ProBeam 360° Proton Therapy System v2.0 (Multiroom)
K231863 · Varian Medical Systems, Inc. · LHN · Oct 6, 2023 · Radiology
Device Facts
Record ID
K231863
Device Name
ProBeam 360° Proton Therapy System v2.0 (Multiroom)
Applicant
Varian Medical Systems, Inc.
Product Code
LHN · Radiology
Decision Date
Oct 6, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ProBeam 360° Proton Therapy System provides protons for precision radiotherapy of lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
Device Story
ProBeam 360° Proton Therapy System v2.0 is a multi-room, compact proton therapy system delivering radiation treatment per physician-prescribed plans. Input: patient treatment plan data. Operation: 226MeV isochronous cyclotron generates protons; beam transport system (steerer, quadrupole, dipole, scanning magnets) directs beam to 2-5 treatment rooms via 90-degree beam ports; beam spot scanning delivers radiation. Output: proton beam for therapy. Used in clinical radiation oncology departments; operated by trained clinicians/technicians. System includes beam scheduler for multi-room management and CBCT/2D imaging for patient positioning. Output affects clinical decision-making by enabling precise dose delivery to target volumes while sparing normal tissue via Bragg peak attenuation. Benefits: compact footprint allows multi-room capacity in smaller facilities.
Clinical Evidence
Bench testing only. No clinical or animal studies were performed. Design verification and validation testing conducted per 21 CFR 820, ISO 13485, ISO 14971, and IEC 62304. Electrical safety and EMC testing performed per FDA-recognized consensus standards.
Technological Characteristics
Isochronous cyclotron (226 MeV) with superconducting coils and helium cryogen cooling. Beam delivery via spot scanning and 360° isocentric gantry. Multi-room configuration (2-5 rooms) with 90-degree beam ports. Patient positioning via 6-axis robotic table. Software-controlled beam scheduling and imaging (2D/CBCT).
Indications for Use
Indicated for patients requiring precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
October 6, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below it.
Varian Medical Systems, Inc. % Peter Coronado Sr. Director, Regulatory Affairs 3100 Hansen Way PALO ALTO CA 94304
# Re: K231863
Trade/Device Name: ProBeam 360 Proton Therapy System v2.0 (Multiroom) Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: LHN Dated: September 6, 2023 Received: September 6, 2023
# Dear Peter Coronado:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Locon Weidner
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
## K231863
Device Name
ProBeam 360° Proton Therapy System v2.0 (Multiroom)
Indications for Use (Describe)
ProBeam 360° Proton Therapy System provides protons for precision radiotherapy of lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "varian" is written in a bold, sans-serif font. Below the logo is the text "A Siemens Healthineers Company" in a smaller font size.
# Premarket Notification - 510(k) Summary
# ProBeam 360° Proton Therapy System version 2.0
#### l. Submitter's Name
Varian Medical Systems 3100 Hansen Way Palo Alto, CA 94304
Contact Name: Dr. Lynn Allman, Senior Director Regulatory Affairs Phone: 650-424-5369 E-mail: submissions.support@varian.com Date Prepared: June 23, 2023
#### II. Device Information
Proprietary Name: ProBeam 360° Proton Therapy System version 2.0 (Multiroom) Common/ Usual Name: Proton Therapy System Classification Name: System, Radiation Therapy, Charged-Particle, Medical Regulation Number: §892.5050 Product Code: LHN
#### III. Predicate Device
ProBeam 360° Proton Therapy System v1.0 (K221791)
#### IV. Reference Device
ProBeam Proton Therapy System v2.0 (K133191)
#### V. Device Description
The ProBeam 360° Proton Therapy System v2.0 (Multiroom) is designed to deliver radiation treatment in accordance with the physician's prescribed treatment plan.
Proton radiation therapy takes advantage of the Bragg peak characteristic of proton attenuation to minimize radiation of normal tissue outside the target volume. Varian markets two lines of proton therapy systems, the standard ProBeam Proton Therapy Systems and the smaller, newer ProBeam 360° Systems. The ProBeam 360° Proton Therapy System line of proton systems includes single-room (cleared in K221791) and multi-room configurations.
