K162797 · Eneura, Inc. · OKP · Jun 26, 2017 · Neurology
Device Facts
Record ID
K162797
Device Name
SpringTMS
Applicant
Eneura, Inc.
Product Code
OKP · Neurology
Decision Date
Jun 26, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5808
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The eNeura Inc. SpringTMS® is indicated for the acute and prophylactic treatment of migraine headache.
Device Story
SpringTMS is a portable, hand-held device for single pulse transcranial magnetic stimulation (sTMS). It delivers a brief, 0.9 Tesla magnetic pulse to the occipital cortex to induce electrical currents, aiming to stop or reduce migraine effects. Designed for patient self-administration in home or office settings. Treatment regimen involves daily prophylactic pulses (4 pulses morning/evening) and acute treatment (3 pulses at onset, repeatable). Device is prescription-only. Healthcare providers use clinical outcomes to monitor treatment efficacy. Benefits include reduction in migraine headache days and decreased reliance on acute medications.
Clinical Evidence
Prospective, multicenter, non-randomized, single-arm observational study (ESPOUSE) with 217 subjects (Safety Data Set). Primary endpoint: mean reduction in migraine headache days over 28 days at 12 weeks. Results showed statistically significant reduction of 2.8 days (FAS, p<0.0001) and 3.0 days (PP, p<0.0001) from baseline. No serious adverse events reported; common AEs included lightheadedness, dizziness, tinnitus, and scalp discomfort.
Technological Characteristics
Portable, hand-held transcranial magnetic stimulator. Delivers 0.9 Tesla single magnetic pulse. Operates via electromagnetic induction to stimulate the occipital cortex. Class II device, Product Code OKP. Contraindicated for patients with metal-containing implants in the head, neck, or upper body (e.g., aneurysm clips, cochlear implants, shunts, stents, metal plates/screws).
Indications for Use
Indicated for acute and prophylactic treatment of migraine headache in patients 18 to 65 years of age with migraine with or without aura.
Regulatory Classification
Identification
A transcranial magnetic stimulator device for headache is a device that delivers brief duration, rapidly alternating, or pulsed, magnetic fields that are externally directed at spatially discrete regions of the brain to induce electrical currents for the treatment of headache.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing must demonstrate electromagnetic compatibility, electrical safety, and thermal safety.
(2) Appropriate verification, validation, and hazard analysis must be performed on the device software and firmware.
(3) The elements of the device that contact the patient must be assessed to be biocompatible.
(4) Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. This includes full characterization of the magnetic pulse output and resulting magnetic field map. This also includes characterization of the sound level of the device during use.
(5) Clinical testing must demonstrate that the device is safe and effective for treating headache in the indicated patient population.
(6) The physician and patient labeling must include the following:
(i) A summary of the clinical performance testing, including any adverse events and complications.
(ii) The intended use population in terms of the types of headaches appropriate for use with the device.
(iii) Information on how to report adverse events and device malfunctions.
(iv) A diagram or picture depicting the proper placement of the device on the user.
1. Appropriate analysis/testing must demonstrate electromagnetic compatibility (EMC), electrical safety, and thermal safety.
2. Appropriate verification, validation, and hazard analysis must be performed on the device software and firmware.
3. The elements of the device that contact the patient must be assessed to be biocompatible.
4. Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. This includes full characterization of the magnetic pulse output and resulting magnetic field map. This also includes characterization of the sound level of the device during use.
5. Clinical testing must demonstrate that the device is safe and effective for treating headache in the indicated patient population.
6. The physician and patient labeling must include the following:
a. A summary of the clinical performance testing, including any adverse events and complications.
b. The intended use population in terms of the types of headaches appropriate for use with the device.
c. Information on how to report adverse events and device malfunctions.
d. A diagram or picture depicting the proper placement of the device on the user.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 26, 2017
eNeura Inc. c/o Larry W. Getlin Consultant 715 North Pastoria Avenue Sunnyvale, California 94085
Re: K162797
Trade/Device Name: Spring TMS Regulation Number: 21 CFR 882.5808 Regulation Name: Transcranial Magnetic Stimulator For Headache Regulatory Class: Class II Product Code: OKP Dated: October 1, 2016 Received: October 4, 2016
Dear Larry Getlin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
William J. Heetderks -S 2017.06.26 13:17:30 -04'00'
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use Statement
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(K) Number (if known)
K162797
Device Name
## eNeura, Inc.- SpringTMS®
° Indications for Use (Describe)
The eNeura Inc. SpringTMS® is indicated for the acute and prophylactic treatment of migraine headache.
Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
## 7. Premarket Notification 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
### 510(k) Number: K162797
#### Applicant Information: 7.1.
| Date Prepared: | May 25, 2017 |
|----------------------|---------------------------------------------------------------|
| Applicant Name: | eNeura, Inc. |
| Address: | 715 North Pastoria Avenue<br>Sunnyvale, CA 94085<br>U.S.A. |
| Phone: | 408-245-6500 |
| FAX: | 408-245-6424 |
| Contact Person: | Larry W. Getlin, Regulatory Consultant<br>lwgetlin@gmail.com. |
| Mobile Number: | (612) 850-8144 |
| Alternative Contact: | Michael A Daniel, Consultant<br>madaniel@clinregconsult.com |
| Mobile Number: | (415) 407-0223 |
#### 7.2. Device Information:
| Device Trade Name: | SpringTMS® |
|---------------------------|--------------------------------------------------------|
| Classification Name(s): | Transcranial magnetic stimulator for migraine headache |
| Product Code/ Regulation: | OKP / 21 CFR§882.5808 |
| Classification: | Class II |
#### 7.3. Predicate Device:
SpringTMS®
#### 7.4. Subject Device Description
The SpringTMS® is a portable, hand-held device that is designed and intended to deliver a brief single pulse of magnetic energy at 0.9 Tesla to the back of the head to induce an electrical current in a portion of the brain called the occipital cortex to stop or lessen the effects of migraine headaches. Since a single pulse of magnetic stimulation is emitted, this method of stimulation is called single pulse transcranial magnetic stimulation or sTMS. The SpringTMS is indicated for the acute treatment and prevention of migraine headache. The device is designed for patient use where treatments are self-administered and can be delivered in a variety of settings including the home or office. The device is intended for prescription use only.
{4}------------------------------------------------
#### 7.5. Intended Use / Indications for Use
The eNeura® Spring TMS® is designed and intended to deliver brief duration, pulsed, magnetic fields that are externally directed at spatially discrete regions of the brain to induce electric currents in the brain (Product Code OKP).
The eNeura Inc. SpringTMS® is indicated for the acute and prophylactic treatment of migraine headache.
#### 7.6. Predicate and Subject Device Comparison Chart
### Comparison to Predicate and Reference Device:
| | Subject Device | Predicate Device | Comparison to predicate<br>device |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| Device Name | SpringTMS | SpringTMS | Same |
| Manufacturer | eNeura Inc. | eNeura Inc. | Same |
| 510(k) # | Not Assigned | K140094 | N/A |
| Regulation<br>Number | 21 CFR§882.5808 | 21 CFR§882.5808 | Same |
| Class | Class II | Class II | Same |
| Device<br>Class/Name | Transcranial magnetic<br>stimulator | Transcranial magnetic<br>stimulator | Same |
| Product Code | OKP | OKP | Same |
| Fundamental<br>scientific<br>technology | Portable, hand-held device<br>that is designed and intended<br>to deliver a brief single pulse<br>of magnetic energy at 0.9<br>Tesla to the back of the head<br>to induce an electrical<br>current in a portion of the<br>brain called the occipital<br>cortex to stop or lessen the<br>effects of migraine<br>headaches. | Portable, hand-held device<br>that is designed and intended<br>to deliver a brief single pulse<br>of magnetic energy at 0.9<br>Tesla to the back of the head<br>to induce an electrical<br>current in a portion of the<br>brain called the occipital<br>cortex to stop or lessen the<br>effects of migraine<br>headaches. | Same |
| Intended Use<br>(From<br>Product Code<br>Description) | Intended to deliver<br>externally directed, pulsed,<br>magnetic fields to induce<br>electric currents in spatially<br>discrete regions of the brains<br>of patients with migraine<br>headache. | Intended to deliver<br>externally directed, pulsed,<br>magnetic fields to induce<br>electric currents in spatially<br>discrete regions of the brains<br>of patients with migraine<br>headache. | Same |
{5}------------------------------------------------
| | Subject Device | Predicate Device | Comparison to predicate device |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Indication for Use | Indications for Use:<br>• The eNeura Inc.<br>SpringTMS® is indicated<br>for the acute and<br>prophylactic treatment of<br>migraine headache. | Indications for Use:<br>• The acute treatment of<br>pain associated with<br>migraine headache with<br>aura. | ADDITIONAL<br>INDICATIONS SUPPORTED<br>BY CLINICAL DATA<br>PROVIDED IN THIS<br>SUBMISSION. |
Table 7.1. Comparison between subject and predicate devices.
