BLT 02.9mm SC, SLA or SLActive, RXD, Loxim, SC Closure Cap and Healing Abutments, SC Temporary Abutments, SC Variobase Abutments, SC CARES Abutment
K162890 · Straumann USA, LLC · DZE · Mar 8, 2017 · Dental
Device Facts
Record ID
K162890
Device Name
BLT 02.9mm SC, SLA or SLActive, RXD, Loxim, SC Closure Cap and Healing Abutments, SC Temporary Abutments, SC Variobase Abutments, SC CARES Abutment
Applicant
Straumann USA, LLC
Product Code
DZE · Dental
Decision Date
Mar 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Straumann dental Implants are intended for oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of patients with missing teeth. Closure Caps and Healing Abutments are intended for use with the Straumann® Bone Level Tapered Implants Ø2.9 mm to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. Straumann® SC Temporary Abutments are intended to be placed into Straumann dental implants to provide support for temporary crowns and bridges. Straumann® SC Variobase® Abutments are intended to be placed into Straumann® Bone Level Tapered Implants Ø2.9 mm to provide support for individual crowns or bridges. Straumann® SC CARES® abutments are intended for use as an aid in prosthetic rehabilitation.
Device Story
System comprises Ø2.9 mm apically tapered endosseous dental implants and associated prosthetic components (closure caps, healing/temporary/Variobase/CARES abutments). Implants placed in jaw by dental professionals; provide foundation for prosthetic restorations. Variobase and CARES abutments utilize CAD/CAM workflow: clinician scans intraoral situation; software designs abutment/restoration; data sent to validated milling center for fabrication. Components facilitate soft tissue healing, temporary stabilization, or final crown/bridge support. Benefits include restoration of chewing function and aesthetics in narrow interdental spaces. System used in dental clinics; output (restoration) affects clinical decision-making by enabling functional tooth replacement.
Clinical Evidence
Bench testing only. No clinical data provided. Performance supported by dynamic fatigue, static strength, and insertion torque testing per FDA guidance for root-form endosseous dental implants. Biocompatibility evaluated per ISO 10993-1. Software verification and validation performed for CAD/CAM design software.
Technological Characteristics
Implants: Titanium-13 Zirconium alloy (Roxolid) with SLA/SLActive surface. Abutments: Titanium Grade 4 or Ti-6Al-7Nb. Connection: Small CrossFit (SC). Dimensions: Ø2.9 mm diameter; 10-14 mm lengths. CAD/CAM workflow for Variobase/CARES abutments. Software: CARES Visual CAD module. Sterilization: Validated reprocessing methods.
Indications for Use
Indicated for oral endosteal implantation in upper/lower jaw for functional/aesthetic rehabilitation of missing teeth. Suitable for immediate/early implantation post-extraction/loss. Supports single or multiple-tooth applications with good primary stability and appropriate occlusal loading. Specifically indicated for reconstruction of missing lower incisors and upper lateral incisors. Components (Closure Caps, Healing/Temporary/Variobase/CARES Abutments) used for soft tissue management, temporary support, or final prosthetic restoration support. Temporary abutments limited to 180 days use.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, composed of three overlapping silhouettes. The silhouettes are arranged to suggest a sense of movement or progress. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2017
Straumann USA, LLC % Jennifer Jackson Director, Regulatory Affairs & Quality 60 Minuteman Road Andover, Massachusetts 01810
Re: K162890
Trade/Device Name: Straumann Ø2.9 mm Bone Level Tapered Implants, SC Closure Cap. SC Healing Abutments, SC Temporary Abutments, SC Variobase Abutments. SC CARES Abutments Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: February 16, 2017 Received: February 17, 2017
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Lori A. Wiggins -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known)
K162890
#### Device Name
Straumann® Ø2.9 mm Bone Level Tapered Implants, SC Closure Cap, SC Healing Abutments, SC Temporary Abutments, SC Variobase Abutments, SC CARES Abutments
#### Indications for Use (Describe)
Straumann® Bone Level Tapered Implants Ø2.9 mm are indicated for oral endosteal implantation in the upper and lower jaw and for the functional and aesthetic oral rehabilitation of patients with missing teeth. Straumann® Bone Level Tapered Implants 02,9 mm can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations are connected to the implants through the corresponding components (abutments).
