MyShoulder Placement Guides

K190738 · Medacta International S.A. · PHX · Dec 13, 2019 · Orthopedic

Device Facts

Record IDK190738
Device NameMyShoulder Placement Guides
ApplicantMedacta International S.A.
Product CodePHX · Orthopedic
Decision DateDec 13, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3660
Device ClassClass 2

Intended Use

The MyShoulder guides are intended to be used as guides specifically designed for the anatomy of a single patient. The humeral guide is used to define the level and orientation of the humeral head resection cut. The glenoidal guide is used to position and orient the K-wire that will subsequently guide the glenoidal reaming. Both anatomical and reverse total joint configurations are acceptable. Both humeral and glenoid guides are suitable for a delto-pectoral approach only. MyShoulder Patient-Matched Guides are intended for use with Medacta Shoulder System and its cleared indications for use. MyShoulder Patient-Matched Guides are intended for single use only.

Device Story

Patient-specific surgical instruments; input: patient CT scan data; process: 3D surgical planning to create anatomical humeral and glenoidal guides; output: physical guides for humeral resection and K-wire positioning for glenoid reaming; used in clinical setting by surgeons; guides facilitate accurate implant placement for Medacta Shoulder System; benefits: improved surgical precision and alignment; single-use; non-sterile.

Clinical Evidence

No clinical studies were conducted. Evidence consists of non-clinical bench testing, including cadaver lab evaluations for accuracy and positioning variability, surgical evaluation, sterilization effect analysis on geometry, and wear testing.

Technological Characteristics

Material: Medical grade nylon for sintering (Polyamide PA 12). Form factor: Patient-specific anatomical guides. Connectivity: None. Sterilization: Provided non-sterile. Software: Used for 3D planning based on CT data.

Indications for Use

Indicated for patients undergoing shoulder joint replacement surgery (anatomical or reverse total joint configurations) requiring patient-specific humeral resection and glenoid reaming guidance via a delto-pectoral approach.

