The MEVION S250-FIT Proton Beam Radiation Therapy Device is intended to deliver proton radiation treatment to patients with localized tumors or any other conditions susceptible to treatment by radiation. When the patient is in the fully seated position, the MEVION S250-FIT is indicated for treatment of patients with localized tumors and other conditions susceptible to treatment by radiation in the sites above the mid-chest or carina.
Device Story
Proton beam radiation therapy system; delivers therapeutic proton beam via pencil beam (spot) scanning. Inputs: treatment plan parameters (prescribed dose/distribution). Transformation: 230 MeV superconducting synchrocyclotron generates beam; beam delivery system ensures prescription accuracy; range shifter plates (boron-carbide ceramic) modulate energy. Output: proton radiation dose to target. Used in clinical settings; operated by trained personnel. Integrates Marie Device (K250970) for upright/perched patient positioning and 3D CT imaging for registration. Features discrete arc planning and remote motion. Healthcare providers use system software to manage treatment delivery; output enables precise tumor irradiation while sparing healthy tissue, potentially improving therapeutic outcomes.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, system/subsystem verification, and treatment planning studies evaluating dosimetric coverage and OAR preservation across various anatomical targets (intracranial, head & neck, thoracic, abdominal, pelvic).
Technological Characteristics
230 MeV superconducting synchrocyclotron; pencil beam scanning modality. Range modulation via boron-carbide ceramic plates. Integrated 3D CT imaging. Patient positioning via 6-axis upright platform. Connectivity: networked system software. Standards: IEC 60601-2-64, IEC 62667, IEC 60601-2-68, IEC 60601-2-44, IEC 60825-1, IEC 60601-1-2. Cybersecurity per ANSI/AAMI SW96:2023.
Indications for Use
Indicated for patients (adults and pediatrics) with localized tumors or conditions susceptible to radiation therapy. When in a fully seated position, indicated for treatment sites above the mid-chest or carina.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
September 12, 2025
Mevion Medical Systems, Inc.
Jason Brown
Director, Quality and Regulatory Affairs
300 Foster St.
Littleton, Massachusetts 01460
Re: K250986
Trade/Device Name: S250-FIT Proton Beam Radiation Therapy Device
Regulation Number: 21 CFR 892.5050
Regulation Name: Medical Charged-Particle Radiation Therapy System
Regulatory Class: Class II
Product Code: LHN
Dated: August 15, 2025
Received: August 15, 2025
Dear Jason Brown:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K250986 - Jason Brown
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250986 - Jason Brown
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250986 | |
| Device Name S250-FIT Proton Beam Radiation Therapy Device | |
| Indications for Use (Describe) The MEVION S250-FIT Proton Beam Radiation Therapy Device is intended to deliver proton radiation treatment to patients with localized tumors or any other conditions susceptible to treatment by radiation. When the patient is in the fully seated position, the MEVION S250-FIT is indicated for treatment of patients with localized tumors and other conditions susceptible to treatment by radiation in the sites above the mid-chest or carina. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250986
Page 1 of 9
Traditional 510(k) Summary
Mevion Medical Systems Inc.
Mevion S250-FIT Proton Beam Radiation Therapy Device
I. Submitter
Mevion Medical Systems Inc.
300 Foster St
Littleton, MA 01460
USA
Contact:
Jason Brown
Jason.Brown@Mevion.com
510(k) Number: K250986
Date: September 10, 2025
II. Device
Device Trade Name: S250-FIT Proton Beam Radiation Therapy Device
Product Classification: Class II
Common Name: Medical charged-particle radiation therapy system
Classification Name: System, Radiation Therapy, Charged-Particle, Medical
Regulation Number: 892.5050
Product Code: LHN
III. Predicate Device
Predicate Device: MEVION S250i Proton Beam Radiation Therapy Device (K172848)
Regulation Number: 892.5050
Product Code: LHN
{5}
K250986
Page 2 of 9
## IV. Reference Device
Reference Device: P-Cure Proton Beam Therapy System (K221996)
Regulation Number: 892.5050
Product Code: LHN
## V. Indications for Use
The MEVION S250-FIT Proton Beam Radiation Therapy Device is intended to deliver proton radiation treatment to patients with localized tumors or any other conditions susceptible to treatment by radiation. When the patient is in the fully seated position, the MEVION S250-FIT is indicated for treatment of patients with localized tumors and other conditions susceptible to treatment by radiation in the sites above the mid-chest or carina.
## VI. Device Description
The MEVION S250-FIT Proton Beam Radiation Therapy Device is a proton beam radiation therapy system that provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose and three-dimensional dose distribution to the prescribed patient treatment site. The MEVION S250-FIT delivers radiation via a pencil beam (spot) scanning modality. In order to reach a target depth of 32cm in the patient, the accelerator is designed to produce a 230MeV beam.
The S250-FIT is comprised of 6 subsystems:
- Beam Generation System – generates the beam and directs it to the beam delivery system.
- Beam Delivery System – ensures that the therapeutic prescription parameters are properly delivered.
