K221791 · Varian Medical Systems, Inc. · LHN · Dec 1, 2022 · Radiology
Device Facts
Record ID
K221791
Device Name
ProBeam 360° Proton Therapy System
Applicant
Varian Medical Systems, Inc.
Product Code
LHN · Radiology
Decision Date
Dec 1, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ProBeam 360° Proton Therapy System provides protons for precision radiotherapy of lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
Device Story
ProBeam 360° Proton Therapy System generates ionizing radiation (protons) for precision radiotherapy. System components include an isochronous cyclotron (226 MeV), beam transport system, rotating gantry, and treatment control room. Operated by clinical staff in a treatment facility. Device delivers radiation according to a predetermined treatment plan using beam spot scanning. Includes 6-axis robotic patient positioning table and nozzle-mounted CBCT for image-guided verification. System supports 2D/3D image acquisition, 3D CBCT reconstruction, and 3D/3D registration for patient setup. Remote network connectivity allows service technicians to perform remote service activities. Enables precise dose delivery to target lesions while sparing surrounding healthy tissue.
Clinical Evidence
No clinical data or animal studies included. Evidence consists of bench testing, including design verification and validation, electrical safety, and electromagnetic compatibility (EMC) testing per FDA-recognized consensus standards.
Technological Characteristics
Isochronous cyclotron (226 MeV), superconducting coils, helium cryogen cooling. Beam delivery via spot scanning. 6-axis robotic patient positioning. Imaging: 2D and 3D CBCT. Connectivity: Remote monitoring/service. Software: Major level of concern. Standards: ISO 13485, ISO 14971, IEC 62304, 21 CFR 820.
Indications for Use
Indicated for precision radiotherapy of lesions, tumors, and conditions anywhere in the body requiring radiation treatment.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
December 1, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
Varian Medical Systems, Inc. % Mr. Peter Coronado Sr. Director, Regulatory Affairs 3100 Hansen Way PALO ALTO CA 94304
Re: K221791
Trade/Device Name: ProBeam 360° Proton Therapy System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: LHN Dated: October 28, 2022 Received: October 31, 2022
Dear Mr. Peter Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Lora D.<br>Weidner -S | Digitally signed by Lor<br>D. Weidner -S<br>Date: 2022.12.01<br>19:38:07 -05'00' |
|-----------------------|----------------------------------------------------------------------------------|
|-----------------------|----------------------------------------------------------------------------------|
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221791
Device Name ProBeam 360° Proton Therapy System
Indications for Use (Describe)
The ProBeam 360° Proton Therapy System provides protons for precision radiotherapy of lesions, turnors, and conditions anywhere in the body where radiation treatment is indicated.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------|--|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |
| <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Varian, a Siemens Healthineers company. The word "Varian" is written in a bold, sans-serif font. Below the word "Varian" is the text "A Siemens Healthineers Company" in a smaller font. The logo is simple and modern.
# Premarket Notification - 510(k) Summary
## ProBeam 360° Proton Therapy System version 1.0
K221791
#### l. Submitter's Name
Varian Medical Systems 3100 Hansen Way Palo Alto, CA 94304
Contact Name: Mr. Peter J. Coronado, Senior Director Regulatory Affairs Phone: 650-424-6320 | Fax: 650-646-9200 E-mail: submissions.support@varian.com Date Prepared: 18 June 2022
#### II. Device Information
Proprietary Name: ProBeam 360° Proton Therapy System version 1.0 Common/ Usual Name: Proton Therapy System Classification Name: System, Radiation Therapy, Charged-Particle, Medical Regulation Number: §892.5050 Product Code: LHN
#### III. Predicate Device
ProBeam Proton Therapy System v2.0 (K133191)
#### IV. Device Description
The ProBeam 360° Proton Therapy System is a device that generates ionizing radiation (protons) in order to deliver radiation therapy in accordance with a predetermined treatment plan.
The ProBeam 360° System version 1.0 is available in single-room, compact configuration and includes the following primary components:
- Cyclotron
- Beam Transport System ●
- Treatment Room with a rotating Gantry
- Treatment Control Room
#### V. Indications for Use
ProBeam 360° Proton Therapy System provides protons for precision radiotherapy of lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
#### VI. Comparison of Technological Characteristics with the Predicate Device
At a high-level, the subject device differs from the predicate device as a result of the following characteristics:
- Smaller Gantry
- . Compact single Gantry room configuration
- Compact Beam Transport System ●
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "Varian" is written in a bold, sans-serif font. Below the word "Varian" is the text "A Siemens Healthineers Company" in a smaller font. The logo is simple and modern.