The Straumann® Bone Level Tapered Implants 02.9 mm are indicated for reconstruction of missing incisors in the lower jaw and lateral incisors in the upper jaw.
Straumann® Closure Caps and Healing Abutments are indicated to be placed in the dental implant after placement in the patient's jaw to protect the inner configuration of the implant and stabilize the soft tissue during the healing process. Closure Caps and Healing Abutments should be used only with the corresponding implant comection.
Straumann® SC Temporary Abutments are indicated for use as an aid in prosthetic rehabilitations. Temporary components can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase. Strauman® SC Temporary Abutments have a maximum duration of usage of 180 days.
Straumann® SC Variobase® abutments are indicated for use as an aid in prosthetic rehabilitations. The prosthetic restoration can be cemented on the Straumann® SC Variobase® prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and form the healing phase. Final abutments and restorations may be placed into occlusion when is fully osseointegrated. All digitally designed copings and/or crowns for use with the Straumann® Variobase® Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.
Straumann® SC CARES® abutments are indicated for single-tooth replacements and multiple tooth restorations. The prosthetic restoration can be cemented or directly veneered/screw-retained.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="font-size:20px;"> </span>Prescription Use (Part 21 CFR 801 Subpart D)</div> |
| <div style="display:flex; align-items:center;"><span style="font-size:20px;"> </span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Straumann® Ø2.9 mm Bone Level Tapered Implants
510(k) Summary
#### 510(k) Summary 5
#### Submitter's Contact Information 5.1
Straumann USA, LLC (on behalf of Institut Straumann AG)
60 Minuteman Road
Andover, MA 01810
| Phone Number: | 1-978-747-2509 |
|---------------|----------------|
|---------------|----------------|
Fax Number: 1-978-747-0023
Contact Person: Jennifer M. Jackson, MS
Date of Submission: March 6, 2017
#### 5.2 Name of the Device
| Trade Names: | Straumann® Ø2.9 mm Bone Level Tapered Implants | |
|----------------------|------------------------------------------------|--|
| | SC Closure Cap | |
| | SC Healing Abutments | |
| | SC Temporary Abutments | |
| | SC Variobase® Abutments | |
| | SC CARES® Abutments | |
| Common Name: | Endosseous Dental Implant | |
| | Endosseous Dental Implant Abutment | |
| Classification Name: | Endosseous Dental Implant | |
| Regulation Number: | §872.3640 | |
| Classification: | Class II | |
| Product Codes: | DZE, NHA | |
### 5.3 Predicate Device(s)
Primary Predicate:
K140878 – Straumann Bone Level Tapered Implant
{5}------------------------------------------------
## Straumann® Ø2.9 mm Bone Level Tapered Implants
510(k) Summary
Reference Devices:
- K120414 OsseoSpeed™ Plus
- K130808 Straumann Healing Abutments, Healing Caps, and Closure Screws
- K092814 Straumann NC Temporary Abutments
- K120822 Straumann CARES Variobase Abutments
- K142890 Straumann Variobase Abutments
- K150203 Medentika CAD/CAM Abutments
- K150899 Straumann CARES TAN Abutments
- K062129 P.004 Implants
#### 5.4 Device Description
The Straumann Ø2.9mm Bone Level Tapered (BLT) Implants are apically tapered implants with an external diameter of Ø2.9 mm and lengths of 10 mm, 12 mm, and 14 mm. The implants are manufactured utilizing the Roxolid material and are finished with either the SLA® or SLActive® surface. The prosthetic platform is identified as SC (Small CrossFit®) which corresponds to a shoulder diameter of Ø2.9 mm.
The closure cap and healing abutments are manufactured from Titanium Grade 4 and are anodized blue for identification purposes. The closure cap is conical and has a height of 0.5 mm. The healing abutments are seated in the implant with a basal screw which is manufactured from TAN. The healing abutments are oval in shape and are available in four different heights ranging between 2.0 mm and 6.5 mm.
The temporary abutments are manufactured from TAN and are anodized blue for identification purposes. The temporary abutments are oval in shape in order to accommodate narrow interdental spaces and are available with three different gingival heights ranging between 1.0 mm and 3.0 mm. The temporary abutments are seated in the implant with a basal screw which is also manufactured from TAN.
