K232675 · Hivox Biotek, Inc. · NUH · Oct 5, 2023 · Neurology
Device Facts
Record ID
K232675
Device Name
Heating Tens, FT-615
Applicant
Hivox Biotek, Inc.
Product Code
NUH · Neurology
Decision Date
Oct 5, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FT-615 is designed for symptomatic relief and management of chronic pain, and for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities and to strain from exercise or normal household work activities. It is also indicated for temporary relief of pain associated with dysmenorthea (menstrual cramps) when used with over-the-counter pain medication, it also provides a heat function intended to temporarily relief of minor aches and pains.
Device Story
Self-adhesive TENS device; provides electrical nerve stimulation via gel pads; includes integrated heat function. User-adjustable intensity (15 levels); TENS and heat functions operate independently or simultaneously. Intended for OTC use; patient self-administers treatment. Device passes electrical currents through skin via 50mm x 56mm gel pads. Requires physician consultation for TENS regime. Benefits include symptomatic pain relief and management of muscle soreness/cramps. Output is electrical stimulation and thermal energy. Device housing and dimensions differ from predicate, but core technology, PCBA, and software remain consistent.
Clinical Evidence
No clinical test data was used to support the decision of substantial equivalence. Bench testing only.
Technological Characteristics
Self-adhesive TENS device with integrated heat function. 15 intensity levels. Materials biocompatible per ISO 10993-1/5/10. Electrical safety/EMC per ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10. Software validated per IEC 62304. Usability per IEC 60601-1-6 and IEC 62366-1. Shelf life per ASTM F1980-16.
Indications for Use
Indicated for symptomatic relief/management of chronic pain and temporary relief of pain in shoulder, waist, back, neck, upper/lower extremities due to exercise or household strain. Also indicated for temporary relief of dysmenorrhea (menstrual cramps) when used with OTC pain medication. Includes heat function for minor aches/pains.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 5, 2023
Hivox Biotek Inc. Leslie Peng Senior Product Manager 5F., No. 123, Xingde Rd., Sanchong Dist. New Taipei City, 24158 Taiwan
# Re: K232675
Trade/Device Name: Heating Tens, FT-615 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Dated: September 1, 2023 Received: September 1, 2023
Dear Leslie Peng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Robert Kang -S
for Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K232675
Device Name HEATING TENS, FT-615
#### Indications for Use (Describe)
The FT-615 is designed for symptomatic relief and management of chronic pain, and for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities and to strain from exercise or normal household work activities. It is also indicated for temporary relief of pain associated with dysmenorthea (menstrual cramps) when used with over-the-counter pain medication, it also provides a heat function intended to temporarily relief of minor aches and pains.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
| 1. Type of Submission | Special |
|-----------------------|------------------------------------------------------------------------------------|
| 2. Date of Summary | 01.09, 2023 |
| 3. Submitter | HIVOX BIOTEK INC. |
| Address: | 5F., No. 123, Xingde Rd., Sanchong Dist., New<br>Taipei City 24158, Taiwan, R.O.C. |
| Phone: | +886-2-8511-2668 |
| Fax: | +886-2-8511-2669 |
| Contact: | Leslie Peng<br>(leslie.peng@hivox-biotek.com) |
#### 4. Identification of the Subject Device
| Proprietary/Trade name: | HEATING TENS |
|------------------------------|------------------------------------------------------------|
| Models: | FT-615 |
| Classification product code: | NUH |
| Regulation number: | 882.5890 |
| Regulation description: | Transcutaneous electrical nerve stimulator for pain relief |
| Review panel: | Neurology |
| Device class: | II |
## 5. Identification of the Predicate Device
| 510(k) number: | K211403 |
|------------------------------|---------------------------------------------------------------|
| Proprietary/Trade name: | HIVOX OTC Electrical Stimulator |
| Models: | FT610-B |
| Classification product code: | NUH |
| Regulation number: | 882.5890 |
| Regulation description: | Transcutaneous electrical nerve stimulator for<br>pain relief |
| Review panel: | Neurology |
| Device class: | II |
## 6. Device Description
The subject device is a self-adhesive TENS device with 15 adjustable intensity levels for pain relief. Moreover, it also provides a heat function which can be used alone, or in conjunction with the TENS function simultaneously. TENS, Transcutaneous
{4}------------------------------------------------
Electrical Nerve Stimulation, refers to the electrical stimulation of nerves through the skin which is an effective method of pain relief. It can be used for self-treatment. Any symptoms that could be relieved using TENS must be checked by your general practitioner who will also give you instruction on how to carry out a TENS self-treatment regime.
