EASYWELL TENS System TENS and Heat, EASYWELL TENS System TENS only
K163166 · Easywell Biomedicals, Inc. · NUH · Aug 11, 2017 · Neurology
Device Facts
Record ID
K163166
Device Name
EASYWELL TENS System TENS and Heat, EASYWELL TENS System TENS only
Applicant
Easywell Biomedicals, Inc.
Product Code
NUH · Neurology
Decision Date
Aug 11, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The relief of pain associated with sore or aching muscles due to strain from exercise or normal household work activities.
Device Story
Easywell TENS System is a wearable, battery-powered device for OTC pain relief. It consists of a stimulator console, electrode patches, conductive hydrogel pads, and an AC/DC adapter. The device delivers biphasic rectangular electrical pulses to the skin to stimulate nerves; select models (EETTA01, EETTL01) also provide heat up to 41.0°C via carbon fiber elements. The user operates the device via two buttons to adjust intensity (10 levels) and power. A 15-minute session is standard. The device is intended for home use by patients to manage muscle pain. The healthcare provider does not interact with the device output directly; the patient uses the device to achieve symptomatic relief from muscle strain.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, performance specifications, and compliance with recognized safety and performance standards (AAMI NS4, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, ISO 10993).
Indicated for the relief of pain associated with sore or aching muscles due to strain from exercise or normal household work activities in adult patients.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Easywell Biomedicals, Inc. Arthur Chen Quality Assurance Manager Suite 2, 2f, No.9, Jhanye 1 Rd. Hsinchu Science Park Hsinchu. TW
Re: K163166
Trade/Device Name: Easywell TENS System (TENS only), Easywell TENS System (TENS and Heat) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH Dated: July 11, 2017 Received: July 13, 2017
Dear Arthur Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely,
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## K163166 - 510(k) Summary
#### I. SUBMITTER
EASYWELL BIOMEDICALS INC. Suite 2, 2F, No.9, Jhanye 1st Rd., Hsinchu Science Park, 30078 Hsinchu, TAIWAN Phone:+886-3-666-9596 FAX: +886-3-666-9697 Contact Person: Richard Hsieh Date:Oct. 04, 2016
#### II. DEVICE
Trade/Device Name: Easywell TENS System (TENS only), Easywell TENS System (TENS and Heat) Common Name: Transcutaneous Electrical Nerve Stimulator (TENS) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulation Class: Class II Product Code: NUH
### III. PREDICATE DEVICE
Omron: Heat Pain Pro (K160115 )
#### IV. DEVICE DESCRIPTION
The EASYWELL TENS System is a Transcutaneous Electrical Nerve Stimulation (TENS) set consisting of a stimulator console, an electrode patch, gel pads and an AC/DC adaptor. used for pain relief. There are two models available over-the-counter- (1) TENS only, and (2) a combination of TENS and heat for a warming sensation up to 41.0℃ (105.8°F). The TENS session lasts 15 minutes with 3 stages of varying frequencies.
The stimulator console is wearable, weights 32g, with one micro-USB port and one LED light indicator. It has two buttons- "+" button to turn on the stimulator console and to increase the TENS intensity level by 1 (10 intensity levels); "." button to decrease the TENS intensity or turn off the stimulator console when it's pressed for 3 seconds. The green LED light is lit up when the stimulator console is activated. The AC/DC adaptor can be connected to the micro-USB port to recharge the Li-ion battery.
The electrode patch is covered by polyester and nylon blended textile and can be connected
{4}------------------------------------------------
to the stimulator console via connectors. Patches with varying shapes and sizes are available to fit different and anatomical positions. Carbon fiber is embedded between the electrode and textile to provide heat conductance and far infrared radiation (model EETTA00 and EETTL00 to provide TENS and heat).
The gel pads are applied to the electrode surface. They are directly in contact with the skin and made from medical grade electrically conductive hydrogel. Thev comply with ISO 10993-5 biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity and ISO 10993-10 biological evaluation of medical devices- Part 10: tests for irritation and skin sensitization.
The AC/DC adapter is medical graded which complies with IEC 60601-1 and IEC 60601-1-2
#### INTENDED USE V.
The relief of pain associated with sore or aching muscles due to strain from exercise or normal household work activities.
## VI. CONFORM STANDARDS
EASYWELL TENS System conforms to the following standard:
- 1. AAMI NS4: 2013 Transcutaneous electrical nerve stimulators
- 2. IEC 60601-1:2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- 3. IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- 4. IEC 60601-1-11:2015 Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- 5. IEC 60601-2-10: 2012 Medical electrical equipment Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
- 6. ISO 10993-1: 2009 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process
- 7. ISO 10993-5: 2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
- 8. ISO 10993-10: 2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
- 9. IEC 62304: 2006 Medical device software -- Software life cycle processes
- 10. EN/ISO 14971: 2012 Medical devices. Application of risk management to medical devices
{5}------------------------------------------------
11. IEC/EN 62133:2013 Safety Test Standard of Li-Ion Cell and Battery
12. UN38.3 Testing-Safety Precautions for the Transportation of Lithium Batteries
VII. Similarities/Differences of the proposed candidate device when compared to the predicate
Omron Heat Pain Pro (K160115)
The subject EASYWELL TENS System is substantially equivalent to the predicate Heat Pain Pro (K160115) marketed by Omron Healthcare, Inc. with similar indications for use. The subject EASYWELL TENS System is intended for pain relief with sore or aching due to exercise or normal household work activities. The predicate Heat Pain Pro is intended for relief of pain associated with sore or aching, muscles of the lower back, arms, legs, shoulder, or foot due to strain from exercise or normal household work activities.