- Reduced-Energy Cyclotron ●
- . Patient Positioning Robot
- CBCT acquisition
- Updated Software Release to support changes listed above ●
## Table 1. Comparison of Subject Device to Predicate Device
| Device | Predicate Device: | Subject Device: | Comparison |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | ProBeam Proton Therapy System<br>(K133191) | ProBeam 360° Proton Therapy<br>System | |
| Intended Use/<br>Indications for Use | ProBeam Proton Therapy System<br>provides protons for precision<br>radiotherapy of lesions, tumors, and<br>conditions anywhere in the body<br>where radiation treatment is<br>indicated. | ProBeam 360° Proton Therapy System<br>provides protons for precision<br>radiotherapy of lesions, tumors, and<br>conditions anywhere in the body where<br>radiation treatment is indicated. | Same |
| Radiation Source | | | |
| Accelerator | Isochronous Cyclotron | Isochronous Cyclotron | Same |
| Type of Coils | Superconducting Coils | Superconducting Coils | Same |
| Cooling Method | Helium Cryogen Cooling | Helium Cryogen Cooling | Same |
| Treatment Particle | Proton | Proton | Same |
| Cyclotron Energy | 250 MeV | 226 MeV | Energy reduced cyclotron to<br>allow compact Gantry design.<br>Substantially equivalent to<br>ProBeam K133191 |
| Nominal Energy | 70 MeV – 240 MeV | 69 MeV – 218 MeV | Reduced maximum beam energy<br>to allow compact Gantry design.<br>Substantially equivalent to<br>ProBeam K133191 |
| Beam Delivery | | | |
| Beam Transport<br>System | Standard beam optical system with<br>quadrupoles and dipole magnets | Standard beam optical system with<br>quadrupoles and dipole magnets | Same |
| Switching System | Yes | None | Single Room system does not<br>require room switching capability<br>as compared to multi-room<br>system ProBeam K133191. |
| Beam Transport<br>Magnets | Yes | Yes | Same |
| Beam Angle<br>Adjustment | Adjustable: Rotating Isocentric Gantry<br>rooms<br>Fixed: Fixed Beam Room | Adjustable: Rotating Isocentric Gantry<br>room | Same as Adjustable in ProBeam<br>K133191 |
| Beam Delivery | Beam Spot Scanning | Beam Spot Scanning | Same |
| Beam Spot Shape | Spot size expressed as 1σ (sigma)<br>value of the gaussian profile of the<br>beam in air at isocenter:<br>$σ$ = 4.0 mm ± 15% at 240 MeV<br>$σ$ = 4.0 mm ± 15% at 140 MeV<br>$σ$ = 5.4 mm ± 15% at 70 MeV | Spot size expressed as 1σ (sigma) value<br>of the gaussian profile of the beam in air<br>at isocenter:<br>$σ$ = 3.8 mm ± 15% at 218 MeV<br>$σ$ = 4.4 mm ± 15% at 140 MeV<br>$σ$ = 5.9 mm ± 15% at 69 MeV | Substantially the same beam<br>characteristics as ProBeam<br>K133191. |
| Beam Field Size | Max: 30cm (x) x 40cm (y) | Max: 25cm (x) x 25cm (y) | Reduced field size as compared<br>to ProBeam K133191. |
| Dose Rate | 2 Gy/l/min | 2 Gy/l/min | Same |
| Physical Characteristics | | | |
| Treatment Room<br>Configuration | Multi-Room | Single Gantry Room Compact<br>configuration | Allows installation in building<br>with small footprint. Single<br>Gantry room is substantially<br>equivalent to Rotating Isocentric |
| Device<br>Characteristic | Predicate Device:<br>ProBeam Proton Therapy System<br>(K133191) | Subject Device:<br>ProBeam 360° Proton Therapy<br>System | Comparison |
| Patient Positioning | Forte 6-AxisRobotic Treatment Table<br>(K122413) | Leoni Orion 6-AxisRobotic Treatment<br>Table (K160518) | Substantially equivalent |
| Maximum Load | 273 Kg (550lbs) | 226 kg (500 lbs) | New patient positioning system<br>as cleared in K160518 supports<br>up to 226 kg. |
| Patient Position<br>Verification System | 2D image acquisition, wall-mounted<br>or nozzle mounted | 2D image acquisition and CBCT, nozzle<br>mounted | Adapted from reference device to<br>allows for 3D image CBCT. Nozzle<br>mounted same as predicate<br>ProBeam K133191. |
| Image Acquisition | Functionality included in P2VA /<br>Imaging Supervisor. Limited to 2D<br>image acquisition. | Functionality included in PVA / Imaging<br>Supervisor. 2D and CBCT (3D) image<br>acquisition. | Allows for 3D image CBCT. See<br>table 2 for comparison to<br>reference device. |
