Neodent Implant System- GM Titanium Base for Bridge
K192229 · Jjgc Industria E Comercio DE Materiais Dentarios S.A. · NHA · Dec 19, 2019 · Dental
Device Facts
Record ID
K192229
Device Name
Neodent Implant System- GM Titanium Base for Bridge
Applicant
Jjgc Industria E Comercio DE Materiais Dentarios S.A.
Product Code
NHA · Dental
Decision Date
Dec 19, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Titanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations or screw-retained single restorations. All digitally designed copings and/or crowns to be used with the Neodent Titanium Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center. The GM Titanium Base for Bridge is indicated for cement or screw-retained multi-unit restorations.
Device Story
Two-piece abutment system for Neodent GM dental implants; supports CAD/CAM customized prosthetic restorations (copings/crowns). Input: digital design data processed via Dental Wings (DWOS) or 3Shape software; manufactured at validated milling center. Output: finished medical device after cementation of superstructure onto titanium base. Used in dental clinics by clinicians. Device provides structural support for prosthetic teeth; restores chewing function. Benefits: customized fit for multi-unit restorations.
Clinical Evidence
Bench testing only. Dynamic fatigue testing per ISO 14801 performed on worst-case implant constructs; results met acceptance criteria. Static torsion testing performed on screws; results met acceptance criteria. Sterilization validated to SAL 1x10^-6 per ISO 11135. Biocompatibility assessed per ISO 10993 series, including cytotoxicity (ISO 10993-5), chemical characterization (ISO 10993-18), and EtO residuals (ISO 10993-7).
Technological Characteristics
Material: Titanium alloy (Ti6Al4v-ELI) per ASTM F136. Design: Conical upper portion with helical micro-grooves in cementable area. Interface: GM implant-to-abutment. Dimensions: Diameters 3.5, 4.5, 5.5 mm; gingival heights 0.8-4.5 mm. Sterilization: Ethylene oxide gas. Single-use. CAD/CAM compatible.
Indications for Use
Indicated for patients requiring dental implant-supported prosthetic restorations. Used for cement-retained single or multi-unit restorations or screw-retained single or multi-unit restorations. Contraindications not specified.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
NC Straumann Variobase Abutment For Bars/Bridges (K151157)
RC Straumann Variobase Abutment For Bars/Bridges (K151157)
NNC Straumann Variobase Abutment For Bars/Bridges (K151157)
RN Straumann Variobase Abutment For Bars/Bridges (K151157)
WN Straumann Variobase Abutment For Bars/Bridges (K151157)
MRI Compatibility For Existing Neodent Implant System (K182620)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration" in a larger font and the word "Administration" in a smaller font.
December 19, 2019
JJGC Industria e Comercio de Materiais Dentarios S.A. % Jennifer Jackson Director of Regulatory Affairs Straumann USA, INC. 60 Minuteman Road Andover, Massachusetts 01810
Re: K192229
Trade/Device Name: Neodent Implant System - GM Titanium Base for Bridge Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 22, 2019 Received: November 25, 2019
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Srinivas Nandkumar, Ph.D. for Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192229
Device Name
Neodent Implant System - GM Titanium Base for Bridge
Indications for Use (Describe)
Titanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations or screw-retained single restorations. All digitally designed copings and/or crowns to be used with the Neodent Titanium Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
The GM Titanium Base for Bridge is indicated for cement or screw-retained multi-unit restorations.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
### ADMINISTRATIVE INFORMATION
| Sponsor | JJGC Indústria e Comércio de Materiais Dentários SA<br>(dba Neodent)<br>Av. Juscelino Kubitschek de Oliveira, 3291<br>Curitiba, Parana, Brazil 81270-200<br>Registration No.: 3008261720<br>Owner/Operator No.: 10031702 |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs,<br>Straumann USA<br>E-mail: jennifer.jackson@straumann.com<br>Telephone (978) 747-2509 |
| Date Prepared | 09/Dec/2019 |
| Preparer / Alternate Contact | Mariana Soares Hartmann<br>Regulatory Affairs Analyst<br>E-mail: mariana.hartmann@neodent.com |
| DEVICE NAME AND CLASSIFICATION | |
| Trade/ Proprietary Name<br>Common Name | Neodent Implant System – GM Titanium Base for Bridge<br>Endosseous dental implant abutment |
| Classification Name | Endosseous dental implant abutment |
| Classification Regulations<br>Product Code | 21 CFR 872.3630, Class II<br>NHA |
| Classification Panel<br>Reviewing Branch | Dental Products Panel<br>Dental Devices Branch |
| PREDICATE DEVICE INFORMATION | |
| Primary Predicate Device | K180536, Neodent Implant System - GM Line, JJGC Indústria e<br>Comércio de Materiais Dentários S.A |
| Reference Predicate Devices | K151157 - NC Straumann Variobase Abutment For Bars/Bridges,<br>RC Straumann Variobase Abutment For Bars/Bridges, NNC |
{5}------------------------------------------------
Straumann Variobase Abutment For Bars/Bridges, RN Straumann Variobase Abutment For Bars/Bridges, WN Straumann Variobase Abutment For Bars/Bridges, STRAUMANN USA, LLC. K182620 - MRI Compatibility For Existing Neodent Implant System, JJGC Indústria e Comércio de Materiais Dentários S.A
#### INDICATIONS FOR USE
Titanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown alone, and is indicated for cement-retained single or multi-unit restorations or screw-retained single restorations. All digitally designed copings and/or crowns to be used with the Neodent Titanium Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
The GM Titanium Base for Bridge is indicated for cement or screw-retained multi-unit restorations.
