ChecQ is indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region.
Device Story
ChecQ is a dental implant stability analyzer consisting of a main unit, charging cradle, and ChecQPEG attachment. It utilizes Resonance Frequency Analysis (RFA) to non-invasively measure implant stability. The device operates by applying magnetic pulses to the ChecQPEG attached to the implant, inducing vibrations. The system captures these free vibrations, processes them via Fast Fourier Transform (FFT) to extract the natural frequency, and calculates an Implant Stability Quotient (ISQ) ranging from 1 to 99. Used in clinical settings by dental professionals, the ISQ output provides a quantitative metric of bone-implant interface stability. This information assists clinicians in evaluating implant integration, supporting clinical decision-making regarding loading protocols and treatment planning, ultimately benefiting patients by ensuring implant success and reducing failure risks.
Clinical Evidence
Bench testing only. Comparative performance testing against the Osstell Beacon predicate using standardized test jigs confirmed equivalent accuracy. Additional non-clinical testing verified design specifications, locking/removal torque, ISQ measurement reliability under simulated clinical conditions, and the impact of barrier sleeves on measurement accuracy.
Technological Characteristics
Dental implant stability analyzer using RFA. Components: main unit, charging cradle, ChecQPEG. Sensing: magnetic field stimulation/vibration capture. Output: ISQ (1-99). Standards: ANSI AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 80601-2-60, IEC 62366-1, IEC 62133-2, ISO 17665-1, IEC 62304, ANSI AAMI ST79, ISO 10993-1/5/10/23, ISO 7405. Sterilization: compliant with ISO 17665-1.
Indications for Use
Indicated for patients requiring dental implant stability assessment in the oral cavity and maxillofacial region.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
March 21, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is an emblem of the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The emblem features a stylized depiction of an eagle or other bird-like figure. The text is arranged in three lines, with "FDA" in a larger font size and a blue square behind it.
DENTIS Co., Ltd. % Soojung Moon CEO Allura Medical Solution, Inc. 5485 Rathdrum Way Antioch, California 94531
Re: K241065
Trade/Device Name: ChecQ (AC100) Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EKX Dated: December 13, 2024 Received: February 27, 2024
Dear Soojung Moon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Image /page/2/Picture/3 description: The image shows the letters FDA in a large, sans-serif font. The letters are light blue and appear to be slightly faded. The background is white, which makes the letters stand out. The letters are arranged horizontally and are evenly spaced.
Bobak Shirmohammadi -S
For Michael E. Adjodha, M.ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K241065
Device Name
ChecQ (AC100)
Indications for Use (Describe)
ChecQ is indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region.
Type of Use (Select one or both, as applicable)
> | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K241065 - 510(k) Summary
March 20, 2025
| 510(k)#: K241065 | | | |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------|--------------------------------------------------------|--------------|
| Contact Detail | | | |
| Applicant Name | | DENTIS Co., Ltd. | |
| Applicant Address | | 99, Seongseoseo-ro, Dalseo-gu Daegu 42718 Korea, South | |
| Applicant Contact Telephone | | +82535893499 | |
| Applicant Contact | | Mr. Gyeong-Seob Kim | |
| Applicant Contact Email | | rudtjq65@dentis.co.kr | |
| Correspondent Name | | Allura Medical Solution, Inc. | |
| Correspondent Address | | 5485 Rathdrum Way Antioch CA 94531 United States | |
| Correspondent Contact Telephone | | +1(407)242796 | |
| Correspondent Contact | | Mr. SOOJUNG MOON | |
| Correspondent Contact Email | | sjmoon@allurameds.com | |
| Device Name | | | |
| Device Trade Name | | ChecQ (AC100) | |
| Common Name | | Dental handpiece and accessories | |
| Classification Name | | Handpiece, Direct Drive, Ac-Powered | |
| Regulation Number | | 872.4200 | |
| Product Code(s) | | EKX | |
| Legally Marketed Predicate Devices | | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | Product Code |
| K181888 | Osstell Beacon | | EKX |
| K143445 | Tellos ISQ Buddy | | EKX |
| Device Description Summary | | | |
| The ChecQ is a dental implant stability analyzer comprising a main unit and a charging cradle that uses | | | |
| Resonance Frequency Analysis (RFA) to measure implant stability non-invasively. It works with the | | | |
| ChecQPEG, which attaches to the dental implant, and measures stability by analyzing resonant frequencies | | | |
| generated by magnetic field stimulation from the probe tip. The stability result is displayed on the main | | | |
| unit's screen as an Implant Stability Quotient (ISQ), ranging from 1 to 99. | | | |
| The RFA method evaluates by applying transverse force to the implant using magnetism, measuring | | | |
| movement, and analyzing the resulting resonance frequency. The resonance frequency depends on the bone- | | | |
| implant gap and is calculated using a formula, with values expressed as ISQ scores (1-99) for clinical | | | |
| interpretation. The ChecQ mechanism converts voltage pulses generated by the DAC of the MCU into | | | |
| magnetic pulses, which induce vibrations in the ChecQPEG magnets. The device captures the free | | | |
vibrations as magnetic pulses, processes them through an FFT to extract the natural frequency, and calculates the ISQ value based on this frequency.
Intended Use/Indications for Use
ChecQ is indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region. Indications for Use Comparison
{5}------------------------------------------------
The Checo has same intended use of the predicate device currently being marketed, and there is no significant difference between two devices that would adversely affect the use.
### Technological Comparison
The device description, software validation, performance testing, electromagnetic compatibility testing, sterilization validation, and biocompatibility assessment demonstrate the substantial equivalence of the subject device to the identified predicate device. The testing and documentation provided to support substantial equivalence of this device has been performed in accordance with the following standards:
- ANSI AAMI ES 60601-1:2005 -
- IEC 60601-1-2 -
- IEC 60601-1-6 -
- IEC 80601-2-60 -
- IEC 62366-1 -
- IEC 62133-2 -
- ISO 17665-1 -
- IEC 62304 -
- ANSI AAMI ST79 -
- ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-23 -
- ISO 7405
Therefore, conclusions drawn from testing demonstrate that ChecQ is substantially equivalent in performance to the predicate devices of K181888.
### Non-Clinical and/or Clinical Tests Summary & Conclusions
The technological comparison and non-clinical testing of the "ChecQPEG" devices have been thoroughly evaluated through self-established test methodologies to ensure their performance and safety. There is no established test standard for these tests: however, they were conducted based on ISO 13485:2016 to ensure compliance with regulatory and quality requirements.
The ChecQ dental implant stability analyzer underwent comparative testing against the predicate device, Osstell Beacon, to assess its accuracy. The self-test methodology involved multiple measurements using standardized test jigs, and accuracy was determined based on the difference between the average values obtained from ChecQ and Osstell Beacon under the same test conditions. The results confirmed that ChecQ exhibited equivalent performance, establishing its functional equivalence with the predicate device.
The ChecQPEG also underwent multiple tests to verify its design and mechanical properties. The dimensional test confirmed that the manufactured parts met the required specifications. The locking and removal torque force test validated the device's stability when attached to an implant fixture. The ISQ (Implant Stability Quotient) measurement test demonstrated the device's reliability in assessing implant stability under simulated clinical conditions. Additional testing was conducted to evaluate the potential effects of the barrier sleeve on ISQ measurements, confirming that the presence of the barrier sleeve did not impact measurement accuracy.
These tests provide comprehensive verification of ChecQ and ChecQPEG's substantial equivalence to predicate devices. The results substantiate their compliance with international standards, demonstrating their suitability for clinical use and regulatory approval.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.