Both the subject EASYWELL TENS System and predicate Heat Pain Pro are available Over-the-Counter (OTC).
Both the subject EASYWELL TENS System and the predicate Heat Pain Pro have similar design specifications. Both use microprocessor technology to control the device and use Lithium-Ion battery as the power source. Both are rechargeable, can automatic shut off, and use the electrode patch with hydrogel.
Both the subject EASYWELL TENS System and the predicate Heat Pain Pro have similar output parameters. Both have 1 output channel. The subject EASYWELL TENS System provides TENS therapy on models EETTA00 and EETTL00, and, provides TENS therapy with heat on models EETTA01 and EETTL01; the predicate Heat Pain Pro provides TENS therapy and TENS therapy alternating heat. Both have similar heat temperature range, the subject EASYWELL TENS System conducts heat at maximum 41°C on models EETTA01 and EETTL01 and the predicate Heat Pain Pro conducts heat at maximum 43°C. Both have biphasic rectangular waveform shape. The subject EASYWELL TENS System has timer range 15 minutes and it's repeatable to extend the range, the predicate Heat Pain Pro has timer range 30 minutes.
The subject EASYWELL TENS System and the predicate Heat Pain Pro comply with the same standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11 and IEC 60601-2-10. The subject EASYWELL TENS System also complies with AMMI NS4.
The similarities between the subject EASYWELL TENS System and the predicate Heat Pain Pro are compared in table 1.
{6}------------------------------------------------
| Feature | Subject | Predicate | | |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|---------------------------------|
| 510(k) Number | - | K160115 | | |
| Manufacturer | EASYWELL Biomedicals, Inc. | Omron<br>Healthcare, Inc. | | |
| Trade Name | EASYWELL TENS System | Omron Heat<br>Pain Pro | | |
| Model Name | EETTA00<br>EETTL00<br>(TENS Only) | EETTA01<br>EETTL01<br>(TENS and Heat) | PM311 | |
| Indications for Use | The EASYWELL TENS System is<br>intended for:<br>The relief of pain associated with sore or<br>aching muscles due to strain from<br>exercise or normal household work<br>activities. | The Omron<br>Heat Pain Pro is<br>intended for:<br>The relief of<br>pain associated<br>with sore or<br>aching, muscles<br>of the lower<br>back, arms, legs,<br>shoulder, or foot<br>due to strain<br>from exercise or<br>normal<br>household work<br>activities. | | |
| Type of Use | Over-the-Counter (OTC) | Over-the-Count<br>er (OTC) | | |
| Design<br>Specification | Controller | Microprocessor | Microprocessor | |
| | Power Source | Lithium-Ion battery | Lithium-Ion<br>battery | |
| | Rechargeable | Yes | Yes | |
| | Automatic Shut Off | Yes | Yes | |
| | Electrode Patch/<br>Gel Pad | Electrode Patch with Hydrogel | Electrode Patch<br>with Hydrogel | |
| Output<br>Parameter | Number of output<br>channels | 1 Channel | 1 Channel | |
| | Output Modes | TENS Therapy<br>only | TENS Therapy with<br>HEAT | |
| | Heat Temperature | N.A. | 41°C Maximum | |
| | Waveform Shape | Biphasic Rectangular | Biphasic<br>Rectangular | |
| | Timer Range | 15 minutes | 30 minutes | |
| Compliance with Standards | IEC60601-1<br>IEC60601-1-2 | IEC60601-1<br>IEC60601-1-2 | | |
| | | | IEC 60601-1-11<br>IEC60601-2-10<br>AAMI NS4 | IEC 60601-1-11<br>IEC60601-2-10 |
| Performance<br>and<br>Specifications | Max.<br>Output<br>Voltage | @500Ω | 58.8V | 67.2V |
| | | @2KΩ | 94V | 86.0V |
| | | @10KΩ | 119V | 95.9V |
| | Max.<br>Output<br>Current | @500Ω | 117.6mA | 134.2mA |
| | | @2KΩ | 45.5mA | 45.0mA |
| | | @10KΩ | 11.9mA | 9.6mA |
| | Duration of primary phase | | 132μs | 32μs |
| | Pulse Duration@500Ω | | 220μs | 96μs |
| | Frequency | | 2.2 to 34.7Hz | 1 to 99Hz |
| | Net Charge @500Ω (per pulse) | | 0μC | 0μC |
| | Maximum Phase Charge @500Ω | | 15.5μC | 4.3μC |
| | Active Surface Area for each Electrode (cm²) | | 1) EW-P000-G Standard 30<br>2) EW-P000-L Lower Back 60 | 25 |
| | | | | |
| | Maximum Current for each Electrode @500Ω(mA) | | 1) EW-P000-G Standard 117.6<br>2) EW-P000-L Lower Back 117.6 | 134.2 |
| | | | | |
| | Maximum Charge Density for each Electrode @500Ω (mA/cm²) | | 1) EW-P000-G Standard 3.92<br>2) EW-P000-L Lower Back 1.96 | 5.38 |
| | | | | |
| | | | | |
| | Maximum Power Density for each Electrode @500Ω (mW/cm²) | | 1) EW-P000-G Standard 230.5<br>2) EW-P000-L Lower Back 115.2 | 361.3 |
| | | | | |
Table 1 Comparison with the predicate Heat Pain Pro (K160115)
{7}------------------------------------------------
## Substantial Equivalence:
EASYWELL TENS System has the same intended use, principles of operation, and similar technological characteristics as predicate devices. Moreover, bench testing contained in this submission and clinical evaluation report supplied demonstrates that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the EASYWELL TENS System is substantially equivalent to the predicate devices.
{8}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.