K230358 · Eneura, Inc. · OKP · May 16, 2023 · Neurology
Device Facts
Record ID
K230358
Device Name
SAVI Dual (TM) Migraine Therapy
Applicant
Eneura, Inc.
Product Code
OKP · Neurology
Decision Date
May 16, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5808
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SAVI Dual™ Migraine Therapy device is indicated for the acute and prophylactic treatment of migraine headache in adolescents (age 12 and older) and adults.
Device Story
Portable, hand-held transcranial magnetic stimulator (sTMS) for migraine treatment. Device delivers brief 0.9 Tesla magnetic pulse to occipital cortex to induce electrical current; stops or lessens migraine effects. Used in home or office settings; self-administered by patient under physician prescription. Device operation requires authorization via inserted SIM chip or wireless cellular connection for a programmed duration corresponding to prescribed months of use. Output is magnetic stimulation; healthcare provider uses device to manage migraine frequency/severity. Benefits include non-invasive, patient-controlled therapy.
Clinical Evidence
Bench testing only. Performance testing verified magnetic pulse characteristics (0.9 Tesla @ 180 µs), field maps, and 5 Gauss line location against predicate. Software verification/validation confirmed delivery of equivalent magnetic pulses. Electrical safety and EMC testing performed per IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11.
Technological Characteristics
Polycarbonate case; integral coil. Energy source: rechargeable lithium-ion battery (100-240V AC input). Output: 0.9 Tesla peak magnetic field @ 180 µs. Connectivity: SIM chip or wireless cellular. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11. Dimensions: 22.4 x 13 x 6.9 cm; 1.4 kg.
Indications for Use
Indicated for acute and prophylactic treatment of migraine headache in adolescents (age 12+) and adults.
Regulatory Classification
Identification
A transcranial magnetic stimulator device for headache is a device that delivers brief duration, rapidly alternating, or pulsed, magnetic fields that are externally directed at spatially discrete regions of the brain to induce electrical currents for the treatment of headache.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing must demonstrate electromagnetic compatibility, electrical safety, and thermal safety.
(2) Appropriate verification, validation, and hazard analysis must be performed on the device software and firmware.
(3) The elements of the device that contact the patient must be assessed to be biocompatible.
(4) Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. This includes full characterization of the magnetic pulse output and resulting magnetic field map. This also includes characterization of the sound level of the device during use.
(5) Clinical testing must demonstrate that the device is safe and effective for treating headache in the indicated patient population.
(6) The physician and patient labeling must include the following:
(i) A summary of the clinical performance testing, including any adverse events and complications.
(ii) The intended use population in terms of the types of headaches appropriate for use with the device.
(iii) Information on how to report adverse events and device malfunctions.
(iv) A diagram or picture depicting the proper placement of the device on the user.
1. Appropriate analysis/testing must demonstrate electromagnetic compatibility (EMC), electrical safety, and thermal safety.
2. Appropriate verification, validation, and hazard analysis must be performed on the device software and firmware.
3. The elements of the device that contact the patient must be assessed to be biocompatible.
4. Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. This includes full characterization of the magnetic pulse output and resulting magnetic field map. This also includes characterization of the sound level of the device during use.
5. Clinical testing must demonstrate that the device is safe and effective for treating headache in the indicated patient population.
6. The physician and patient labeling must include the following:
a. A summary of the clinical performance testing, including any adverse events and complications.
b. The intended use population in terms of the types of headaches appropriate for use with the device.
c. Information on how to report adverse events and device malfunctions.