There are three components to the Straumann® SC Variobase™ Abutments:
- Straumann® SC Variobase™ Abutments (Ti-base) ●
- Prosthetic restoration (coping and/or crown) .
- Basal Screw .
{6}------------------------------------------------
# Straumann® Ø2.9 mm Bone Level Tapered Implants
### 510(k) Summary
The Straumann® SC Variobase® Abutments are manufactured from TAN. The abutments are oval in shape to accommodate narrow interdental spaces and are available with three different gingival heights ranging between 1.0 mm and 3.0 mm. The abutments will be delivered with the corresponding basal screw.
The following is an overview of the possible prosthetic restoration (coping and/or crown) materials:
- Cast materials: .
- o Type 4 metals (ISO 22674)
- Base metal alloys (e.g., cobalt-chromium (CoCr)) o
- Noble metal alloys (e.g., gold alloy) o
- Press materials: ●
- o IPS e.max® Press Ceramic (K120053)
- . Digital materials:
- o coron®
- o zerion® LT
- o polycon® ae
All digitally designed copings and/or crowns for use with the Straumann® Variobase® Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.
The Straumann® SC CARES® Abutments are customized abutments manufactured from TAN. The abutments are designed by the customer by scanning the intraoral situation and designing of the shape with the CAD module of the integrated software. The design data is then sent to Straumann where the fabrication of the customized abutment is performed. The TAN alloy is capable of being directly veneered once the abutment is subject to a heat treatment step by the dental laboratory technician. A finished crown may also be cemented to the subject abutments rather than by direct veneer. The abutments will be delivered with the corresponding basal screw.
{7}------------------------------------------------
## Straumann® Ø2.9 mm Bone Level Tapered Implants
510(k) Summary
#### 5.5 Intended Use
The Straumann dental Implants are intended for oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of patients with missing teeth.
Closure Caps and Healing Abutments are intended for use with the Straumann® Bone Level Tapered Implants Ø2.9 mm to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process.
Straumann® SC Temporary Abutments are intended to be placed into Straumann dental implants to provide support for temporary crowns and bridges.
Straumann® SC Variobase® Abutments are intended to be placed into Straumann® Bone Level Tapered Implants Ø2.9 mm to provide support for individual crowns or bridges. Straumann® SC CARES® abutments are intended for use as an aid in prosthetic rehabilitation.
### 5.6 Indications for Use
Straumann® Bone Level Tapered Implants ø2.9 mm are indicated for oral endosteal implantation in the upper and lower jaw and for the functional and aesthetic oral rehabilitation of patients with missing teeth. Straumann® Bone Level Tapered Implants ø2.9 mm can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth or multiple-tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations are connected to the implants through the corresponding components (abutments).
The Straumann® Bone Level Tapered Implants ø2.9 mm are indicated for reconstruction of missing incisors in the lower jaw and lateral incisors in the upper jaw.
Straumann® Closure Caps and Healing Abutments are indicated to be placed in the dental implant after placement in the patient's jaw to protect the inner configuration of the implant and to form, maintain and stabilize the soft tissue during the healing process. Closure Caps and Healing Abutments should be used only with the corresponding implant connection.
{8}------------------------------------------------
### Straumann® Ø2.9 mm Bone Level Tapered Implants
### 510(k) Summary
Straumann® SC Temporary Abutments are indicated for use as an aid in prosthetic rehabilitations. Temporary components can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase. Straumann® SC Temporary Abutments have a maximum duration of usage of 180 days. Straumann® SC Variobase® abutments are indicated for use as an aid in prosthetic rehabilitations. The prosthetic restoration can be cemented on the Straumann® SC Variobase® prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and form the soft tissue during the healing phase. Final abutments and restorations may be placed into occlusion when the implant is fully osseointegrated. All digitally designed copings and/or crowns for use with the Straumann® Variobase® Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.
Straumann® SC CARES® abutments are indicated for single-tooth replacements and multiple tooth restorations. The prosthetic restoration can be cemented or directly veneered/screw-retained.
