K243812 · Volta Medical · DQK · May 9, 2025 · Cardiovascular
Device Facts
Record ID
K243812
Device Name
Volta AF-Xplorer
Applicant
Volta Medical
Product Code
DQK · Cardiovascular
Decision Date
May 9, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
AI/ML
Intended Use
The Volta AF-Xplorer assists operators in the real-time manual or automatic annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia.
Device Story
Volta AF-Xplorer is a machine and deep learning-based software system used by electrophysiologists in clinical settings to assist in real-time annotation of 3D cardiac maps. It accepts intra-cardiac multipolar electrogram (EGM) signals via compatible EP recording systems (analog) or 3D mapping systems (digital). The software analyzes these signals to identify regions exhibiting spatiotemporal dispersion. Outputs are displayed on a monitor, highlighting areas of interest; in digital mode, the system can automatically tag these regions within the 3D mapping system. By identifying dispersed EGMs, the device aids clinicians in targeting specific atrial regions during catheter ablation procedures. This real-time guidance is intended to support the identification of potential ablation targets, potentially improving procedural outcomes for patients with persistent atrial fibrillation.
Clinical Evidence
Evidence provided via the Tailored-AF trial (NCT04702451), a multicenter, randomized, controlled trial (n=370) comparing tailored ablation (PVI + dispersion targeting via VX1 AI) vs. conventional PVI-only. Primary endpoint: freedom from AF >30s at 12 months. Results: 88% success in Tailored arm vs. 70% in Anatomical arm (p<0.0001). Secondary endpoints showed no significant difference in freedom from any atrial arrhythmia (76% vs 71%) or quality of life. Tailored arm had higher minor complication rates (8.0% vs 3.2%) and longer procedure/ablation times.
Technological Characteristics
System comprises a computing platform (Intel Core i7-7700, 32GB RAM, Debian-based Linux) and software. Inputs: analog/digital intra-cardiac signals. Connectivity: unidirectional analog (custom cable + Advantech PCI-1713U converter) or bidirectional digital (Ethernet). Output: real-time visualization of EGM dispersion and cycle length. Algorithm: machine/deep learning-based signal processing. Non-sterile, reusable.
Indications for Use
Indicated for patients with atrial fibrillation or atrial tachycardia undergoing electrophysiological mapping of the human atria.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
K220786 — STAR Apollo Mapping System · Rhythm Al, Ltd. · Dec 1, 2022
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 9, 2025
Volta Medical
% Kristin Duggan
Partner
Hogan Lovells US LLP
555 Thirteenth Street, NW
Washington, District of Columbia 20004
Re: K243812
Trade/Device Name: Volta AF-Xplorer
Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable Diagnostic Computer
Regulatory Class: Class II
Product Code: DQK
Dated: December 11, 2024
Received: April 10, 2025
Dear Kristin Duggan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243812 - Kristin Duggan
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243812 - Kristin Duggan
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MARCO CANNELLA -S
for
- Aneesh Deoras
- Assistant Director
- Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices
- Office of Cardiovascular Devices
- Office of Product Evaluation and Quality
- Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243812 | |
| Device Name Volta AF-Xplorer | |
| Indications for Use (Describe) The Volta AF-Xplorer assists operators in the real-time manual or automatic annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K243812 510(k) SUMMARY
VOLTA MEDICAL's
Volta AF-Xplorer
Submitter
Volta Medical
65 Avenue Jules Cantini
13006 Marseille
France
Phone: +33 7 68 02 54 99
Contact Person: Paola MILPIED
Date Prepared: December 11, 2024
Name of Device: Volta AF-Xplorer
Common or Usual Name: Cardiac Mapping System
Classification Name: Programmable Diagnostic Computer
Regulatory Class: 21 C.F.R § 870.1425
Product Code: DQK
Predicate Device
Volta Medical, Volta AF-Xplorer (K232616)
Reference Device
Volta Medical, VX1 (K201298)
Device Description
The Volta AF-Xplorer is a machine and deep learning based-algorithm designed to assist operators in the real-time manual or automatic annotation of 3D anatomical and electrical maps of the human heart for the presence of electrograms (EGMs) exhibiting spatio-temporal dispersion, i.e., dispersed EGMs.