d. A diagram or picture depicting the proper placement of the device on the user.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
May 16, 2023
eNeura, Inc. % Larry Getlin Regulatory Consultant eNeura, Inc 2690 Pheasant Road Orono, Minnesota 55331
Re: K230358
Trade/Device Name: SAVI Dual (TM) Migraine Therapy Regulation Number: 21 CFR 882.5808 Regulation Name: Transcranial Magnetic Stimulator For Headache Regulatory Class: Class II Product Code: OKP Dated: February 10, 2023 Received: February 10, 2023
Dear Larry Getlin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jitendra V. Virani -S
CDR Jitendra Virani Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
## 510(k) Number (if known)
## K230358
Device Name
## SAVI Dual™ Migraine Therapy
Indications for Use (Describe)
The SAVI Dual™ Migraine Therapy device is indicated for the acute and prophylactic treatment of migraine headache in adolescents (age 12 and older) and adults.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
B PSC Publishing Services (301) 443-6740
Over-The-Counter Use (21 CFR 801 Subpart C)
{3}------------------------------------------------
| Date Prepared: | 16-May-23 |
|-------------------------------------------------------------------------|----------------------------------------------------------------------|
| eNeura® Inc<br>101 West Dickman St.<br>Suite 900<br>Baltimore, MD 21230 | |
| Official Contact: | David Rosen - President and CEO<br>drosen@eneura.com<br>970-481-2492 |
| Submission Correspondent: | Larry W. Getlin |
| Proprietary or Trade Name: | SAVI Dual Migraine Therapy |
| Regulation Code: | 21 CFR§882.5808 |
| Regulation Name(s): | Transcranial Magnetic Stimulator for Headache |
| Device Class: | Class II |
| Product Code: | OKP |
| Predicate Device:<br>Reference Device: | K182976 – SpringTMS<br>K161663 – sTMS mini |
## Modification:
The subject device is similar to the predicate device except for:
- Hardware Modifications ●
- Software (firmware) updates
## Device Description:
The SAVI Dual Migraine Therapy device is a portable, hand-held device that delivers a brief single pulse of magnetic energy at 0.9 Tesla to the back of the head to induce an electrical current in a portion of the brain, called the occipital cortex, to stop or lessen the effects of migraine headaches. Since a single pulse of magnetic stimulation is emitted, this method of stimulation is called single pulse transcranial magnetic stimulation or sTMS. The SAVI Dual is indicated for the acute and prophylactic treatment of migraine headache in adolescents (age 12 and older) and adults. The device is designed for patient use where treatments are self-administered and can be delivered in a variety of settings including the home or office. The device is intended for prescription use only. To use the device, the user must either insert a SIM chip to use the device for a programmed duration or the user may connect wirelessly via a cellular connection for the programmed duration. The programmed duration corresponds to the prescribed months of use.
## Indications for Use:
The SAVI Dual™ Migraine Therapy is indicated for the acute and prophylactic treatment of migraine headache in adolescents (age 12 and older) and adults.
{4}------------------------------------------------
| Feature | SAVI Dual™ Migraine<br>Therapy Device | SpringTMS®<br>Predicate Device | sTMS mini<br>Reference Device | Substantial<br>Equivalence<br>Rationale |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | eNeura, Inc. | eNeura, Inc. | eNeura, Inc. | -- |
| 510(k) Number | New Submission | K182976 | K161663 | -- |
| Indications for<br>Use | The SAVI Dual™<br>Migraine Therapy device<br>is indicated for the acute<br>and prophylactic treatment<br>of migraine headache in<br>adolescents (age 12 and<br>older) and adults. | The eNeura Inc.<br>SpringTMS® is indicated<br>for the acute and<br>prophylactic treatment of<br>migraine headache in<br>adolescents (age 12 and<br>older) and adults. | The eNeura® sTMS mini<br>is indicated for the acute<br>treatment of pain<br>associated with migraine<br>headache with aura. | Indications for Use are<br>identical to the<br>predicate device. |
