K171284 · Prosidyan, Inc. · MQV · Jun 30, 2017 · Orthopedic
Device Facts
Record ID
K171284
Device Name
FIBERGRAFT BG Matrix Bone Graft Substitute
Applicant
Prosidyan, Inc.
Product Code
MQV · Orthopedic
Decision Date
Jun 30, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
FIBERGRAFT® BG Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT® BG Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT® BG Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine.
Device Story
FIBERGRAFT® BG Matrix is a resorbable, osteoconductive bone void filler. Composition includes 45S5 bioactive glass (M-45 granules, MS-45 microspheres) and bovine type I collagen. Device is hydrated with autogenous bone marrow aspirate (BMA) and mixed with autograft in a 1:1 ratio before implantation. Intended for surgical use by physicians to fill non-structural bony voids or gaps in the skeletal system, including the posterolateral spine. Collagen component provides a single-piece form factor for ease of handling during surgery. Device resorbs over time and is replaced by host bone during the healing process. Benefits include providing a scaffold for bone growth in osseous defects.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (physical/chemical properties, biocompatibility per ISO 10993, endotoxin testing) and a prospective animal study. Rabbit study (n=53) evaluated performance in critical-sized cancellous bone defects in the posterolateral spine over 26 weeks. Endpoints included radiographic, histological, histomorphometric, and biomechanical assessments. Results demonstrated substantial equivalence to the predicate and positive control.
Technological Characteristics
Bioactive, osteoconductive, resorbable bone graft substitute. Materials: 45S5 bioactive glass (M-45 granules, MS-45 microspheres) and bovine type I collagen. Form factor: single-piece matrix. Energy source: none. Connectivity: none. Sterilization: not specified. Software: none.
Indications for Use
Indicated for bony voids or gaps not intrinsic to skeletal stability in the skeletal system, specifically the posterolateral spine. Applicable to surgically created osseous defects or traumatic bone injuries. Must be used with autogenous bone marrow aspirate and autograft.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 30, 2017
Prosidyan, Inc. % Ms. Janice M. Hogan Regulatory Counsel Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103
Re: K171284
Trade/Device Name: FIBERGRAFT® BG Matrix Bone Graft Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV Dated: May 1, 2017 Received: May 1, 2017
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
510(k) Number (if known)
K171284
Device Name
FIBERGRAFT® BG Matrix Bone Graft Substitute
Indications for Use (Describe)
FIBERGRAFT® BG Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT® BG Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT® BG Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)
□ Over-The-Counter Use (21 CFR 801
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per
response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
# 510(k) SUMMARY
# Prosidyan, Inc.'s FIBERGRAFT® BG Matrix - Bone Graft Substitute
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Prosidvan. Inc. 30 Technology Drive Warren. NJ 07059 Phone: (610)-945-5640 Facsimile: (908) 396-1151 Contact Person: Charanpreet S. Bagga
Date Prepared: June 29, 2017
#### Name of Device and Name
FIBERGRAFT® BG Matrix Bone Graft Substitute
Common or Usual Name Bone Void Filler
#### Classification Name/CFR Requlation/Product Code
Resorbable Calcium Salt Bone Void Filler, 21 CFR 888.3045, product code MQV
#### Predicate Device
- Stryker/Orthovita Inc, Vitoss Bioactive Bimodal (K103173) .
# Reference Device
- NovaBone Products LLC, Novabone MacroFORM (K140946) ●
# Intended Use / Indications for Use
FIBERGRAFT® BG Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT® BG Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT® BG Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine.
# Device Description
FIBERGRAFT® BG Matrix product is composed of 45S5 bioactive glass components (M-45 granules, MS-45 microspheres) and bovine type I collagen. The product is intended to be hydrated with BMA and mixed with autograft in a 1:1 ratio.
# Technological Characteristics
BG Matrix is a bioactive osteoconductive, resorbable, biocompatible bone graft substitute. The product is composed of 45S5 bioactive glass components (M-45 granules, MS-45 microspheres) and bovine type I collagen. After hydrating with BMA, the BG Matrix must be mixed with autograft.
{4}------------------------------------------------
The same M-45 granules have previously been 510(k) cleared in Prosidyan's FIBERGRAFT BG Morsels (K141956, K132805, and K151154) and FIBERGRAFT BG Putty (K143533). The MS-45 microspheres included in the proposed device have also been cleared in the company's FIBERGRAFT BG Puttv (K143533),
The addition of collagen creates a single piece bone graft in a shape and size that is clinically relevant for ease of use during implantation.
#### Performance Data
The performance of the FIBERGRAFT® BG Matrix has been established by undertaking physical and chemical property evaluation studies, functional performance animal studies and biocompatibility tests. The physical and chemical property studies confirmed the in vitro functionality and bioactivity of the BG Matrix. The in vitro bioactivity test results have not been correlated to clinical performance. The biocompatibility of the FIBERGRAFT® BG Matrix is demonstrated by ISO 10993 testing and the long history of clinical use of the bioactive glass material for the same intended use. In addition, the BG Matrix is composed of the same bioactive glass material and the same type and duration of patient contact as the predicates. Packaging evaluations, shelf life testing and real time aging testing were performed with passing results. Bacterial endotoxin testing was performed using the limulus amebocyte lysate (LAL) method and showed that the device meets the endotoxin limits of established quidelines.
The FIBERGRAFT® BG Matrix product was evaluated in a rabbit study to further support device performance for its indications for use. The FIBERGRAFT® BG Matrix product was compared to its predicate device as well as controls. The animal study evaluated device performance in critical sized cancellous bone in the posterolateral spine of 53 skeletally mature rabbits. The performance was evaluated using radiographic, histological, histomorphometric, and biomechanical data. Testing of the FIBERGRAFT® BG Matrix in the rabbit model is representative of the indications for use and range of anatomical sites proposed for the subject device. The results of the study through 26 week follow up demonstrated that the FIBERGRAFT® BG Matrix device performs substantially equivalently to the predicate device and positive control, and any minor technological differences between the device groups do not raise new types of safety or effectiveness concerns.
Therefore, performance testing demonstrated that the FIBERGRAFT® BG Matrix device functions as intended and meets the requirements of class II bone void fillers as compared to the predicate device.
# Substantial Equivalence
As demonstrated in performance testing, the FIBERGRAFT® BG Matrix has the same intended use and similar indications, technological characteristics, and principles of operation as its predicate and reference devices. The minor technological differences between BG Matrix and its predicate device do not raise any new issues of safety or effectiveness. The data demonstrate that BG Matrix is substantially equivalent to the predicate device.
# Conclusion
FIBERGRAFT® BG Matrix is an osteoconductive, resorbable, biocompatible bone graft substitute composed of bioactive glass, mixed with Type I collagen. The FIBERGRAFT® BG Matrix is substantially equivalent to its predicate device for its intended use as a synthetic bone void filler.
{5}------------------------------------------------
Performance testing, including in vivo data, demonstrated that the device functions as intended
without raising new safety or effectiveness questions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.