TheraFace Mask

K230293 · Therabody, Inc. · OHS · Jun 9, 2023 · General, Plastic Surgery

Device Facts

Record IDK230293
Device NameTheraFace Mask
ApplicantTherabody, Inc.
Product CodeOHS · General, Plastic Surgery
Decision DateJun 9, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Red Light is intended to treat full face wrinkles Blue Light is intended to treat mild to moderate inflammatory acne Red + Infrared Light is intended to treat full face wrinkles

Device Story

TheraFace Mask is a wearable, face-shaped LED device for OTC use. It utilizes 648 LEDs to emit red (633nm), blue (415nm), or combined red/infrared (633nm/830nm) light to treat facial wrinkles and inflammatory acne. The device includes 17 vibration motors (8 around eyes, 9 on head) offering three patterns (continuous, breathe, wave) for relaxation. Powered by two internal rechargeable lithium-ion batteries, it is operated via two physical buttons. Users wear the mask to receive light therapy; vibration can be used concurrently or independently. The device provides non-invasive light energy to the skin surface. Benefits include wrinkle reduction and acne management. It is designed for home use by the patient.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-5, -10, -11, -23), electrical safety (IEC 60601-1, -1-2, -1-11, -2-57), photobiological safety (IEC 62471), battery safety (IEC 62133), and usability/cleaning validation.

Technological Characteristics

Materials: PC+ABS. Light source: 648 LEDs. Wavelengths: Red (633±10nm), Blue (415±10nm), IR (830±10nm). Power: 2x 3.7V 1500mAh Li-ion batteries, USB-C/A charging. Form factor: Face-shaped mask. Connectivity: None. Software: Microprocessor-controlled. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60601-2-57, IEC 62471, ISO 10993.

