Catalyst Bone Void Filler is indicated for filling bone voids or defects of the skeletal system (i.e., the posterolateral spine, intervertebral disc space, extremities, and pelvis) that are not intrinsic to the stability of the bony structure. These osseous defects may be surgically created or as a result of traumatic injury to the bone. Catalyst Bone Void Filler is a bone graft putty that is resorbed and replaced with bone during the healing process. Catalyst Bone Void Filler must be used with morselized autograft bone at a ratio of 1:1 by volume in the posterolateral spine. When used in intervertebral body fusion procedures, Catalyst Bone Void Filler must be used with morselized autograft bone at a ratio of 1:1 by volume with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Device Story
Osteoconductive, resorbable, porous bone void filler; composed of 5.8 wt% silicon-substituted calcium phosphate granules in resorbable polymer gel (30% granules, 70% gel). Supplied as sterile, moldable putty in 1-10 cc syringes. Implanted by surgeons in OR settings for skeletal defects. Functions as bone graft extender; resorbed and replaced by bone during healing. Provides consistent, rapid bone ingrowth via high surface area porous granules. No mixing or setting time required. Clinical benefit: facilitates bone healing in orthopedic/spinal procedures.
Clinical Evidence
No new clinical data. Relies on non-clinical testing from prior clearance (K193075), including chemical composition, physical properties, sterilization, material-mediated pyrogenicity, bacterial endotoxin, shelf life, biocompatibility, and animal testing in a posterolateral spine fusion model.
Technological Characteristics
100% nanosynthetic calcium phosphate bone void filler; 5.8 wt% silicon-substituted calcium phosphate granules; resorbable polymer gel binder. Form factor: moldable putty. Sterilization: radiation. Standalone device, non-electronic.
Indications for Use
Indicated for filling non-structural bone voids or defects in the skeletal system (posterolateral spine, intervertebral disc space, extremities, pelvis) resulting from surgery or trauma. Requires 1:1 volume ratio with autograft in posterolateral spine and intervertebral body fusion procedures.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym "FDA" stacked above the words "U.S. FOOD & DRUG ADMINISTRATION".
September 15, 2023
OssDsign AB Jonas Aberg Regulatory Affairs Manager Rapsgatan 23 A Uppsala, SE 754 50 Sweden
Re: K232315
Trade/Device Name: Catalyst Bone Void Filler Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: August 2, 2023 Received: August 2, 2023
Dear Mr. Aberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Sara S. Thompson -S
For
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K232315
Device Name
Catalyst Bone Void Filler
Indications for Use (Describe)
Catalyst Bone Void Filler is indicated for filling bone voids or defects of the skeletal system (i.e., the posterolateral spine, intervertebral disc space, extremities, and pelvis) that are not intrinsic to the stability of the bony structure. These osseous defects may be surqically created or as a result of traumatic injury to the bone. Catalyst Bone Void Filler is a bone graft putty that is resorbed and replaced with bone during the healing process. Catalyst Bone Void Filler must be used with morselized autograft bone at a ratio of 1:1 by volume in the posterolateral spine. When used in intervertebral body fusion procedures, Catalyst Bone Void Filler must be used with morselized autograft bone at a ratio of 1:1 by volume with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Device Trade Name: | Catalyst Bone Void Filler |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | OssDsign AB<br>Rapsgatan 23 A<br>SE 754 50 Uppsala<br>Sweden |
| Contact: | Jonas Aberg<br>Regulatory Affairs Manager<br>OssDsign AB<br>Rapsgatan 23A, 754 50 Uppsala<br>Sweden<br>Phone: +1(410) 775 5465<br>Email: ja@ossdsign.com |
| Prepared by: | MCRA, LLC<br>803 7th Street, NW, 3rd Floor<br>Washington, DC 20001<br>Office: 202.552.5800 |
| Date Prepared: | August 2, 2023 |
| Classification: | 21 CFR §888.3045, Resorbable calcium salt bone void filler<br>device |
| Class: | II |
| Product Code: | MQV |
| Primary Predicate: | NuVasive Attrax Putty (NuVasive, Inc., K203714) |
| Reference Device: | Osteo³ ZP Putty (Sirakoss Ltd., K193075) |
#### Indications For Use:
Catalyst Bone Void Filler is indicated for filling bone voids or defects of the skeletal system (i.e., the posterolateral spine, intervertebral disc space, extremities, and pelvis) that are not intrinsic to the stability of the bony structure. These osseous defects may be surgically created or as a result of traumatic injury to the bone. Catalyst Bone Void Filler is a bone graft putty that is resorbed and replaced with bone during the healing process. Catalyst Bone Void Filler must be used with morselized autograft bone at a ratio of 1:1 by volume in the posterolateral spine. When used in intervertebral body fusion procedures, Catalyst Bone Void Filler must be
{4}------------------------------------------------
used with morselized autograft bone at a ratio of 1:1 by volume with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
#### Device Description:
OssDsign Catalyst is an osteoconductive, resorbable, porous, 100% nanosynthetic calcium phosphate bone void filler. OssDsign Catalyst contains 5.8 wt% silicon-substituted calcium phosphate granules suspended in a resorbable polymer gel. The final, finished OssDsign Catalyst is 30 wt% granules and 70 wt% polymer gel. The high surface area porous granules have been designed to deliver consistent and rapid bone ingrowth, remodeling, and cellmediated resorption during the bone healing process. The aqueous polymer gel phase binds the highly porous granules into a moldable, pliable formulation which enables OssDsign Catalyst to be implanted directly from the packaging without any further gelation, mixing or graft setting time. OssDsign Catalyst is provided sterile to the end user in 1 cc, 2.5 cc, 5 cc, and 10 cc syringe volumes.
