The device is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors.
Device Story
The EliteCore™ Full Core Biopsy Device is a sterile, disposable, mechanically actuated instrument used to obtain full core tissue specimens. It features a stainless steel cutting cannula with a spoon, a rotating coring cannula, and a stylet housed within a plastic body. The device includes a variable throw feature allowing users to select throw settings between 13 and 25 mm. It is operated by a clinician to perform biopsies on soft tissues such as the liver, kidney, prostate, spleen, lymph nodes, and various tumors. The device is manually triggered to capture tissue samples. By providing a full core specimen, it assists healthcare providers in obtaining sufficient tissue for diagnostic evaluation, potentially improving the accuracy of disease diagnosis.
Clinical Evidence
Bench testing only. Testing included package inspection, burst testing, device inspection, dimensional verification of throw settings, firing force measurements, and tissue sampling efficacy in chicken and liver models. Results confirmed the device meets performance specifications and is equivalent to the predicate device.
Technological Characteristics
Materials include stainless steel (cannula, spoon, stylet) and plastic (housing). The device is a mechanically actuated, disposable biopsy instrument. It features a variable throw mechanism (13-25 mm). No electronic components, software, or energy sources are utilized. Sterilization method is not specified.
Indications for Use
Indicated for obtaining biopsies from soft tissues including liver, kidney, prostate, spleen, lymph nodes, and various soft tissue tumors in patients requiring tissue sampling.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{1}------------------------------------------------
K112945
NOV - 3 2011
Image /page/0/Picture/2 description: The image shows the word "INRAD" in bold, black letters. To the left of the word is a black square with a white diagonal line running from the top left corner to the bottom right corner. A small "R" with a circle around it is in the top right corner of the image.
# September 30, 2011
#### Contact Information
Ryan Goosen President Phone: 616-301-7800, ext. 105 Fax: 616-301-7799 E-mail: rgoosen@inrad-inc.com
#### Company Information
Inrad, Inc 4375 Donker Court SE Kentwood, MI 49512 Phone: 616-301-7800 616-301-7799 Fax:
# Device Name(s)
EliteCore™ Full Core Biopsy Device EliteCore™ Full Core Biopsy Device w/HiLiter®
#### Device Summary
| Trade or Proprietary Name: | EliteCore™ Full Core Biopsy Device | |
|----------------------------|-----------------------------------------------------------------------------------|--|
| | EliteCore™ Full Core Biopsy Device w/HiLiter® | |
| Common or Usual Name: | Biopsy Instrument | |
| Classification Name: | Gastroenterology-urology biopsy instrument<br>(21 CFR 876.1075, Product code KNW) | |
# Name of Predicate(s) or Legally Marketed Device(s)
K090611– Revolution® Full Core Biopsy Device
{2}------------------------------------------------
#### Device Description
The EliteCore™ Full Core Biopsy Device is a sterile, disposable device which features a stainless steel cutting cannula with spoon, rotating coring cannula and stylet. The device is comprised of a plastic housing that contains the mechanically actuated components. The key performance attribute of the EliteCore™ Full Core Biopsy Device is the capability of obtaining a full core specimen. A variable throw feature allows the user to choose a throw setting ranging from 13 to 25 mm.
#### Indications for Use
The device is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors.
#### Substantial Equivalence
The EliteCore™ Full Core Biopsy Device has the same intended use as the Revolution® Full Core Biopsy Device with the exception of the omission of the word breast. The device and the predicate device have the same technological characteristics in terms of design and materials.
### Performance Testing Summary
Performance testing confirms that the quality of samples obtained with the EliteCore™ Full Core Biopsy Device is equivalent to that of the predicate device as shown below:
| Test | Involved | How Equivalent |
|---------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| -Package inspection | Visually inspect packaging<br>for damage, loose components<br>tip protector secure, and variable<br>throw set at 22mm. | Identical testing<br>performed on predicate<br>device w/ equivalent<br>results. |
| -Burst test | Measure pressure at which<br>sealed tray fails. | Identical testing<br>performed on predicate<br>device. w/ equivalent<br>results. |
| -Device inspection | Visually inspect device<br>for damage | Identical testing<br>performed on predicate<br>device w/ equivalent<br>results. |
{3}------------------------------------------------
K112945
#### -Dimensional
-Tissue test
-Tissue test cont.
Measure variable throw desired vs. actual.
Measure variable throw after dry fires, desired vs. actual.
Measure force to fire front triggers (RH / LH)
Measure force to fire rear trigger.
Able to take tissue sample in chicken and liver.
Identical testing performed on predicate device w/ equivalent results.
Identical testing performed on predicate device w/ equivalent results.
Identical testing performed on predicate device w/ equivalent results.
Additional testing that was not performed on predicate.
Identical testing performed on predicate device w/ equivalent results.
Able to take tissue samples at various throw distances.
Measure 1st sample as compared
to last sample for sample weight
and verify no loss of effectiveness.
Identical testing performed on predicate device w/ equivalent results.
Identical testing performed on predicate device w/ equivalent results.
Mates with adjustable coaxial Identical testing And able to take sample. performed on predicate
Section 5 - 3
{4}------------------------------------------------
K//2945
device w/ equivalent results.
Identical testing. performed on predicate device w/ equivalent results.
Identical testing performed on predicate device w/ equivalent results.
ldentical testing performed on predicate device w/ equivalent results.
-Verify weld integrity
Visually inspect device for damage along weld joints after functional testing.
-Dimensional
Measure cannula, spoon, and stylet for migration.
Measure the joint strength for the cannula, spoon, and stylet.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines representing the staff and a snake winding around it. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Ryan Goosen President Inrad Inc. 4375 Donker CT SE GRAND RAPIDS MI 49512
NOV - 3 2011
Re: K112945
Trade/Device Name: EliteCore™ Full Core Biopsy Device EliteCore™ Full Core Biopsy Device w/HiLiter® Regulation Number: 21 CFR§ 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II Product Code: KNW Dated: September 30, 2011 Received: October 4, 2011
Dear Mr. Goosen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{6}------------------------------------------------
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Huker Lemur MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{7}------------------------------------------------
#### SECTION 4: INDICATIONS FOR USE STATEMENT
# Indications for Use
12945
510(k) Number (if known):
Device Name:
Indications for Use:
EliteCore™ Full Core Biopsy Device EliteCore™ Full Core Biopsy Device w/HiLiter®
The device is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
)Division Sloחס-ח(
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K112945
Section 4 - 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.