K233992 · Precision AI Pty, Ltd. · QHE · Aug 5, 2024 · Orthopedic
Device Facts
Record ID
K233992
Device Name
Precision AI Surgical Planning System (PAI-SPS)
Applicant
Precision AI Pty, Ltd.
Product Code
QHE · Orthopedic
Decision Date
Aug 5, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K233992 · Aug 5, 2024
Precision AI Surgical Planning System (PAI-SPS)
Precision AI Pty, Ltd.
Clinical case series (post-operative CT scans)
A post-market clinical case series of 35 subjects was conducted to evaluate the measurement accuracy of the Precision AI Surgical Planning System via post-operative CT.
Post-market evaluation; Clinical case series; Measurement accuracy
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Post-market evaluation of a clinical case series; Clinical case series
Post-market evaluation of a clinical case series of 35 subjects.
—
Indications for Use
The Precision Al Planning Software is intended to be used as a pre-surgical planner for simulation of surgical interventions for shoulder ioint arthroplasty. The software is used to assist in the positioning of shoulder components by creating a 3D bone construct of the joint and allows the surgeon to visualize, measure, reconstruct, annotate and edit pre-surgical plan data. The software leads to the generation of a surgery report along with a pre-surgical plan data file which can be used as input data to design the Precision Al Shoulder Guide and Biomodels. The Precision Al Planning System Guides and Biomodels are intended to be used as patientspecific surgical instruments to assist in the intraoperative positioning of shoulder implant components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-imaging scans. The Glenoid Guide is used to place the k-wire and the Humeral Guide is used to place humeral pins for humeral head resection. The Precision Al Guides and Biomodels are indicated for single use only. The Precision Al Surgical Planning System is indicated for use on adult patients that have been consented for shoulder joint arthroplasty. Both humeral and glenoid guides are suitable for a deltopectoral approach only. The Precision Al Surgical Planning System is indicated for total and reverse shoulder arthroplasty using the following Enovis implant systems and their compatible components: - AltiVate Anatomic Shoulder System (K162024) - AltiVate Anatomic Augmented Glenoid (K213387) - Turon Shoulder System (K111629) - Reverse Shoulder Prosthesis (K051075, K092873, K111629)
Device Story
System comprises pre-surgical planning software and patient-specific 3D-printed surgical guides/biomodels. Input: preoperative CT scans. Software creates 3D bone constructs; allows surgeon to visualize, measure, and simulate implant positioning. Output: surgery report, plan data file, and CAD files for guides. Guides manufactured via 3D printing (sintering) using biocompatible Polyamide-12. Used in clinical settings by surgeons to assist intraoperative positioning of shoulder components (glenoid k-wire placement, humeral pin placement for resection). Guides are single-use. System facilitates surgical planning and improves accuracy of implant placement via patient-specific anatomical referencing.
Clinical Evidence
Post-market clinical case series of 35 subjects evaluated measurement accuracy via post-operative CT. Non-clinical testing included biocompatibility, dimensional stability, packaging, drop/compression/wear testing, and software V&V. Composite bone model and cadaveric testing performed. Results demonstrate safety and effectiveness comparable to predicates.
Technological Characteristics
System includes software and 3D-printed patient-specific guides. Guides made of biocompatible medical-grade Polyamide-12 (nylon). Software utilizes non-adaptive machine/deep learning for semi-automatic segmentation and landmark identification. Single-use hardware. Deltopectoral approach only.
Indications for Use
Indicated for adult patients consented for total or reverse shoulder joint arthroplasty. Guides are for deltopectoral approach only. Compatible with specific Enovis implant systems.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{1}------------------------------------------------
August 5, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Precision AI Pty Ltd Sara Baniadam Quality and Regulatory Manager Suite 18, 36 Agnes Street Fortitude Valley, Queensland Australia
Re: K233992
Trade/Device Name: Precision AI Surgical Planning System (PAI-SPS) Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: QHE, PHX, KWS Dated: July 5, 2024 Received: July 5, 2024
Dear Sara Baniadam:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image contains the text "Farzana" and "Sharmin -S" stacked on top of each other. The text is written in a simple, sans-serif font and is black. The background is white, and there is a faint watermark in the background.