Regulatory Classification

Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,” (vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 13, 2019 Medacta International SA % Chris Lussier Director, Quality and Regulatory Medacta USA 3973 Delp Street MEMPHIS TN 38118 Re: K190738 Trade/Device Name: MyShoulder Placement Guides Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: PHX, KWS, HSD, MBF Dated: November 13, 2019 Received: November 14, 2019 Dear Chris Lussier: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Michael C. Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) ### K190738 Device Name MyShoulder™ Placement Guides ### Indications for Use (Describe) The MyShoulder guides are intended to be used as guides specifically designed for the anatomy of a single patient. The humeral guide is used to define the level and orientation of the humeral head resection cut. The glenoidal guide is used to position and orient the K-wire that will subsequently guide the glenoidal reaming. Both anatomical and reverse total joint configurations are acceptable. Both humeral and glenoid guides are suitable for a delto-pectoral approach only. MyShoulder Patient-Matched Guides are intended for use with Medacta Shoulder System and its cleared indications for use. MyShoulder Patient-Matched Guides are intended for single use only. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 2.0 510(k) Summary # K190738 #### I. Submitter Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66 Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Date Prepared: Mar 21, 2019 Date Revised: November 13, 2019 #### II. Device | Device Proprietary Name: | MyShoulder™ Placement Guides | |--------------------------|-------------------------------------------------------------------| | Common or Usual Name: | Shoulder Prosthesis System | | Classification Name: | Shoulder joint metal/polymer semi-constrained cemented prosthesis | | Primary Product Code: | PHX | | Secondary Product Code: | KWS, HSD, MBF | | Regulation Number: | 21 CFR 888.3660, 21 CFR 888.3690, and 21 CFR 888.3670 | | Device Classification | II | #### III. Predicate Device Substantial equivalence is therefore claimed to the following devices: - > Primary Predicate: BLUEPRINT Patient Specific Instrumentation (K143374), Wright-Tornier - Reference Device: A - Anatomic Shoulder Prosthesis (K170910), Medacta International SA - Shoulder Reverse System (K170452), Medacta International SA ● - MySpine Pedicle Screw Placement Guides (132788), Medacta International SA ● #### IV. Device Description The MyShoulder" Placement Guides are patient specific surgical instruments that allow for guided shoulder implants placement of both Medacta Shoulder Reverse System (K170452) and Medacta Anatomic Shoulder Prosthesis (K170910) and the additional products cleared under K181826 (Glenoid Polyaxial Non-locking screws), K180089 (Short Humeral Stem) and K171058 (Threaded Glenoid Baseplate). {4}------------------------------------------------ The MyShoulder™ Placement guides uses the patient's CT Scan to perform a 3D planning of the glenohumeral joint replacement surgery. Then, following this planning the humeral and glenoidal guides are created. The MyShoulder™ Placement guides are anatomical guides for a single patient anatomy. The MyShoulder™ Placement guides subject of this submission are comprised of the following products: - 1. MyShoulder™ humeral guide: it will allow to perform an accurate humeral resection - 2. MyShoulder™ glenoidal guide: it will provide a precise alignment reference that will guide the glenoidal reaming and final implant positioning. The MyShoulder™ Placement guides are manufactured from medical grade nylon for sintering (Polyamide PA 12) which is identical to Medacta reference device MySpine Pedicle Screw Placement Guides (132788). The MyShoulder™ Placement guides are single use, external communicating devices with limited (<24 hours) contact duration and are provided in non-sterile version only. The MyShoulder™ Placement guides are substantially equivalent to BLUEPRINT Patient Specific K143374), Wright-Tornier, Medacta reference devices Shoulder Reverse System (K170452) Anatomic Shoulder Prosthesis (K170910), and MySpine Pedicle Screw Placement Guides (132788). #### V. Indications for Use The MyShoulder guides are intended to be used as guides specifically designed for the anatomy of a single patient. The humeral guide is used to define the level and orientation of the humeral head resection cut. The glenoidal guide is used to position and orient the K-wire that will subsequently guide the glenoidal reaming. Both anatomical and reverse total joint configurations are acceptable. Both humeral and glenoid guides are suitable for a delto-pectoral approach only. MyShoulder Patient-Matched Guides are intended for use with Medacta Shoulder System and its cleared indications for use. MyShoulder Patient-Matched Guides are intended for single use only. #### VI. Comparison of Technological Characteristics MyShoulder glenoid guides are substantially equivalent to Wright-Tornier BLUEPRINT Patient Specific Instrumentation (K143374) for what concerns indications for use, contraindications, design, material and manufacturing process. MyShoulder guides have substantially equivalent indications for use and contraindications of the reference predicate devices Medacta Shoulder Reverse System (K170452) and Medacta Anatomic Shoulder Prosthesis (K170910). The technological characteristics including materials, contact duration, software, manufacturing process, device usage, packaging, labeling, and shelf-life of the MyShoulder™ Placement guides are similar to those of the reference device MySpine Pedicle Screw Placement Guides (132788). {5}------------------------------------------------ The safety and effectiveness of the MyShoulder™ Placement guides are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification. #### Performance Data VII. Risks were identified based on the proposed design and testing was conducted to mitigate those risks. Based on the risk analysis, testing was conducted according to written protocols with acceptance criteria. The following performance data was provided in support of the substantial equivalence determination: Non-Clinical Studies: - . Characterization Tests - Cadaver Lab Test Report 1; o - Surgical Evaluation Test Report 2; and o - MyShoulder accuracy and positioning variability cadaver test Test Report 4 o - MyShoulder accuracy test Test Report 5 o - Performance Tests ● - o Effects of the Sterilization on the geometry of the MyShoulder Glenoidal Guide -Test Report 3 - Wear Test, according to RPO-01.028.009 o Clinical Studies: - No clinical studies were conducted. . ## VIII. Conclusion Based on the above information, the MyShoulder™ Placement guides are substantially equivalent to the identified predicate and reference devices. Substantial equivalence has been demonstrated through a comparison of indication for use, design, and technological characteristics, as well as performance evaluations. The MyShoulder™ Placement guides are as safe and effective as the predicate device BLUEPRINT Patient Specific Instrumentation (K143374), Wright-Tornier and the reference devices Medacta Shoulder Reverse System (K170452), Medacta Anatomic Shoulder Prosthesis (K170910) and MySpine Pedicle Screw Placement Guides (K132788).
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%