K201298 · Volta Medical · DQK · Sep 16, 2020 · Cardiovascular
Device Facts
Record ID
K201298
Device Name
VX1
Applicant
Volta Medical
Product Code
DQK · Cardiovascular
Decision Date
Sep 16, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
Intended Use
The VX1 assists operators in the real-time manual annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia. The clinical significance of utilizing the VX1 software to help identify areas with intra-cardial electrograms exhibiting spatiotemporal dispersion for catheter ablation of atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations.
Device Story
VX1 is a machine and deep learning-based software system assisting clinicians in real-time manual annotation of 3D atrial maps. It identifies dispersed intra-cardiac atrial electrograms (DEs) during atrial fibrillation or tachycardia. Input: intra-cardiac atrial electrograms from standard mapping and coronary sinus catheters via data acquisition systems (LabSystem Pro or MacLab CardioLab) and an Advantech PCI-1713U analog-to-digital converter. Operation: software analyzes signals to cue operators to regions harboring DEs and estimates cycle length. Output: graphical display on a computer monitor located outside the sterile field. Used in cardiac electrophysiology labs by physicians. Benefits: provides real-time identification of signal dispersion patterns to assist in mapping procedures; does not direct treatment or affect arrhythmia outcomes.
Clinical Evidence
Clinical study (OUS) involved 300 patients across 8 centers and 18 operators. Results showed VX1 implementation did not increase procedure risk or time and demonstrated performance comparable to the predicate's CFAE module in locating signal disruptions. Non-clinical evidence included a reader study (14,370 electrograms) showing high agreement between experts and VX1, a period estimation study (2,550 electrograms) where VX1 outperformed AC and FFT algorithms, and 5-fold cross-validation on 275,020 electrograms (accuracy >89%).
Technological Characteristics
Programmable diagnostic computer; signal processing-based atrial mapping system. Uses off-the-shelf PC, monitor, and Advantech PCI-1713U analog-to-digital converter. Compatible with standard electrophysiology catheters (1mm electrode size, 2-3mm spacing). Software utilizes machine and deep learning models. Standalone local application, not network-connected during commercial use. Operates via keyboard/mouse input.
Indications for Use
Indicated for use by operators to assist in real-time manual annotation of 3D anatomical and electrical maps of human atria to identify multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
K232616 — Volta AF-Xplorer · Volta Medical · Sep 27, 2023
K251518 — Volta AF-Xplorer II · Volta Medical · Sep 12, 2025
K243812 — Volta AF-Xplorer · Volta Medical · May 9, 2025
K251344 — OptiMap System · Boston Scientific Corporation · May 28, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 16, 2020
Volta Medical % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, Suite 2300 Philadelphia, Pennsylvania 19103
Re: K201298
Trade/Device Name: VX1 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable diagnostic computer Regulatory Class: Class II Product Code: DOK Dated: August 17, 2020 Received: August 17, 2020
Dear Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Fellman Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K201298
Device Name
VX1
Indications for Use (Describe)
The VX1 assists operators in the real-time manual annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia.
The clinical significance of utilizing the VX1 software to help identify areas with intra-cardial electrograms exhibiting spatiotemporal dispersion for catheter ablation of atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number
{3}------------------------------------------------
# 510(k) SUMMARY VOLTA MEDICAL's VX1 K201298
#### Submitter
Volta Medical 29 Boulevard de Louvain 13008 Marseille France Phone: (+33) 6 52 57 96 94
Contact Person: Théophile Mohr Durdez Date Prepared: September 15, 2020
| Name of Device: | VX1 |
|-----------------------|------------------------------------------------------------|
| Common or Usual Name: | Cardiac mapping system |
| Classification Name: | Programmable Diagnostic Computer |
| Regulatory Class: | 21 C.F.R. § 870.1425 |
| Product Code: | DQK |
| Predicate Devices: | Biosense Webster, Inc.'s CARTO® XP V10 System<br>(K093566) |
### Device Description:
The VX1 "" is a machine and deep learning based-algorithm designed to assist operators in the real-time manual annotation of 3D anatomical and electrical maps of the human atria for the presence of electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia", i.e., dispersed intra-cardiac atrial electrograms (DEs). The device works with all existing 510(k) cleared catheters that meet specific dimension requirements and with two data acquisition systems: the LabSystem Pro Acquisition System (Boston Scientific) (K141185) and the MacLab CardioLab Acquisition System (General Electric) (K130626). A connection cable is used to connect the data acquisition system with an Advantech PCI-1713U analog-to-digital converter, which transmits the acquired information to a nearby computer that hosts the VX1 software. The computer and its attached display are located outside the sterile operating room area. The VX1 software analyzes the patient's electrograms to cue operators in real-time to intra-cardiac electrograms of interest for atrial regions harboring DEs as well as a cycle length estimation from electrograms recorded with the mapping and the coronary sinus catheters. The results of the analysis are graphically presented on the attached computer display.