| Operating<br>Principle | Induces electrical current in region near coil Transcranial Evoked response Stimulation on the occipital cortex | Induces electrical current in region near coil Transcranial Evoked response Stimulation on the occipital cortex | Induces electrical current in region near coil Transcranial Evoked response Stimulation on the occipital cortex | N/A (same) |
| Design | Time varying magnetic field Non-invasive | Time varying magnetic field Non-invasive | Time varying magnetic field Non-invasive | N/A (same) |
| Use<br>Authorization | With the SAVI Dual™<br>Migraine Therapy device,<br>the user may connect<br>either with the SIM chip<br>just as the predicate or the<br>user may connect<br>wirelessly via a cellular<br>connection for the<br>programmed duration. The<br>programmed duration<br>corresponds to the<br>prescribed months of use.<br>The device remains under<br>physician prescription. | The user must insert a SIM<br>chip to use the device for a<br>programmed duration.<br>The programmed duration<br>corresponds to the<br>prescribed months of use.<br>The SIM chip is only<br>available under physician<br>prescription. | The user must insert a<br>SIM chip to use the<br>device for a programmed<br>duration. The<br>programmed duration<br>corresponds to the<br>prescribed months of use.<br>The SIM chip is only<br>available under physician<br>prescription. | Subject device allows<br>for a wireless<br>connection /<br>communication for<br>user to use the device<br>for the programmed<br>duration in addition to<br>the use of SIM chip.<br>No new issues of<br>safety or efficacy are<br>raised with this<br>change. |
| Display | LED indicators | LCD display | LED indicators | Subject device has the<br>same display as the<br>reference device. No<br>new issues of safety or<br>efficacy are raised<br>with this change. |
| Feature | SAVI Dual™ Migraine<br>Therapy Device | SpringTMS®<br>Predicate Device | sTMS mini<br>Reference Device | Substantial<br>Equivalence<br>Rationale |
| Magnetic Field | 0.9 Tesla Peak @ 180 µs<br>(total magnetic energy<br>140J) | 0.9 Tesla Peak @ 180 µs<br>(total magnetic energy<br>140J) | 0.9 Tesla Peak @ 180 µs<br>(total magnetic energy<br>140J) | N/A (same) |
| Current | 4 mA/cm2 induced at 1 cm | 4 mA/cm2 induced at 1 cm | 4 mA/cm2 induced at 1 cm | N/A (same) |
| Electrical Power | Internally powered with<br>rechargeable embedded<br>lithium ion battery pack.<br>Battery pack charger mains<br>input -100-240V AC,<br>50/60 Hz, output 12 V DC | Internally powered with<br>rechargeable embedded<br>lithium ion battery pack.<br>Battery pack charger mains<br>input -100-240V AC,<br>50/60 Hz, output 12 V DC | Internally powered with<br>rechargeable embedded<br>lithium ion battery pack.<br>Battery pack charger<br>mains input -100-240V<br>AC, 50/60 Hz, output 12<br>V DC | N/A (same) |
| Materials | Hand held portable<br>stimulator in<br>polycarbonate case<br>(integral coil) | Hand held portable<br>stimulator in<br>polycarbonate case<br>(integral coil) | Hand held portable<br>stimulator in<br>polycarbonate case<br>(integral coil) | N/A (same) |
| Environment of<br>Use | Home-use and where the<br>operator is | Home-use and where the<br>operator is | Home-use and where the<br>operator is | N/A (same) |
| Dimensions and<br>Weight | 8.8 in. (22.4 cm) long<br>5.1 in. (13 cm) wide<br>2.7 in. (6.9 cm) deep<br>3.2 lb. (1.4 kg) | 9 in. (23 cm) long<br>5 in. (13 cm) wide<br>3 in. (8 cm) deep<br>3.8 lb. (1.7 kg) | 8.8 in. (22.4 cm) long<br>5.1 in. (13 cm) wide<br>2.7 in. (6.9 cm) deep<br>3.2 lb. (1.4 kg) | Subject device is<br>identical in size as the<br>reference device<br>(sTMS mini) No new<br>issues of safety or<br>efficacy are raised<br>with this change. |
| Compliance to<br>Standards | IEC 60601-1-2<br>IEC 60601-1<br>IEC 60601-1-11 | IEC 60601-1-2<br>IEC 60601-1<br>IEC 60601-1-11 | IEC 60601-1-2<br>IEC 60601-1<br>IEC 60601-1-11 | N/A (same) |
## Table 5.1 – Comparison – Subject vs. Predicate and Reference Devices
{5}------------------------------------------------
{6}------------------------------------------------
## Substantial Equivalence Discussion
## Intended Use/ Indications for Use
The indications for use are identical to those which were cleared for the predicate device, SpringTMS, under K182976.
## Patient Population
The patient population is identical to that cleared for the predicate device, SpringTMS, under K182976.