- Hardwired Safety System (HSS) – provides for system and beam delivery interlocking without the use of software
- Patient Positioning System – The Marie Device from Leo Cancer Care (K250970) allows for accurate and efficient positioning of the patient in a seated or perched position for treatment using an Upright Patient Positioner and 3D CT Scanner for Treatment Planning and Patient Registration.
- Structural Support/Alignment System – supports the beam generation and delivery systems and allows the fixed beam delivery to the single point in space (i.e., the Isocenter)
{6}
K250986
Page 3 of 9
- System Software – controls the above subsystems (except the HSS) and provides interfaces to the system for the end-user.
## VII. Comparison of Technological Characteristics with the Predicate
The Subject device is substantially equivalent to the previously cleared predicate device. Both systems provide a therapeutic proton beam for clinical treatment and have the same indications for use. Both systems are designed to deliver a proton beam from a treatment plan with a prescribed dose and dose distribution to a prescribed patient-based target. The systems have comparable specifications, with differences in the beam angle adjustment, patient positioner, and imaging. These differences between the systems do not raise different questions of safety or effectiveness and can be deemed substantially equivalent to the predicate device. The subject device positions the patient in a seated and perched position with the Leo Cancer Care Marie (K250970) device compared to the supine position used with the predicate system. These differences do not raise questions of safety or effectiveness and testing demonstrates substantially equivalent performance compared to the predicate. A table comparing the key features of the subject and predicate devices is provided below.
| Device Feature | Subject Device | Predicate Device | SE Comparison |
| --- | --- | --- | --- |
| Trade Name | S250-FIT Proton Beam Radiation Therapy Device | S250i Proton Beam Radiation Therapy Device | - |
| Manufacturer | Mevion Medical Systems, Inc. | Mevion Medical Systems, Inc. | - |
| 510(k) | K250986 | K172848 | - |
{7}
K250986
Page 4 of 9
| Intended Use/Indications | The MEVION S250-FIT Proton Beam Radiation Therapy Device is intended to deliver proton radiation treatment to patients with localized tumors or any other conditions susceptible to treatment by radiation. When the patient is in the fully seated position, the MEVION S250-FIT is indicated for treatment of patients with localized tumors and other conditions susceptible to treatment by radiation in the sites above the mid-chest or carina. | The MEVION S250i System is intended to deliver proton radiation therapy treatment to patients with localized tumors or other conditions susceptible to treatment by radiation. | The difference is that the S250-FIT Proton Beam Radiation Therapy Device includes a clarification, which specifies positional limitations for treatment sites for patients in the ‘fully seated’ position. This clarification does not alter the intended therapeutic use of the device. This clarification does not affect the safety and effectiveness of the device when used as labeled. |
| --- | --- | --- | --- |
| Patient Population | Adults, Pediatrics | Adults, Pediatrics | The Subject Device and the Predicate include both Adults and Pediatrics. |
| Proton Accelerator | 230 MeV superconducting synchrocyclotron | 230 MeV superconducting synchrocyclotron | The subject device and predicate device are identical, having same particle type, energy level, acceleration mechanism, beam delivery, magnetic field generation, and operational mode. |
| Usable Energy Range | 30 - 230 MeV | 30 - 230 MeV | The subject device and predicate device have identical usable energy range. |
| Beam Delivery Modality | Pencil Beam Scanning | Pencil Beam Scanning | The subject device and predicate device both utilize pencil beam scanning. |
| Beam Range in patient | 0.8 g/cm² – 32.2 g/cm² | 0.8 g/cm² – 32.2 g/cm² | The subject device and predicate device have identical beam range in patients. |
{8}
K250986
Page 5 of 9
| Dose Rate | 2Gy/Min | 2Gy/Min | The dose rate for both the subject device and predicate device is 2Gy/Min. |
| --- | --- | --- | --- |
| Spot Distal Dose Falloff | The 80% - 20% distal dose falloff of any pristine Bragg peak in water is 0.5 cm or less | The 80% - 20% distal dose falloff of any pristine Bragg peak in water is 0.5 cm or less. | The Spot Distal Dose Falloff is identical for both the subject device and the predicate device. |
| Beam Spot Size | ≤ 0.5 cm for maximum energy (depth) and no greater than 1.5 cm for minimum energy (depth) | ≤ 0.5 cm for maximum energy (depth) and no greater than 1.5 cm for minimum energy (depth) | The beam spot size for the predicate and subject devices are identical. |
| Spot Range Accuracy | Within ± 0.1 g/cm² over all operational parameters of the system. | Within ± 0.1 g/cm² over all operational parameters of the system. | The spot range accuracy is identical for the predicate and the subject device. |
| Range Selection | Boron-carbide ceramic Range Shifter Plates | Polycarbonate Range Shifter Plates | There are minor changes to the range shifter from the S250i to the S250-FIT. Changes include modification to the Range Shifter plate material from polycarbonate to boron-carbide ceramic (with exception to two smallest plates). |
| Energy Modulation | Boron-carbide ceramic Range Shifter Plates | Polycarbonate Range Shifter Plates | There are minor changes to the range shifter from the S250i to the S250-FIT. Changes include modification to the Range Shifter plate material from polycarbonate to boron-carbide ceramic (with exception to two smallest plates). |
{9}
K250986
Page 6 of 9
| Patient Positioner | 6-axis Platform Mounted Upright Positioner | Multi-Axis Supine Positioner | The Subject Device and the predicate both feature multi-axis patient positioners. The Subject Device positions the patient upright. The Predicate Device positions the patient supine. |
| --- | --- | --- | --- |
| Maximum Load | 330 lbs | 400 lbs | The subject device and predicate device have differing maximum loads. |
| Imaging | Integrated CT supplied by Leo Cancer Care, Ltd. | 2D X-ray | Both subject and predicate devices utilize X-ray Imaging for Image Guided Radiotherapy. |
| Discrete Arc Planning | Yes | Yes; but not automatic sequencing of fields | Discrete Arc Planning is a new feature within the S250-FIT that was not present in the predicate device. |
| Remote Motion | Yes | No | Remote motion is a new feature within the S250-FIT device that was not present within the predicate device. |
## VIII. Summary of Performance Testing (Non-Clinical)
System and subsystem verification testing was performed to address aspects of the device design that have changed or that may have been impacted by the changes to the device design.
