K242686 · Jjgc Indústria E Comércia DE Materiais Dentários S.A. · NHA · Dec 4, 2024 · Dental
Device Facts
Record ID
K242686
Device Name
Neodent Implant System
Applicant
Jjgc Indústria E Comércia DE Materiais Dentários S.A.
Product Code
NHA · Dental
Decision Date
Dec 4, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NeoConvert Solution: Instruments and components used to perform the procedure of converting removable prostheses into temporary fixed prostheses on GM line implants installed in the maxilla and/or mandible. Direct Screw to MUA: The Direct Screw to MUA is suitable for attaching the multiple restoration milled from a CAD/CAM library to the prosthetic abutment (mini or micro abutments). GM Attachment TiN 30°: This product is used to stabilize implant-mucosa supported removable full prostheses (retained on the implant and supported on the mucosa), on implants installed in the maxilla or mandibula. It can be used in immediate or conventional rehabilitation procedures.
Device Story
Neodent Implant System additions include NeoConvert Solution, Direct Screw to MUA, and GM Attachment TiN 30°. NeoConvert converts removable temporary prostheses into screw-retained fixed temporary prostheses. Direct Screw to MUA fixes digitally designed multi-unit milled restorations directly onto mini/micro abutments, bypassing passive seating cylinders. GM Attachment TiN 30° stabilizes implant-mucosa-supported removable full prostheses. Devices are used in dental clinics by clinicians; components are compatible with existing Neodent implants. Personalized copings for Direct Screw to MUA are manufactured at a Straumann-validated milling center. These components provide additional options for multi-unit rehabilitation, improving chewing function and patient aesthetics.
Clinical Evidence
Bench testing only. Dynamic fatigue testing per ISO 14801 and torsion testing demonstrated performance equal or superior to predicates. Biocompatibility evaluated per ISO 10993-1, -5, and -18. Sterilization validated per ISO 11135 and ISO 17665. No clinical data presented.
Technological Characteristics
Materials: Titanium alloy (ASTM F136). Components: NeoConvert (screw, coping, distal bar), Direct Screw to MUA, GM Attachment TiN 30°. Sterilization: Ethylene Oxide (SAL 1x10^-6) or moist heat (steam) for non-sterile components. Connectivity: None (mechanical prosthetic components). Manufacturing: CAD/CAM library-based milling for personalized copings.
Indications for Use
Indicated for patients requiring dental implants in the upper or lower jaw to support prosthetic devices (artificial teeth) to restore chewing function. Applicable for single-stage or two-stage procedures, including temporary or screw-retained multi-unit restorations, and immediate loading when primary stability and occlusal loading conditions are met.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
MRI Compatibility for Existing Neodent Implant System (K182620)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 4, 2024
JJGC Indústria e Comércia de Materiais Dentários S.A. % Jennifer Jackson Sr. Director, Regulatory Affairs and Quality Straumann USA. LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K242686
Trade/Device Name: Neodent Implant System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: September 6, 2024 Received: September 6, 2024
Dear Jennifer Jackson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known) K242686
Device Name Neodent Implant System
### Indications for Use (Describe)
NeoConvert Solution:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The NeoConvert Solution may be used with single-stage or two-stage procedures, for temporary multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
#### Direct Screw to MUA:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The Direct Screw to MUA may be used with single-stage or two-stage procedures, for screw-retained multi-unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. All digitally designed personalized copings to be used with the Direct to Multi Unit Abutment Screw are intended to be sent to Straumann for manufacture at a validated milling center.