The ProBeam 360° System Multiroom (subject device) introduces the multi-room, compact configuration and includes the following primary components:
- Cyclotron (226MeV)
- Beam Transport System (energy selection system and beam transport system)
- Two (2) to five (5) Treatment Rooms: Rotating Isocentric Gantry room with attached scanning nozzle; and a treatment table
- One Treatment Control Room for each treatment room within the chosen configuration
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "varian" is written in a bold, sans-serif font. Below the word "varian" is the text "A Siemens Healthineers Company" in a smaller font. The logo is in black and white.
#### VI. Indications for Use
ProBeam 360° Proton Therapy System provides protons for precision radiotherapy of lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
The indications for use and intended use of the subject device is the same as the predicate device.
#### VII. Comparison of Technological Characteristics with the Predicate Device
The ProBeam 360° System version 2.0 (Multiroom) introduces the multi-room configuration to the ProBeam 360° line of proton therapy systems. The ProBeam 360° System Multiroom is substantially the same as the predicate device, the ProBeam 360° Proton Therapy System v1.0 (ProBeam 360° System v1.0) as cleared in K221791. The systems share the same intended use, compact design, and equivalent technological characteristics. The significant difference between ProBeam 360° System v2.0 and ProBeam 360° System v1.0 is in the configuration of the device.
ProBeam 360° System version 2.0 (Multiroom) adds a multi-room, modular design wherein a customer site may choose to construct a system of two (2) to five (5) treatment rooms with a rotating gantry. The multi-room configuration allows users to deliver proton beam to a maximum of five (5) treatment rooms within one facility, as compared to a single-room system which would require multiple facilities to achieve the same treatment capacity. The multi-room technology has been derived from the reference device, the standard ProBeam Proton Therapy System v2.0 (ProBeam PTS v2.0) as cleared under K133191. The ProBeam 360° System Multiroom combines the compact design of the ProBeam 360° System v1.0 (K221791) with the modular, multi-room design of the ProBeam PTS v2.0 (K133191).
The multi-room configuration adds 14 new risks, modifies three (3) existing risks, triggers verification of two (2) existing risk control measures, and is implemented through three (3) primary design changes:
- Addition of Beam Ports
- . Software change to facilitate beam request scheduling across multi room configuration
- Compact Gantry Room offered as multi-room system (from 2 to 5 treatment rooms)
In addition, the multi-room configuration introduces the following non-significant changes:
- . New Scanning Nozzle System (SNS) High Voltage (HV) power supply due to obsolescence
- . Fixed Defects inherited from the prior ProBeam release
Those primary changes affect four (4) device characteristics as identified in Tables 1 and 2.
- Beam Ports
- Treatment Room Configuration
- Software Release
- Beam Scheduler
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "varian" in a bold, sans-serif font. The letters are all lowercase and black. The word is centered in the image and takes up most of the space. The background is white.
A Siemens Healthineers Company
| | | | Table 1. Comparison of Subject Device to ProBeam 360° System version 1.0 Predicate Device | |
|--------------------------------------|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Device Characteristic | Predicate Device:<br>ProBeam 360° Proton Therapy<br>System v1.0<br>K221791 | Subject Device:<br>ProBeam 360° Proton Therapy<br>System v2.0 | Comparison |
| Intended Use/ Indications for<br>Use | | ProBeam 360° Proton Therapy<br>System provides protons for<br>precision radiotherapy of lesions,<br>tumors, and conditions anywhere<br>in the body where radiation<br>treatment is indicated. | ProBeam 360° Proton Therapy<br>System provides protons for<br>precision radiotherapy of lesions,<br>tumors, and conditions anywhere<br>in the body where radiation<br>treatment is indicated. | Same |
| Radiation Source | | | | |
| | Accelerator | Isochronous Cyclotron | Isochronous Cyclotron | Same |
| | Type of Coils | Superconducting Coils | Superconducting Coils | Same |
| | Cooling Method | Helium Cryogen Cooling | Helium Cryogen Cooling | Same |
