The Mac-Lab System is intended for acquiring, filtering, digitizing, amplifying, measuring and calculating, displaying, recording and monitoring of clinical data from adult and pediatric patients. The Mac-Lab System is configurable. Clinical data includes: ECG waveforms, heart rate, pulse oximetry (SpO2), respiration rate, CO2 (EtCO2), temperature, hemodynamic measures [e.g. valve gradients and areas, cardiac output, shunts, Fractional Flow Reserve (FFR), invasive and noninvasive blood pressure] Physiological parameters such as diastolic, systolic, and mean pressures, and heart rate are derived from the signal data, displayed and recorded. The data is entered manually or acquired via interfaced devices and/or information systems and may be used for report generation. Procedural information and optional anatomical and physiological imaging and data devices may be interfaced (e.g. X-ray, ultrasound, patient monitors and information systems). The Mac-Lab System can display, store and annotate images previously acquired and stored by other systems. Data may be provided to other systems via multiple formats (e.g. HL7, DICOM, Analog outputs). Data may be received from other devices via multiple formats (e.g. DICOM, Analog inputs). Optional accessories for hardware and software include research tools to be used exclusively outside active patient care settings. The purpose of the research tools is to assist researchers or clinicians in developing algorithms. The Mac-Lab System does not have alarms, does not generate energy delivered to the patient, does not administer drugs and does not perform any life-supporting or life-sustaining functions. The Mac-Lab System is not intended for use on unattended patients, or in situations where diagnostic arrhythmia detection is required. The Mac-Lab System provides the ability to transmit patient data for storage, analysis and viewing at distributed locations within a clinical facility via network connectivity. The Mac-Lab System also functions as a stand-alone device. The Mac-Lab System is used in a variety of hospital and clinical settings including interventional laboratories (e.g. cardiac catheterization and radiology), operating room environments, and pre and post areas all under the direct supervision of licensed healthcare practitioners who are responsible for interpreting the data. The CardioLab System is intended for acquiring, filtering, digitizing, amplifying, measuring and calculating, displaying, recording and monitoring of clinical data from adult and pediatric patients. The CardioLab System is configurable. Clinical data includes: ECG waveforms, intracardiac signals, stimulus data, ablation data, pulse oximetry (SpO2), respiration rate, CO2 (EtCO2), temperature, and invasive and noninvasive blood pressure. Physiological parameters such as diastolic, systolic, mean pressures, heart rate, and cycle length are derived from the signal data, displayed and recorded. The data is entered manually or acquired via interfaced devices and/or information systems and may be used for report generation. Procedural information and optional anatomical and physiological imaging and data devices may be interfaced [e.g. X-ray, ultrasound, mapping systems, ablation generators (e.g. RF and cryogenic)], stimulators, patient monitors and information systems. The CardioLab System can display, store and annotate images previously acquired and stored by other systems. Data may be provided to other systems via multiple formats (e.g. HL7, DICOM, Analog outputs). Data may be received from other devices via multiple formats (e.g. DICOM, Analog inputs). Optional accessories for hardware and software include research tools to be used exclusively outside active patient care settings. The purpose of the research tools is to assist researchers or clinicians in developing algorithms. Optional accessories for hardware and software includes a waveform simulator to be used exclusively outside active patient care settings. The waveform simulator may be used for training, demonstration without a patient attached, and as a troubleshooting tool on the CardioLab System. The CardioLab System does not have alarms, does not generate energy delivered to the patient, does not administer drugs and does not perform any life-supporting or life-sustaining functions. The CardioLab System is not intended for use on unattended patients, or in situations where diagnostic arrhythmia detection is required. The CardioLab System provides the ability to transmit patient data for storage, analysis and viewing at distributed locations within a clinical facility via network connectivity. The CardioLab System also functions as a stand-alone device. The CardioLab System is used in a variety of hospital and clinical settings including interventional laboratories (e.g. electrophysiology and cardiac catheterization), operating room environments, and pre and post areas all under the direct supervision of licensed healthcare practitioners who are responsible for interpreting the data.