#### GM Attachment TiN 30°:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The GM Attachment TiN 30° may be used with single-stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
### Neodent Implant System
510(k) Summary
# 510(k) Summary
### Submitter's Contact Information
| Submitter: | Straumann USA, LLC<br>60 Minuteman Road<br>Andover, MA 01810, USA<br>Registration No.: 1222315 Owner/Operator No.: 9005052 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | On the behalf of: |
| | JJGC Indústria e Comércio de Materiais Dentários S.A (dba Neodent)<br>Av. Juscelino Kubitschek de Oliveira, 3291<br>Curitiba, Paraná, Brazil 81270-200<br>Registration No.: 3008261720<br>Owner/Operator No.: 10031702 |
| Contact Person: | Jennifer M. Jackson, MS, RAC<br>Sr. Director, Regulatory Affairs and Quality<br>Phone Number: +1 978 747-2509<br>Fax Number: +1 978 747-0023<br>Email: jennifer.jackson@straumann.com |
| Prepared By & | Leticia Milani |
| Alternate Contact: | Regulatory Affairs Analyst<br>JJGC Indústria e Comércio de Materiais Dentários SA |
| Date Prepared: | December 2, 2024 |
| Name of the Device | |
| Trade Names: | Neodent Implant System |
| Common Name: | Endosseous dental implant abutment |
| Classification Name: | Endosseous dental implant abutment |
| Regulation Number: | 21 CFR 872.3630 |
| Device Classification: | II |
| Product Code(s): | NHA |
| Classification Panel: | Dental |
{6}------------------------------------------------
### Neodent Implant System
510(k) Summary
#### Predicate Device(s)
#### NeoConvert Solution:
Primary Predicate:
K163194 Neodent Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A)
Reference Devices:
| K101945 | Neodent Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A. |
|---------|-------------------------------------------------------------------------------|
|---------|-------------------------------------------------------------------------------|
K193234 Nuvo IF Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A)
#### Direct Screw to MUA:
Primary Predicate:
K163194 Neodent Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A)
#### Reference Devices:
| K162890 | Straumann SC Variobase Abutments (Straumann USA, LLC) |
|---------|-----------------------------------------------------------------------------------------|
| K180536 | Neodent Implant System - GM Line (JJGC Indústria e Comércio de Materiais Dentários S.A) |
| K192229 | GM Titanium Base for Bridge (JJGC Indústria e Comércio de Materiais Dentários S.A) |
| K202452 | NobelProcera Zirconia Implant Bridge (Nobel Biocare AB) |
| K203309 | NUVO CF Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A) |
| K203542 | Mini Abutment 60°(JJGC Indústria e Comércio de Materiais Dentários S.A) |
| K220251 | GM Narrow Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A) |
| K223638 | Helix Short Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A) |
#### GM Attachment TiN 30°:
Primary Predicate:
- K163194 Neodent Implant System (JJGC Indústria e Comércio de Materiais Dentários S.A)
#### Reference Devices:
Neodent Implant System – GM Line (JJGC Indústria e Comércio de Materiais Dentários S.A) K173902
- K192229 GM Titanium Base for Bridge (JJGC Indústria e Comércio de Materiais Dentários S.A)
#### MRI:
K182620 MRI Compatibility for Existing Neodent Implant System
{7}------------------------------------------------
### Neodent Implant System
510(k) Summary
### Device Description
This premarket notification includes the following additions to the Neodent Implant System:
- . NeoConvert Solution
- . Direct Screw to MUA
- GM Attachment TiN 30° .
The Neodent Multi-Unit Abutments proposed on this submission are similar to devices already cleared in previous submissions of Neodent Implant System, according to predicate devices described above. This submission intends to expand the portfolio of Neodent Abutments with these new solutions for multi-unit restorations, to provide more treatment options to customers.
The NeoConvert Solution is a technique comprised by some abutments (not subject of this submission) used for converting removable temporary prosthesis into a screw-retained fixed temporary prosthesis.
The Direct Screw to MUA is a prosthetic component designed to fix planned and digitally designed multi-unit milled restorations directly onto mini or micro abutments, eliminating the need for using the passive seating cylinder to restore chewing function.
The GM Attachment TiN 30° is a prosthetic component used to stabilize implant-mucosa-supported removable full prosthesis (retained on the implant and supported on the mucosa), on implants installed in the maxilla or mandible.
Although all these three new solutions are compatible with the same implants and components of Neodent Implant System and used for multiple unit rehabilitation, they are not used in conjunction with each other.
#### Intended Use
NeoConvert Solution:
Instruments and components used to perform the procedure of converting removable prostheses into temporary fixed prostheses on GM line implants installed in the maxilla and/or mandible.
#### Direct Screw to MUA:
The Direct Screw to MUA is suitable for attaching the multiple restoration milled from a CAD/CAM library to the prosthetic abutment (mini or micro abutments).