{4}------------------------------------------------
### Intended Use / Indications for Use:
The VX1 assists operators in the real-time manual annotation of 3D anatomical and electrical maps of human atria for the presence of multipolar intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during atrial fibrillation or atrial tachycardia.
The clinical significance of utilizing the VX1 software to help identify areas with intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion for catheter ablation of atrial arrhythmias, such as atrial fibrillation, has not been established by clinical investigations.
### Summary of Technological Characteristics Compared to the Predicate Device:
The VX1 device and its predicate (the CARTO® XP V10 System; K093566) both work with standard electrophysiology catheters to aid in mapping the atria. The predicate device contains several functional modules that are not included in the VX1 device; however, as described in the table below. these functions can generally be performed by another (compatible) device that can be used during the same procedure that includes the VX1 device. Whether these functions are bundled in one device (as with the predicate device) or performed by separate devices during the same procedure does not affect the subject device performance nor raise different questions of safety or effectiveness. Because mapping can be used independently, the provision of a separate functionality in the VX1 versus an integrated functionality in the predicate does not raise new types of questions or otherwise adversely impact performance, as confirmed in the submitted testing.
In particular, the VX1 performs an equivalent function to the CFAE Module of the predicate device. Both devices aid operators by assisting in annotating complex electrical maps of the atria, and both devices process and output information via a computer and display that are operated by use of a keyboard / mouse. Additionally, the signals identified by the VX1 device are of the same nature (multipolar atrial electrograms) as those identified by the CFAE module of CARTO, as demonstrated in the definition of dispersion which partly relies on the definition of CFAE. Both devices also have the same input (intra-cardiac signals) and output (associated dispersion / fractionation).
The CARTO CFAE module: however, performs analysis of individual per-procedural data sets, whereas VX1 includes analytical parameters that pertain to previous similar procedures. Specifically, VX1 methods rely on conceptualizing a discriminant algorithm, which, during training, drew on anonymized information from a very large database of 1.5 second snippets of multipolar intra-cardiac atrial electrograms located on a dedicated data server. In actual commercial use, however, the trained machine and deep learning models are incorporated into the stand alone local software application which is not connected to a network.
Like the predicate device, the VX1 assists in the real-time annotation of complex 3D anatomical and electrical maps of the human atria, including the presence of multipolar electrograms. Thus, both the subject and predicate device have the same intended use,
The indications for use of the CARTO® XP V10 System, while broader, specify that it "displays Complex Fractionated Atrial Electrograms (CFAE)." Spatiotemporal dispersed Intra-cardiac atrial electrograms are a non-exclusive subset of CFAEs in which the
{5}------------------------------------------------
fractionation occurs in a non-simultaneous fashion at neighboring electrode locations (temporal dispersion) and are organized in well-defined clusters (spatial dispersion). Thus, the predicate CFAE technology relates to the same cardiac phenomena for which the VX1 is indicated for use.
Neither device is intended for directing treatment or affecting the outcome of any particular heart arrhythmia.
The difference in indications therefore does not alter the intended diagnostic effect, and therefore the first criterion for a finding of substantial equivalence is satisfied.
