K250970 · Leo Cancer Care · JAK · Jul 25, 2025 · Radiology
Device Facts
Record ID
K250970
Device Name
Marie
Applicant
Leo Cancer Care
Product Code
JAK · Radiology
Decision Date
Jul 25, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The Marie Imaging System is indicated for the acquisition of CT images and the precise positioning of human patients to facilitate delivery of external beam radiation when integrated with a separate therapy treatment delivery device.
Device Story
Marie System is a CT imaging and patient positioning system for upright radiotherapy. It acquires CT images and provides 6-degree-of-freedom positioning to facilitate external beam radiation therapy. The system consists of a CT scanner (helical scan capability) and a chair-based patient support system. It is operated by healthcare professionals in a clinical setting. The device images patients in an upright position, enabling precise treatment planning and reproducible patient setups. The system translates and rotates around the patient to acquire helical scans. Output consists of CT images used by clinicians for treatment planning and image-guided radiation therapy. By enabling upright positioning, the device offers an alternative to conventional supine treatment setups, potentially improving workflow and patient comfort for specific cancer sites.
Clinical Evidence
Bench testing only. Performance evaluated against AAPM Report No. 233 (HU accuracy, uniformity, geometric efficiency), AAPM TG-179 (resolution), and AAPM TG-66 (contrast-to-noise, spatial integrity). Compliance with IEC 60601-1, 60601-2-44, 60601-2-68, and other consensus standards confirmed basic safety and essential performance.
Technological Characteristics
Computed tomography X-ray system; 84 cm bore diameter; 6-degree-of-freedom servo-motor-based patient positioning system; chair-based patient support; helical scan capability; 120kV, 250mA operation. Software developed per IEC 62304 (Class C). Connectivity and integration with external therapy delivery devices. Complies with IEC 60601-1, 60601-2-44, and 60601-2-68.
Indications for Use
Indicated for human patients requiring CT imaging and precise positioning to facilitate external beam radiation therapy. Used by healthcare professionals for upright patient imaging, treatment planning, and immobilization.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 25, 2025
Leo Cancer Care
% Nickie Power
Head of Regulatory and Quality
7921 UW Health Court
MIDDLETON, WI 53562
Re: K250970
Trade/Device Name: Marie
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: JAK
Dated: March 31, 2025
Received: June 30, 2025
Dear Nickie Power:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250970 - Nickie Power
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250970 - Nickie Power
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250970 | |
| Device Name | |
| Marie | |
| Indications for Use (Describe) | |
| The Marie Imaging System is indicated for the acquisition of CT images and the precise positioning of human patients to facilitate delivery of external beam radiation when integrated with a separate therapy treatment delivery device. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Traditional 510(k) SUMMARY
Leo Cancer Care Marie System
I. SUBMITTER
Leo Cancer Care Inc
7921 UW Health Court
Middleton WI 53562
USA
Contact: Nickie Power
Nickie.Power@leocancercare.com
Establishment Registration # 3017500061
510(k) Number: K250970
Date: 28 March 2025
II. DEVICE
Trade Name: Marie
Product Classification: Class II
Common Name: CT Imaging and Patient Positioning System (for use with external Beam Treatment delivery system)
Regulation Number: 21 CFR 892.1750
Regulation Description: Computed tomography x-ray system
Product Code: JAK
III. PREDICATE
Predicate device: P-ARTIS K160611
Regulation Number: 21 CFR 892.1750 (Computed Tomography X-ray System)
Product Code: JAK
{5}
IV. INDICATIONS FOR USE
The Marie Imaging System is indicated for the acquisition of CT images and the precise positioning of human patients to facilitate delivery of external beam radiation when integrated with a separate therapy treatment delivery device.
V. DEVICE DESCRIPTION
The Marie System is intended to acquire CT images and enable the precise positioning of human patients to facilitate delivery of external beam radiation when integrated with a separate therapy treatment delivery device.
The Marie System is intended to be used by healthcare professionals to image patients in an upright position rather than conventional supine treatments, to enable precise treatment planning and patient positioning for radiotherapy.
Specifically, it is intended to:
- Image the patient to provide image-guided radiation therapy
- Image the patient to acquire images for the purpose of treatment planning
- Immobilize patients in an upright position for upright radiotherapy.
The Marie System is comprised of two major sub-systems: a computed tomography (CT) imaging system that performs pretreatment imaging and treatment simulation in the upright positions and a beam agnostic, patient positioning system that supports the patient in the upright positions.
The Marie Imaging System is used with compatible devices for treatment delivery and patient immobilization. The positioning system is designed with six degrees of freedom of motion and a patient positioning system to provide the desired posture for each cancer site to achieve accurate, reproducible patient setups, while the imaging system acquires helical scans by translating and rotating up the patient.