{8}------------------------------------------------
### Neodent Implant System
510(k) Summary
### GM Attachment TiN 30°:
This product is used to stabilize implant-mucosa supported removable full prostheses (retained on the implant and supported on the mucosa), on implants installed in the maxilla or mandibula. It can be used in immediate or conventional rehabilitation procedures.
### Indications for Use
### NeoConvert Solution:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The NeoConvert Solution may be used with single-stage or two-stage procedures, for temporary multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
### Direct Screw to MUA:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The Direct Screw to MUA may be used with single-stage or two-stage procedures, for screw-retained multi-unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. All digitally designed personalized copings to be used with the Direct to Multi Unit Abutment Screw are intended to be sent to Straumann for manufacture at a validated milling center.
### GM Attachment TiN 30°:
The Neodent Implant System is intended to be surqically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. The GM Attachment TiN 30° may be used with single-stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
{9}------------------------------------------------
## Neodent Implant System
510(k) Summary
### Technological Characteristics
The technological characteristics of the subject devices are compared to the primary predicate and reference devices in the following tables:
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K242686<br>Neodent Implant System | K163194<br>Neodent Implant System | K101945<br>Neodent Implant System | K193234<br>Nuvo IF Implant System |
| Indications<br>for Use | The Neodent Implant System is intended<br>to be surgically placed in the bone of the<br>upper or lower jaw to provide support for<br>prosthetic devices, such as artificial teeth,<br>to restore chewing function.<br>The<br>NeoConvert Solution may be used with<br>single-stage or two-stage procedures, for<br>temporary multiple unit restorations, and<br>may be loaded immediately when good<br>primary stability is achieved and with<br>appropriate occlusal loading. | The Neodent Implant System is intended to<br>be surgically placed in the bone of the upper<br>or lower jaw to provide support for prosthetic<br>devices, such as artificial teeth, to restore<br>chewing function. It may be used with single-<br>stage or two-stage procedures, for single or<br>multiple unit restorations, and may be loaded<br>immediately when good primary stability is<br>achieved and with appropriate occlusal<br>loading. | The Neodent Implant System is intended to be<br>surgically placed in the bone of the upper or<br>lower jaw to provide support for prosthetic<br>devices, such as artificial teeth, to restore<br>chewing function. It may be used with single-<br>stage or two-stage procedures, for single or<br>multiple unit restorations, and may be loaded<br>immediately when good primary stability is<br>achieved and with appropriate occlusal loading.<br>Multiple tooth applications may be rigidly<br>splinted. | The Implant System is intended to be<br>surgically placed in the maxilla or mandible to<br>provide support for prosthetic devices such<br>as artificial teeth in order to restore chewing<br>function. It may be used with single-stage or<br>two-stage procedures, for single or multiple<br>unit restorations, and may be loaded<br>immediately when good primary stability is<br>achieved and with appropriate occlusal<br>loading. Multiple tooth applications may be<br>rigidly splinted. |
| Material | Titanium alloy, according to ASTM F136. | Titanium alloy, according to ASTM F136. | Titanium alloy, according to ASTM F136. | Titanium alloy, according to ASTM F136. |
| Overall<br>dimensions | NeoConvert Screw:<br>● Laying diameter: 1.4 mm<br>● Height: 3.7 mm<br><br>NeoConvert Coping:<br>● Laying diameter: 4.8 mm<br>● Height: 5.0 and 6.0 mm<br><br>NeoConvert Distal Bar:<br>● Laying diameter: 4.8 mm<br>● Height: 5 mm | Mini Conical Abutment Coping Screw:<br>● Laying diameter: 1.4 mm<br>● Height: 3.7 mm | Mini Conical Abutment Coping:<br>● Laying diameter: 4.8 mm<br>● Height: 12 mm | Nuvo Distal Bar:<br>● Laying diameter: 5.2 mm<br>● Height: 10 mm |