| | Volta Medical<br>VX1 | CARTO® XP System with<br>CARTOXPRESS™ Module<br>(V10)<br>K093566 | Comments |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation | 21 C.F.R. § 870.1425<br>Programmable<br>Diagnostic Computer<br>Product Code: DQK | 21 C.F.R. § 870.1425<br>Programmable Diagnostic<br>Computer<br>Product Code: DQK | Same |
| Indications<br>for Use | The VX1 assists<br>operators in the real-<br>time manual<br>annotation of 3D<br>anatomical and<br>electrical maps of<br>human atria for the<br>presence of multipolar<br>intra-cardiac atrial<br>electrograms exhibiting<br>spatiotemporal<br>dispersion during atrial<br>fibrillation or atrial<br>tachycardia.<br>The clinical<br>significance of utilizing<br>the VX1 software to<br>help identify areas with<br>intra-cardiac atrial<br>electrograms exhibiting<br>spatiotemporal<br>dispersion for catheter<br>ablation of atrial<br>arrhythmias, such as<br>atrial fibrillation, has<br>not been established<br>by clinical<br>investigations. | The intended use of the<br>CARTO® XP System with<br>CARTOXPRESS™ Module<br>(V10) is catheter-based<br>atrial and ventricular<br>mapping.<br>The CARTO® XP System<br>with CARTOXPRESS™<br>Module (V10) allows real-<br>time display of cardiac<br>maps in a number of<br>different formats. Maps<br>may be displayed as<br>cardiac electrical activation<br>maps, cardiac electrical<br>propagation maps, cardiac<br>electrical potential maps,<br>impedance maps, and<br>cardiac chamber geometry<br>maps. The acquired patient<br>signals, including body<br>surface EGG and intra-<br>cardiac electrograms may<br>also be displayed in real-<br>time on the system display<br>screen.<br>CARTO® XP System with<br>CARTOXPRESS™ Module<br>(V10) includes the<br>CARTOMERGE® capability<br>to import, register, and | Like the predicate device,<br>the VX1 assists in the real-<br>time annotation of complex<br>3D anatomical and<br>electrical maps of the<br>human atria, including the<br>presence of multipolar<br>electrograms. Thus, both<br>the subject and predicate<br>device have the same<br>intended use.<br>The indications for use of<br>the CARTO® XP V10<br>System, while broader,<br>specify that it "displays<br>Complex Fractionated<br>Atrial Electrograms<br>(CFAE)." Spatiotemporal<br>dispersed intra-cardiac<br>atrial electrograms are a<br>non-exclusive subset of<br>CFAEs in which the<br>fractionation occurs in a<br>non-simultaneous fashion<br>at neighboring electrode<br>locations (temporal<br>dispersion) and are<br>organized in well-defined<br>clusters (spatial<br>dispersion). Thus, the<br>predicate CFAE technology<br>relates to the same cardiac |
| | Volta Medical<br>VX1 | CARTO® XP System with<br>CARTOXPRESSTM Module<br>(V10)<br>K093566 | Comments |
| | | merge CT or MRI structural<br>images with CARTO Maps<br>physiological information<br>and real-time catheter<br>navigation. CARTO® XP<br>System with<br>CARTOXPRESSTM Module<br>(V10) also allows the<br>integration of intra-cardiac<br>echo (ICE) 2D images to<br>provide 3D combined maps.<br>CARTO® XP System with<br>CARTOXPRESSTM Module<br>(V10) includes Fast<br>Anatomical Mapping (FAM)<br>that is a method for quick<br>creation of cardiac<br>anatomical volumes using<br>catheters with magnetic<br>location sensors, supports<br>the LASSO® NAV Catheter<br>with location sensors,<br>displays Complex<br>Fractionated Atrial<br>Electrograms (CFAE), and<br>adds Pace-Mapping<br>Software (PaSo) ECG<br>signal correlation tool. | phenomena for which the<br>VX1 is indicated for use.<br>Neither device is intended<br>for directing treatment or<br>affecting the outcome of<br>any particular heart<br>arrhythmia.<br>Additional CARTO® XP<br>System with<br>CARTOXPRESSTM Module<br>(V10) functions can be<br>accomplished by other<br>(compatible) devices used<br>in the same procedure as<br>the VX1.<br>Thus, the difference in<br>indications does not alter<br>the intended diagnostic<br>effect. |
| System Type | Signal processing-<br>based atrial mapping<br>system. | Electromagnetic and<br>catheter-based atrial and<br>ventricular mapping system. | Similar catheter-based<br>systems. The lack of<br>ventricular mapping in the<br>subject device does not<br>affect its performance<br>relative to its atrial mapping<br>functionality. Further, use<br>of the VX1 does not<br>preclude the use of other<br>mapping systems during<br>the same procedure which<br>can perform broad<br>mapping features (e.g.,<br>ventricular mapping). The<br>VX1 simply provides<br>complementary additional<br>information regarding<br>dispersed intra-cardiac<br>atrial electrograms to the<br>user in addition to the |
| | Volta Medical<br>VX1 | CARTO® XP System with<br>CARTOXPRESS™ Module<br>(V10)<br>K093566 | Comments |
| | | | information provided by<br>traditional mapping<br>systems. Thus, there are<br>no different questions of<br>safety or effectiveness<br>raised by the VX1 system. |
| Display(s)<br>Control | Color monitor<br>Standard keyboard /<br>mouse | Color monitor<br>Standard keyboard / mouse | Same<br>Same |