The Volta AF-Xplorer device is a non-sterile reusable medical device, composed of a computing platform and a software application. Volta AF-Xplorer works with all existing 510(k)-cleared catheters that meet specific dimension requirements and with one of the three specific data acquisition systems:
- Two compatible EP recording systems: the LabSystem Pro EP Recording System (Boston Scientific) (K141185) or the MacLab CardioLab EP Recording System (General Electric) (K130626),
- a 3D mapping system: EnSite X 3D mapping system (Abbott) (K221213).
1 of 7
{5}
K243812
A connection cable is used to connect the corresponding data acquisition system to the Volta AF-Xplorer system, depending on the type of communication used:
- Unidirectional analog communication with the EP recording systems via a custom-made cable (two different variants: DSUB, Octopus) and an Advantech PCI-1713U analog-to-digital converter, which acquires analog data, digitizes it, and transmits the digital signals to the computer that hosts the Volta AF-Xplorer software.
- Bidirectional digital communication with the EnSite 3D mapping system via an ethernet cable (four different lengths: 20, 10, 5 or 2m) which transmits the digital signals directly to the computer.
The computer and its attached display are located outside the sterile operating room area. The Volta AF-Xplorer software analyzes the patient's electrograms to cue operators in real-time to intra-cardiac electrograms of interest for atrial regions harboring dispersed electrograms as well as a cycle length estimation from electrograms recorded with the mapping and the coronary sinus catheters. The results of the analysis are graphically presented on the attached computer display and/or on a secondary medical screen or on an operating room widescreen. The identified regions of interest are either manually (all configurations) or automatically (only available in digital bidirectional communication with the EnSite X 3D mapping system) tagged in the corresponding 3D mapping system.
## Intended Use / Indications for Use
The Volta AF-Xplorer assists operators in the real-time manual or automatic annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia.
Differences between subject and predicate intended use:
The intended use / indications for use of the subject and predicate devices are identical with the exception that the qualifying language related to the clinical significance of the device has been removed based upon the results of the Tailored-AF clinical study which provide new clinical evidence related to the device safety, effectiveness and performance.
The updated indications for use statement and additions to the product labeling do not alter the intended use of the device as an electrophysiological evaluation tool and do not affect, or raise different questions of, the device safety and effectiveness.
## Summary of Technological Characteristics
The Volta AF-Xplorer device is technologically identical to its Volta AF-Xplorer predicate (K232616).
Volta AF-Xplorer and predicate are software programs that work with standard electrophysiology catheters to aid in mapping the heart. Volta AF-Xplorer and its predicate aid operators by assisting in annotating complex electrical maps of the heart, and process and output information via a computer and display that are operated by use of a keyboard / mouse. Volta AF-Xplorer and its predicate have the same input (intra-cardiac multipolar signals) and the same output (associated dispersion).
2 of 7
{6}
K243812
Volta AF-Xplorer and its predicate support electrophysiologists in the manual annotation of dispersed areas using an unidirectional analog communication. Volta AF-Xplorer and its predicate provide the ability to connect to a specific 3D mapping system through a bidirectional digital communication, which enables the operator to use the device's function for automatically tagging regions of interest.
A table comparing the key features of the subject and predicate devices is provided below.
| | Volta Medical Volta AF-Xplorer (Subject) | Volta Medical Volta AF-Xplorer (K232616) |
| --- | --- | --- |
| Regulation | 21 C.F.R. § 870.1425 | SAME |
| Classification Name | Programmable Diagnostic Computer | SAME |
| Product Code | DQK | SAME |
| Indications for Use | The Volta AF-Xplorer assists operators in the real-time manual or automatic annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia. | The Volta AF-Xplorer assists operators in the real-time manual or automatic annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia. The clinical significance of utilizing the Volta AF-Xplorer software to help identify areas with intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion for catheter ablation of atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations. |
| System Type | Signal processing based atrial mapping system | SAME |
| Primary Feature | Displays and analyzes electrical maps such as intra-cardiac electrograms in real-time using machine learning and signal processing techniques | SAME |