#### 7.7. Testing Completed
In addition to previously completed in vitro and in vivo testing (reference K140094), eNeura completed a prospective, multicenter, observational, clinical study demonstrating device safety and effectiveness in treating and preventing and migraine headaches.
| Test Device | eNeura SpringTMS® Device | |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Study Design | A prospective, non-randomized, single arm, multi-center observational<br>study designed to evaluate the use of the SpringTMS system in<br>reducing the frequency of headache days against a performance goal. | |
| Number of<br>Patients | A total of 263 subjects were consented between December 2014 and<br>March 2016. 229 of these subjects completed a Baseline Diary and 220<br>were confirmed by the sites to be eligible for participation. There were<br>217 subjects that were assigned Spring TMS devices and these subjects<br>comprise the Safety Data Set. There were 179 subjects who began<br>treatment and completed a Month 1 treatment Diary, but 47 of these<br>subjects did not meet the definition of a Migraine Headache Day<br>(minimum requirement of at least 4 days with moderate to severe<br>headache pain for at least 4 hours at baseline). Thus 132 of these subjects<br>complied with the protocol requirements based upon headache day<br>definition. This was the Full-Analysis Data Set (FAS) described below.<br>There were 117 of these subjects that went on to finish treatment and<br>completed both baseline and Month 3 diaries. This was the Completed<br>Cases data set (CC). Of these subjects 95 complied with the protocol<br>instructions regarding use of the device. This was the per Protocol (PP)<br>data set. | |
| Number of Sites | Seven (7) | |
| Duration of Trial | Each provisionally enrolled subject was followed for 1 month (28±5<br>days) to establish a baseline number of headache days (HD) and<br>confirm final eligibility and then each confirmed enrolled patient was<br>followed for 3 consecutive months (12±1 weeks) of treatment followed<br>by a final assessment. | |
| Treatment | Patients were instructed to treat daily using the following protocol: | |
| Regimen | 1. From the start, treat with 4 Pulses each morning and evening:<br>2 consecutive pulses wait 15 minutes and repeat the 2<br>consecutive pulses. | |
| | 2. Additionally, the patient may treat an acute attack with:<br>3 sequential pulses (early) at the onset of migraine pain<br>Wait 15 minutes, if needed treat with additional 3 pulses<br>Wait 15 minutes, if needed treat with additional 3 pulses<br>Patients may rescue with acute medication 30 minutes after the first<br>three pulses are delivered. | |
| Performance Goal | The originally proposed performance goal for the primary endpoint,<br>representing the mean reduction in headache days for a controlled<br>study, assumed an approximate 80% chronic headache population and<br>was PGMD = -5.3 days. | |
| | This performance goal was amended ahead of un-blinding and the<br>final analysis to PGD=-0.633 in order to reflect the actual population of<br>~20% chronic and ~80% episodic enrollment. | |
| Primary<br>Effectiveness<br>Endpoint | Mean reduction in headache days over a 28-day period at 12±1 weeks<br>(i.e., weeks 9 through 12 from start of treatment period). The null and<br>alternative hypotheses for this endpoint are: | |
| | H0: $μ$ T ≥ PGMD versus Ha: $μ$ T < PGD | |
| | PGD = - 0.633 days. | |
| | where $μ$ T is the mean reduction in migraine headache days from<br>baseline at 12±1 weeks in the treated population and PGD is the<br>Performance Goal. | |
| | Headache day is defined as ≥4 hours of headache pain which at any<br>time reaches moderate or severe intensity. | |
| Secondary<br>Effectiveness<br>Endpoints | 1. The percentage of subjects who had at least a 50% reduction in<br>headache days | |
| | 2. The reduction from baseline in the days of medications use to<br>acutely treat migraine headaches | |