| | Treatment Particle | Proton | Proton | Same |
| | Cyclotron Energy | 226 MeV | 226 MeV | Same |
| | Nominal Energy | 69 MeV – 218 MeV | 69 MeV – 218 MeV | Same |
| Beam Delivery | | | | |
| | Gantry<br>Beam Line | Beam line with steerer,<br>quadrupole, dipole and scanning<br>magnets. | Beam line with steerer,<br>quadrupole, dipole and scanning<br>magnets. | Same |
| Beam Transport<br>System | Beam Ports | None | Beam Ports (assembly magnets,<br>beam stop, and beam diagnostic) | Beam ports allow for<br>beam assignment to<br>selected treatment<br>room(s) in a multi-room<br>configuration. Refer to<br>Table 3 for comparison<br>to reference device. |
| Beam Angle Adjustment | | Adjustable: Rotational type<br>isocentric 360° Gantry | Adjustable: Rotational type<br>isocentric 360° Gantry | Same |
| | Beam Delivery | Beam Spot Scanning | Beam Spot Scanning | Same |
| Beam Spot Shape | | Spot size expressed as 1σ (sigma)<br>value of the gaussian profile of the<br>beam in air at isocenter:<br>$σ$ = 3.8 mm ± 15% at 218 MeV<br>$σ$ = 4.4 mm ± 15% at 140 MeV<br>$σ$ = 5.9 mm ± 15% at 69 MeV | Spot size expressed as 1σ (sigma)<br>value of the gaussian profile of the<br>beam in air at isocenter:<br>$σ$ = 3.8 mm ± 15% at 218 MeV<br>$σ$ = 4.4 mm ± 15% at 140 MeV<br>$σ$ = 5.9 mm ± 15% at 69 MeV | Same |
| Beam Field Size | | Max: 25cm (x) x 25cm (y) | Max: 25cm (x) x 25cm (y) | Same |
| | Dose Rate | 2 Gy/l/min | 2 Gy/l/min | Same |
| Physical Characteristics | | | | |
| Treatment Room Configuration | | Single Gantry Room Compact<br>configuration | Two (2) to five (5) gantry rooms | Modular, configurable<br>system allows for<br>multiple treatment<br>rooms in one<br>installation. Refer to<br>Table 3 for comparison<br>to reference device. |
| Patient Positioning | | Leoni Orion 6-AxisRobotic<br>Treatment Table (K160518) | Leoni Orion 6-AxisRobotic<br>Treatment Table (K160518) | Same |
| Maximum Load | | 226 kg (500 lbs) | 226 kg (500 lbs) | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Varian" in a bold, sans-serif font. The letters are black and closely spaced together. The image appears to be a logo or brand name.
**A Siemens Healthineers Company**
| Device Characteristic | Predicate Device:<br>ProBeam 360° Proton Therapy<br>System v1.0<br>K221791 | Subject Device:<br>ProBeam 360° Proton Therapy<br>System v2.0 | Comparison |
|-----------------------------------------|-----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient Position Verification<br>System | 2D and image acquisition CBCT,<br>nozzle mounted | 2D and image acquisition CBCT,<br>nozzle mounted | Same |
| Image Acquisition | Functionality included in PVA /<br>Imaging Supervisor. 2D and CBCT<br>(3D) image acquisition. | Functionality included in PVA /<br>Imaging Supervisor. 2D and CBCT<br>(3D) image acquisition. | Same |
| Software | | | |
| Version | ProBeam 360° version 1.0 | ProBeam 360° version 2.0 | Support multi-room<br>configuration of the<br>system including the<br>Beam Scheduler. Refer<br>to Table 3 for<br>comparison to reference<br>device. |
| Beam Priority | Beam priority can be assigned to<br>the single treatment room | Beam priority is able to be<br>assigned and prioritized in each<br>treatment room prior to beam<br>delivery. | Substantially equivalent<br>to ProBeam 360° v1.0<br>(K221791) |
| Beam Scheduler | Beam request applies only to<br>single room | Beam request must be scheduled<br>among multi-room configuration<br>(2 to 5 treatment rooms) | Allows for beam<br>assignment to selected<br>treatment room(s) in<br>multi-room<br>configuration. Refer to<br>Table 3 for comparison<br>to reference device. |
| Network Connectivity | Remote Monitoring and Limited<br>Remote control | Remote Monitoring and Limited<br>Remote control | Same |
| 2D/3D Match | Integration of 2D/3D algorithm<br>into P2VA | Integration of 2D/3D algorithm<br>into P2VA | Same |
| CBCT Imaging | 3D CBCT reconstruction algorithm | 3D CBCT reconstruction<br>algorithm | Same |
| 3D/3D Match | 3D/3D registration algorithm for<br>patient setup | 3D/3D registration algorithm for<br>patient setup | Same |
