K182074 · Collagen Matrix, Inc. · MQV · Mar 21, 2019 · Orthopedic
Device Facts
Record ID
K182074
Device Name
Mineral Collagen Composite Bioactive Moldable
Applicant
Collagen Matrix, Inc.
Product Code
MQV · Orthopedic
Decision Date
Mar 21, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Mineral Collagen Composite Bioactive Moldable combined with either autogenous bone marrow or autograft with saline is indicated for bony voids or gaps, that are not intrinsic to the stability of the bony structure; Mineral Collagen Composite Bioactive Moldable can also be used with autograft as a bone graft extender. The device is to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process.
Device Story
Mineral Collagen Composite Bioactive Moldable is an osteoconductive, porous bone graft matrix; composed of anorganic bovine bone mineral, 45S5 bioactive glass, and type I bovine collagen. Device is supplied sterile and moldable; hydrated with autogenous bone marrow or saline/autograft before use. Intended for surgical implantation by physicians to fill bony voids or gaps in the skeletal system (e.g., posterolateral spine). Acts as a scaffold for bony ingrowth; resorbs over time and is replaced by new bone tissue. Benefits include providing a bioactive, moldable filler that supports natural bone healing processes.
Clinical Evidence
No human clinical data provided. Evidence consists of non-clinical bench testing and an in vivo animal study. Animal study (rabbit posterolateral spine fusion model) compared subject device to Vitoss BA and autograft control, demonstrating equivalent performance. Bench testing included biocompatibility (ISO 10993), physicochemical characterization (density, mineral content, absorbency, pH), viral inactivation, material-mediated pyrogenicity (USP <151>), and endotoxin testing (ANSI/AAMI ST72:2011). Bioactivity confirmed via apatite layer formation in simulated body fluid.
Technological Characteristics
Composition: Anorganic bovine bone mineral (ASTM F1581), 45S5 bioactive glass (ASTM F1538), and type I bovine collagen. Form: Moldable strip or cylindrical matrix. Sterilization: Gamma irradiation (ISO 11137, SAL 10^-6). Biocompatibility: ISO 10993 compliant. Non-pyrogenic. Single-use.
Indications for Use
Indicated for patients requiring treatment of bony voids or gaps not intrinsic to skeletal stability, or as a bone graft extender with autograft. Applicable to surgically created osseous defects or traumatic bone injuries in the skeletal system, including the posterolateral spine.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Predicate Devices
Vitoss® Bioactive Foam Bone Graft Substitute (K083033)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square are the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" below it.
March 21, 2019
Collagen Matrix, Inc. Gloria Zuclich Director, Regulatory Affairs 15 Thornton Road Oakland, New Jersey 07436
Re: K182074
Trade/Device Name: Mineral Collagen Composite Bioactive Moldable Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV Dated: February 18, 2019 Received: February 19, 2019
Dear Ms. Zuclich:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurence D. Coyne -S
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K182074
Device Name
Mineral Collagen Composite Bioactive Moldable
#### Indications for Use (Describe)
Mineral Collagen Composite Bioactive Moldable combined with either autogenous bone marrow or autograft with saline is indicated for bony voids or gaps, that are not intrinsic to the stability of the bony structure; Mineral Collagen Composite Bioactive Moldable can also be used with autograft as a bone graft extender.
The device is to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
| 1. | Applicant Information | | |
|----|----------------------------------------------------------------------|---------------------------------------------------------------------------------|--|
| | Applicant Name: | Collagen Matrix, Inc. | |
| | Address: | 15 Thornton Road<br>Oakland, New Jersey 07436 | |
| | Telephone: | (201) 405-1477 | |
| | Fax: | (201) 405-1355 | |
| | Contact Person: | Gloria Zuclich<br>Director of Regulatory Affairs<br>gzuclich@collagenmatrix.com | |
| | Date Prepared: | March 19, 2019 | |
| 2. | Name of the Device | | |
| | Device Trade Name: | Mineral Collagen Composite Bioactive Moldable | |
| | Device Common Name(s): | Bone Void Filler<br>Bone Graft Matrix<br>Bone Graft Substitute | |
| | Device Classification Name: | Filler, Bone Void, Calcium Compound<br>888.3045<br>MQV<br>Class II | |
| 3. | Legally Marketed Devices to Which Substantial Equivalence is Claimed | | |
| | Primary Predicate Device: | Vitoss® Bioactive Foam Bone Graft Substitute<br>Orthovita, Inc.<br>K083033 | |
| | Predicate Device: | MASTERGRAFT® Strip/Putty<br>Medtronic Sofamor Danek USA, Inc.<br>K140375 | |
### 4. Description of the Device
Mineral Collagen Composite Bioactive Moldable Bone Graft Matrix is composed of anorganic bone mineral, bioactive glass, and type I collagen that can be molded to fit the bone defect. It is an osteoconductive, bioactive, porous implant that allows for bony ingrowth across the graft site. The bone graft matrix is slowly resorbed and replaced by new bone tissue during the natural healing process.
The anorganic bone mineral component of the bone graft matrix is a natural, porous bone graft material produced by removal of all organic components from bovine bone. The composition of the anorganic bone mineral meets ASTM F1581 standard specifications for composition of
{4}------------------------------------------------
anorganic bone for surgical implants. The bioactive glass component of the device is made of 45S5 Bioactive Glass and meets ASTM F1538 standard specifications for glass and glass ceramics biomaterials for implantation. The purified type I collagen is derived from bovine deep flexor Achilles tendon.
The product is available in various sizes and is provided sterile, non-pyrogenic, and for single use only.