| | 3. The reduction from baseline in the HIT6 impact questionnaire | |
| | 4. The reduction from baseline in the days with headache for more<br>than 4 hours with any pain intensity | |
| | 5. Migraine is defined as >4 hours of headache pain which at any<br>time reached moderate or severe intensity. | |
| Primary Safety<br>Endpoint | The proportion of patients experiencing any adverse event in aggregate<br>and by event. | |
| Inclusion Criteria | 1. Patients 18 to 65 years of age; | |
| | 2. Patients able to understand and communicate in English; | |
| | 3. Migraine with or without aura; | |
| | 4. 4-25 headache days per month (confirmed by 1-month baseline<br>diary, minimum of 5 complete headache-free days/month); | |
| Exclusion Criteria | 5. Understand and willing to provide diary and survey data.<br>Subjects will be excluded from participating in this trial if they meet any of the following criteria | |
| | 1. Severe co-existing disease having a life expectancy of less than 1 year; | |
| | 2. Currently involved in any other investigational clinical trials that have not completed their primary endpoint or that may interfere with the SpringTMS study results; | |
| | 3. Mental impairment or other conditions which may not allow the subject to understand the nature, significance and scope of the study and to cooperate with follow-up requirements; | |
| | 4. Known drug and/or alcohol addiction or use of illicit substances; | |
| | 5. Patients with epilepsy or history of seizure; | |
| | 6. Severe active major depression or major psychiatric illness; | |
| | 7. Concurrent use of other neurostimulation devices (Cefaly®, TENS, implantable devices); | |
| | 8. Use of Botox® within past 4 months; | |
| | 9. Extracranial nerve block (e.g. occipital, supraorbital) within past 3 months; | |
| | 10. Use of Cefaly for prevention within past month; | |
| | 11. Patients with metal containing implants as follows: | |
| | The SpringTMS may <b>not</b> be used in patients who have metals, conductive materials, or metal-containing implants in their head, neck or upper body. Patients with implants that are affected by a magnetic field should not use the SpringTMS. Examples of such implants include: | |
| | ● Aneurysm clips or coils<br>● Cochlear implants<br>● Cerebral spinal fluid shunts<br>● Bullets or pellets lodged in the head or upper body<br>● Filters<br>● Electrodes | ● Radioactive seeds<br>● Magnetically programmable shunt valves<br>● Stents<br>● Metal plates, screws, staples or sutures in skull, neck, shoulders, arms or hands<br>● Metallic artificial heart valves<br>● Facial tattoos with metallic ink |
| | Dental implants, fillings, or other dental appliances are okay and are not affected by the device. | |
| | Note: although not explicitly excluded, safety and effectiveness have not been established in pregnant women. Please defer to the judgment of the investigator when considering the eligibility of this population. | |
| Final Enrollment Decision | 1. Satisfactory completion of Baseline Patient Diary (~80% or 22 out of 28 days) | |
| Subject Demographics | 106 Female / 132 Total (80.3% Female)<br>Age (Range 16-65 years, Mean 42.8 years) | |
7.8. Clinical Summary
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| | Baseline # of migraine days/month (Range 4-21 days,<br>Mean 9.06 days, Median 9.0 days) | | | |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|--------------------|---------------------------------------------------------------------------------|
| Primary Safety<br>Endpoint results | Approximately 29% of the 217 subjects included in the Safety Dataset reported<br>experiencing at least one adverse event in this study. No subject had events that could be<br>determined to be serious adverse events. None of the events required treatment. Adverse<br>events as described below are the same as those reported in previous studies. | | | |
| | Adverse Events Reported in the ESPOUSE Study (greater than 2%) |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.