For subsystems or components that have been modified or are impacted by the S250-FIT configuration, Mevion repeated the original test protocols used for the Mevion S250i for testing S250-FIT to demonstrate substantially equivalent safety and performance.
Treatment Plans evaluating the clinical adequacy of the subject device and the predicate were evaluated for Intracranial, Head & Neck, Thoracic, Abdominal and Pelvic
{10}
K250986
Page 7 of 9
(for perched and max-perched positions) targets. Radiation beam delivery was determined not to be compromised by organ positional or physiological changes resulting from upright patient orientation. Treatment Plans were reviewed for dosimetric coverage, preservation of OARs and overall clinical acceptability. The results of this study demonstrated substantial equivalence.
Standards Testing was completed in accordance with the following FDA-recognized standards:
## Ion Beam
- IEC 60601-2-64 Edition 1.0 2014-09, Particular requirements for the basic safety and essential performance of light ion beam medical electrical equipment
- IEC 62667 Edition 1.0 2017-08, Medical electrical equipment - Medical light ion beam equipment – Performance characteristics
## X-Ray/CT
- IEC 60601-2-68:2014 Particular requirements for the basic safety and essential performance of X-ray-based image-guided radiotherapy equipment for use with electron accelerators, light ion beam therapy equipment and radionuclide beam therapy equipment
- IEC 60601-2-44:2016 Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography
- IEC 60601-1-3:2008 + A1:2013+A2:2021 General Requirements for radiation protection in diagnostic x-ray equipment
## Lasers
- IEC 60825-1 Edition 3, Safety of Laser Products – Part 1 Equipment Classification and Requirements
Interoperability testing was conducted following FDA Guidance, Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices - Guidance for Industry and Food and Drug Administration Staff, Issued September 2017. ANSI/AAMI/UL 2800-1:2022 including parts -1,-2, and -3 will be consulted for guidance.
Cybersecurity testing was conducted in accordance with ANSI/AAMI SW96:2023 and follow FDA Draft Guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions Draft Guidance for Industry and Food and Drug Administration Staff, Issued April 8, 2022.
{11}
K250986
Page 8 of 9
Software development and testing was performed in accordance with IEC 62304 Edition 1.1 2015-06 and FDA Guidance Document, Content of Premarket Submissions for Device Software Functions - Guidance for Industry and Food and Drug Administration Staff, Issued June 14, 2023.
Human Factors and Usability Testing was conducted according to IEC 62366-1:2015+A1:2020, IEC 60601-1-6 Edition 3.2 2020-07, FDA Guidance, Applying Human Factors and Usability Engineering to Medical Devices - Guidance for Industry and Food and Drug Administration Staff, Issued February 2016 and FDA Draft Guidance, Content of Human Factors Information in Medical Device Marketing Submissions.
## System Testing
System level standards testing was conducted to show system compliance to:
- ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 [Incl. AMD2:2021], Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
- IEC 60601-1-2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-2-64:2014 Medical electrical equipment - Part 2-64: Particular requirements for the basic safety and essential performance of light ion beam medical electrical equipment
## Clinical Testing
- Not applicable.
## IX. Substantial Equivalence Conclusion
The Mevion S250-FIT Proton Beam Radiation Therapy Device is substantially equivalent to the previously cleared S250i Proton Beam Radiation Therapy system. Both systems provide a therapeutic proton beam for clinical treatment and have the same intended use. Both systems are designed to deliver a proton beam set with a prescribed dose and dose distribution to a prescribed patient-based target, consistent with the facility's treatment planning system (TPS). The systems have the same beam performance specifications and utilize the same accelerator. Differences between the systems do not raise different questions of safety or effectiveness. To demonstrate substantial equivalence as well as safety and effectiveness, the same type of testing conducted to support the cleared S250i has been performed to verify and validate the Mevion S250-
{12}
K250986
Page 9 of 9
FIT Proton Beam Radiation Therapy Device. Thus, the Mevion S250-FIT Proton Beam Radiation Therapy System is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.