Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix
K240424 · Collagen Matrix, Inc. · MQV · Mar 12, 2024 · Orthopedic
Device Facts
Record ID
K240424
Device Name
Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix
Applicant
Collagen Matrix, Inc.
Product Code
MQV · Orthopedic
Decision Date
Mar 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix is intended for use as a bone void filler for voids or gaps, that are not intrinsic to the stability of the bony structure. The device is to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, pelvis, intervertebral disc space, and posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process. In the posterolateral spine and intervertebral disc space, Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix is combined with either autogenous bone marrow or autograft with saline and can also be used with autograft as a bone graft extender. When used in intervertebral body fusion procedures, Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Device Story
Osteoconductive, bioactive, porous bone graft matrix; composed of anorganic bovine bone mineral, 45S5 bioactive glass, and type I bovine collagen. Moldable implant packed into bony voids or gaps by surgeons; resorbs over time and is replaced by new bone tissue. Used in orthopedic and spinal procedures; provides scaffold for bony ingrowth. No electronic components or software.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on identical technological characteristics, materials, and performance data previously established for predicate devices K231942 and K182074.
Technological Characteristics
Materials: Anorganic bovine bone (ASTM F1581), 45S5 Bioactive Glass (ASTM F1538), and purified type I bovine collagen. Form factor: Moldable, porous matrix. Sterilization: Validated method (SAL 10^-6). Non-pyrogenic. No software or electronic components.
Indications for Use
Indicated for patients requiring bone void filling in the skeletal system (extremities, pelvis, intervertebral disc space, posterolateral spine) for non-structural gaps or voids resulting from surgery or trauma. Used with autogenous bone marrow or autograft in spine applications; must be used with an FDA-cleared intervertebral body fusion device for intervertebral body fusion procedures.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Mineral Collagen Composite Bioactive Extra Moldable (K231942)
Reference Devices
Mineral Collagen Composite Bioactive Moldable (K182074)
Submission Summary (Full Text)
{0}------------------------------------------------
March 12, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Collagen Matrix, Inc. Daniel Fernandez Regulatory Affairs, Specialist 15 Thornton Road Oakland, New Jersey 07436
## Re: K240424
Trade/Device Name: Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: February 12, 2024 Received: February 13, 2024
Dear Daniel Fernandez:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Sara S. Thompson -S
For
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K240424
Device Name
Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix
#### Indications for Use (Describe)
Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix is intended for use as a bone void filler for voids or gaps, that are not intrinsic to the stability of the bony structure. The device is to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, pelvis, intervertebral disc space, and posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process.
In the posterolateral spine and intervertebral disc space, Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix is combined with either autogenous bone marrow or autograft with saline and can also be used with autograft as a bone graft extender. When used in intervertebral body fusion procedures, Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) SUMMARY
| 1. | Applicant Information | | |
|----|----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|--|
| | Applicant Name:<br>Address:<br>Telephone:<br>Fax: | Collagen Matrix, Inc.<br>15 Thornton Road<br>Oakland, New Jersey 07436<br>(201) 405-1477<br>(201) 405-1355 | |
| | Contact Person: | Daniel Fernandez<br>Regulatory Affairs, Specialist<br>dfernandez@regenity.com | |
| | Date Prepared: | March 6, 2024 | |
| 2. | Name of the Device | | |
| | Device Trade Name: | Mineral Collagen Composite Bioactive Moldable<br>Bone Graft Matrix | |
| | Device Common Name(s): | Bone Void Filler<br>Bone Graft Matrix<br>Bone Graft Substitute | |
| | Device Classification Name: | Filler, Bone Void, Calcium Compound<br>888.3045<br>MQV<br>Class II | |
| 3. | Legally Marketed Devices to Which Substantial Equivalence is Claimed | | |
| | Predicate Device: | Catalyst Bone Void Filler<br>OssDsign Av<br>K232315 | |
| | Secondary Predicate Device: | Mineral Collagen Composite Bioactive Extra<br>Moldable<br>Collagen Matrix, Inc<br>K231942 | |
| | Reference Device: | Mineral Collagen Composite Bioactive Moldable<br>Collagen Matrix, Inc<br>K182074 | |
## 4. Description of the Device
Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix is composed of anorganic bone mineral, bioactive glass, and type I collagen that can be molded to fit the bone defect. It is an osteoconductive, bioactive, porous implant that allows for bony ingrowth across the graft site.
{5}------------------------------------------------
The bone graft matrix is slowly resorbed and replaced by new bone tissue during the natural healing process.
The anorqanic bone mineral component of the bone graft matrix is a natural, porous bone graft material produced by removal of all organic components from bovine bone. The composition of the anorganic bone mineral meets ASTM F1581 standard specifications for composition of anorganic bone for surgical implants. The bioactive glass component of the device is made of 45S5 Bioactive Glass and meets ASTM F1538 standard specifications for glass and glass ceramics biomaterials for implantation. The purified type I collagen is derived from bovine Achilles tendon.
The product is available in various sizes and is provided sterile, non-pyrogenic, and for single use only.
## 5. Intended Use / Indications for Use
Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix is intended for use as a bone void filler for voids or gaps, that are not intrinsic to the bony structure. The device is to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, pelvis, intervertebral disc space, and posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process.
In the posterolateral spine and intervertebral disc space, Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix is combined with either autogenous bone marrow or autograft with saline and can also be used with autograft as a bone graft extender. When used in intervertebral body fusion procedures. Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
### 6. Technological Comparison
The subject device has equivalent indications, design principles, and performance characteristics as the predicate device. OssDsign Catalyst Bone Void Filler (510(k) K232315) and secondary predicate Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix (K231942).
The subject device has the same basic design characteristics and technological characteristics (i.e design, material, chemical composition, principle of operation) as the secondary predicate device, K231942 and reference device K182074, with the same specification range.
Any differences in the technological characteristics between the subject and predicate devices do not raise new issues or concerns of safety or efficacy.
### 7. Non-Clinical and/or Clinical Test Summary and Conclusions
The purpose of this submission is to expand the indications for use of the Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix to include use in the intervertebral disc space in conjunction with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
{6}------------------------------------------------
Performance testing of the subject device remains unchanged as that submitted for the secondary predicate (K231942) and reference device (K182074). There is no change to the device characteristics and specifications, manufacturing processes or composition resulted from the expanded indications, and the subject device remains identical to the cleared secondary predicate and reference devices K231942 and K182074.
There are no changes to the validated sterilization method and sterility assurance level (SAL 1x10t0) remains the same as documented in the secondary predicated device (K231942) and reference device (K182074).
The subject device is non-pyrogenic. There are no changes to the product and the biocompatibility data remains the same as that submitted in the secondary predicate device (K231942) and reference device (K182074).
Biocompatibility testing of the subject device was not required, as there are no changes to the product and the performance data as submitted for the secondary predicate device (K231942) and the reference device (K182074).
Additional animal testing was not required to determine substantial equivalence. The animal testing conducted and submitted in secondary predicate K231942 and reference device K182074 is applicable for the subject device.
Clinical performance data was not required to determine substantial equivalence.
## 8. Conclusion
The subject device and the predicate devices have intended use, similar technological characteristics, and are made of similar materials. The subject device and predicate devices share similar packaging and sterilization methods. The data included and referenced in this submission demonstrates substantial equivalence of the subject device to the predicate devices. Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix is as safe, effective, and performs as well as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.