### Table 1. Comparison of Subject Device to ProBeam 360° System version 1.0 Predicate Device
## Table 2. Comparison of Subject Device to Reference Device Multi-Room Configuration
| Device Characteristic | Reference Device<br>ProBeam Proton Therapy System v2.0<br>K133191 | Subject Device<br>ProBeam 360° Proton Therapy System<br>v2.0 | Comparison |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for Use | ProBeam Proton Therapy System<br>provides protons for precision<br>radiotherapy of lesions, tumors, and<br>conditions anywhere in the body where<br>radiation treatment is indicated. | ProBeam 360° Proton Therapy System<br>v2.0 provides protons for precision<br>radiotherapy of lesions, tumors, and<br>conditions anywhere in the body where<br>radiation treatment is indicated. | Same as ProBeam PTS v2.0<br>(K133191) |
| Beam Delivery | | | |
| Beam Ports | 60-degree Beam Ports (assembly<br>magnets, beam stop, and beam<br>diagnostic) | 90-degree Beam Ports (assembly<br>magnets, beam stop, and beam<br>diagnostic) | Substantially the same as<br>ProBeam PTS v2.0<br>(K133191). The 90-degree<br>bend facilitates the smaller<br>footprint of the ProBeam<br>360° v2.0. |
| Physical Characteristics | | | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "Varian" is written in a bold, sans-serif font. Below the word "Varian" is the text "A Siemens Healthineers Company" in a smaller font.
| Table 2. Comparison of Subject Device to Reference Device Multi-Room Configuration | | | |
|------------------------------------------------------------------------------------|--|--|--|
|------------------------------------------------------------------------------------|--|--|--|
| Device Characteristic | Reference Device<br>ProBeam Proton Therapy System v2.0<br>K133191 | Subject Device<br>ProBeam 360° Proton Therapy System<br>v2.0 | Comparison |
|---------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Treatment Room<br>Configuration | Two (2) to five (5) rotational type<br>isocentric gantry rooms; or fixed beam<br>rooms | Two (2) to five (5) rotational type<br>isocentric gantry rooms | Substantially the same as<br>ProBeam PTS v2.0<br>(K133191) for the gantry<br>room. Gantry rooms in the<br>ProBeam 360° System<br>have a smaller footprint. |
| Software | | | |
| Beam Scheduler | Beam request must be scheduled<br>among multi-room configuration (2 to 5<br>treatment rooms) | Beam request must be scheduled<br>among multi-room configuration (2 to 5<br>treatment rooms) | Same as ProBeam PTS v2.0<br>(K133191) |
#### VIII. Summary of Performance Testing (Non-Clinical Testing)
The ProBeam 360° System version 2.0 (Multiroom) and its corresponding software version have undergone formal design verification and design validation testing. Design verification and design validation testing demonstrates that the ProBeam 360° System Multiroom performs as intended and meets its essential performance. The software for the subject device is considered to have a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
Hardware and software design verification and design validation testing was performed according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 Quality Management System Standard, ISO 14971 Risk Management Standard, and IEC 62304 Software Life Cycle Process standard. Test results demonstrate conformance to applicable requirements specifications and assure hazard safeguards function properly.
Software design verification and design validation testing was conducted, and documentation is provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 2005). Electrical Safety and Electromagnetic Compatibility (EMC) testing was performed for the ProBeam 360° System Multiroom. The system conforms to FDA recognized consensus standards for electrical safety and electromagnetic compatibility.
No animal studies or clinical tests have been included in this submission.
#### IX. Determination of Substantial Equivalence to the Predicate Device
A subset of technological characteristics and features of the subject device differs from the predicate device. These differences are all considered to be enhancements of the predicate, which facilitate the compact, multi-room design of the subject device, ProBeam 360° System Multiroom.
The intended use and indications for use are the same as the predicate device. Further, there are no changes in the principle of operation of the devices. The Verification and Validation demonstrates that the device is as safe and effective as the predicate. Varian therefore believes the data demonstrates that the ProBeam 360° System v2.0 is substantially equivalent to the predicate device, ProBeam 360° System v1.0 (K221791).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.