The ENVOY® Distal Access (DA) Guiding Catheter is intended for use in the peripheral, coronary, and neurovasculature for the intravascular introduction of interventional/diagnostic devices.
Device Story
Variable stiffness, braided catheter; large non-tapered lumen; hydrophilic outer coating; PTFE-lined inner lumen. Distal segment flexible for navigation; radiopaque marker band 3mm from distal tip. Available with preshaped tips. Used in peripheral, coronary, and neurovasculature; operated by physicians. Facilitates intravascular passage of interventional/diagnostic devices (e.g., guide wires). Provides access to distal vasculature; aids visualization under fluoroscopy. Peel-away introducer included for sheath insertion.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including visual inspection, tensile strength, coating integrity, leakage, stiffness, track testing, and biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, hemocompatibility).
Technological Characteristics
Variable stiffness braided catheter; stainless steel reinforcement; PTFE inner liner; hydrophilic outer coating. Radiopaque marker band (Pellethane with filler). 6F outer diameter; 0.071" inner diameter. Sterilized via Ethylene Oxide (EtO). Complies with ISO 594-1/2 (Luer), ISO 10555-1 (catheters), and ISO 10993 (biocompatibility).
Indications for Use
Indicated for use in peripheral, coronary, and neurovasculature for intravascular introduction of interventional/diagnostic devices.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
### SECTION 005 510(k) Summary [As Required By 21 CFR 807.92(a)]
. APR 2 4 2014
| <b>A. Sponsor</b> | |
|----------------------------------------------------------|----------------|
| <b>Submitter's Name:</b> Codman & Shurtleff, Inc. | |
| <b>Address:</b> 325 Paramount Drive<br>Raynham, MA 02767 | |
| Primary Contact: | Hannah Foley |
| Telephone: | (305) 265-6810 |
| Fax: | (508) 977-6979 |
| Secondary Contact: | Amarilys Machado |
|--------------------|------------------|
| Telephone: | (305) 265-6869 |
| Fax: | (508) 977-6979 |
- B. Date Prepared: January 10, 2014
#### C. Device Name and Classification:
| Proprietary Name: | ENVOY® Distal Access (DA) Guiding Catheter |
|----------------------|---------------------------------------------------|
| Common/Usual Name: | Catheter, Percutaneous |
| Classification Name: | Percutaneous Catheter (21 CFR 870.1250), Class II |
| Product Code: | DQY |
#### D. Predicate Devices
This 510(k) submission provides pre-market notification for the ENVOY® Distal Access (DA) Guiding Catheter line extension and modifications. The proposed line extension and modifications have not altered the fundamental technology of the predicate devices or the devices' intended use.
| Table 1: Prior 510(k) Clearances | | | | | |
|----------------------------------|--------------|------------------------------------------|-----------------------------|--------------|--------------------------------------------------------------------|
| 510(k)<br>Number | Date Cleared | Name | Manufacturer | Product Code | Predicate For: |
| Predicate<br>K120229 | 02/24/2012 | ENVOY® Distal Access<br>Guiding Catheter | Codman &<br>Shurtleff, Inc. | DQY | Intended Use Design<br>Materials<br>Manufacturing<br>Sterilization |
| Predicate<br>K093184 | 11/06/2009 | ENVOY Guiding Catheters | Codman &<br>Shurtleff. Inc. | DQY | Packaging |
{1}------------------------------------------------
#### E. Device Description
The ENVOY Distal Access (DA) Guiding Catheter is a variable stiffness, braided catheter with a large non-tapered lumen that facilitates the intravascular passage of interventional devices. The distal segment is flexible for navigation into distal vasculature. The catheter has an outer hydrophilic coating on the outer surface that reduces friction during manipulation in the vessel. The lubricious PTFE lined inner lumen is designed to facilitate movement of the guide wires and other devices. A luer fitting located on the end of the catheter hub can be used to attach accessories. The distal section of the catheter is radiopaque to aid visualization under fluoroscopy, and the distal tip is clearly distinguished by a radiopaque marker band which is approximately 3mm from the distal tip. The catheter is available with preshaped tips to facilitate positioning. A peel away introducer is included to facilitate insertion into the sheath.