#### 5.7 Technological Characteristics
#### 5.7.1 Straumann® Ø2.9 mm Bone Level Tapered Implants
The technological principles are the same for the subject and primary predicate devices. Both the subject device and primary predicate device are indicated for oral endosteal implantation in the upper and lower jaw and for the functional and aesthetic oral rehabilitation of patients with missing teeth. The subject devices are specifically indicated for the lateral incisor in the upper jaw and the lower incisors. The indications for use for the subject devices are within the scope of the indications for use for the primary predicate device. The narrower indications do not change the intended use. The main differences in the technological characteristics between the subject device and the primary predicate device are the implant diameter (Ø2.9 mm vs. Ø3.3, Ø 4.1 and Ø 4.8 mm) and the implant to abutment connection (Small CrossFit® (SC) versus Narrow CrossFit® (NC) and Regular CrossFit® (RC)). The technological characteristics of the subject devices are compared to the primary and reference predicate in Table 1.
{9}------------------------------------------------
# Straumann® Ø2.9 mm Bone Level Tapered Implants
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE<br>DEVICE | REFERENCE<br>PREDICATE DEVICE |
|--------------------------------------|-------------------------------------------|-------------------------------------------------|----------------------------------------|
| K Number | Subject Device | K140878 | K120414 |
| Material | Titanium-13 Zirconium<br>alloy (Roxolid®) | Titanium-13 Zirconium<br>alloy (Roxolid®) | Commercially pure<br>titanium, grade 4 |
| Surface<br>Treatment | SLA® and SLActive® | SLA® and SLActive® | OsseoSpeed™ |
| Implant to<br>Abutment<br>Connection | Small CrossFit® (SC) | Narrow CrossFit® (NC)<br>Regular CrossFit® (RC) | Six position connection |
| Implant<br>Diameter | Ø2.9 mm | Ø3.3, Ø4.1, and Ø4.8 mm | Ø3.0 mm |
| Implant Length | 10, 12, and 14 mm | 8, 10, 12, and 14 mm | 8, 9, 11, 13, and 15 mm |
| Implant Design | Straight implant with apical<br>taper | Straight implant with apical<br>taper | Straight implant with apical<br>taper |
#### 510(k) Summary
Table 1 – Comparison of subject device versus primary and reference predicate devices (Dental Implants)
#### 5.7.2 Straumann® Closure Cap and Healing Abutments
The technological principles and intended use are identical for the subject and reference predicate devices. The intended use is to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. The main differences in the technological characteristics between the subject and reference predicate devices are the shape and diameter of the devices. The subject devices are dimensionally smaller to accommodate the smaller diameter implant. The technological characteristics of the subject devices are compared to the reference predicate in Table 2.
{10}------------------------------------------------
# Straumann® Ø2.9 mm Bone Level Tapered Implants
| FEATURE | PROPOSED DEVICE | REFERENCE PREDICATE<br>DEVICE |
|--------------------------------------------------------------------|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| K Number | Subject Device | K130808 |
| Material | Titanium Grade 4<br>Ti-6Al-7Nb (Basal Screws) | Titanium Grade 4 |
| Implant to<br>Abutment<br>Connection | Small CrossFit® (SC) | Narrow CrossFit® (NC) |
| Diameter or<br>Minor Oval<br>Dimension/<br>Major Oval<br>Dimension | Closure Cap:<br>Ø2.4 mm<br>Healing Abutments:<br>3.3/4.3 mm<br>3.55/4.86 mm<br>3.6/5.0 mm | Closure Cap:<br>Ø2.77 and Ø3.05 mm<br>Healing Abutment:<br>Ø3.6 |
| Overall Length | Closure Cap:<br>5.8 mm<br>Healing Abutment (including basal screw):<br>7.0, 8.5, 10.0, 11.5 mm | Closure Cap:<br>6.25 and 6.8 mm<br>Healing Abutment:<br>8.1, 9.6, 11.1 mm |
| Gingival Heights | Closure Cap:<br>0.5 mm<br>Healing Abutments:<br>2.0, 3.5, 5.0 and 6.5 mm | Closure Cap:<br>0 mm and 0.5 mm<br>Healing Abutments:<br>2.0, 3.5, and 5.0 mm |
510(k) Summary
### Table 2 – Comparison of subject device versus reference device (SC Closure Cap and SC Healing Abutments)
#### Temporary Abutments 5.7.3
The technological principles and intended use are identical for the subject and reference predicate devices. The devices are intended to be placed into Straumann dental implants to provide support for temporary crowns and bridges. The main differences in the technological characteristics between the subject and reference predicate devices are the shape and diameter of the devices. The subject devices are dimensionally smaller to accommodate the smaller diameter implant. The technological characteristics of the subject devices are compared to the reference predicate in Table 3.