Device Story
Hemodynamic and electrophysiology recording systems (Mac-Lab, CardioLab, ComboLab, SpecialsLab) acquire, filter, digitize, and display physiological signals (ECG, SpO2, CO2, pressure, intracardiac signals). Used in interventional labs and ORs by clinicians to monitor patients and generate reports. Inputs include direct patient sensors, interfaced imaging/ablation equipment, and manual data entry. System performs signal processing and parameter calculation (e.g., cardiac output, FFR, cycle length). Outputs displayed on-screen and stored for clinical review. Supports network transmission for distributed viewing. Benefits include real-time visualization and documentation of procedural data to assist clinical decision-making. Version 6.9.5 adds Patient Data Module (PDM) support, CARTO system integration, and enhanced waveform/medication reporting.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence consists of bench testing, including risk analysis, design reviews, module verification, system integration testing, and simulated-use validation.
Technological Characteristics
Hemodynamic/EP recording system; configurable hardware/software. Supports TRAM and PDM data acquisition modules. Connectivity via network (HL7, DICOM) or standalone. Software-based signal processing and parameter calculation. No energy delivery or life-support functions.
Indications for Use
Indicated for adult and pediatric patients undergoing hemodynamic and electrophysiology monitoring in interventional labs, ORs, and clinical settings. Not for unattended patients or diagnostic arrhythmia detection.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
Mac-Lab, CardioLab, ComboLab, and SpecialsLab System (K111200)
Submission Summary (Full Text)
{0}------------------------------------------------
K130626 pg 1 of 5
Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circle. The letters and the circle are in black, and the background is white.
MAY 1 2013
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | March 7, 2013 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare<br>(GE Medical Systems Information Technologies, Inc.)<br>8200 West Tower Avenue<br>Milwaukee, WI 53223 |
| Primary Contact Person: | Ms. Carol Alloian<br>Regulatory Leader<br>GE Healthcare<br>(GE Medical Systems Information Technologies, Inc.)<br>Telephone: 224 280-1008<br>Fax: 847 589 8524 |
| Secondary Contact<br>Person: | Mr. Philip Malca<br>Regulatory Affairs Director<br>GE Healthcare,<br>(GE Medical Systems Information Technologies, Inc.)<br>Telephone: 33(0) 1 3070 4207<br>Fax: 33(0) 1 3070 4399 |
| Device Trade Name: | Mac-Lab, CardioLab, ComboLab, SpecialsLab Recording Systems<br>v6.9.5 |
| Common/Usual Name: | Hemodynamic and Electrophysiology (EP) Recording Systems |
| Classification Names: | 21 CFR 870.1425 Computer, Diagnostic Programmable |
| Product Code: | DQK |
| Predicate Device(s): | K111200 Mac-Lab, CardioLab, ComboLab, and SpecialsLab System |
| Device Description: | Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording<br>Systems are hemodynamic and electrophysiology (EP) recording<br>systems. |
| | The product will be available in the following configurations: Mac-Lab<br>System, CardioLab System, SpecialsLab System, or a combination of<br>both Mac-Lab and CardioLab marketed as the ComboLab System.<br>The product designated as SpecialsLab is the same as the Mac-Lab<br>System with the exception that it will support fewer options. The<br>SpecialsLab System performs the same intended use as the Mac-<br>Lab, executes the same software, and runs on the same hardware. |
| | The Mac-Lab, CardioLab, and ComboLab Recording Systems are<br>each available in several configurations ranging from basic to<br>advanced functionality. |
| Intended Use: | Mac-Lab |
| | The Mac-Lab System is intended for acquiring, filtering, digitizing,<br>amplifying, measuring and calculating, displaying, recording and<br>monitoring of clinical data from adult and pediatric patients. The<br>Mac-Lab System is configurable. Clinical data includes: ECG<br>waveforms, heart rate, pulse oximetry (SpO2), respiration rate, CO2<br>(EtCO2), temperature, hemodynamic measures [e.g. valve gradients<br>and areas, cardiac output, shunts, Fractional Flow Reserve (FFR),<br>invasive and noninvasive blood pressure] Physiological parameters<br>such as diastolic, systolic, and mean pressures, and heart rate are<br>derived from the signal data, displayed and recorded. The data is<br>entered manually or acquired via interfaced devices and/or<br>information systems and may be used for report generation.<br>Procedural information and optional anatomical and physiological<br>imaging and data devices may be interfaced (e.g. X-ray, ultrasound, |
| | patient monitors and information systems). The Mac-Lab System can<br>display, store and annotate images previously acquired and stored by<br>other systems. Data may be provided to other systems via multiple<br>formats (e.g. HL7, DICOM, Analog outputs). Data may be received<br>from other devices via multiple formats (e.g. DICOM, Analog inputs). |