#### F. Indications for Use
The ENVOY Distal Access (DA) Guiding Catheter is intended for use in the peripheral, coronary, and neuro vasculature for the intravascular introduction of interventional/diagnostic devices.
## G. Summary of Technological Characteristics of the Proposed Device to the Predicate Device
The proposed ENVOY® Distal Access (DA) Guiding Catheter is substantially equivalent to the predicate ENVOY® Distal Access (DA) Guiding Catheter. No new technological characteristics are being introduced with the proposed device.
The proposed ENVOY® Distal Access (DA) Guiding Catheter has the same intended use, sterilization process, function, mechanism of action, and clinical utility as the predicate ENVOY® Distal Access (DA) Guiding Catheter, and is similar in regards to design, material, and manufacturing. The proposed ENVOY DA was shown to be substantially equivalent to the predicate device through comparison of indications for use, function, operating principle, bench testing, biocompatibility, and materials. A summary table including characteristics of the proposed device compared with those of the predicate device is provided in Table 2A and 2B.
{2}------------------------------------------------
1
| Description | Predicate Device:<br>ENVOY® DA Guiding Catheter<br>(K120229) | This Submission:<br>ENVOY® DA Guiding Catheter |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Envoy Distal Access Guiding Catheter is<br>intended for use in the peripheral, coronary and<br>neuro vasculature for the intravascular introduction<br>of interventional/ diagnostic devices | Same as predicate |
| Device Description | The ENVOY Distal Access (DA) Guiding Catheter<br>is a variable stiffness, braided catheter with a large<br>non-tapered lumen that facilitates the intravascular<br>passage of interventional devices. The distal<br>segment is flexible for navigation into distal<br>vasculature. The catheter has an outer hydrophilic<br>coating that reduces friction during manipulation in<br>the vessel. The lubricious PTFE lined inner lumen<br>is designed to facilitate movement of the guide<br>wires and other devices. A luer fitting located on<br>the end of the catheter hub can be used to attach<br>accessories. The distal section of the catheter is<br>radiopaque to aid visualization under fluoroscopy,<br>and the distal tip is clearly distinguished by a<br>radiopaque marker. The catheter is available with<br>preshaped tips to facilitate positioning. A peel away<br>introducer is included to facilitate insertion into the<br>sheath. | The ENVOY Distal Access (DA) Guiding Catheter is a<br>variable stiffness, braided catheter with a large non-<br>tapered lumen that facilitates the intravascular passage<br>of interventional devices. The distal segment is flexible<br>for navigation into distal vasculature. The catheter has<br>an outer hydrophilic coating that reduces friction<br>during manipulation in the vessel. The lubricious PTFE<br>lined inner lumen is designed to facilitate movement of<br>the guide wires and other devices. A luer fitting located<br>on the end of the catheter hub can be used to attach<br>accessories. The distal section of the catheter is<br>radiopaque to aid visualization under fluoroscopy, and<br>the distal tip is clearly distinguished by a radiopaque<br>marker band which is approximately 3mm from the<br>distal tip. The catheter is available with preshaped tips<br>to facilitate positioning. A peel away introducer is<br>included to facilitate insertion into the sheath. |
| Product Code | DQY | Same as predicate |
| Classification | 21 CFR 870.1250, Class II | Same as predicate |
| Length/ Working<br>Length (cm) | 95cm & 105cm | Same as predicate |
| Catheter Inner<br>Diameter | 0.071" (1.8mm) | Same as predicate |
| Catheter Outer<br>Diameter | 6.0F (0.082"/2.0mm) | Same as predicate |
| Shapes | Straight<br>Multi-purpose D (MPD) | Straight<br>Multi-purpose D (MPD)<br>Multi-purpose C (MPC) |
| Transition Segment<br>Length | Standard Transition Segment Length<br>(Standard) | Standard Transition Segment Length<br>Standard<br>Shorter Transition Segment Lengths for Extra<br>Backup Support XB |
| Distal Tip<br>Radiopaque Marker | Pellethane Compound with Radiopaque Filler | Pellethane Compound with Radiopaque Filler & Metal<br>Marker Band |
| Reinforcing Member<br>(Braid) | Stainless Steel | Same as predicate |
| Liner | PTFE Liner | Same as predicate |
| Sterilization | EtO | Same as predicate |
| Product Shelf-Life | 1 year | 3 years |
Table 2B provides a packaging comparison between the existing ENVOY® Distal Access (DA) Guiding Catheter, the ENVOY Guiding Catheter and proposed ENVOY® Distal Access (DA) Guiding Catheter.