| Inputs<br>Required | Intra-cardiac atrial<br>electrograms that are<br>recorded with<br>catheters, amplified<br>and redirected to an<br>acquisition system on<br>which they are<br>displayed. | ECG and electrogram data<br>from the patient captured<br>with proprietary and non-<br>proprietary catheters and<br>patches. | Comparable inputs to<br>those captured by the<br>CARTO XP System (e.g.,<br>ECG and EGM data) can<br>be captured by compatible<br>devices used during the<br>same procedure that<br>includes VX1.<br>Whether these functions<br>are bundled in one device<br>(as with the predicate<br>device) or performed by<br>separate devices during<br>the same procedure does<br>not affect the subject<br>device performance nor<br>raise different questions of<br>safety or effectiveness.<br>Use of the VX1 does not<br>preclude the use of other<br>mapping systems during<br>the same procedure which<br>can provide appropriate<br>input (ECG and patches). |
| Principal<br>Mapping<br>Output | Displays adjudications<br>as visual cues after<br>analyzing intra-cardiac<br>atrial electrograms in<br>real-time using signal<br>processing techniques. | Displays anatomical and<br>electrical maps such as<br>activation and voltage maps<br>of the human heart in real-<br>time using magnetic<br>navigation techniques and<br>ECG-EGM analysis. | Use of the VX1 does not<br>preclude the use of other<br>mapping systems during<br>the same procedure which<br>can perform these<br>additional mapping<br>features. Thus, there is no<br>effect on safety or<br>effectiveness. |
| Map Types<br>Generated | Real-time Dispersed<br>Electrogram (DE)<br>subtype of multipolar<br>electrogram map- The<br>operator is provided<br>with display of the | Real-time 3D cardiac<br>maps including: cardiac<br>electrical activation maps,<br>cardiac electrical<br>propagation maps,<br>impedance maps, and | VX1 provides real-time<br>adjudications of dispersion<br>without providing<br>corresponding 3D mapping<br>information. The use of the<br>VX1 does not preclude the |
| | Volta Medical<br>VX1 | CARTO® XP System with<br>CARTOXPRESS™ Module<br>(V10)<br>K093566 | Comments |
| | electrode locations<br>where dispersed or<br>non-dispersed<br>electrograms have<br>been recorded during<br>atrial fibrillation or atrial<br>tachycardia. | cardiac chamber geometry<br>maps.<br>Complex Fractionated Atrial<br>Electrograms (CFAE) are<br>displayed on maps that are<br>colored according to the<br>duration and repetitions of<br>fragmented electrograms. | use of other mapping<br>systems during the same<br>procedure which can<br>perform 3D cardiac<br>maps including: cardiac<br>electrical activation maps,<br>cardiac electrical<br>propagation maps,<br>impedance maps, and<br>cardiac chamber geometry<br>maps. |
| Compatible<br>Catheters | Any compatible<br>catheter meeting the<br>requirements listed<br>below:<br><b>Compatibility<br/>Requirements:</b><br>Mapping catheter:<br>- Electrode size : 1 mm<br>- Inter-electrode<br>Spacing : 2 - 3 mm<br>- Number of selected<br>dipoles : 10<br>Coronary sinus<br>catheter:<br>- Electrode size: 1 mm<br>- Inter-<br>electrode Spacing : 2 -<br>3 mm<br>- Number of selected<br>dipoles : 5 | Supports the Biosense<br>Webster NAVISTAR® family<br>of catheters with built-in<br>magnetic sensor for real-<br>time navigation, including<br>the LASSO® NAV Catheter<br>(K093376) with location<br>sensors and the ability to<br>display the catheter's<br>circular loop based on the<br>position of magnetic single<br>axial sensors (SAS) and<br>provide selective pacing<br>through the loop electrodes.<br>Supports the CARTO<br>SOUNDSTAR 3D<br>ultrasound catheter; a<br>location sensing<br>Intracardiac Echo (ICE)<br>catheter with an acoustic<br>array embedded in the<br>catheter tip that allows the<br>acquisition of real time<br>ultrasound images. The<br>location sensor enables the<br>accurate location of the<br>U/S-observed intracardiac<br>anatomies in the CARTO<br>XP EP Navigation System<br>spatial coordinates. | The CARTO proprietary<br>and non-proprietary<br>catheters perform identical<br>electromagnetic sensing<br>functions to the catheters<br>that are compatible for use<br>with VX1.<br>Other systems' catheters<br>may be used to provide<br>real time navigation and/or<br>catheter loop display,<br>selective pacing through<br>the loop electrodes, or<br>ultrasound integration.<br>The difference in specific<br>catheters does not affect<br>the device performance or<br>raise different questions of<br>safety or effectiveness. |
| Hardware<br>Design and<br>Materials | Off-the-shelf analog /<br>digital converter,<br>computer and monitor,<br>connection cable,<br>acquisition system,<br>proprietary software | Off-the-shelf information<br>technology (IT) hardware:<br>computer and monitor,<br>proprietary catheters,<br>patient interface unit and<br>accessories. | Equivalent, but with certain<br>compatible vs. proprietary<br>equipment (e.g.,<br>catheters). |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
## Performance Data - Nonclinical Tests:
The Volta Medical VX1 was subjected to extensive non-clinical testing including rigorous verification and validation testing. The Company additionally conducted the following studies:
(1) A Reader Study to evaluate the level of inter-operator agreement after unlimited-time physician visual intra-cardiac atrial DE annotation and determine whether algorithm-enabled classification acceptably correlates with physician annotation. A dataset of 14,370 electrograms were adjudicated by three independent experts and VX1 for the presence or absence of dispersed intra-cardiac atrial electrograms. The study demonstrated that the expert readers both strongly agreed with one another as to the presence or absence of dispersed intra-cardiac atrial electrograms in the evaluated electrogram recordings and that they, similarly, independently agreed with the VX1 adjudication of the presence or absence of intra-cardiac atrial DEs.