#### SUBJECT DEVICE DESCRIPTIONS
The GM Titanium Base for Bridge are prosthetic components to be installed onto GM implants to support the final prosthesis; the GM Titanium Base for Bridge is a two-piece abutment used as a base when fabricating a CAD/CAM customized restoration. The planning and milling of the customized superstructures should be made using the validated Dental Wings Operating System (DWOS) or 3Shape Software. All digitally designed copings and/or crowns for use with the GM Titanium Base for Bridge are intended to be manufactured as a validated milling center. The limits for customization are stated in the GM Titanium Base for Bridge Instructions for Use. The superstructure produced through the CAD/CAM System will compose the second part of the two-piece abutment; the assembly becomes a finished medical device after cementation on the GM Titanium Base for Bridge.
The main characteristics of the subject devices are the following:
- Intended for single use;
- Provided sterile via ethylene oxide gas;
- . Manufactured of titanium alloy (Ti6Al4v-ELI) per ASTM F136;
- Conical format available in different diameters, height of cementable area and gingival height;
- Screw-retained;
- Provided in rotational (non-indexed) version supporting multi-unit restorations;
- Provided with an implant-to-abutment interface compatible with the GM implants of the Neodent Implant System.
{6}------------------------------------------------
## TECHNOLOGICAL CHARACTERISTIC COMPARISON TABLE
| | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | EQUIVALENCE DISCUSSION |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | K192229<br>Neodent Implant System - GM Titanium<br>Base for Bridge<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | K180536<br>Neodent Implant System - GM Line<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | K151157<br>NC Straumann Variobase Abutment For<br>Bars/Bridges, RC Straumann Variobase<br>Abutment For Bars/Bridges, NNC<br>Straumann Variobase Abutment For<br>Bars/Bridges, RN Straumann Variobase<br>Abutment For Bars/Bridges, WN<br>Straumann Variobase Abutment For<br>Bars/Bridges<br>Straumann USA, LLC | Equivalent<br>The subject devices bring specific<br>information regarding the restoration<br>type indicated to be used with them but<br>the general indications are the same for<br>both subject and the GM Exact Titanium<br>Base abutments of the predicate device. |
| | Titanium Base Abutment is a titanium<br>base placed onto Neodent dental<br>implants to provide support for<br>customized prosthetic restorations. It is<br>used with a coping and crown, or crown<br>alone, and is indicated for cement-<br>retained single or multi-unit restorations<br>or screw-retained single restorations. All<br>digitally designed copings and/or crowns<br>to be used with the Neodent Titanium<br>Base Abutment System are intended to<br>be sent to Straumann for manufacture at<br>a validated milling center.<br>The GM Titanium Base for Bridge is<br>indicated for cement or screw-retained<br>multi-unit restorations. | GM Helix Implants and conventional<br>abutments:<br>The Neodent Implant System is intended<br>to be surgically placed in the bone of the<br>upper or lower jaw to provide support<br>for prosthetic devices such as artificial<br>teeth, to restore chewing function. It<br>may be used with single-stage or two-<br>stage procedures, for single or multiple<br>unit restorations, and may be loaded<br>immediately when good primary stability<br>is achieved and with appropriate occlusal<br>loading.<br>GM Exact Titanium Block for Medentika<br>Holder:<br>GM Exact Titanium Block for Medentika<br>Holder is a titanium abutment to be used<br>in fabricating a full custom abutment and<br>placed onto Neodent dental implants to<br>provide support for customized<br>prosthetic restorations. The GM Exact<br>Titanium Block for Medentika Holder<br>abutments are indicated for screw-<br>retained single restorations or cement-<br>retained single or multi-unit restorations.<br>All digitally designed abutments for use<br>with the GM Exact Titanium Block for<br>Medentika Holder are intended to be<br>sent to Straumann for manufacture at a<br>validated milling center. | Straumann® Variobase™ prosthetic<br>components directly connected to the<br>endosseous dental implants are<br>indicated for use as an aid in prosthetic<br>rehabilitations. The prosthetic<br>restoration (bridge or overdenture) can<br>be cemented on the Straumann®<br>Variobase™ prosthetic components. A<br>temporary restoration can be used prior<br>to the insertion of the final components<br>to maintain, stabilize and form the soft<br>tissue during the healing phase. They<br>may not be placed into occlusion. Final<br>abutments and restorations may be<br>placed into occlusion when the implant is<br>fully osseointegrated. | |
| | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | |
| | K192229<br>Neodent Implant System - GM Titanium<br>Base for Bridge<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | K180536<br>Neodent Implant System - GM Line<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | K151157<br>NC Straumann Variobase Abutment For<br>Bars/Bridges, RC Straumann Variobase<br>Abutment For Bars/Bridges, NNC<br>Straumann Variobase Abutment For<br>Bars/Bridges, RN Straumann Variobase<br>Abutment For Bars/Bridges, WN<br>Straumann Variobase Abutment For<br>Bars/Bridges<br>Straumann USA, LLC | EQUIVALENCE DISCUSSION |
| | | GM Exact Titanium Base abutments: | | |
| | | Titanium Base Abutment is a titanium<br>base placed onto Neodent dental<br>implants to provide support for<br>customized prosthetic restorations. It is<br>used with a coping and crown, or crown<br>alone, and is indicated for cement-<br>retained single or multi-unit restorations,<br>or screw-retained single restorations. All<br>digitally designed copings and/or crowns<br>for use with the Neodent Titanium Base<br>Abutment System are intended to be<br>sent to Straumann for manufacture at a<br>validated milling center.<br>Titanium Base C for GM Exact | | |
| | | abutments:<br>The Titanium Base C for GM Exact<br>abutments is a titanium component that<br>is placed over Neodent implants to<br>provide support for custom prosthetic<br>restorations, such as copings or crowns.<br>It is indicated for single-tooth screw-<br>retained restorations. All digitally<br>designed copings and/or crowns for use<br>with the Titanium Base C for GM Exact<br>abutments are to be designed using<br>Sirona inLab software or Sirona CEREC<br>Software and manufactured using a<br>Sirona CEREC or inLab MC X or MC XL<br>milling unit. | | |
| Implant-Abutment<br>interface | GM | GM | N/A | Same as Primary Predicate |
| | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | |
| | K192229<br>Neodent Implant System - GM Titanium<br>Base for Bridge<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | K180536<br>Neodent Implant System - GM Line<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | K151157<br>NC Straumann Variobase Abutment For<br>Bars/Bridges, RC Straumann Variobase<br>Abutment For Bars/Bridges, NNC<br>Straumann Variobase Abutment For<br>Bars/Bridges, RN Straumann Variobase<br>Abutment For Bars/Bridges, WN<br>Straumann Variobase Abutment For<br>Bars/Bridges<br>Straumann USA, LLC | EQUIVALENCE DISCUSSION |
| | Straight | Straight | Straight | Equivalent |
| Design | Conical upper portion<br>Helical micro grooves in the cementable<br>portion | Cylindrical upper portion<br>Helical micro grooves in the cementable<br>portion | Conical upper portion<br>Helical micro grooves in the cementable<br>portion | The coronal design of the subject device<br>is equivalent to that of the reference<br>device K151157. |
| Reusable | No | No | No | Same |
| Diameter (Ø) | 3.5, 4.5, and 5.5 mm | 5.5 mm | 3.3 mm | Equivalent<br>The reference devices (K151157) have a<br>minimum diameter of 3.3 mm which is<br>similar to the subject devices. Dynamic<br>fatigue testing was performed with the<br>new worst-case design of the subject<br>devices, the results indicate performance<br>suitable for the intended use. |
| Gingival Height<br>(mm) | 0.8, 1.5, 2.5, 3.5, and 4.5 | 0.8, 1.5, 2.5, 3.5, 4.5, and 5.5 | N/A | Equivalent<br>Subject devices are within the range<br>established by the primary predicate. |
| Maximum<br>Angulation | 30° | 30° | 30° | Same |
| Abutment Height | 4 mm | 4 mm | 3.5 – 5.2 mm | Same as Primary Predicate |
| Material | TAV (ASTM F136) | TAV (ASTM F136) | Ti-6Al-7Nb (TAN) | Same as Primary Predicate |
| Sterilization<br>Method | Ethylene Oxide to an SAL of 1x10-6 | Ethylene Oxide to an SAL of 1x10-6 | Non-Sterile | Same as Primary Predicate |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
The subject devices and the primary predicate device (K180536) have the same Indications for Use, similar design, similar range of gingival height, same raw material and same sterilization method.
#### PERFORMANCE DATA
Dynamic fatigue test per ISO 14801 was performed to determine the fatigue strength for the worst-case implant construct assembled with GM Titanium Base for Bridge, according to FDA Guidance. The abutments were tested with three different restoration materials and all of them are in accordance with the acceptance criteria.
Torsion testing was performed to evaluate the Titanium Base Screws under static torsional loading. The results met the acceptance criteria.
Sterilization of the subject abutments via ethylene oxide gas using the overkill method has been validated according to the requirements of ISO 1135. A minimum Sterility Assurance Level (SAL) of 1 x 10th has been validated.
Ethylene oxide residuals have been assessed per ISO 10993-7. Residuals are within accepted limits.
Biological Safety Assessment guided by ISO 10993-1.
Cytotoxicity testing was performed per ISO 10993-5.
Chemical characterization was performed per ISO 10993-18.
Biocompatibility sample preparation was performed per ISO 10993-12.
#### CONCLUSION
The subject devices and the primary predicate devices K180536 have the same intended use, similar designs and technological characteristics, similar range of gingival height, same sterilization method and are made of the same materials. The data included in this submission demonstrate that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.