·
{3}------------------------------------------------
| Table 2B: Predicate Packaging Comparison Information | | | |
|------------------------------------------------------|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Description | Predicate Device Packaging:<br>ENVOY® DA Guiding Catheter<br>(K120229) | Predicate Device Packaging:<br>ENVOY® Guiding Catheter<br>(K093184) | This Submission:<br>Proposed Device Packaging<br>ENVOY® DA Guiding Catheter |
| Dimensions | | | |
| Pouch Size | 49.21" x 3.95" | 46.5" x 4.0" | 49.91" x 4.09" |
| Carton Size | 47"x 4" x 0.75" | 47" x 4" x 0.75" | Same as K120229 |
| Mounting Card | 46"x 2.75" x 0.012" | 42.75" x 2.687" | Same as K120229 |
| Material | | | |
| Pouch Material | Top Web: MRM4820PU (RLE-<br>005) 48GA PET/.002 LDPE<br>Bottom Web: Uncoated 1073B<br>Tyvek | Top Web: MRM002075 (M-2075)<br>48GA PET/.002 LDPE<br>Bottom Web: Uncoated 1073B Tyvek | Same as K093184 |
| Carton Material | .024" Clay Coated Solid Bleached<br>Surface | .024" Clay Coated Solid Bleached<br>Surface | Same as K120229 |
| Mounting Material | 0.012" Clay Coated Solid<br>Bleached Sulfate | Clay Coated Solid Bleached Sulfate | Same as K120229 |
| Sterilization | | | |
| Sterilization | EtO | Same | Same as predicate |
| Shelf-Life | | | |
| Packaging<br>Shelf-Life | 1 year | 3years | Same as K093184 |
#### H. Summary of Nonclinical Testing
The proposed ENVOY® Distal Access (DA) Guiding Catheter has the same intended use, sterilization process, function, mechanism of action, and clinical utility as the predicate ENVOY® Distal Access (DA) Guiding Catheter, and is similar in regard to design, material, and manufacturing. The testing conducted to assess the line extension and modifications include performance assessments per the following recognized standards:
| Table 3: Performance Standards | |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Standard/Guidance/Directive | Description |
| BS EN ISO 11607-1: 2009 | Packaging for Terminally Sterilized Medical Devices - Part 1: Requirements for<br>Materials. Sterile Barrier Systems and Packaging Systems |
| BS EN ISO 11135-1: 2007 | Sterilization of Healthcare products Ethylene Oxide: Part 1 Requirements for<br>development, validation and routine control of a sterilization process for medical devices |
| BS EN ISO 10993-7: 2008 | Biological Evaluation of medical devices: Ethylene oxide sterilization residuals - Part 7 |
| BS EN ISO 10555-1: 2009 | Sterile. single use intravascular catheters Part 1: General requirements; Sterile |
| ISO 594-1: 1986 (E) | Conical fitting with a 6% (Luer) taper for syringes, needles and certain other medical<br>equipment: Part 1 - General Requirements |
| ISO 594-2 : 1998 (E) | Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical<br>equipment: Part 2 - Local fittings |
| AAMI / ANSI HE75 : 2009 | Human Factors Engineering - Design of Medical Devices |
| BS EN 62366: 2008 | Medical Devices. Application of usability engineering to Medical Devices |
| BS EN ISO 10993-1: 2009 | Biological evaluation of medical devices: Evaluation & Testing - Part 1 |
| BS EN ISO 10993-5: 2009 | Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity |
| BS EN ISO 10993-4: 2009 | Biological evaluation of medical devices Part 4: Selection for tests for interactions with<br>blood |
| USP <661> (2013) | Containers - Plastic. Physicochemical Tests |
| ISO 10993-10: 2010 | Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization |
| ISO 10993-11: 2009 | Biological evaluation of medical devices Part11: Test for systemic toxicity |
| ISO 10993-3: 2009 | Biological evaluation of medical devices Part 3: Test for genotoxicity, carcinogenicity,<br>and reproductive toxicity |
{4}------------------------------------------------
#### Bench Testing
Results of verification and validation testing that was conducted on the proposed ENVOY® Distal Access (DA) Guiding Catheter demonstrates that it performs as designed, is suitable for its intended use, is substantially equivalent to the predicate device and therefore, does not raise any new questions of safety and effectiveness. Appropriate testing was identified based on a review of the products' risk analyses and previous validation and verification testing.