(2) A period estimation algorithm testing to assess the performance of the VX1 software against two alternative period estimation approaches: a sophisticated auto-correlation (AC) algorithm and a Fast Fourier Transform (FFT) with Blackman window processing across a dataset of 2,550 electrograms. Overall, the VX1 period estimation algorithm outperformed both the FFT-based and the AC-based algorithms for evaluating noisy and non-periodic electrograms and for estimating the period on the period estimation dataset.
(3) A k-fold cross-validation of VX1 algorithm was performed on the training database of 275,020 1.5 second annotated electrograms. 5-fold cross validation was performed on the model with a resulting accuracy of over 89%.
(4) Limited time annotation testing in which 1,020 electrograms were classified by two annotating experts who were allowed unlimited time to make a classification determination and reach a consensus as to the presence or absence of intra-cardiac atrial dispersion. VX1 performance was compared to the performance of 28 cardiac electrophysiologists who underwent a full-day seminar on spatiotemporal dispersion. The cardiac electrophysiologists each classified 1,020 electrograms as either dispersed or non-dispersed with 17 seconds allotted for each classification. The VX1 algorithm demonstrated superior performance (agreement to the annotating experts) to the cardiac electrophysiologists in identifying dispersed and non-dispersed intra-cardiac atrial electrograms.
# Performance Data - Clinical Tests:
An OUS clinical study was performed to evaluate the performance of the VX1 in the detection of intra-cardiac atrial electrograms exhibiting spatiotemporal dispersion during a catheter ablation procedure for atrial fibrillation or atrial tachycardia. The study was aimed at evaluating the usability and efficacy of VX1 for detecting DEs and involved 8 centers. 18 operators, and 300 patients. The results indicate that the implementation of VX1 in a cardiac electrophysiology laboratory is not associated with additional risk or procedure time. Also,
{10}------------------------------------------------
the VX1 device demonstrated comparable performance in locating electrical signal transmission pattern disruptions as the predicate's CFAE module.
# Conclusions:
The VX1 device and its predicate both work with standard electrophysiology catheters to aid in mapping the human atria. The predicate device contains several functional modules that are not included in the VX1 device; however, these functions can generally be performed by another device that can be used during the same procedure as the VX1 device. Because mapping can be used independently, the provision of a separate functionality in the VX1 versus an integrated functionality in the predicate does not raise new types of questions or otherwise adversely impact performance, as confirmed in the submitted testing.
Specifically, the VX1 device is substantially equivalent to the CARTO® XP System with CARTOXPRESS™ Module. In particular, the VX1 performs an equivalent function to the CFAE Module of the predicate device, and works with mapping systems and compatible catheters to form a system which is equivalent to the predicate system. The VX1 and CFAE module have similar technological characteristics and principles of operation. The signals identified by the VX1 device are of the same nature (multipolar electrograms), and have the same input (intra-cardiac atrial signals) and output (associated dispersion / fractionation). The VX1 has the same intended use and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended use of the device as an electrophysiological evaluation tool and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the VX1 and its predicate device raise no new issues of safety or effectiveness. Performance data, as described above, demonstrate that the VX1 device is as safe and effective as the predicate. Thus, the VX1 device is substantially equivalent.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.