SENDERO MAX Delivery Catheter

K242644 · Okami Medical, Inc. · DQY · Dec 20, 2024 · Cardiovascular

Device Facts

Record IDK242644
Device NameSENDERO MAX Delivery Catheter
ApplicantOkami Medical, Inc.
Product CodeDQY · Cardiovascular
Decision DateDec 20, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Intended Use

The SENDERO MAX Delivery Catheter is intended for the peripheral vasculature for the infusion of diagnostic and therapeutic agents.

Device Story

SENDERO MAX Delivery Catheter is a single-lumen, variable-stiffness catheter designed for peripheral vasculature access. Device features radiopaque distal marker, Luer-lock proximal hub, and hydrophilic shaft coating to reduce friction. Operated by clinicians using standard interventional techniques (e.g., guidewire) under fluoroscopic guidance. Once positioned, device serves as conduit for infusion of diagnostic or therapeutic agents. Benefits include facilitated navigation and targeted delivery of agents to peripheral sites.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including dimensional/visual inspection, coating thickness/lubricity, simulated use, coating integrity, compatibility, burst pressure (ISO 10555-1), tensile strength (ISO 10555-1), kink diameter, torque, hub testing (ISO 80369-7), particulate, flowrate (ISO 10555-1), radiopacity (ASTM F640), packaging validation, and biocompatibility (ISO 10993-1).

Technological Characteristics

Single-lumen, variable-stiffness catheter with hydrophilic coating. Features radiopaque distal marker and proximal Luer-lock hub. Standards: ISO 10555-1 (burst pressure, tensile strength, flowrate), ISO 80369-7 (hub), ASTM F640 (radiopacity), ISO 10993-1 (biocompatibility). Non-powered, manual operation.

Indications for Use

Indicated for infusion of diagnostic and therapeutic agents into the peripheral vasculature.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized caduceus symbol. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in a larger, bold blue font, and "ADMINISTRATION" in a smaller blue font below. December 20, 2024 Okami Medical, Inc. Jill Delsman Vice President RA/QA 8 Argonaut, Suite 100 Aliso Viejo, California 92656 Re: K242644 Trade/Device Name: SENDERO MAX Delivery Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: August 31, 2024 Received: September 3, 2024 Dear Jill Delsman: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely, Digitally signed by Lydia S. Lydia S. Glaw -S Date: 2024.12.20 Glaw -S 11:21:54 -05'00' Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K242644 Device Name SENDERO MAX Delivery Catheter Indications for Use (Describe) The SENDERO MAX Delivery Catheter is intended for the peripheral vasculature for the infusion of diagnostic and therapeutic agents. X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Okami Medical. The word "Okami" is written in a teal sans-serif font. Below "Okami" is the word "MEDICAL" in a smaller, gray sans-serif font with a registered trademark symbol. To the right of the "i" in "Okami" is a red circle with a yellow outline. ## 510(k) SUMMARY | Sponsor | Okami Medical, Inc.<br>8 Argonaut, Suite 100<br>Aliso Viejo, CA 92656<br>USA | |----------------------|------------------------------------------------------------------------------| | Contact | Jill Delsman<br>949-446-9710<br>jilld@okamimedical.com | | Device Trade Name | SENDERO MAX Delivery Catheter | | Common Name | Catheter | | Classification Name | Catheter, Percutaneous | | Regulation Number | 870.1250 | | Product Code | DQY | | Predicate Device | K240384 – SENDERO Microcatheter – Product Code DQY | | Reference Device | K140080 – Envoy Distal Access Guiding Catheter – Product Code DQY | | 510 (k) Summary Date | December 4. 2024 | ### Device Description Summary The SENDERO MAX Delivery Catheter is a line extension of the SENDERO family of delivery catheters which includes the predicate SENDERO Microcatheter (K240384). The SENDERO MAX Delivery Catheter is a single lumen, variable stiffness catheter with a radiopaque marker on the shaped distal end and a Luer-lock hub on the proximal end. A hydrophilic coating on the catheter shaft reduces friction during navigation through the vasculature. The device is delivered to the target location using standard interventional techniques (e.g. use of guide wire, etc.) under fluoroscopic guidance. Once at the target location, the lumen of the device allows for the introduction of diagnostic and therapeutic agents into the peripheral vasculature. ### Intended Use / Indications for Use The SENDERO MAX Delivery Catheter is intended for the peripheral vasculature for the infusion of diagnostic and therapeutic agents. ### Indications for Use Comparison The indication for use statement between the primary predicate and the subject device is identical. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Okami Medical. The word "Okami" is written in a teal sans-serif font, with a red circle with a yellow outline above the "i". Below the word "Okami" is the word "MEDICAL" in a smaller, gray sans-serif font, with the registered trademark symbol next to it. ## Comparison of Technological Characteristics The Okami Medical SENDERO MAX Delivery Catheter has the same or similar technological characteristics when compared to the primary and reference predicates (e.g. intended use, principle of operation, fundamental design characteristics, materials, etc.). All devices share the same intended use and principle of operation, namely the placement of the device to the target site by percutaneous means under fluoroscopy and introduction of diagnostic and/or therapeutic agents into the peripheral vasculature. All devices are single lumen, variable stiffness catheters with radiopaque markers on the distal end and a Luer-lock hub on the proximal end. All the devices also have hydrophilic coatings on the catheter shaft to reduce friction during navigation through the vasculature. Each of the fundamental design characteristics of the Okami Medical SENDERO MAX Delivery catheter are also present in the primary predicate or reference predicate. Any minor design differences between the primary predicate device and the subject device are covered by the reference predicate. ### Non-clinical Performance Testing Performance testing was conducted for the SENDERO MAX Delivery Catheter to demonstrate the suitability of the device for its intended use. The results of the testing indicate that the SENDERO MAX is substantially equivalent to the predicate device. Performance testing included: - Dimensional/Visual Inspection - Coating Thickness and Lubricity - Simulated Use Testing - Catheter Coating Integrity - Compatibility Testing - Burst Pressure Testing (per ISO 10555-1) - Catheter Tensile Strength (per ISO 10555-1) - Kink Diameter Testing - Torque Testing - Hub Testing (per ISO 80369-7) - Particulate Testing - Flowrate Testing (per ISO 10555-1) - . Radiopacity Testing (per ASTM F640) - Packaging Validation - Biocompatibility Evaluation (per ISO 10993-1) No clinical studies were required. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for Okami Medical. The word "Okami" is written in a teal sans-serif font, with a red circle with a yellow outline above the "i". Below "Okami" is the word "MEDICAL" in a smaller, gray sans-serif font, with the registered trademark symbol next to it. ### Conclusions The SENDERO MAX Delivery Catheter is substantially equivalent to the predicate in terms of intended use, principle of operation, technological characteristics, and performance characteristics. The non-clinical performance data submitted in this premarket notification clearly supports the substantial equivalence of the SENDERO MAX to the predicate, and demonstrates the subject device should perform as intended when used as instructed.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...