#### 5. Intended Use
Mineral Collagen Composite Bioactive Moldable combined with either autogenous bone marrow or autograft with saline is indicated for bony voids or gaps, that are not intrinsic to the stability of the bony structure; Mineral Collagen Composite Bioactive Moldable can also be used with autograft as a bone graft extender.
The device is to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process.
## 6. Summary/Comparison of Technical Characteristics
| Feature | Mineral Collagen<br>Composite Bioactive<br>Bone Graft Matrix<br>(This submission) | VITOSS Bioactive Foam<br>Bone Graft Substitute*<br>K083033 | MASTERGRAFT Strip/Putty<br>K140375 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | Mineral Collagen Composite<br>Bioactive Moldable<br>combined with either<br>autogenous bone marrow or<br>autograft with saline is<br>indicated for bony voids or<br>gaps, that are not intrinsic to<br>the stability of the bony<br>structure; Mineral Collagen<br>Composite Bioactive<br>Moldable can also be used<br>with autograft as a bone<br>graft extender.<br>The device is to be gently<br>packed into bony voids or<br>gaps of the skeletal system<br>(i.e., posterolateral spine).<br>These defects may be<br>surgically created osseous<br>defects or osseous defects<br>created from traumatic injury<br>to the bone. The device<br>resorbs and is replaced with<br>bone during the healing<br>process. | Vitoss Bioactive Foam Bone<br>Graft Substitute is intended for<br>use as a bone void filler for<br>voids or gaps that are not<br>intrinsic to the stability of the<br>bony structure. Vitoss<br>Bioactive Foam is indicated for<br>use in the treatment of<br>surgically created osseous<br>defects or osseous defects<br>created from traumatic injury to<br>the bone.<br>Vitoss Bioactive Foam Bone<br>Graft Substitute is intended to<br>be used for filling bony voids or<br>gaps of the skeletal system<br>(i.e. the extremities, pelvis, and<br>spine, which includes<br>posterolateral fusion<br>procedures), and may be<br>combined with saline,<br>autogenous blood, and/or<br>bone marrow. Following<br>placement in the bony void or<br>gap, the scaffold resorbs and<br>is replaced with bone during<br>the healing process. | MASTERGRAFT® Putty<br>combined with either<br>autogenous bone marrow,<br>and/or sterile water, and/or<br>autograft is indicated as a<br>bone void filler for bony voids<br>or gaps that are not intrinsic to<br>the stability of the bony<br>structure. Additionally,<br>MASTERGRAFT® Putty can<br>be used with autograft as a<br>bone graft extender.<br>MASTERGRAFT® Putty is to<br>be gently packed into bony<br>voids or gaps of the skeletal<br>system (e.g., the posterolateral<br>spine, pelvis, ilium, and/or<br>extremities). These defects<br>may be surgically created<br>osseous defects or osseous<br>defects created from traumatic<br>injury to the bone.<br>MASTERGRAFT® Putty<br>resorbs and is replaced with<br>bone during the healing<br>process. |
| Physical Form | Strip or Cylindrical matrix<br>(moldable upon hydration) | Foam Pack (moldable upon<br>hydration) | Strip or Putty Matrix<br>(Putty is moldable upon) |
{5}------------------------------------------------
| | | Foam Strip | hydration) |
|-------------------------|--------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Color | White to off-white | White to off-white | White to off-white |
| Material<br>Composition | • Calcium phosphate<br>(anorganic bone mineral)<br>• Bioactive glass<br>• Type I bovine collagen | • β-Tricalcium phosphate<br>• Bioactive glass<br>• Type I bovine collagen | • β-Tricalcium phosphate and<br>Hydroxyapatite<br>• Type I bovine collagen |
| Product Sizes | 2.5 cc – 40 cc | 1 cc – 10 cc | 0.75 cc – 43 cc |
| Biocompatibility | Biocompatible<br>ISO 10993 | Biocompatible<br>ISO 10993 | Biocompatible<br>ISO 10993 |
| Sterility | Sterile, SAL 10-6<br>Gamma irradiation<br>ISO 11137 | Sterile, SAL 10-6<br>Gamma irradiation | Sterile, SAL 10-6<br>Gamma irradiation |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Non-pyrogenic |
| Single Use/<br>Reuse | Single use only | Single use only | Single use only |
## Nonclinical Tests Submitted
In vivo and in vitro testing of the subject device was conducted to demonstrate substantial equivalence of the subject device to its predicate devices. The performance of the device in posterolateral spine fusion rabbit model was compared to the performance of the predicate device, Vitoss BA and to the performance of the Autograft control. The results of the animal study demonstrate performance substantially equivalent to the predicate device Vitoss BA and performance substantially equivalent to autograft when used as an autograft extender.
The materials are safe and biocompatible as demonstrated through ISO 10993 Biological Evaluation of medical Devices. The device is considered bioactive based on in vitro testing, that show apatite laver formation on the surface of the bioactive glass particles following immersion in simulated body fluid (SBF). In vitro characterization studies included evaluation of physical properties such as density, mineral content, absorbency and an evaluation of physicochemical properties such as pH. The characterization test results of the subject device were equivalent to those of the predicate devices. Viral inactivation studies were performed to ensure the viral safety of the product.
Material-mediated pyrogenicity testing was performed on the finished device in accordance with USP<151>. Bacterial endotoxin testing performed using the limulus amebocyte lysate (LAL) method showed that the device meets the endotoxin limits in accordance with ANSI/AAMI ST72:2011.
## Conclusions Drawn from Non-clinical Studies
The conclusions drawn from the nonclinical tests demonstrate that the device is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.