| | Optional accessories for hardware and software include research<br>tools to be used exclusively outside active patient care settings. The<br>purpose of the research tools is to assist researchers or clinicians in<br>developing algorithms. |
| | The Mac-Lab System does not have alarms, does not generate<br>energy delivered to the patient, does not administer drugs and does<br>not perform any life-supporting or life-sustaining functions. The Mac-<br>Lab System is not intended for use on unattended patients, or in<br>situations where diagnostic arrhythmia detection is required. |
| | The Mac-Lab System provides the ability to transmit patient data for<br>storage, analysis and viewing at distributed locations within a clinical |
| | |
| facility via network connectivity. The Mac-Lab System also functions<br>as a stand-alone device. The Mac-Lab System is used in a variety of<br>hospital and clinical settings including interventional laboratories (e.g.<br>cardiac catheterization and radiology), operating room environments,<br>and pre and post areas all under the direct supervision of licensed<br>healthcare practitioners who are responsible for interpreting the data. | |
| CardioLab | |
| The CardioLab System is intended for acquiring, filtering, digitizing,<br>amplifying, measuring and calculating, displaying, recording and<br>monitoring of clinical data from adult and pediatric patients. The<br>CardioLab System is configurable. Clinical data includes: ECG<br>waveforms, intracardiac signals, stimulus data, ablation data, pulse<br>oximetry (SpO2), respiration rate, CO2 (EtCO2), temperature, and<br>invasive and noninvasive blood pressure. Physiological parameters<br>such as diastolic, systolic, mean pressures, heart rate, and cycle<br>length are derived from the signal data, displayed and recorded. The<br>data is entered manually or acquired via interfaced devices and/or<br>information systems and may be used for report generation. | |
| Procedural information and optional anatomical and physiological<br>imaging and data devices may be interfaced [e.g. X-ray, ultrasound,<br>mapping systems, ablation generators (e.g. RF and cryogenic)],<br>stimulators, patient monitors and information systems. The CardioLab<br>System can display, store and annotate images previously acquired<br>and stored by other systems. Data may be provided to other systems<br>via multiple formats (e.g. HL7, DICOM, Analog outputs). Data may be<br>received from other devices via multiple formats (e.g. DICOM, Analog<br>inputs). | |
| Optional accessories for hardware and software include research<br>tools to be used exclusively outside active patient care settings. The<br>purpose of the research tools is to assist researchers or clinicians in<br>developing algorithms. | |
| Optional accessories for hardware and software includes a waveform<br>simulator to be used exclusively outside active patient care settings.<br>The waveform simulator may be used for training, demonstration<br>without a patient attached, and as a troubleshooting tool on the<br>CardioLab System. | |
| The CardioLab System does not have alarms, does not generate<br>energy delivered to the patient, does not administer drugs and does<br>not perform any life-supporting or life-sustaining functions. The<br>CardioLab System is not intended for use on unattended patients, or<br>in situations where diagnostic arrhythmia detection is required. | |
| The CardioLab System provides the ability to transmit patient data for<br>storage, analysis and viewing at distributed locations within a clinical<br>facility via network connectivity. The CardioLab System also functions<br>as a stand-alone device. The CardioLab System is used in a variety<br>of hospital and clinical settings including interventional laboratories | |
| Technology: | (e.g. electrophysiology and cardiac catheterization), operating room<br>environments, and pre and post areas all under the direct supervision<br>of licensed healthcare practitioners who are responsible for<br>interpreting the data.<br><br><b>ComboLab</b><br>The ComboLab System is the combination of both the Mac-Lab and<br>CardioLab Systems. The ComboLab System allows the user to run<br>either the Mac-Lab System or the CardioLab System, although only<br>one system may be used at a time. The ComboLab System executes<br>the same software and runs on the same hardware in the same<br>environments as the Mac-Lab and CardioLab Systems.<br><br><b>SpecialsLab</b><br>The SpecialsLab System executes the same software and runs on<br>the same hardware in the same environments as the Mac-Lab<br>System. Products designated as a SpecialsLab System support fewer<br>options than the Mac-Lab system. |
| | The proposed Mac-Lab, CardioLab, ComboLab, and SpecialsLab<br>Recording Systems employ the same fundamental scientific<br>technology as the predicate devices. |
| | The proposed Mac-Lab, CardioLab, ComboLab, SpecialsLab<br>Recording Systems v 6.9.5 adds an additional data acquisition<br>alternative. In addition to TRAM, the Patient Data Module (PDM) is<br>now supported - inclusive of software, service tools, and hardware<br>(PDM and base station). |