{6}
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE
| Characteristic | Leo Cancer Care Marie system | P-Cure P-ARTIS System – K160611 |
| --- | --- | --- |
| Intended Use | The Marie Imaging System is indicated for the acquisition of CT images and the precise positioning of human patients to facilitate delivery of external beam radiation when integrated with a separate therapy treatment delivery device. | The P-ARTIS CT scanner is a Computed Tomography X-Ray System intended to produce images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. The system may include signal analysis and display equipment, patient, and equipment supports, components and accessories. The system is intended for scanning patients while seated |
| Indications for Use | The Marie Imaging System is indicated for the acquisition of CT images and the precise positioning of human patients to facilitate delivery of external beam radiation when integrated with a separate therapy treatment delivery device. | The P-ARTIS CT scanner is a Computed Tomography X-Ray System intended to produce images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. The system may include signal analysis and display equipment, patient, and equipment supports, components and accessories. The system is intended for scanning patients while seated |
| CT scanner | Acquire CT images of patients to facilitate delivery of external beam radiation | Acquire CT images of patients to facilitate delivery of external beam radiation |
| Image Acquisition | Axial and Multi-slice helical scanning, 2D/3D reconstruction | Axial and Multi-slice helical scanning, 2D/3D reconstruction |
| Imaging system central bore diameter | 84 cm | 85 cm |
| Image resolution at 120kV, 250mA | 4.0 lp/cm @ 50% MTF | 4.3 lp/cm @ 50% MTF |
| | 7.9 lp/cm @ 10% MTF | 8.0 lp/cm @ 10% MTF |
| | 10.0 lp/cm @ 2% MTF | 15.0 lp/cm @ 0% MTF |
| Image Slice Thickness | 0.625, 1.0, 2.0, 4.0, 5.0 mm | 2.0, 3.0, 4.0, 5.0, 8.0 mm |
| Field of View | 62cm | 60cm |
| Imaging Angle | 0-15 degrees | 20 degrees |
| Gantry Scan Plane | Up and down over patient | Up and down over patient |
{7}
| Patient load (maximum) | 150 kg | 375 kg |
| --- | --- | --- |
| Patient Positioning Motion axis | 6 degrees of freedom | Same as predicate |
| Motion drives | Computer controlled, servo motor based multi axis motion system. | Robotic positioning system (Computer controlled, servo motor system) |
| Patient orientation | Chair based patient support | Patient table couch or other approved patient support device |
| Speed | X: 0 – 50mm/sec
Y: 0 – 50mm/sec
Z: 0 – 25mm/sec
Pitch & Roll: 2 degrees per sec
360 rotation: 6 degrees per sec | 0.1 m/s and 6 degrees per sec |
| Accuracy | X: ±0.3mm
Y: ±0.3mm
Z: ±0.7mm
Iso-rotation: ±0.1deg.
Pitch & Roll: ±0.5deg | +/-0.5mm and +/- 0.2 degrees |
| Rotation | Full 360-degree rotation | Full 360-degree rotation |
| Patient support range of pitch | The backrest has 15-degrees from vertical | Support table has +/- 15 degrees of pitch |
{8}
VII. SUMMARY OF PERFORMANCE TESTING (NON-CLINICAL)
Design verification and performance testing were carried out in accordance with design controls of FDA’s Quality System Regulation (21 CFR §820.30), applicable ISO 13485 Quality Management System requirements, ISO 14971 risk management requirements, IEC 62304 requirements for software life-cycle processes, and FDA guidance1.
Non-clinical testing was performed to evaluate device performance and functionality against design and risk management requirements at sub-system and system levels. Leo Cancer Care utilised the test methods and performance criteria outlined in the table below to demonstrate performance to current guidance.
| Source | Performance Requirements |
| --- | --- |
| AAPM REPORT NO. 233 - Performance Evaluation of Computed Tomography Systems | - HU Accuracy
○ Water: ± 7 HU
○ Acrylic: ± 10 HU
○ Polyethylene: ± 10 HU
○ Bones: ± 60 HU
○ Air: ± 17 HU
- Image uniformity ± 15 HU
- Geometric Efficiency >70% |
| Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179 | - Image resolution Line pairs: ≥10 lp/cm |
| Quality assurance for computed-tomography simulators and the computed tomography-simulation process: Report of the AAPM Radiation Therapy Committee Task Group No. 66 | - Contrast to Noise: ≥1.0
- Spatial Integrity (in-plane): ± 1 mm over at least 100 mm |
Software verification testing was conducted and documented in accordance with FDA guidance for “Enhanced Documentation”, (Class C per IEC 62304).
Basic safety and essential performance have been satisfied through conformance with recognized consensus standards for medical devices listed below.
{9}
| Standard | Title |
| --- | --- |
| ISO 14971:2019 Third Edition 2019-12 | Medical Devices – Application of risk management to medical devices |
| IEC 60601-1: Edition 3.2, 2020 | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance |
| IEC 60601-2-68: Edition 1.0, 2014 | Medical electrical equipment - Part 2-68: Particular requirements for the basic safety and essential performance of X-ray-based image-guided radiotherapy equipment for use with electron accelerators light ion beam therapy equipment and radionuclide beam therapy equipment |
| IEC 60601-2-44: Edition 3.2, 2016 | Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography |
| IEC 60601-1-2 Edition 4.1 (2020-09) | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -- Requirements and tests |
| IEC 62304:2016 | Software life-cycle processes |
| IEC 60601-1-6:2020 Edition 3.2 | Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability |
| ANSI AAMI IEC 62366-1:2020 Edition 1.1 | Medical Devices – Part 1: Application of Usability Engineering to Medical Devices |
VIII. SUBSTANTIAL EQUIVALENCE CONCLUSION
The Marie System is substantially equivalent to the predicate devices in terms of intended use and technological characteristics. The devices safety and performance have been addressed by non-clinical testing in conformance with pre-determined performance criteria, FDA guidance, and recognized consensus standards.
The differences between Marie System and the predicate device raise no new efficacy or safety concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.