{11}------------------------------------------------
# Straumann® Ø2.9 mm Bone Level Tapered Implants
| FEATURE | PROPOSED DEVICE | REFERENCE PREDICATE DEVICE |
|--------------------------------------------------------------------|-------------------------|----------------------------|
| K Number | Subject Device | K092814 |
| Material | Ti-6Al-7Nb | Ti-6Al-7Nb |
| Implant to<br>Abutment<br>Connection | Small CrossFit® (SC) | Narrow CrossFit® (NC) |
| Diameter or<br>Minor Oval<br>Dimension/<br>Major Oval<br>Dimension | 3.3/4.3 mm | Ø3.5 mm |
| Overall Length | 13.2, 14.2, and 15.2 mm | 13.75 mm |
| Gingival Heights | 1.0, 2.0, and 3.0 mm | 1.0 mm |
510(k) Summary
Table 3 - Comparison of subject device versus reference predicate device (SC Temporary Abutments)
#### 5.7.4 SC Variobase® Abutments
The technological principles and intended use are identical for the subject and reference predicate devices. The devices are intended to be placed into dental implants to provide support for individual crowns or bridges. The subject devices are specifically indicated for the lateral incisor in the upper jaw and the lower incisors. The indications for use for the subject devices are within the scope of the indications for use for the reference predicate devices. The narrower indications do not change the intended use.
The main differences in the technological characteristics between the subject and reference predicate devices are the shape and diameter of the devices. The subject devices are dimensionally smaller to accommodate the smaller diameter of the implant. The subject devices also have only two retention features compared to the Straumann reference predicate device which has four retention features. The Medentika reference predicate has one retention feature. The technological characteristics of the subject devices are compared to the reference predicate in Table 4.
{12}------------------------------------------------
# Straumann® Ø2.9 mm Bone Level Tapered Implants
| FEATURE | PROPOSED DEVICE | REFERENCE<br>PREDICATE DEVICES | REFERENCE<br>PREDICATE DEVICES |
|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| K Number | Subject Device | K142890/K120822 | K150203<br>R Series |
| Material | Ti-6Al-7Nb | Ti-6Al-7Nb | Ti-6Al-4V |
| Implant to Abutment<br>Connection | Small CrossFit® (SC) | Narrow CrossFit® (NC)<br>Regular CrossFit® (RC)<br>Narrow Neck CrossFit®<br>(NNC)<br>Regular Neck (RN)<br>Wide Neck (WN) | 2.5mmD Internal<br>Hexagon |
| Diameter or Minor<br>Oval Dimension/<br>Major Oval<br>Dimension | 3.3/4.3 mm | Ø3.8 - 7.0 mm | Ø3.5 – 5.7 mm |
| Overall Abutment<br>Height | 6.7 – 8.7 mm | 5.9 – 8.9 mm | 6.15 – 7.54 mm |
| Coping/ Crown<br>Material | Traditional Workflow:<br>Type 4 metals (ISO<br>22674)<br>IPS e.max® Press<br>Ceramic<br>Digital Workflow:<br>coron® (permanent)<br>zerion® LT (permanent)<br>polycon® ae (temporary) | Traditional Workflow:<br>Type 4 metals (ISO<br>22674)<br>IPS e.max® Press<br>Ceramic<br>Digital Workflow:<br>IPS e.max® CAD<br>Ceramic (permanent)<br>coron® (permanent)<br>zerion® (permanent)<br>polycon® ae (temporary) | Digital Workflow:<br>zerion® (permanent) |
| Design Workflow | Wax-up or Open CAD | Wax-up or Open CAD<br>CARES® Visual | CARES® Visual |
| Manufacturing<br>Workflow | Traditional casting or<br>pressing<br>or<br>Straumann Milling | Traditional casting or<br>pressing<br>or<br>Straumann Milling | Straumann Milling |
| Mode of Action | Screw-retained or<br>cement retained | Screw-retained or<br>cement retained | Screw-retained or<br>cement retained |
#### 510(k) Summary
Table 4 - Comparison of subject device versus reference device (SC Variobase Abutments)
#### 5.7.5 SC CARES® Abutments
The technological principles and intended use are identical for the subject and reference predicate devices. The devices are intended for use as an aid in prosthetic rehabilitation. The main differences in the technological characteristics between the subject and reference predicate devices are the diameter and the dimension for the abutment to implant connection. The technological characteristics of the subject devices are compared to the reference predicate in Table 5.