OssDsign Catalyst is identical to the currently marketed product Osteo3 ZP Putty (K193075). The cleared Osteo3 ZP Putty was launched in the USA in October 2021 and marketed under the name OssDsign Catalyst.
#### Predicate Device:
OssDsign AB submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices. Catalyst Bone Void Filler is substantially equivalent in intended use, indications, basic design, function, and performance to the following predicate device, which has been determined by FDA to be substantially equivalent to pre-amendment devices:
Primary Predicate: NuVasive Attrax Putty, NuVasive, Inc., K203714
#### Reference Device:
The Catalyst Bone Void Filler subject device substantially equivalent with respect to design principles, materials, manufacturing process, and mechanical performance to the following reference device, which has been determined by FDA to be substantially equivalent to preamendment devices:
Reference Device: Osteo3 ZP Putty, Sirakoss Ltd., K193075
### Performance Testing Summary:
The purpose of this submission is to expand the indications for use of the Catalyst Bone Void Filler device to include use in the intervertebral disc space in conjunction with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Because the subject device is identical to the previously cleared reference device (K193075), it was determined that existing testing data is adequate to support the expanded indications proposed in the current submission.
{5}------------------------------------------------
In support of the prior clearance (K193075), non-clinical testing data were submitted according to the guidance documents Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device (issued June 2003) and Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile (issued January 2016). The non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included chemical composition, physical properties, sterilization, material-mediated pyrogenicity, bacterial endotoxin, sterile barrier shelf life, product shelf life, biocompatibility, and animal testing.
#### Substantial Equivalence:
The subject device was demonstrated to be substantially equivalent to the NuVasive Attrax Putty (K203714) primary predicate device with respect to indications, design principles, and performance.
The subject device and primary predicate device both incorporate calcium phosphate materials within a resorbable polymeric binder or scaffold, are similar in physical form (putty), and are provided in multiple size options. The subject and primary predicate devices are provided sterile using similar radiation sterilization methods and are intended for single-patient and single-use.
The reference device, Osteo3 ZP Putty (K193075), is referenced in support of the subject device material composition, manufacturing processes, and mechanical performance testing in the posterolateral spine fusion animal model.
The subject device, the primary predicate device (K203714), and the reference device (K193075) have the same intended use, the same product classification and product code (MOV) and have similar Indications for Use statements. The subject device, the primary predicate device, and the reference device are indicated for use as stand-alone bone void fillers in the extremities and pelvis. The subject device and reference device are indicated for use with autograft bone (as a bone graft extender) in the posterolateral spine; the primary predicate device is indicated for standalone or optional use with autograft bone in the posterolateral spine. The subject device is indicated for use in the intervertebral disc space when used with autograft bone and an intervertebral body fusion device cleared by FDA for use with a bone void filler; the primary predicate device is indicated for standalone use in the intervertebral disc space when used with an intervertebral body fusion device cleared by FDA for use with a bone void filler; the reference device is not indicated for use in the intervertebral disc space. Although the subject device and the primary predicate device have slightly different indications for use language, these differences do not change the intended use as a bone void filler. Notably, the indications for use language pertaining to the intervertebral disc space retains the identical configurations for use as those specified for the respective device's posterolateral spine indication (i.e., mixed with autograft or standalone).
The performance of the subject Catalyst Bone Void Filler has previously been assessed at the time of the prior clearance (K193075).
{6}------------------------------------------------
Any differences in the technological characteristics between the subject and predicate devices do not raise new issues or concerns of safety or efficacy.
#### Conclusion:
The subject device and the primary predicate device have the same intended use, nearly identical indications for use, are similar in physical form (putty), and have similar technological characteristics. The subject and predicate devices are provided sterile, are sterilized using similar methods, and are intended for single-patient and single-use. The data included in this submission demonstrate substantial equivalence to the predicate device and reference device listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.