| | Additionally the proposed v 6.9.5 includes software application<br>enhancements to existing features and functions, including the<br>following:<br>Receive and store movies and additional snapshots from<br>CARTO™ system Provide expanded functionality to allow currently available<br>signal data to be requested on-demand Allow association of user defined author names to entries<br>recorded in the procedure log Additional window on Mac-Lab to allow a duplicate display of<br>recorded pressure measurements. Enhanced presentation of the Medication Summary windows Enhanced existing waveform export to include ablation data<br>export and added additional example algorithm Enhanced Service Utilities Support for additional versions of third party software for<br>Antivirus and Backup solutions Hardware and software updates to address normal<br>technological advancement and obsolescence issues |
| Determination of<br>Substantial Equivalence: | Summary of Non-Clinical Tests: |
| | The Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems and their applications comply with voluntary standards as detailed in Section 9, 11 and 17 of this premarket submission. The following quality assurance measures were applied to the development of the system: |
| | Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation) |
| | Summary of Clinical Tests: |
| | The subject of this premarket submission, Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems and their applications did not require clinical studies to support substantial equivalence. |
| Conclusion: | GE Healthcare considers the Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems and their applications to be as safe, as effective, and performance is substantially equivalent to the predicate device(s). |
{1}------------------------------------------------
K130626 pg 2ofs
Image /page/1/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized and appear to be hand-drawn or calligraphic, giving the logo a classic and recognizable look.
{2}------------------------------------------------
K130626 Pg 30f5
## GE Healthcare 510(k) Premarket Notification Submission Mac-Lab, CardioLab, ComboLab, SpecialsLab v 6.9.5
Image /page/2/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. There are three small, rounded shapes evenly spaced around the outside of the circle, giving the logo a distinctive and recognizable appearance. The logo is black and white.
{3}------------------------------------------------
## GE Healthcare 510(k) Premarket Notification Submission Mac-Lab, CardioLab, ComboLab, SpecialsLab v 6.9.5
Image /page/3/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are in black, contrasting with the white background.
{4}------------------------------------------------
K130626 pg 5 of 5
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. There are small, stylized flourishes or dots placed around the circle, enhancing the logo's design.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
May 1. 2013
GE Healthcare C/O Philip Malca Regulatory Affairs Director 8200 West Tower Avenue Milwaukee, Wisconsin 53223
Re: K130626
> Trade/Device Name: Mac-Lab, Cardio-Lab, ComboLab, and SpecialsLab Recording Systems
Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable Diagnostic Computer
Regulatory Class: Class II Product Code: DQK
Dated: March 29, 2013
- Received: April 2, 2013
Dear Mr. Malca:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Philip Malca
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
OwenPEaris -S
for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### GE Healthcare 510(k) Premarket Notification Submission Mac-Lab, CardioLab, ComboLab, SpecialsLab v 6.9.5
#### 510(k) Number (if known): To be Assianed
Device Name: Mac-Lab, CardioLab, ComboLab, SpecialsLab Recording System(s)
#### Indications for Use: Mac-Lab
The Mac-Lab System is intended for acquiring, filtering, digitizing, amplifying, measuring and calculating, displaying, recording and monitoring of clinical data from adult and pediatric patients. The Mac-Lab System is configurable. Clinical data includes: ECG waveforms, heart rate, pulse oximetry (SpO2), respiration rate, CO2 (EtCO2), temperature, hemodynamic measures [e.g. valve gradients and areas, cardiac output, shunts, Fractional Flow Reserve (FFR), invasive and noninvasive blood pressure] Physiological parameters such as diastolic. systolic, mean pressures, and heart rate are derived from the signal data, displayed and recorded. The data is entered manually or acquired via interfaced devices and/or information systems and may be used for report generation.
Procedural information and optional anatomical and physiological imaging and data devices may be interfaced (e.g. X-ray, ultrasound, patient monitors and information systems). The Mac-Lab System can display, store and annotate images previously acquired and stored by other systems. Data may be provided to other systems via multiple formats (e.g. HL7, DICOM. Analog outputs). Data may be received from other devices via multiple formats (e.g. DICOM, Analog inputs).