Digitally signed by Farzana Sharmin -S Date: 2024.08.05 20:01:43 -04'00
Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K233992
Device Name
Precision AI Surgical Planning System (PAI-SPS)
Indications for Use (Describe)
Software
The Precision Al Planning Software is intended to be used as a pre-surgical planner for simulation of surgical interventions for shoulder ioint arthroplasty. The software is used to assist in the positioning of shoulder components by creating a 3D bone construct of the joint and allows the surgeon to visualize, measure, reconstruct, annotate and edit pre-surgical plan data. The software leads to the generation of a surgery report along with a pre-surgical plan data file which can be used as input data to design the Precision Al Shoulder Guide and Biomodels.
Hardware
The Precision Al Planning System Guides and Biomodels are intended to be used as patientspecific surgical instruments to assist in the intraoperative positioning of shoulder implant components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-imaging scans.
The Glenoid Guide is used to place the k-wire and the Humeral Guide is used to place humeral pins for humeral head resection.
The Precision Al Guides and Biomodels are indicated for single use only.
The Precision AI Surgical Planning System is indicated for use on adult patients that have been consented for shoulder joint arthroplasty. Both humeral and glenoid guides are suitable for a deltopectoral approach only.
The Precision Al Surgical Planning System is indicated for total and reverse shoulder arthroplasty using the following Enovis implant systems and their compatible components:
- AltiVate Anatomic Shoulder System (K162024)
- AltiVate Anatomic Augmented Glenoid (K213387)
- Turon Shoulder System (K111629)
- Reverse Shoulder Prosthesis (K051075, K092873, K111629)
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
ver-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Precision AI. The logo consists of a circular target-like design made of dots, with horizontal and vertical lines intersecting at the center. Below the design, the words "PRECISION AI" are written in a simple, sans-serif font.
## 1. Submitter
| 510(k) Sponsor | Precision Al Pty Ltd |
|-----------------------|--------------------------------------------------------------------------|
| Address | Suite 18, 36 Agnes Street, Fortitude Valley, Queensland, 4006, Australia |
| Correspondence Person | Sara Baniadam<br>Quality and Regulatory Manager |
| Contact Information | Email: Regulatory@precisionai.com.au |
| Submission Date | December 18, 2023 |
### 2. Subject Device
| Proprietary Name | Precision Al Surgical Planning System (PAI-SPS) |
|--------------------------|---------------------------------------------------------------------------------|
| Common Name | Surgical Planning Software and Patient Specific Surgical Guide |
| Classification Name | Single/multiple component metallic bone fixation appliances and<br>accessories. |
| Regulation Number | 21 CFR 888.3660 |
| Regulation Name | Shoulder joint metal/polymer semi-constrained cemented prosthesis |
| Product Code | QHE |
| Additional Product Codes | PHX, KWS |
| Regulatory Class | II |
## 3. Predicate Devices
Substantial equivalence is therefore claimed to the following devices:
- > Primary Predicate:
- Materialise Shoulder System (K212569)
- > Secondary Predicate:
- Medacta MyShoulder Placement Guide (K190738)
- A Reference Device:
- Formus Hip (K213272) ●
- . PeekMed Web (K222767)
### 4. Device Description
Precision Al Surgical Planning System (PAI-SPS) is a patient-specific medical device that is designed to be used to assist the surgeon in the placement of shoulder components during total anatomic and reverse shoulder replacement surgery. This can be done by generating a pre-surgical shoulder plan and, if requested
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Precision AI. The logo consists of a circular pattern of dots arranged in concentric circles, with a crosshair in the center. The text "PRECISION AI" is written in a sans-serif font below the circular pattern.
by the surgeon, by manufacturing a patient-specific guide and models to transfer the surgical plan to surgery.
The device is a system composed of the following:
- a software component, Precision Al Surgical Planning Software which will create a 3D construct of the patient's joint for the surgeon to plan the operatively. The surgeon will be able to visualise the movement of the diseased joint and determine mechanical failings. They will then be able to place the virtual shoulder replacement in different positions and decide which position gives the patient the best result. Once the surgeon has decided on the best position, the software will generate a CAD file for a Patient Specific Guide.