| 3D Location Technology | Electroanatomic location is performed by another commercially available navigation system. In bidirectional digital communication, 3D Location is shared by the 3D Mapping System with Volta AF-Xplorer | SAME |
| Compatible Acquisition Systems | • LabSystem Pro EP Recording System (Boston Scientific) • CardioLab EP Recording System (GE) • EnSite X 3D Mapping System (Abbott) | SAME |
{7}
K243812
| | Volta Medical Volta AF-Xplorer
(Subject) | Volta Medical Volta AF-Xplorer
(K232616) |
| --- | --- | --- |
| Compatible Catheters | Any compatible mapping and ablation catheter | SAME |
| Display(s) | Color monitor | SAME |
| Multi-Display Support | Yes, duplicate display on a secondary medical screen or on an operating room widescreen | SAME |
| Control | Standard keyboard / mouse | SAME |
| Display Timing | Real-time | SAME |
| Inputs Required | Analog or digital Intra-cardiac signals In digital mode, 3D locations of corresponding electrodes bipoles | SAME |
| Output | Presence or absence of electrogram dispersion at each electrode bipole under consideration In digital mode, 3D locations of corresponding electrodes bipoles
Computed values of mapping and reference cycle length | SAME |
| Duration of Electrogram Recordings | 1.5 Seconds | SAME |
| Output Display | The system generates color coded symbol(s) that indicates to the operator that the area under investigation is one exhibiting dispersion In bidirectional digital communication, validated dispersion area can also be automatically displayed in the 3D mapping system as tags in the 3D atrial shell | SAME |
| Signal Information Displayed | Acquired patient signals, including body surface ECG and intra-cardiac EGMs | SAME |
| Computing Platform | Computer with Intel Core i7-7700 CPU (8MB Cache, up to 4.20 GHz, RAM 32 GB), with integrated analog/digital converter PCI card and TPM (Trusted Platform Module)
Debian-based Linux OS | SAME |
| Hardware Design and Materials | Computing platform, proprietary software algorithm, monitor, mouse/keyboard, custom-made analog connection cable, ethernet cable, acquisition system | SAME |
4 of 7
{8}
K243812
# Performance Data
No bench or animal testing was conducted in support of this submission.
The company also provided the Tailored-AF study results published in Nature Medicine Journal (Deisenhofer et al) and performed on VX1, the substantially equivalent predicate device to Volta AF-Xplorer, to support the updated indications for use and product labeling.
The Tailored-AF Trial (NCT04702451) was an international, multicenter, randomized, controlled, single-blind, superiority trial in which patients with drug-refractory persistent atrial fibrillation were randomly assigned in a 1:1 ratio to either (1) a tailored ablation procedure targeting areas of spatiotemporal dispersion detected by VX1 artificial intelligence software (predecessor to Volta AF-Xplorer) in addition to pulmonary vein isolation (PVI) (the "Tailored" group; n=187), or (2) to a conventional pulmonary vein isolation-only procedure (the "Anatomical" group; n=183). The trial was conducted at 26 centers across 5 countries in Europe and the United States. A total of 51 operators performed the catheter ablation procedures.
The primary effectiveness endpoint was freedom from documented atrial fibrillation lasting more than 30 seconds after a 3-month blanking period through 12 months after one ablation procedure with or without anti-arrhythmic drugs (AADs). Freedom from any atrial arrhythmia was addressed in secondary effectiveness endpoints. The use of anti-arrhythmic medications was allowed during the first 3 months after the initial ablation (the post-ablation "blanking period"), after which their use was discouraged. All study subjects were followed for 12 months with scheduled visits at 3, 6, and 12 months. Clinical assessments, 12-lead electrocardiograms and 24-hour Holter monitoring recordings were obtained at each visit. All patients received a 6-lead Kardia portable monitor and were asked to perform and transmit rhythm recordings weekly and any time they had symptoms for the complete study duration. Patients with recurrent atrial fibrillation after the blanking period were allowed to start or resume the use of anti-arrhythmic medications; they were also allowed to undergo a repeat ablation using the same procedure to which they were initially randomly assigned. A total of 13 patients dropped out of the study before the end of the blanking period (7 in the Tailored arm and 6 in the Anatomical arm) and were not included in the analyzed modified Intention-To-Treat ("mITT") population.
The proportion of subjects who were on a Class 1 or 3 AAD at study enrollment, hospital discharge, and the 3-month visit was similar between the two study arms. However, a greater proportion of subjects in the Tailored arm were on a Class 1 or 3 AAD at the 6-month visit (24% in Tailored arm vs. 20% in Anatomical arm) and the 12-month visit (25% in Tailored arm vs. 23% in Anatomical arm), due at least in part to the need for managing atrial tachycardia that occurred more often in the Tailored arm after the blanking period.