Indications for Use

Indicated for treatment of full face wrinkles (Red Light or Red + Infrared Light) and mild to moderate inflammatory acne (Blue Light) in adults.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 9, 2023 Therabody, Inc. % Thomas Padula Vice President Regulatory Compliance Schiff & Company, Inc. 583 Mountain Avenue North Caldwell, New Jersey 07006 Re: K230293 Trade/Device Name: TheraFace Mask Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS, OLP Dated: May 10, 2023 Received: May 10, 2023 Dear Thomas Padula: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Jianting Wang -S Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use Submission Number (if known) K230293 Device Name TheraFace Mask Indications for Use (Describe) - · Red Light is intended to treat full face wrinkles - · Blue Light is intended to treat mild to moderate inflammatory acne - · Red + Infrared Light is intended to treat full face wrinkles Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary (as required by 807.92) # K230293 Date: June 6, 2023 (1) SUBMITTER: THERABODY, Inc. 6100 Wilshire Blvd Suite 200 Los Angeles, CA 90048 Registration Number: 3012386142 FEI Number: 3012386142 Contact person: CJ Frederick, III Telephone: 484-888-1290 Email: cifrederick@therabodycorp.com Date prepared: May 18, 2023 #### Application Correspondent: Contact Person: Thomas Padula Company: Schiff & Company, Inc. Address: 583 Mountain Avenue, North Caldwell, NJ 07006 Tel: 201-317-8810 Email: thomaspadula@schiffandcompany.com (2) DEVICE NAME: Trade Name: TheraFace Mask Common Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Classification Name: Light Based Over the Counter Wrinkle Reduction Device Classification: Class II Review Panel: General & Plastic Surgery Regulation Number: 21 CFR 878.4810 Product Code: OHS, OLP - (3) PREDICATE DEVICE(S): Substantial equivalence is based on following legally marketed devices. | Device Name and<br>Model | MZ Skin LightMAX<br>Supercharged LED Mask 2.0 | LED Therapy Device | RED Light Device | |--------------------------|-----------------------------------------------|----------------------------|-------------------------------| | 510(k) Number | K213184 (Primary Predicate<br>device) | K192295 (Predicate device) | K162489 (Reference<br>device) | | Product Code | OHS, OLP | OHS, OLP | OHS | | Regulation Number | 878.4810 | 878.4810 | 878.4810 | | Regulation Class | II | II | II | {4}------------------------------------------------ # (4) DESCRIPTION OF THE DEVICE: The TheraFace Mask device is a lightweight device which uses specified wavelengths of LED light. For LED light irradiation function, the device produces light in the red-light region of the spectrum (633±10nm), combination of IR and Red light (830nm±10nm & 633 ± 10 nm), or in the blue light region of the spectrum (415±10nm). The TheraFace Mask device is shaped like a human face and is designed to be "one size fits most." There are two physical buttons located on the mask; one controls the LED function and the other controls the vibration. The 648 LEDs in the device are powered by two internal lithium-ion rechargeable batteries which are charged via USB Type C or A cable with power adaptor. Red light mode: In Red light irradiation mode, the device utilizes Light Emitting Diodes to emit The output is one wavelength with a narrow spectral bandwidth in 633±10nm. red light. lt provides narrow bands of red-light energy to facial skin and is intended to treat full-face wrinkles. Blue light mode: In blue light irradiation mode, the device utilizes Light Emitting Diodes to emit blue light. The output is one wavelength with a narrow spectral bandwidth in 415±10nm. It provides narrow bands of blue light energy to facial skin and is intended to treat mild to moderate inflammatory acne. Red+ IR mode: When the device is operated in the red combined with infrared light mode, it emits LED light in the RED (633 nm±10nm) and IR (830 nm±10nm) spectrum on facial skin. lt is intended to treat full face wrinkles. Vibration mode: The device can drive 8 vibration motors around the eyes and 9 vibration motors on the top and back of the head in different vibration speeds. There are 3 different vibration patterns; continuous mode, breathe mode, and wave mode. Vibration is included for general relaxation purposes. # (5) INDICATIONS FOR USE: The device can work in multiple modes as described below, with the corresponding indications for use: - •Red Light is intended to treat full face wrinkles - ·Blue Light is intended to treat mild to moderate inflammatory acne - ·Red + Infrared Light is intended to treat full face wrinkles - (6) COMPARISON WITH PREDICATE DEVICES: Following table is a comparison of TheraFace Mask and predicate/reference devices. TheraFace Mask is substantially equivalent in terms of the technological characteristics, {5}------------------------------------------------ features, specifications, materials, mode of operation and indications for use, to MZ Skin LightMAX Supercharged LED Mask 2.0 K213184 (Primary Predicate Device), LED Therapy Device, K192295 (Secondary Predicate device), and RED Light Device K162489 (Reference device), cleared for marketing under 510(K). Comparison in Detail(s): Substantially Equivalent (SE) Comparison | Item | Proposed Device | Primary Predicate Device<br>K213184 | Secondary Predicate Device<br>K192295 | Reference Device<br>K162489 | Remark | |--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Code | OLP, OHS | OLP, OHS | OHS, OLP | OHS | (Please provide Justification for differences) | | Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | SAME | | Class | Class 2 | Class 2 | Class 2 | Class 2 | SAME | | Indication for Use | •Red Light is intended to treat full face wrinkles<br>•Blue Light is intended to treat mild to moderate inflammatory acne<br>•Red + Infrared Light is intended to treat full face wrinkles | The MZ Skin LightMAX Supercharged LED Mask 2.0 is an over-the-counter device intended to emit energy in the red and blue region of the light spectrum, specifically indicated to treat mild to moderate acne vulgaris of the face. The MZ Skin LightMAX Supercharged LED Mask 2.0 is an over-the-counter device intended to emit energy in the red and Near Infra-red spectrum and is intended for the use in the treatment of full face wrinkles. | The red light is intended for the treatment of periorbital wrinkles and the blue light is intended for the treatment of the mild to moderate inflammatory acne. The device is indicated for adults only. | The RED Light Device is an OTC device indicated to emit energy in the red and IR region of the spectrum for use in the dermatology for the treatment of periorbital wrinkles. | No differences in Indications for Use for Red Light, Blue Light or Red Light + Near Infra-Red Light. Of note is that while the primary predicate uses the term "acne vulgaris," this is a form of inflammatory acne and therefore the IFU for blue light are identical for all devices. | Table 1 General Comparison {6}------------------------------------------------ | Prescription/OTC | OTC | OTC | OTC | OTC | SAME | |------------------|-----|-----|-----|-----|------| |------------------|-----|-----|-----|-----|------| # Table 2 Performance Comparison | Item | Proposed Device | Primary<br>Predicate<br>Device K213184 | Secondary<br>Predicate Device<br>K192295 | Reference<br>Device K162489 | Remark | |--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Power Source | 5-15V DC 2.5A max<br>powered by 2 Li-Ion<br>Batteries 3.7V<br>1500mAh) is charged<br>via Universal USB<br>charger cord or fast<br>charger adaptor | Lithium-ion<br>battery powered<br>controller.<br>Power Supply<br>charges the<br>batter by direct<br>plug-in adapter<br>(2 or 3 pin input<br>socket and wall<br>plug. Power<br>cable is<br>connected to the<br>controller by a<br>standard micro-<br>USB A-C<br>Connector. | 5.V DC 2.0 A<br>Powered by direct<br>plug-in adapter:<br>Input 100-240V<br>AC, 50/60 Hz, 0.5A<br>Max., Output 5.0V<br>DC 2.0A | Adaptor:<br>100~240V AC<br>50/60Hz<br><br>Lithium Battery: 2<br>x 3.7V | Note 1 | | Software/Firm<br>ware/Microprocessor<br>Control? | Yes | Yes | Yes | Yes | SAME | | Power<br>(mW/cm²) | Red: 73 $\pm$ 5mW/cm²<br>Blue: 64 $\pm$ 5mW/cm²<br>Red+IR: 73 $\pm$ 5mW/cm²/<br>55 $\pm$ 5mW/cm² | Blue/Red:<br>44mW/cm²<br><br>Red/NIR:<br>29mW/cm² | Red light: 80 $\pm$ 10%<br>Blue light: 50 $\pm$ 10% | 125mW/cm²<br><br>70mW/cm²<br>(633nm); 55<br>mW/cm²<br>(830nm) | Reference device<br>has the<br>same/similar<br>power density as<br>proposed device<br>and has been<br>cleared therefore<br>proposed device is<br>safe and effective. | | Dose (J/cm²) | Red 13.14 +/-0.9 J/cm²<br>Blue: 11.52 +/-0.9 J/cm²<br>Red+IR: 11.52 +/- 0.9<br>J/cm² | Red 9.6J/cm² &<br>11J/cm²<br>Blue: 16.8J/cm²<br>NIR: 7J/cm² | Not available | Not available | | | Wavelength | Red: 633 $\pm$ 10nm<br>Blue: 415 $\pm$ 10nm<br>Red+IR:633nm $\pm$ 10nm/830 $\pm$ 10 nm | Blue light:<br>415nm $\pm$ 10nm<br>Red light: 630nm $\pm$ 10nm<br>NIR: 830+10nm | Blue light: 415nm<br>$\pm$ 5nm<br>Red light: 630nm<br>$\pm$ 5nm | Red: 633 +5nm<br>IR: 830 +5nm | Note 2<br><br>SAME for Blue<br>Light & Red+IR<br>(aka IR or NIR)<br>Light | {7}------------------------------------------------ | Treatment<br>Duration | LED: 3 minutes each<br>light mode for a total of 9<br>minutes per treatment,<br>recommended to use 2<br>to 5 times per week.<br><br>Vibration: accompanies<br>LED treatments or can<br>be used without LED's<br>active. 3 vibration<br>patterns, 5 minutes each,<br>for a total of 15 minutes.<br>During blue light<br>treatment mode,<br>vibration is not active<br>around the eyes.<br>Vibration is included for a<br>more relaxing<br>experience. | 10 minutes per<br>treatment.<br><br>Acne: 4x<br>Weekly, 6 weeks<br><br>Wrinkles: 5x<br>Weekly, 6 weeks | 3-5 minutes each<br>time, twice<br>a week | For the first<br>month (4 weeks),<br>treatment should<br>be performed 3<br>times a week for<br>15-20 minutes<br>each time (5-7<br>minutes on each<br>treatment zone). | Note 3 | |-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------| | Main Materials | PC+ABS | Silicone | PC+ABS | ABS + Stainless<br>Steel | SAME and Similar | # Table 3 Safety Comparison | Item | Proposed Device | Primary Predicate<br>Device #1<br>K213184 | Secondary Predicate<br>Device<br>K192295 | Reference Device<br>K162489 | Remark | |---------------------------|--------------------------------------------------------------------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------|---------------------------------------------------------------|--------| | Electrical Safety | Complies with IEC<br>60601-1,<br>IEC 60601-1-11 | Comply with IEC<br>60601-1,<br>IEC 60601-1-11 | Comply with IEC<br>60601-1,<br>IEC 