- Precision Al Surgical Guides, which are patient-specific guides and models will be manufactured if the surgeon requests patient-specific guides to transfer the surgical plan to surgery. Once the CAD model is generated from the planning software, the model is sent to a 3D printer which will then print the guide out of a biocompatible medical grade Nylon material for sintering (Polyamide-12) which has an established usage for similar application. The specific design of the guide will be customised to the individual patient as well as depending on the particular anatomy it will be applied to. Precision Al Patient Specific Guides are intended for single use only.
### 5. Indications for Use
#### Software
The Precision Al Planning Software is intended to be used as a pre-surgical planner for simulation of surgical interventions for shoulder joint arthroplasty. The software is used to assist in the positioning of shoulder components by creating a 3D bone construct of the joint and allows the surgeon to visualize, measure, reconstruct, annotate and edit pre-surgical plan data. The software leads to the generation of a surgery report along with a pre-surgical plan data file which can be used as input data to design the Precision Al Shoulder Guide and Biomodels.
#### Hardware
The Precision Al Planning System Guides and Biomodels are intended to be used as patient-specific surgical instruments to assist in the intraoperative positioning of shoulder implant components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-imaging scans.
The Glenoid Guide is used to place the k-wire and the Humeral Guide is used to place humeral pins for humeral head resection.
The Precision Al Guides and Biomodels are indicated for single use only.
The Precision Al Surgical Planning System is indicated for use on adult patients that have been consented for shoulder joint arthroplasty. Both humeral and glenoid guides are suitable for a delto-pectoral approach only.
The Precision Al Surgical Planning System is indicated for total and reverse shoulder arthroplasty using the following Enovis implant systems and their compatible components:
- AltiVate Anatomic Shoulder System (K162024)
- AltiVate Anatomic Augmented Glenoid (K213387)
- Turon Shoulder System (K111629)
- Reverse Shoulder Prosthesis (K051075, K092873, K111629)
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo for a company called "PRECISION AI". The logo features a stylized target symbol above the company name. The target symbol is made up of concentric circles of dots, with crosshairs extending from the center.
## 6. Comparison of Technological Characteristics
The Precision AI Surgical Planning System is comprised of Surgical Planning software and physical patientmatched Surgical Guides. Likewise, the primary predicate, Materialise Shoulder System (K212569), comprises of surgery planning software and physical surgical guides.
The secondary predicate, Medacta MyShoulder Placement Guide (K190738), is a patient-matched Surgical Guide only, similar to Precision Al Surgical Guide.
Precision Al Surgical Guide is substantially equivalent to predicate devices, Materialise Shoulder System (K212569) and Medacta MyShoulder Placement Guide (K190738) for what concerns indications for use and contraindications as well as technological characteristics including materials, manufacturing process, device usage, packaging, labelling, and shelf-life.
Precision Al Surgical Planning Software incorporates non-adaptive machine or deep learning algorithms trained for the purpose of semi-automatic segmentation and landmark identification of image scans similar to the listed reference devices.
## 7. Performance Data
The following non-clinical performance studies have been conducted on the subject devices:
Characterization tests:
- Biocompatibility Evaluation
- Dimensional Stability Testing Post Cleaning and Sterilisation
- Packaging and Transportation Testing
- Drop (Impact) Testing ●
- Compression Testing
- Wear (Debris) Testing
- Software Verification and Validation Testing
Performance testing
- Composite Bone Model Testing
- Cadaveric Testing ●
In addition, a post-market evaluation of a clinical case series of 35 subjects was conducted to evaluate the measurement accuracy of the subject device via post-operative CT. The study was performed in Australia under ethics committee approval and according to GCP.
The non-clinical and clinical performance testing indicates that the subject device is as safe, as effective, and performs as well as the predicate device.
### 8. Conclusion
Based on the above information, the Precision Al Surgical Planning System is substantially equivalent to the identified predicate device.
Substantial equivalence has been demonstrated through a comparison of indication for use, design, and technological characteristics, as well as performance evaluations.
None of the technological characteristics differences raised any new question of safety and effectiveness, therefore the Precision Al Surgical Planning System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.