| Version | Version 2.0 | ProBeam 360° version 1.0 | Supports changes for compact<br>Gantry, patient positioning table,<br>and operating system updates. |
| Network<br>Connectivity | None | Remote Monitoring and Limited Remote<br>control | Remote network connectivity<br>enables the Varian service<br>technicians to perform remote<br>service activities. |
| 2D/3D Match | Integration of 2D/3D algorithm into<br>P2VA | Integration of 2D/3D algorithm into<br>P2VA | Same |
| CBCT Imaging | None | 3D CBCT reconstruction algorithm | Allows for 3D image CBCT. See<br>table 2 for comparison to<br>reference device. |
| 3D/3D Match | Not included | 3D/3D registration algorithm for<br>patient setup | Allows for 3D image CBCT. See<br>table 2 for comparison to<br>reference device |
{5}------------------------------------------------
varian
A Siemens Healthineers Company
A Siemens Healthineers Company
### Table 1. Comparison of Subject Device to Predicate Device
## Table 2. Comparison of Subject Device to Reference Device
| Device Characteristic | Reference Device:<br>TrueBeam System<br>(K123291) | Subject Device:<br>ProBeam 360 Proton Therapy<br>System | Comparison |
|-----------------------|------------------------------------------------------|---------------------------------------------------------|------------|
| Imaging | | | |
| Patient Position | 3D CBCT | 3D CBCT | Same |
| Verification System | | | |
| Image Acquisition | Functionality included in PVA/<br>Imaging Supervisor | Functionality included in PVA/ Imaging<br>Supervisor | Same |
| CBCT Imaging | 3D CBCT reconstruction<br>algorithm | 3D CBCT reconstruction algorithm | Same |
| 3D/3D Match | 3D/3D registration algorithm for<br>patient setup | 3D/3D registration algorithm for<br>patient setup | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Varian, a Siemens Healthineers Company. The word "varian" is written in a bold, sans-serif font. Below the word "varian" is the text "A Siemens Healthineers Company" in a smaller, sans-serif font. The logo is black and white.
#### VII. Summary of Performance Testing (Non-Clinical Testing)
Design verification and design validation testing demonstrates that the ProBeam 360° System v1.0 performs as intended and meets its essential performance defined as: the functionality enabling its intended use to deliver the radiation dose as prescribed for treatment.
The software for the subject device is considered to have a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
Hardware and software design verification and design validation testing was performed according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 quality Management System standard, ISO 14971 Risk Management Standard, and IEC 62304 Software Life Cycle Process standard.
Test results showed conformance to applicable requirements specifications and assured hazard safeguards functioned properly. Software design verification and design validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
Electrical safety and electromagnetic compatibility (EMC) testing were performed for the ProBeam 360° System. The system conforms to FDA recognized consensus standards for electrical safety and electromagnetic compatibility.
No animal studies or clinical tests have been included in this pre-market submission.
#### VIII. Determination of Substantial Equivalence to the Predicate Device
A subset of technological characteristics and features of the subject device differs from the predicate device. These differences are all considered to be enhancements of the predicate, which facilitate the compact, single-room design of the subject device, ProBeam 360° System.
The Intended Use and indications for use are substantially the same as the predicate device. Further, there are no changes in the principle of operation of the devices. The Verification and Validation demonstrates that the device is as safe and effective as the predicate. Varian therefore believes the data demonstrates that the ProBeam 360° System v1.0 is substantially equivalent to the predicate device, ProBeam Proton Therapy System v.2.0 (K133191).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.