60601-1-11 | Comply with IEC<br>60601-1, IEC 60601-<br>1-11, IEC60601-2-57 | SAME | | Photobiological<br>Safety | Complies with IEC<br>62471<br>Complies with IEC<br>60601-2-57 | Comply with IEC<br>62471 | Comply with IEC<br>62471 | IEC 62471 | SAME | | EMC | Complies with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | IEC 60601-1-2 | SAME | | Biocompatibility | Complies with ISO<br>10993-1,<br>ISO 10993-5 and<br>ISO 10993-10<br>ISO 10993-11<br>ISO 10993-23 | Comply with ISO<br>10993-1,<br>ISO 10993-5 and<br>ISO 10993-10 | Comply with ISO<br>10993-1,<br>ISO 10993-5 and ISO<br>10993-10 | Comply with ISO<br>10993-1 | SAME | | Label and | Conforms to FDA | Conforms to FDA | Conforms to FDA | Conforms to FDA | SAME | {8}------------------------------------------------ | Labeling | Regulatory<br>Requirements | Regulatory<br>Requirements | Regulatory<br>Requirements | Regulatory<br>Requirements | | |----------|----------------------------|----------------------------|----------------------------|----------------------------|--| |----------|----------------------------|----------------------------|----------------------------|----------------------------|--| #### Difference Analysis: (Please explain any item that is not "SAME") Note 1: The proposed device is battery powered, similar to the reference devices. The battery in the proposed device supports a longer battery life as well as fast charging. This does not affect the safety, effectiveness, or indications for use of the proposed device. # Note 2: The target LED wavelength and variance range of the proposed, predicate, and reference device (variance ranges not provided) differ by just 3nm for Red Light at the high and low end of the output spectrum, with the proposed device being slightly higher. This represents a very minor difference and not one to affect equivalence. Furthermore, the proposed device has passed testing according to IEC60601-2-57. There is no effect on safety, effectiveness, or intended use of the proposed device. # Note 3: Predicate device and reference devices do not have vibration. Proposed device is used for 7 less minutes per light treatment compared to the predicate device and is either identical or less than reference devices with respect to treatment time per light mode. # Substantially Equivalent (SE) Conclusion Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicates K213184 and K192295 and Reference Device K162489. # (7) PERFORMANCE STANDARDS APPLIED: A series of studies were completed to demonstrate the substantial equivalence of TheraFace Mask to the predicate/reference devices. All testing was conducted in accordance with and in conformance to applicable device regulations and guidance. Results of all testing demonstrate the device is non-toxic, is comparable to other currently marketed devices and is substantially equivalent to legally marketed predicates and included: #### Biocompatibility ISO 10993-5:2009, biological evaluation of medical devices - part 5: tests for In Vitro Cytotoxicity Test (CSTBB2022120087) ISO 10993-10 :2010, Biological evaluation of medical devices - part 10: Skin Sensitization Test (CSTBB2022120404R1) {9}------------------------------------------------ ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: - Acute Systemic Toxicity Test (CSTBB2022110675R1) - Material-mediated Pyrogens Test (CSTBB2022120016R1) ISO 10993-23:2021, Biological evaluation of medical devices - Part 23: Tests for Intradermal Reactivity (CSTBB2022110602R1 # Electrical Safety and Electromagnetic Compatibility IEC60601-1:2005 +CORR.1:2006+ CORR.2:2007+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (CHTSM22120256) IEC 60601-1-2: 2014-02, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - collateral standard: Electromagnetic Compatibility. (CHTEM22120257) IEC 60601-1-11 :2015, Medical electrical equipment -- part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home health care environment. (CHTSM22120260) IEC 60601-2-57: 2011 for use in conjunction with IEC 60601-1:2005, Medical electrical equipment - Part 2: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use. (CHTSM22120259) IEC 62471:2006, Photobiological safety of lamps and lamp systems. (CHTSM22120258) IEC 62133 Edition 2.0 2012-12, Secondary Cells And Batteries Containing Alkaline Or Other Non-Acid Electrolytes - Safety Requirements For Portable Sealed Secondary Cells, And For Batteries Made From Them, For Use In Portable Applications (PNC221027191 01001) (8) PERFORMANCE TESTING BENCH Light power density test report (FP221125050384) {10}------------------------------------------------ Usability Study Report DES-013.3 Device Temperature Range Testing Final Report Apr 16, 2023 Cleaning and Disinfection Testing AAMI TIR30:2011/(R)2016 - A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices: Tests for Cleaning (CSTBB22100180) AAMI TIR12:2010-Designing, testing and labeling reusable medical devices for reprocessing in health care facilities-Section 5: Tests for Disinfection (CSTBB22100180) (9) PERFORMANCE TESTING CLINICAL There were no clinical studies performed. (10) CONCLUSION: TheraFace Mask has the same indications for use and technology characteristics as the predicate/reference devices. TTheraFace Mask is as safe, as effective, and performs as well as the predicate/reference devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...