Same | Grit blasted cannula | | Throw Length | Variable 10-30mm | Variable 13-25mm | Variable 13-25mm | Fixed | | Mechanism of Throw Adjustment | Thumb Wheel | Same | Same | N/A | | Mechanism of Arming | Dual Stroke | Same | Same | Single Stroke | | Firing Modes | Semi or Automatic | Same | Same | Single Action | | Packaging Configuration | Tyvek / PE pouch | Tyvek / PETG Tray | Tyvek / PETG Tray | Same | | Sterile | Yes | Same | Same | Same | | Method of Sterilization | Ethylene Oxide | Same | Same | Same | | Single-Use, Disposable | Yes | Same | Same | Same | {9} VII. SUMMARY OF VERIFICATION DATA AND VERIFICATION TEST CONCLUSIONS Non-Clinical Bench Performance Testing was conducted on the INRAD, Inc. ExpressCore Full Core Biopsy Device [subject device] to confirm that the device meets all system requirements and is substantially equivalent (SE) to the INRAD, Inc. Revolution Full Core Biopsy Device (K090611) [predicate device]. The following Verification Data was provided in support of the substantial equivalence (SE) determination. - Performance Testing – Simulated Use / Bench - Usability of the Device - Biopsy Sample Collection - Finger Cannula Tensile Test - Spoon Cannula Tensile Test - Stylet Tensile Test - Representative Tissue Sampling - Biocompatibility - Sterilization - Packaging - Shelf Life VIII. SUBSTANTIAL EQUIVALENCE SUMMARY / CONCLUSIONS: The data generated from the results of the Verification Testing, Design Validation, and evaluations, along with a side-by-side comparison of the technological characteristics of design, function, and materials between the subject device and the predicate and reference devices, demonstrate that the INRAD, Inc. ExpressCore Full Core Biopsy Device [subject device] is as safe and effective and performs as well as the INRAD, Inc. Revolution Full Core Biopsy Device (K090611) [predicate device]. The similar technological and performance characteristics for the proposed INRAD, Inc. ExpressCore Full Core Biopsy Device [subject device] have been assessed to be substantially equivalent to the predicate device, and the slight differences do not raise concerns of safety and effectiveness when compared to the predicate or reference devices. Therefore, the INRAD, Inc. ExpressCore Full Core Biopsy Device [subject device] is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...