| Commercially<br>Supplied | NeoConvert Coping and Screw:<br>One cavity blister with five units.<br><br>NeoConvert Distal Bar:<br>Two cavity blister with one unit and one<br>screw. | Mini Conical Abutment Coping Screw:<br>One cavity blister with one unit. | Mini Conical Abutment Coping:<br>Two cavity blister with one unit and one screw. | Nuvo Distal Bar:<br>Two cavities blister with one unit and one<br>screw. |
| Single use | Yes | Yes | Yes | Yes |
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | |
| K Number | K242686<br>Neodent Implant System | K163194<br>Neodent Implant System | K101945<br>Neodent Implant System<br>K193234<br>Nuvo IF Implant System | |
| Sterilization<br>Method | Provided sterile via Ethylene Oxide. The<br>coping and screw, if opened and not<br>used, must be terminally sterilized by<br>user via moist heat with cycle parameters<br>validated to a SAL of 1x10-6. | Provided sterile via Ethylene Oxide. | Provided sterile via Ethylene Oxide. Terminally<br>sterilized by user via moist heat with cycle<br>parameters validated to a SAL of 1x10-6.<br>Provided sterile via Ethylene Oxide. | |
#### Table 1 – Table of Substantial Equivalence NeoConvert
{10}------------------------------------------------
# Neodent Implant System
510(k) Summary
{11}------------------------------------------------
# Neodent Implant System
### 510(k) Summary
### Table 2 – Table of Substantial Equivalence Direct Screw to MUA (Zygomatic Implants)
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | | | | | | | | | | | | | |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|
| K Number | K242686<br>Neodent Implant System | K163194<br>Neodent Implant System | K202452<br>NobelProcera Zirconia Implant Bridge | K203542<br>Mini Abutment 60° | | | | | | | | | | | | |
| Indications<br>for Use | The Neodent Implant System is intended to<br>be surgically placed in the bone of the<br>upper or lower jaw to provide support for<br>prosthetic devices, such as artificial teeth,<br>to restore chewing function. The Direct<br>Screw to MUA may be used with single-<br>stage or two-stage procedures, for screw-<br>retained multi-unit restorations, and may be<br>loaded immediately when good primary<br>stability is achieved and with appropriate<br>occlusal loading. All digitally designed<br>personalized copings to be used with the<br>Direct to Multi Unit Abutment Screw are<br>intended to be sent to Straumann for<br>manufacture at a validated milling center. | The Neodent Implant System is intended to<br>be surgically placed in the bone of the upper<br>or lower jaw to provide support for prosthetic<br>devices, such as artificial teeth, to restore<br>chewing function. It may be used with single-<br>stage or two-stage procedures, for single or<br>multiple unit restorations, and may be loaded<br>immediately when good primary stability is<br>achieved and with appropriate occlusal<br>loading. | The NobelProcera Zirconia Implant Bridge<br>are indicated for use as a bridge anatomically<br>shaped and/or framework in the treatment of<br>partially or totally edentulous jaws for the<br>purpose of restoring chewing function. | The Mini Conical Abutments are indicated for<br>use with Zygomatic Implants, in case of<br>severe jaw resorption, in order to restore<br>patient aesthetics and chewing function. It<br>may be used with single-stage or two-stage<br>procedures, for multiple unit restorations, and<br>may be loaded immediately when good<br>primary stability is achieved and with<br>appropriate occlusal loading. | | | | | | | | | | | | |
| Material | Titanium alloy, according to ASTM F136. | Titanium alloy, according to ASTM F136. | Titanium alloy, according to ASTM F136. | Titanium alloy, according to ASTM F136. | | | | | | | | | | | | |
| Overall<br>dimensions | Diameter of the screw head: 2.2 mm<br>Height of screw head: 1.7 mm<br>Total height: 4.1 mm | Diameter of the screw head: 1.9 mm<br>Diameter of the external thread: 1.4 mm<br>Height of screw head: 1.7 mm<br>Total height: 3.7 mm | Diameter of the screw head: 2.0 mm<br>Total height: 4.0 mm | | | | | | | | | | | | | |
| Commercially<br>Supplied | One cavity blister with five units. | One cavity blister with one units. | | | | | | | | | | | | | | |
| Single Use | Yes | Yes | Yes | Yes | | | | | | | | | | | | |