The following Verification and Validation tests were conducted to verify the modified design ENVOY® Distal Access (DA) Guiding Catheter:
- . Visual Inspection
- . Catheter Shape
- Catheter Dimensional Verification ◆
- � Tensile Strength Testing
- Hydrophilic Coating Integrity ◆
- System Liquid Leakage Testing .
- Delamination of PTFE Liner Testing .
- Lateral Stiffness Testing .
- Linear Stiffness Testing
- Back-Up Support
- Track Testing .
- Sheath Introducer Compatibility .
- Hub Luer Taper ●
- System Air Leakage ●
- . Shaft Peel Strength
The following Packaging Validation Testing was conducted with the proposed ENVOY® Distal Access (DA) Guiding Catheter:
- Visual Inspection .
- Dve Leak
- Seal Strength .
The following Biocompatibility Testing was conducted with the proposed ENVOY® Distal Access (DA) Guiding Catheter:
- . In Vitro Cytotoxicity – MEM Elution
- · · Sensitization Guinea Pig Maximization
- Intracutaneous/Irritation Reactivity .
- Acute Systemic Toxicity .
{5}------------------------------------------------
#### Page 28 of 109 K140080
- Material Mediated Pyrogenicity .
- . Genotoxicity
- o · In Vitro Bacterial Mutagenicity Ames Assay
- o In Vitro Mouse Lymphoma Assay with extended treatment
- o In Vivo Mouse Micronucleus Assay
- Hemocompatibility .
- In Vitro Hemolysis Direct & Extract Method o
- Partial Thromboplastin Time (PTT) o
- Complement Activation C3a & SC5b-9 Assay 0
- In Vivo Thrombogenicity o
- . Physicochemical Aqueous Extraction Tests
In addition, the new sterile pouch packaging material was evaluated for in vitro cytotoxicity and was tested per ISO 10993-5 (Biological Evaluation of Medical Devices - Part 5: Tests for in vitro cytotoxicity.
The following Sterilization Test was conducted with the ENVOY® Distal Access (DA) Guiding Catheter:
. USP Limumus Amebocyte Lysate (LAL) Test - Kinetic Chromogenic Method
#### I. Animal Testing
No animal studies were required as appropriate verification and validation of the catheter and packaging modifications were achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing.
#### J. Summary of Clinical testing:
No clinical studies were required as appropriate verification and validation of the catheter and packaging modifications were achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing.
#### Conclusion:
Based upon the design, materials, function, intended use, and the non-clinical testing performed by Codman it is concluded that the proposed ENVOY® Distal Access (DA) Guiding Catheter is substantially equivalent to the current ENVOY® Distal Access (DA) Guiding Catheter (K120229), and therefore, does not raise any new questions of safety and effectiveness.
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three diagonal lines that resemble a bird in flight. The caduceus is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring. MD 20993-0002
April 24, 2014
Codman & Shurtleff. Inc. Hannah Foley Regulatory Affairs Specialist II 325 Paramount Dr. Raynham, MA 02767-0350 US
Re: K140080
> Trade/Device Name: ENVOY Distal Access (DA) Guiding Catheter, 6F, 95cm & 105cm. Straight & XB. MPD & XB. MPC & XB Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: March 26, 2014 Received: March 27, 2014
Dear Ms. Folev.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Ms. Foley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Bram D. Zuckerm
> Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{8}------------------------------------------------
Indications for Use
K140080 510(k) Number (if Known):
Device Name: ENVOY® Distal Access (DA) Guiding Catheter
#### Indications for Use:
The ENVOY® Distal Access (DA) Guiding Catheter is intended for use in the peripheral, coronary, and neurovasculature for the intravascular introduction of interventional/diagnostic devices.
Prescription Use: _ X (Part 21 CFR 801Subpart D)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
:
AND/OR
Over-The-Counter Use: (Part 21 CFR 807 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Bram D. Zuckerman -S 2014.04.24 16:56:39 -04'00'
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.