{13}------------------------------------------------
# Straumann® Ø2.9 mm Bone Level Tapered Implants
510(k) Summary
| FEATURE | PROPOSED DEVICE | REFERENCE PREDICATE DEVICE |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | Subject Device | K150899 |
| Material | Ti-6Al-7Nb | Ti-6Al-7Nb |
| Implant to<br>Abutment<br>Connection | Small CrossFit® (SC) | Narrow CrossFit® (NC) |
| Abutment<br>coronal design | CADCAM design process.<br>Designs controlled by material-specific<br>design limits in the CARES Visual CAD<br>software, model verification performed by<br>the CAM software and milling blank<br>dimensions used by the Straumann<br>milling center. | CADCAM design process.<br>Designs controlled by material-specific<br>design limits in the CARES Visual CAD<br>software, model verification performed by<br>the CAM software and milling blank<br>dimensions used by the Straumann<br>milling center. |
| CAD Design<br>Limits | Maximum Angulation 30° | Maximum Angulation 30° |
| | Emergence Offset 0.1 mm | Emergence Offset 0.1 mm |
| | Emergence Angle 65° | Emergence Angle 65° |
| | Minimum Thickness 0.4 mm | Minimum Thickness 0.4 mm |
| | Smooth Distance 0.5 mm | Smooth Distance 0.5 mm |
| Coping/ Crown | Cemented | Cemented |
| | Directly veneerable | Directly veneerable |
Table 5 - Comparison of subject device versus reference device (SC CARES Abutments)
#### 5.8 Performance Testing
The following performance data were provided in support of the substantial equivalence determination.
#### Sterilization Validation 5.8.1
The sterilization process for the SC Temporary Abutments, SC Variobase Abutments, and SC CARES Abutments as recommended in the labeling was validated according to applicable recommendations in the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015".
#### 5.8.2 Biocompatibility Testing
Table 32ISO 10993-1:2009 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and the biocompatibility evaluation flow chart according to the FDA Guidance document "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process', Guidance for Industry and Food and Drug
{14}------------------------------------------------
# Straumann® Ø2.9 mm Bone Level Tapered Implants
510(k) Summary
Administration Staff, Document issued on: June 16, 2016" for each of the subject devices.
The subject devices have the identical nature of body contact, contact duration, material formulation, manufacturing processes, and sterilization methods compared to the primary and reference predicate devices. No new issues of biocompatibility are raised for the subject devices. Therefore, no additional biocompatibility testing was required.
### 5.8.3 Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided according to the FDA guidance documents "Class II Special Controls Guidance Document: Optical Impression Systems for Computer Assisted Design and Manufacturing (CAD/CAM) of Dental Restorations" and "General Principles of Software Validation; Final Guidance for Industry and FDA Staff". The software for this device was considered as a "moderate" level of concern as a failure in the Straumann CARES Visual software could result in a non-fitting or poor-fitting restoration or framework.
#### 5.8.4 Bench Testing
Dynamic fatigue, static strength, and insertion torque tests were conducted according to the FDA guidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" and demonstrated the Straumann® Ø2.9 mm Bone Level Tapered implants, SC Temporary Abutments, SC Variobase Abutments, and SC CARES abutments are equivalent to the predicate and reference devices.
### 5.9 Conclusion
The documentation submitted in this premarket notification demonstrates the Straumann® Ø2.9 mm Bone Level Tapered implants, SC Closure Caps and Healing Abutments, SC Temporary Abutments, SC Variobase Abutments, and SC CARES Abutments are substantially equivalent to the primary predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.