TENS device works by passing electrical currents over the skin via a set of gel pads. As a transfer medium, the gel pads are subject to natural wear and tear, and must be replaced when they stop providing sufficient contact or the main unit no longer sticks to the skin completely. Failure to replace the gel pad may lead to skin irritation as a result of heightened current density in particular areas.
This device is only compatible with the 50 mm x 56 mm gel pads which are the OTC medical device cleared by FDA under K132588, and come with the device.
# 7. Intended Use / Indications for Use of the Device
The subject device is designed for symptomatic relief and management of chronic pain, and for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities and lower extremities due to strain from exercise or normal household work activities. It is also indicated for temporary relief of pain associated with dysmenorrhea (menstrual cramps) when used with over-the-counter pain medication. In addition, it also provides a heat function intended to temporarily relief of minor aches and pains.
## 8. Non-clinical Testing
A series of safety and performance tests, as follows, were conducted on the subject device in accordance with FDA recognized consensus standards and/or guidance:
- Shelf life (ASTM F1980-16)
- Biocompatibility (ISO 10993-1 Edition 4.0, ISO 10993-5 Edition 3.0 and ISO 10993-10 Edition 3.0)
- Software validation (IEC 62304 Edition 1.1)
- Electromagnetic compatibility and electrical safety (ANSI/AAMI ES60601-1:2015/(R)2012, IEC 60601-1-2 Edition 4.0, IEC 60601-1-11 Edition 2.0 and IEC 60601-2-10 Edition 2.1)
- Function test (Guidance Document for Powered Muscle Stimulator 510(K)s. Document issued on: June 9, 1999)
- Usability test (IEC 60601-1-6 Edition 3.1 and IEC 62366-1 Edition 1.0)
{5}------------------------------------------------
All the test results demonstrate the subject device, HEATING TENS (FT-615), meets the requirements of its pre-defined acceptance criteria and intended use, and its substantially equivalent to the predicate device.
# 9. Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
# 10. Comparison of Differences and Substantial Equivalence Determination
The subject device, HEATING TENS (FT-615), submitted in this 510(k) file is substantially equivalent in intended use, design, functions, technology, operation principles, materials, PCBA, software, electrode area and performance to the cleared HIVOX OTC Electrical Stimulator (FT610-B)(K211403). Differences between the devices cited in the following do not raise any new issue of substantial equivalence.
| Subject device | Predicate device | Substantial Equivalence<br>Determination | |
|----------------|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| K232675 | K211403 | | |
| HEATING TENS | HIVOX OTC Electrical<br>Stimulator | N/A | |
| FT-615 | FT610-B | | |
| Housing | | The housing is different but<br>all of them material are same.<br>It does not affect the intended<br>use, biocompatibility or alter<br>the fundamental scientific<br>technology of subject device. | |
| | Dimensions | | Slight different in device<br>dimensions because the<br>design of product appearance.<br>However, it does not affect<br>the usability of the device. |
# 11. Discussion
Based on the comparison information in our submission, we can determine that the subject device is similar to the predicate device in most aspect. The subject device has the same intended use, provided the same functions by the same operating principle as the predicate device, except housing and dimensions. The subject device has same material, design, technology, PCBA, software, electrode area and performance to
{6}------------------------------------------------
demonstrate these differences would not adversely impact the safety and effectiveness of the subject devices. The differences between the subject devices and the predicate device would not raise any problem in substantial equivalence claims.
# 12. Conclusion
After analyzing a series of non-clinical test results we have ensure that all of our design outputs meet the specified requirements of inputs, and also the final product meets the user needs. Thus, we can reasonably believe that the subject devices, HEATING TENS (FT-615), is substantially equivalent to the predicate device in safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.