Optional accessories for hardware and software include research tools to be used exclusively outside active patient care settings. The purpose of the research tools is to assist researchers or clinicians in developing algorithms.
The Mac-Lab System does not have alarms, does not generate energy delivered to the patient, does not administer drugs and does not perform any life-supporting or life-sustaining functions. The Mac-Lab System is not intended for use on unattended patients, or in situations where diagnostic arrhythmia detection is required.
The Mac-Lab System provides the ability to transmit patient data for storage, analysis and viewing at distributed locations within a clinical facility via network connectivity. The Mac-Lab System also functions as a stand-alone device. The Mac-Lab System is used in a variety of hospital and clinical settings interventional laboratories (e.g. cardiac catheterization and radiology), operating room environments, and post areas all under the direct supervision of licensed healthcare practitioners who are responsible for interpreting the data.
Prescription Use x AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use_ (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Owen P. Faris -S
2013.05.01 15:45:04
-04'00'
FM000031
{8}------------------------------------------------
### GE Healthcare
Image /page/8/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. There are decorative swirls around the outside of the circle. The logo is black and white.
## 510(k) Premarket Notification Submission Mac-Lab, CardioLab, ComboLab, SpecialsLab v 6.9.5
#### 510(k) Number (if known): To Be Assigned
Device Name: Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems Indications for Use:
### SpecialsLab
The SpecialsLab System executes the same software and runs on the same hardware in the same environments as the Mac-Lab System. Products designated as a SpecialsLab System support fewer options than the Mac-Lab system.
### ComboLab
The ComboLab System is the combination of both the Mac-Lab and CardioLab Systems. The ComboLab System allows the user to run either the Mac-Lab System or the CardioLab System, although only one system may be used at a time. The ComboLab System executes the same software and runs on the same hardware in the same environments as the Mac-Lab and CardioLab Systems.
Prescription Use X
AND/OR
(Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Owen P. Faris -S
2013.05.01 15:45:41
-04'00'
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The logo is black and white, with the letters and border in black and the background in white.
#### 510(k) Number (if known): To Be Assigned
Device Name: Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording System(s)
#### Indications for Use: CardioLab
The CardioLab System is intended for acquiring, filtering, digitizing, amplifying, measuring and calculating, displaying, recording and monitoring of clinical data from adult and pediatric patients. The CardioLab System is configurable. Clinical data includes: ECG waveforms, intracardiac signals, stimulus data, ablation data, pulse oximetry (SpO2), respiration rate, CO2 (EtCO2), temperature, and invasive and noninvasive blood pressure. Physiological parameters such as diastolic, systolic, mean pressures, heart rate, and cycle length are derived from the signal data, displayed and recorded. The data is entered manually or acquired via interfaced devices and/or information systems and may be used for report generation.
Procedural information and optional and physiological imaging and data devices may be interfaced [e.g. X-ray, ultrasound, mapping systems, ablation generators (e.g. RF and cryogenic)], stimulators, patient monitors and information systems. The CardioLab System can display, store and annotate images previously acquired and stored by other systems. Data may be provided to other systems via multiple formats (e.g. HL7, DICOM, Analog outputs). Data may be received from other devices via multiple formats (e.g. DICOM, Analog inputs).
Optional accessories for hardware and software include research tools to be used exclusively outside active patient care settings. The purpose of the research tools is to assist researchers or clinicians in developing algorithms.
Optional accessories for hardware and software include a waveform simulator to be used exclusively outside active patient care settings. The waveform simulator may be used for training, demonstration without a patient attached, and as a troubleshooting tool on the CardioLab System.
The CardioLab System does not have alarms, does not generate energy delivered to the patient, does not administer drugs and does not perform any life-sustaining functions. The CardioLab System is not intended for use on unattended patients, or in situations where diagnostic arrhythmia detection is required.
The CardioLab System provides the ability to transmit patient data for storage, analysis and viewing at distributed locations within a clinical facility via network connectivity. The CardioLab System also functions as a stand-alone device. The CardioLab System is used in a variety of hospital and clinical settings including interventional laboratories (e.g. electrophysiology and cardiac catheterization), operating room environments, and pre and post areas all under the direct supervision of licensed healthcare practitioners who are responsible for interpreting the data.
Prescription Use x AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Owen P. Faris -S
2013.05.01
15:45:25 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.