## Design and Technology
The SAVI Dual™ Migraine Therapy device, like the predicate device, is a portable, hand-held device that delivers a brief single pulse of magnetic energy at 0.9 Tesla to the back of the head to induce an electrical current in a portion of the brain, called the occipital cortex, to stop or lessen the effects of migraine headaches. Since a single pulse of magnetic stimulation is emitted, this method of stimulation is called single pulse transcranial magnetic stimulation or sTMS. The device is designed for patient use where treatments are self-administered and can be delivered in a variety of settings including the home or office. The device is intended for prescription use only.
To use the subject device, the user must either insert a SIM chip to use the device for a programmed duration or the user may connect wirelessly via a cellular connection for the programmed duration. The programmed duration corresponds to the prescribed months of use. The predicate device use authorization by the patient is done by inserting only a SIM chip into the device. Addition of the wireless cellular connection for subject device does not raise any new issues of safety or efficacy.
Thus, the design and technological characteristics of the SAVI Dual are substantially equivalent to the predicate device, SpringTMS (K182976).
## Principles of Operation
The principle of operation remains unchanged to the predicate including that each device:
- Induces electrical current in region near coil
- Transcranial
- . Evoked response
- Stimulation on the occipital cortex ●
## Use Authorization:
With the SAVI Dual™ Migraine Therapy device, the user may connect either with the SIM chip just as the predicate or the user may connect wirelessly via a cellular connection for the programmed duration. The programmed duration corresponds to the prescribed months of use. The device remains under physician prescription only. Though the predicate device use authorization only allowed for use by the patient inserting a SIM chip into the device, this technology modification does not raise any new issues of safety or efficacy.
## Environment of Use
The environment of use is identical to that cleared for the predicate device, SpringTMS under K 182976.
## Non-Clinical Testing
{7}------------------------------------------------
All necessary performance testing was conducted on the proposed SAVI Dual to support a determination of substantial equivalence to the predicate device. The testing performed is provided in the section below:
| Testing Type | Test Description | Results Supporting Substantial Equivalence |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Bench<br>Testing | Magnetic Pulse<br>Characteristics vs. Time | Both devices have the same specification for<br>magnetic pulse shape and both devices tested<br>within specification. No new issues of safety or<br>efficacy have been raised. The measured rate of<br>change of the magnetic field is substantially<br>equivalent. |
| | Magnetic Pulse Field Map | No new issues of safety or efficacy have been<br>raised. The Magnetic Pulse Field Maps for the<br>subject device and the predicate device are<br>substantially equivalent. |
| | Location of 5 Gauss Line | No new issues of safety or efficacy have been<br>raised. The location of the 5 Gauss line for the<br>subject device and the predicate device are<br>substantially equivalent. |
| Software Verification<br>Validation Testing | SAVI Dual Software Testing | The SAVI Dual software was tested against<br>requirements of the Software Requirements<br>Specification (SRS) and no new issues of safety<br>or efficacy have been raised. The software<br>requirements specify device operations that result<br>in the delivery of a magnetic pulse that is<br>substantially equivalent to the magnetic pulse of<br>the predicate device. |
| Electromagnetic<br>Compatibility and<br>Electrical Safety | Testing in accordance with<br>the following standards:<br>• IEC 60601-1-1<br>• IEC 60601-1-2<br>• IEC 60601-1-11 | The subject device met all acceptance criteria. No<br>new issues of safety or efficacy have been raised.<br>Therefore, the SAVI Dual is substantially<br>equivalent to the predicate. |
| Table 5.2 Non-Clinical Performance Testing and Substantial Equivalence Support | | |
|--------------------------------------------------------------------------------|--|--|
|--------------------------------------------------------------------------------|--|--|
The collective results of performance testing demonstrate that the materials chosen, the manufacturing processes, and design of the SAVI Dual meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bemonstrates that the SAVI Dual does not raise new questions of safety or effectiveness when compared to the predicate device.
## Substantial Equivalence Conclusion
The SAVI Dual is considered by eNeura to be substantially equivalent to the predicate device. Any differences do not present different questions of safety or effectiveness than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.