| Sterilization<br>Method | Provided non-sterile. Must be sterilized<br>before installation via moist heat (steam),<br>using either gravity or vacuum, with an<br>exposure time of 3 minutes at 132°C<br>(270°F). | | Provided non-sterile. Must be sterilized<br>before installation via moist heat (steam),<br>using either gravity or vacuum, at 132°C<br>(270°F) or 134°C (273°F). | | | | | | | | | | | | | |
| Top Half<br>Material | | | | Material Minimum thickness N!ce® HT 0.4 mm N!ce® LT 0.4 mm N!ce® XT 0.4 mm Ticon® 0.4 mm Coron® 0.4 mm | | | | | | | | | | | | |
{12}------------------------------------------------
# Neodent Implant System
### 510(k) Summary
### Table 3 – Table of Substantial Equivalence Direct Screw to MUA (GM Implants)
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE PREDICATE DEVICE | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K242686<br>Neodent Implant System | K163194<br>Neodent Implant System | K192229<br>GM Titanium Base for Bridge<br>K203309<br>NUVO CF Implant System | |
| Indications<br>for Use | The Neodent Implant System is intended to<br>be surgically placed in the bone of the upper<br>or lower jaw to provide support for prosthetic<br>devices, such as artificial teeth, to restore<br>chewing function. The Direct Screw to MUA<br>may be used with single-stage or two-stage<br>procedures, for screw-retained multi-unit<br>restorations, and may be loaded immediately<br>when good primary stability is achieved and<br>with appropriate occlusal loading. All digitally<br>designed personalized copings to be used<br>with the Direct to Multi Unit Abutment Screw<br>are intended to be sent to Straumann for<br>manufacture at a validated milling center. | The Neodent Implant System is intended to<br>be surgically placed in the bone of the<br>upper or lower jaw to provide support for<br>prosthetic devices, such as artificial teeth,<br>to restore chewing function. It may be used<br>with single-stage or two-stage procedures,<br>for single or multiple unit restorations, and<br>may be loaded immediately when good<br>primary stability is achieved and with<br>appropriate occlusal loading. | Titanium Base Abutment is a titanium base<br>placed onto Neodent dental implants to provide<br>support for customized prosthetic restorations.<br>It is used with a coping and crown, or crown<br>alone, and is indicated for cement-retained<br>single or multi-unit restorations or screw-<br>retained single restorations. All digitally<br>designed copings and/or crowns to be used<br>with the Neodent Titanium Base Abutment<br>System are intended to be sent to Straumann<br>for manufacture at a validated milling center.<br>The GM Titanium Base for Bridge is indicated<br>for cement or screw-retained multi-unit<br>restorations. | The CF Titanium Base for Bridge is a<br>titanium abutment placed onto dental<br>implants to provide support for customized<br>prosthetic restorations (multiple structures).<br>It is indicated for multi-unit restorations,<br>cement-retained or screw-retained in<br>aesthetic areas on implants installed in the<br>maxilla or mandible. |
| Material | Titanium alloy, according to ASTM F136. | Titanium alloy, according to ASTM F136. | Titanium alloy, according to ASTM F136. | Titanium alloy, according to ASTM F136. |
| Overall<br>dimensions | Diameter of the screw head: 2.2 mm<br>Height of screw head: 1.7 mm<br>Total height: 4.1 mm | Diameter of the screw head: 1.9 mm<br>Diameter of the external thread: 1.4 mm<br>Height of screw head: 1.7 mm<br>Total height: 3.7 mm | Diameter: 3.5mm<br>Cementable area height 4.0mm<br>Gingival Height: 0.8 mm | Diameter: 3.5 mm<br>Cementable area height: 4.5 mm<br>Gingival height: 0.5 mm |
| Commercially<br>Supplied | One cavity blister with five units. | One cavity blister with one units. | Two cavity blister with one unit and one screw. | Two cavity blister with one unit and one<br>screw. |
| Single Use | Yes | Yes | Yes | Yes |
| Sterilization<br>Method | Provided non-sterile. Must be sterilized<br>before installation via moist heat (steam),<br>using either gravity or vacuum, with an<br>exposure time of 3 minutes at 132°C (270°F). | | Provided sterile by Ethylene Oxide. Must be<br>sterilized before installation via moist heat<br>(steam), using either gravity or vacuum, at<br>132°C (270°F). | Provided sterile by Ethylene Oxide. Must be<br>sterilized before installation via moist heat<br>(steam), using either gravity or vacuum, at<br>132°C (270°F). |
| Top Half<br>Material | Material | Minimum thickness | Material…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.