Regarding the primary effectiveness endpoint, of the 180 mITT patients in the Tailored arm who underwent the Tailored procedure and remained in the study, 158 (88%) demonstrated treatment success with respect to the primary effectiveness endpoint as compared to 124 (70%) patients who demonstrated treatment success in the Anatomical mITT population (n=177). The 18% difference in primary effectiveness success between the two study arms was statistically significant (log-rank p<0.0001). The proportion of primary effectiveness successes who were completely off Class 1/3 AADs throughout the effectiveness evaluation period (i.e., after the 3-month blanking period through the 12-month visit) was lower in the Tailored arm than in the Anatomical arm (91/158 or 58% in Tailored
5 of 7
{9}
K243812
arm vs. 82/124 or 66% in Anatomical arm) due at least in part to the use of AADs for managing atrial tachycardia that occurred more often in the Tailored arm after the blanking period. The proportion of primary effectiveness successes who were on a Class 1 or 3 AAD throughout the effectiveness evaluation period was similar between the two study arms (30/158 or 19% in Tailored arm vs. 21/124 or 17% in Anatomical arm).
Regarding secondary effectiveness endpoints, in the mITT population, more patients in the Tailored arm experienced freedom from any atrial arrhythmia (including atrial fibrillation and atrial tachycardia) at 12 months after one or two ablation procedures, with or without the use of anti-arrhythmic medications after a 3-month blanking period, although this difference was not statistically significant (76% in Tailored arm vs. 71% in Anatomical arm). There was no difference between the two study arms in the proportion of subjects free of any atrial arrhythmia recurrence at 12 months after one procedure, with or without the use of antiarrhythmic medications after a 3-month blanking period (60% vs. 60%).
The number of repeat procedures considered in the multiple-procedure endpoint was greater in the Tailored arm than in the Anatomical arm (47 vs. 34 in the Tailored and Anatomical arms, respectively).
Patient's quality of life assessed by using the AFEQT and SF-36 questionnaires improved significantly at 12 months post-index procedure from baseline, independent of the ablation strategy. The differences in the improvement in the AFEQT and SF-36 scores from baseline to 12 months post-index procedure were not statistically significant between the Tailored and Anatomical arms.
The rate of death, cerebrovascular events, or serious treatment-related adverse event at 12 months post procedure, which was a secondary endpoint of the study did not differ statistically between the two study arms (8/187 or 4.3% in the Tailored arm vs. 5/183 or 2.7% in the Anatomical arm). There was no difference between the two study arms in the proportion of subjects with at least one major procedure-related complications (5/187 or 2.7% in Tailored arm vs. 5/183 or 2.7% in Anatomical arm). However, the rate of subjects with at least one minor procedure-related complication is numerically higher in the Tailored arm (8.0%) (driven by more events of fluid overload and pericardial complications) compared to the Anatomical arm (3.2%) due presumably to the longer procedure and radiofrequency (RF) ablation times in the Tailored arm. Indeed, the procedure and ablation time were twice as long in the Tailored arm (178 ± 60 minutes vs. 92 ± 36 minutes for procedure time and 42 ± 17 minutes vs. 20 ± 11 minutes for ablation time).
In summary, the results of the Tailored-AF study showed that ablation of areas with dispersed EGMs detected by the VX1 device, a previous version of the Volta AF-Xplorer that uses the same primary algorithm, in addition to PVI was superior to PVI alone in terms of freedom from AF at one-year follow-up in patients with drug-refractory symptomatic persistent AF, while additional ablation of areas with dispersed EGMs did not result in greater treatment success in terms of freedom from any atrial arrhythmia at one-year follow-up in this population. The latter was due to more occurrences of atrial tachycardia following the tailored ablation procedure targeting areas with dispersed EGMs in addition to PVI. Moreover, compared to PVI alone, the tailored ablation procedure was associated with more minor procedure-related complications (fluid overload and pericardial complications), and longer procedure and ablation times due to additional ablation of areas with dispersed EGMs.
6 of 7
{10}
K243812
# Conclusions
The subject Volta AF-Xplorer device is as safe and effective as the previously cleared Volta AF-Xplorer device (K232616). The Volta AF-Xplorer has the same intended uses and similar indications, and identical technological characteristics, and principles of operation as its predicate device.
Performance data demonstrate the appropriateness of updating the indications for use and product labeling. The updated indications for use statement and additions to the product labeling do not alter the intended use of the device as an electrophysiological evaluation tool and do not affect, or raise different questions of, the device safety and effectiveness.
Thus, the Volta AF-Xplorer is substantially equivalent to its predicate device.
7 of 7
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.